510(k) · Cleared
Recalled · 2012DA Vinci S Surgical System-V1.1, Model IS2000
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Intuitive Surgical, Inc. · NAY System, Surgical, Computer Controlled Instrument
Cleared on Dec 1, 2006 as substantially equivalent to K050369 (Intuitive Surgical, Inc., 2005) and 3 other predicates after a 38-day review. Its family tree reaches back 9 generations to 1994. 1 of the 268 devices in its family, counting itself, has a Class I recall. 224 later devices descend from it. It has been recalled 4 times.
- Decided
- Dec 1, 2006
- Review
- 38 days
- Class
- Class II
- Product code
- NAY
- Recalls
- 4
- Predicates
- 4
Its longest line of descent runs back 9 generations to K930667 (1994). 43 ancestors in all, the oldest K811390 (1981). 224 later devices descend from it, 7 citing it directly. 1 of the 268 devices in this family, counting this one, has a Class I recall.
- Root · 9 generations backK930667 Reusable Laparoscopic InstrumentsSnowden-Pencer · 1994 · Recalled
- 5 generations between
- 3 generations backK050005 Intuitive Surgical Monopolar Curved Scissors, Model 400179; Tip…Intuitive Surgical, Inc. · 2005 · Recalled
- 2 generations backK050404 Intuitive Surgical DA Vinci Surgical System and Endoscopic InstrumentsIntuitive Surgical, Inc. · 2005
- 1 generation backK050802 Modification to Intuitive Surgical DA Vinci Surgical System and…Intuitive Surgical, Inc. · 2005
- This deviceK063220 DA Vinci S Surgical System-V1.1, Model IS2000Intuitive Surgical, Inc. · 2006
- Descendant · cites it directly · 211 of its ownK081137 Intuitive Surgical DA Vinci SI Surgical System: Model IS3000Intuitive Surgical, Inc. · 2009 · Recalled
- Descendant · cites it directly · 153 of its ownK090993 Intuitive Surgical Endoscopic Instrument Control Systems, Models…Intuitive Surgical, Inc. · 2009 · Recalled
- Descendant · 2 generations on · 148 of its ownK123463 Intuitive Surgical DA Vinci SI Surgical System SmartpedalsIntuitive Surgical, Inc. · 2012 · Recalled
- Descendant · 2 generations on · 124 of its ownK131861 DA Vinci Surgical System, Endowrist Instruments and AccessoriesIntuitive Surgical, Inc. · 2014 · Recalled
- Descendant · 2 generations on · 104 of its ownK101077 DA Vinci Fluorescence Imaging Vision System, Model FF 100Intuitive Surgical, Inc. · 2011 · Recalled
Show all 9 generations
- Root · 9 generations backK930667 Reusable Laparoscopic InstrumentsSnowden-Pencer · 1994 · Recalled
- 8 generations backK960400 Diamond-Touch and Micro Diamond-Touch Instruments/Diamond-Line…Snowden-Pencer · 1996
- 7 generations backK990144 Intuitive Surgical Endoscopic Instruments, Intuitive Surgical…Intuitive Surgical, Inc. · 2000 · Recalled
- 6 generations backK021036 Intuitive Surgical DA Vinci Surgical System, Model IS1000Intuitive Surgical, Inc. · 2002 · Recalled
- 5 generations backK040237 Intuitive Surgical DA Vinci Endoscopic Instrument Control System and…Intuitive Surgical, Inc. · 2004
- 4 generations backK043153 Intuitive Surgical DA Vinci Surgical System and Endoscopic…Intuitive Surgical, Inc. · 2004
- 3 generations backK050005 Intuitive Surgical Monopolar Curved Scissors, Model 400179; Tip…Intuitive Surgical, Inc. · 2005 · Recalled
- 2 generations backK050404 Intuitive Surgical DA Vinci Surgical System and Endoscopic InstrumentsIntuitive Surgical, Inc. · 2005
- 1 generation backK050802 Modification to Intuitive Surgical DA Vinci Surgical System and…Intuitive Surgical, Inc. · 2005
- This deviceK063220 DA Vinci S Surgical System-V1.1, Model IS2000Intuitive Surgical, Inc. · 2006
Showing 5 of 224 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K223095 Senhance Surgical System3 generations on · 2023
Most influential descendants
Showing 5 of 224 · See all- K081137Intuitive Surgical DA Vinci SI Surgical System: Model IS3000Intuitive Surgical, Inc. · 2009 · cites it directly211descendants
- K090993Intuitive Surgical Endoscopic Instrument Control Systems, Models IS1200, IS2000 and IS3000Intuitive Surgical, Inc. · 2009 · cites it directly153descendants
- K123463Intuitive Surgical DA Vinci SI Surgical System SmartpedalsIntuitive Surgical, Inc. · 2012 · 2 generations on148descendants
- K131861DA Vinci Surgical System, Endowrist Instruments and AccessoriesIntuitive Surgical, Inc. · 2014 · 2 generations on124descendants
- K101077DA Vinci Fluorescence Imaging Vision System, Model FF 100Intuitive Surgical, Inc. · 2011 · 2 generations on104descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −9 | K930667 | Reusable Laparoscopic Instruments | Snowden-Pencer | 1994 | 15 |
| −8 | K960400 | Diamond-Touch and Micro Diamond-Touch Instruments/Diamond-Line… | Snowden-Pencer | 1996 | — |
| −7 | K990144 | Intuitive Surgical Endoscopic Instruments, Intuitive Surgical Endoscopic… | Intuitive Surgical, Inc. | 2000 | 3 |
| −6 | K021036 | Intuitive Surgical DA Vinci Surgical System, Model IS1000 | Intuitive Surgical, Inc. | 2002 | 18 |
| −5 | K040237 | Intuitive Surgical DA Vinci Endoscopic Instrument Control System and… | Intuitive Surgical, Inc. | 2004 | — |
| −4 | K043153 | Intuitive Surgical DA Vinci Surgical System and Endoscopic Instruments, Models… | Intuitive Surgical, Inc. | 2004 | — |
| −3 | K050005 | Intuitive Surