AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature Sterilization System, V-PRO s2 Low Temperature Sterilization System, V-PRO 60 Low Temperature Sterilization System, RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP and 7053HP Single Chamber Washer Disinfector

K222849510(k) SpecialCleared Sep 30, 20229-day reviewClass II21 CFR 876.1500

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Steris · NVE Washer, Cleaner, Automated, Endoscope

Cleared on Sep 30, 2022 as substantially equivalent to K222543 (Steris, 2022) and 1 other predicate after a 9-day review. Its family tree reaches back 15 generations to 1994.

Decided
Sep 30, 2022
Review
9 days
Class
Class II
Product code
NVE
Recalls
None
Predicates
2

Its longest line of descent runs back 15 generations to K930667 (1994). 74 ancestors in all, the oldest K811390 (1981). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2019: 4 other ancestors2021: 1 other ancestor2018: 1 other ancestor2017: 1 other ancestor2012: 2 other ancestors2016: 1 other ancestor2002: 4 other ancestors2013: 1 other ancestor2005: 4 other ancestors2001: 6 other ancestors2011: 4 other ancestors2009: 2 other ancestors2007: 1 other ancestor1998: 3 other ancestors1997: 6 other ancestors2004: 3 other ancestors1995: 2 other ancestors1994: 2 other ancestors1993: 2 other ancestors1992: 1 other ancestor2006: 1 other ancestor2003: 1 other ancestor1999: 2 other ancestors1991: 2 other ancestors1988: 1 other ancestor1981: 1 other ancestorK930667 · Reusable Laparoscopic Instruments · Snowden-Pencer · 1994 · recalledK960400 · Diamond-Touch and Micro Diamond-Touch Instruments/Diamond-Line Instruments/Diamond-Port (Access Parts) · Snowden-Pencer · 1996K990144 · Intuitive Surgical Endoscopic Instruments, Intuitive Surgical Endoscopic Instrument Control System · Intuitive Surgical, Inc. · 2000 · recalledK021036 · Intuitive Surgical DA Vinci Surgical System, Model IS1000 · Intuitive Surgical, Inc. · 2002 · recalled×5: 5 generations cleared close together, K040237 (2004), K043153 (2004), K050005 (2005), K050404 (2005), K050802 (2005) K063220 · DA Vinci S Surgical System-V1.1, Model IS2000 · Intuitive Surgical, Inc. · 2006 · recalledK073112 · DA Vinci Sonicpro Cleaning System, Model 400253 · Intuitive Surgical, Inc. · 2008K190081 · RAS 12 Rack, RAS 12 Long Rack; RAS Cycle of AMSCO 7052HP/7053HP Single Chamber Washer/Disinfector · STERIS Corporation · 2019K200577 · RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP/7053HP Single Chamber Washer/Disinfector · STERIS Corporation · 2020K203199 · RAS 12 Rack / RAS 12 Long Rack, used in RAS Cycle of AMSCO 7052HP / 7053HP Single Chamber Washer-Disinfector · STERIS Corporation · 2020K222543 · AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature Sterilization System, V-PRO s2 Low Temperature Sterilization System, RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP and 7053HP Single Chamber Washer Disinfector · Steris · 202219801990200020102020
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 2 other ancestors4 generations back: 2 other ancestors5 generations back: 2 other ancestors6 generations back: 3 other ancestors7 generations back: 12 other ancestors8 generations back: 13 other ancestors9 generations back: 7 other ancestors10 generations back: 6 other ancestors11 generations back: 5 other ancestors12 generations back: 3 other ancestors K222543 · AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature Sterilization System, V-PRO s2 Low Temperature Sterilization System, RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP and 7053HP Single Chamber Washer Disinfector · Steris · 2022K203199 · RAS 12 Rack / RAS 12 Long Rack, used in RAS Cycle of AMSCO 7052HP / 7053HP Single Chamber Washer-Disinfector · STERIS Corporation · 2020K200577 · RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP/7053HP Single Chamber Washer/Disinfector · STERIS Corporation · 2020K190081 · RAS 12 Rack, RAS 12 Long Rack; RAS Cycle of AMSCO 7052HP/7053HP Single Chamber Washer/Disinfector · STERIS Corporation · 2019K073112 · DA Vinci Sonicpro Cleaning System, Model 400253 · Intuitive Surgical, Inc. · 2008K063220 · DA Vinci S Surgical System-V1.1, Model IS2000 · Intuitive Surgical, Inc. · 2006 · recalledK050802 · Modification to Intuitive Surgical DA Vinci Surgical System and Endoscopic Instruments · Intuitive Surgical, Inc. · 2005K050404 · Intuitive Surgical DA Vinci Surgical System and Endoscopic Instruments · Intuitive Surgical, Inc. · 2005K050005 · Intuitive Surgical Monopolar Curved Scissors, Model 400179; Tip Cover Accessory, Model 400180 · Intuitive Surgical, Inc. · 2005 · recalledK043153 · Intuitive Surgical DA Vinci Surgical System and Endoscopic Instruments, Models IS1200 & IS1000 · Intuitive Surgical, Inc. · 2004K040237 · Intuitive Surgical DA Vinci Endoscopic Instrument Control System and Endoscopic Instruments · Intuitive Surgical, Inc. · 2004K021036 · Intuitive Surgical DA Vinci Surgical System, Model IS1000 · Intuitive Surgical, Inc. · 2002 · recalledK990144 · Intuitive Surgical Endoscopic Instruments, Intuitive Surgical Endoscopic Instrument Control System · Intuitive Surgical, Inc. · 2000 · recalledK960400 · Diamond-Touch and Micro Diamond-Touch Instruments/Diamond-Line Instruments/Diamond-Port (Access Parts) · Snowden-Pencer · 1996K930667 · Reusable Laparoscopic Instruments · Snowden-Pencer · 1994 · recalled−15−10−50
  1. Root · 15 generations backK930667 Reusable Laparoscopic InstrumentsSnowden-Pencer · 1994 · Recalled
  2. 11 generations between
  3. 3 generations backK200577 RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP/7053HP…STERIS Corporation · 2020
  4. 2 generations backK203199 RAS 12 Rack / RAS 12 Long Rack, used in RAS Cycle of AMSCO 7052HP /…STERIS Corporation · 2020
  5. 1 generation backK222543 AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature…Steris · 2022
  6. This deviceK222849 AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature…Steris · 2022
Show all 15 generations
  1. Root · 15 generations backK930667 Reusable Laparoscopic InstrumentsSnowden-Pencer · 1994 · Recalled
  2. 14 generations backK960400 Diamond-Touch and Micro Diamond-Touch Instruments/Diamond-Line…Snowden-Pencer · 1996
  3. 13 generations backK990144 Intuitive Surgical Endoscopic Instruments, Intuitive Surgical…Intuitive Surgical, Inc. · 2000 · Recalled
  4. 12 generations backK021036 Intuitive Surgical DA Vinci Surgical System, Model IS1000Intuitive Surgical, Inc. · 2002 · Recalled
  5. 11 generations backK040237 Intuitive Surgical DA Vinci Endoscopic Instrument Control System and…Intuitive Surgical, Inc. · 2004
  6. 10 generations backK043153 Intuitive Surgical DA Vinci Surgical System and Endoscopic…Intuitive Surgical, Inc. · 2004
  7. 9 generations backK050005 Intuitive Surgical Monopolar Curved Scissors, Model 400179; Tip…Intuitive Surgical, Inc. · 2005 · Recalled
  8. 8 generations backK050404 Intuitive Surgical DA Vinci Surgical System and Endoscopic InstrumentsIntuitive Surgical, Inc. · 2005
  9. 7 generations backK050802 Modification to Intuitive Surgical DA Vinci Surgical System and…Intuitive Surgical, Inc. · 2005
  10. 6 generations backK063220 DA Vinci S Surgical System-V1.1, Model IS2000Intuitive Surgical, Inc. · 2006 · Recalled
  11. 5 generations backK073112 DA Vinci Sonicpro Cleaning System, Model 400253Intuitive Surgical, Inc. · 2008
  12. 4 generations backK190081 RAS 12 Rack, RAS 12 Long Rack; RAS Cycle of AMSCO 7052HP/7053HP…STERIS Corporation · 2019
  13. 3 generations backK200577 RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP/7053HP…STERIS Corporation · 2020
  14. 2 generations backK203199 RAS 12 Rack / RAS 12 Long Rack, used in RAS Cycle of AMSCO 7052HP /…STERIS Corporation · 2020
  15. 1 generation backK222543 AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature…Steris · 2022
  16. This deviceK222849 AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature…Steris · 2022
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−15K930667Reusable Laparoscopic InstrumentsSnowden-Pencer199415
−14K960400Diamond-Touch and Micro Diamond-Touch Instruments/Diamond-Line…Snowden-Pencer1996—
−13K990144Intuitive Surgical Endoscopic Instruments, Intuitive Surgical Endoscopic…Intuitive Surgical, Inc.20003
−12K021036Intuitive Surgical DA Vinci Surgical System, Model IS1000Intuitive Surgical, Inc.200218
−11K040237Intuitive Surgical DA Vinci Endoscopic Instrument Control System and…Intuitive Surgical, Inc.2004—
−10K043153Intuitive Surgical DA Vinci Surgical System and Endoscopic Instruments, Models…Intuitive Surgical, Inc.2004—
−9K050005Intuitive Surgical Monopolar Curved Scissors, Model 400179; Tip Cover…Intuitive Surgical, Inc.20052
−8K050404Intuitive Surgical DA Vinci Surgical System and Endoscopic InstrumentsIntuitive Surgical, Inc.2005—
−7K050802Modification to Intuitive Surgical DA Vinci Surgical System and Endoscopic…Intuitive Surgical, Inc.2005—
−6K063220DA Vinci S Surgical System-V1.1, Model IS2000Intuitive Surgical, Inc.20064
−5K073112DA Vinci Sonicpro Cleaning System, Model 400253Intuitive Surgical, Inc.2008—
−4K190081RAS 12 Rack, RAS 12 Long Rack; RAS Cycle of AMSCO 7052HP/7053HP Single Chamber…STERIS Corporation2019—
−3K200577RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP/7053HP Single…STERIS Corporation2020—
−2K203199RAS 12 Rack / RAS 12 Long Rack, used in RAS Cycle of AMSCO 7052HP / 7053HP…STERIS Corporation2020—
−1K222543AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature Sterilization System…Steris2022—
0 (this device)K222849AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature Sterilization System…Steris2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198810
199120
199210
199320
199420
199520
199760
199830
199920
200160
200240
200310
200430
200540
200610
200710
200920
201140
201220
201310
201610
201710
201810
201940
202110
FIG. 1 — Lineage of K222849, 1994–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NVE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NVE device

  1. #1 K222543 AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low…score 10.8 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K222543AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature Sterilization System, V-PRO s2…
Summary page 13
…n Name: Sterilizer, Steam Classification Number: 21 CFR 880.6880 Product Code: FLE, PEC 2. Predicate Device K222543 AMSCO 600 Steam Sterilizer, V-PRO maX 2 and s2 Low Temperature Sterilization Systems, RAS 12 Rack, RAS 1…
Steris2022—
*K190917V-PRO 60 a Low Tempurature Strilization Systems, V-PRO s2 Low Temperature Sterilization…
Summary page 28
…e: Sterilizer, Ethylene Oxide Gas Classification Number: 21 CFR 880.6860 Product Code: MLR 9. Predicate Device K190917 V-PRO 60 Low Temperature Sterilization Systems, V-PRO s2 Low Temperature Sterilization Systems 10. Description…
STERIS Corporation2019—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 4 companies; a new generation was cleared every 1.4 years on median.
  • 15 of the 74 ancestors have a recall of any class.

How does its category compare?

All NVE devices

NVE Washer, Cleaner, Automated, Endoscope: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.

FDA decisions per year · NVE

2in 2022

2008: 12019: 12020: 22022: 2
2007high 2 (2020) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NVE

Too few NVE 510(k)s in any year since 2007 to give a median.

Recalls per year · NVE

No recall names NVE since 2007.

Adverse event reports per year · NVE

1in 2025

2008: 202009: 2192010: 62012: 32015: 32016: 12017: 22018: 22019: 42020: 22025: 1
2007high 219 (2009) · low 0 (2007)2025

MAUDE reports naming product code NVE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—020
20090—0219
20100—06
20110—00
20120—03
20130—00
20140—00
20150—03
20160—01
20170—02
20180—02
20191—04
20202—02
20210—00
20222—00
20230—00
20240—00
20250—01
20260—0—

What is it indicated for?

The AMSCO 600 Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles (Table 1): The Automated Load and Unload System (ALUS) provides semi-automated loading and unloading from an AMSCO 600 steam sterilizer when a cycle is complete. Alternatively, the ALUS may also be used to provide automatic unloading only in combination with manual…

Read the full statement (1,858 words)

The AMSCO 600 Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles (Table 1):

The Automated Load and Unload System (ALUS) provides semi-automated loading and unloading from an AMSCO 600 steam sterilizer when a cycle is complete. Alternatively, the ALUS may also be used to provide automatic unloading only in combination with manual loading. The ALUS can start a cycle automatically when equipped with the optional bar code reader.

The V-PRO maX 2 Low Temperature Sterilization System using VAPROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and temperature, suitable for processing medical devices without leaving toxic residues.

Each Cycle can sterilize non-lumened instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors. Only stainless steel or titanium diffusion-restricted spaces should be processed in the Non Lumen Cycle and the Fast Non Lumen Cycle.

The Non Lumen Cycle can sterilize:

Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.

The validation studies were conducted using a validation load consisting of two instrument trays for a total weight of 50 lbs (22.7 kg).

The Fast Non Lumen Cycle can sterilize:

Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.

The validation studies were conducted using a validation load consisting of one pouched instrument tray for a total weight of 11 lbs (5 kg).

The Flexible Cycle can sterilize:

Single or dual lumen surgical flexible endoscopes (such as those used in ENT, Urology and Surgical Care) and bronchoscopes in either of the two configurations:

1. Two flexible endoscopes with a light cord (if not integral to endoscope) and mat with no additional load. The flexible endoscopes may contain either:

• A single lumen that is ≥ 1 mm internal diameter (ID) and ≤ 1050 mm in length

• Or two lumens with:

▪ One lumen that is ≥ 1 mm ID and ≤ 990 mm in length

▪ And the other lumen that is ≥ 1 mm ID and ≤ 850 mm in length

The validation studies were conducted with two flexible endoscopes, each packaged into a tray with silicone mat and light cord (if not integral to endoscope).

2. One flexible endoscope with a light cord (if not integral to endoscope) and mat and additional non-lumened instruments. The flexible endoscope may contain either:

• A single lumen that is ≥ 1 mm ID and ≤ 1050 mm in length

• Or two lumens with:

▪ One lumen that is ≥ 1 mm ID and ≤ 990 mm in length

▪ And the other lumen is ≥ 1 mm ID and ≤ 850 mm in length.

The validation studies were conducted with a flexible endoscope in a tray with silicone mat and light cord (if not integral to endoscope). Also included in the load were an additional instrument tray and one pouch for a total weight of 24 lbs (11 kg).

The Lumen Cycle can sterilize:

Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

• Single channeled devices with a stainless lumen that is ≥ 0.77 mm ID and ≤ 500 mm in length

• Single channeled devices with a stainless lumen that is ≥ 1.8 mm ID and ≤ 542 mm in length

• Dual channeled devices with stainless lumens that are ≥ 0.77 mm ID and ≤ 527 mm in length

• Triple channeled devices with stainless lumens that are either:

≥ 1.2 mm ID and ≤ 275 mm in length

≥ 1.8 mm ID and ≤ 310 mm in length

or

≥ 2.8 mm ID and ≤ 317 mm in length

Validation testing for all lumen sizes was conducted using a maximum of 20 lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of two instrument trays and two pouches for a total weight of 19.65 lbs (8.9 kg).

The V-PRO s2 Low Temperature Sterilization System using VAPROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues.

Each Cycle can sterilize non-lumened instruments and instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

The V-PRO s2 Sterilizer Non Lumen Cycle can sterilize:

Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.

The validation studies were conducted using a validation load consisting of one instrument tray for a total weight of 25 lbs (11.3 kg).

The V-PRO s2 Sterilizer Fast Cycle can sterilize:

Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

• Single or dual channeled devices with stainless steel lumens that are

≥ 0.77 mm (~1/32”) internal diameter (ID) and ≤ 410 mm (~16-9/64”) in length

≥ 1.8 mm (~5/64") ID x ≤ 542 mm (~21-5/16") in length

• Triple channeled devices with stainless steel lumens that are either:

≥ 1.2 mm (~3/64”) ID and ≤ 275 mm (~10-53/64”) in length

≥ 1.8 mm (~5/64”) ID and ≤ 310 mm (~12-13/64”) in length

or

≥ 2.8 mm (~7/64”) ID and ≤ 317 mm (12-31/64”) in length

Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.

Validation testing for all lumen sizes was conducted using a maximum of eight (8) lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. Validation testing was conducted using a validation load consisting of one pouched instrument tray and two pouched devices outside of the tray with a total weight of 4.0 lbs (~1.8kg).

The V-PRO s2 Sterilizer Flexible Cycle can sterilize:

One surgical flexible endoscope (such as those used in ENT, Urology and Surgical Care) or bronchoscope with light cord (if not integral to endoscope), mat, and additional load.

• The flexible endoscope may be a single or dual lumen device with lumens that are ≥ 1 mm ID and ≤ 990 mm in length

• Additional load, up to 11 lb (5 kg) can include stainless steel lumens with the following dimensions

≥ 0.76 mm (~1/32”) ID and ≤ 233 mm (~9 11/64”) in length

≥ 1.0 mm (~3/64”) ID and ≤ 254 mm (~10”) in length

≥ 1.8 mm (~5/64”) ID and ≤ 542 mm (~21-5/16”) in length

The validation studies were conducted using a validation load consisting of two instrument trays. One tray contained one flexible endoscope, with silicone mat, instrument organizers and light cord (if not integral to scope), and the second tray contained additional load and twelve (12) stainless steel lumens for a total load weight of 11 lbs (5 kg). Hospital loads should not exceed the maximum number of lumens validated by this testing.

The V-PRO s2 Sterilizer Lumen Cycle can sterilize:

Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

• Single or dual channeled devices with stainless steel lumens that are

≥ 0.77 mm (~1/32”) ID and ≤ 410 mm (16-9/64”) in length

≥ 1.8 mm (~5/64) ID x ≤ 542 mm (~21-5/16") in length

• Triple channeled devices with stainless steel lumens that are either:

≥ 1.2 mm (~3/64”) ID and ≤ 275 mm (~10-53/64”) in length

≥ 1.8 mm (~5/64”) ID and ≤ 310 mm (~12-13/64”) in length

or

≥ 2.8 mm (~7/64”) ID and ≤ 317 mm (12-31/64”) in length

Validation testing for all lumen sizes was conducted using a maximum of twelve (12) stainless steel lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of one instrument tray and two pouches for a total weight of 11 lbs (5.0 kg).

The V-PRO 60 Low Temperature Sterilization System using VAPROX HC Sterilant is intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) medical devices in Healthcare Facilities. The preprogrammed sterilization cycles operate at low pressure and low temperature, suitable for processing medical devices without leaving toxic residues.

Each Cycle can sterilize non-lumened instruments and instruments with diffusion-restricted spaces such as the hinged portion of forceps and scissors.

The V-PRO 60 Sterilizer Non Lumen Cycle can sterilize:

Non-lumened instruments including non-lumened general medical instruments, non-lumened rigid, semi-rigid and flexible endoscopes.

The validation studies were conducted using a validation load consisting of one instrument tray for a total weight of 25 lbs (11.3 kg).

The V-PRO 60 Sterilizer Flexible Cycle can sterilize:

One surgical flexible endoscope (such as those used in ENT, Urology and Surgical Care) or bronchoscope with light cord (if not integral to endoscope), mat, and additional load.

• The flexible endoscope may be a single or dual lumen device with lumens that are ≥ 1 mm (~3/64”) internal diameter (ID) and ≤ 990 mm (38-63/64”) in length

• Additional load, up to 11 lb (5 kg) can include stainless steel lumens with the following dimensions

≥ 0.76 mm (~1/32”) ID and ≤ 233 mm (~9-11/64”) in length

≥ 1.0 mm (~3/64”) ID and ≤ 254 mm (10”) in length

≥ 1.8mm (~5/64”) ID and ≤ 542 mm (~21-5/16”) in length

The validation studies were conducted using a validation load consisting of two instrument trays. One tray contained one flexible endoscope, with silicone mat, instrument organizers and light cord (if not integral to scope), and the second tray contained additional load and twelve (12) stainless steel lumens for a total load weight of 11 lbs (5 kg). Hospital loads should not exceed the maximum number of lumens validated by this testing.

The V-PRO 60 Sterilizer Lumen Cycle can sterilize:

Medical devices (including single, dual and triple channeled rigid and semi-rigid endoscopes) with the following configurations:

• Single or dual channeled devices with stainless steel lumens that are

≥ 0.77 mm (~1/32”) ID and ≤ 410 mm (16-9/64”) in length

≥ 1.8 mm (~5/64") ID x ≤ 542 mm (~21-5/16") length

• Triple channeled devices with stainless steel lumens that are either:

≥ 1.2 mm (~3/64”) ID and ≤ 275 mm (~10-53/64”) in length

≥ 1.8 mm (~5/64”) ID and ≤ 310 mm (~12-13/64”) in length

or

≥ 2.8 mm (~7/64”) ID and ≤ 317 mm (12-31/64”) in length

Validation testing for all lumen sizes was conducted using a maximum of twelve (12) stainless steel lumens per load. Hospital loads should not exceed the maximum number of lumens validated by this testing. The validation studies were performed using a validation load consisting of one instrument tray and two pouches for a total weight of 11 lbs (5.0 kg).

The RAS Racks are used in the RAS Cycle of the AMSCO 7052HP Single-Chamber Washer/Disinfector and the AMSCO 7053HP Single-Chamber Washer/Disinfector for the effective cleaning, rinsing, intermediate level disinfection and drying of reusable da Vinci® X/Xi and S/Si EndoWrist® instruments.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K222849: AMSCO 600 Steam Sterilizer, V-PRO maX 2 Low Temperature Sterilization System, V-PRO s2 Low Temperature Sterilization System, V-PRO 60 Low Temperature Sterilization System, RAS 12 Rack, RAS 12 Long Rack, RAS Cycle of the AMSCO 7052HP and 7053HP Single Chamber Washer Disinfector.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K222849

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026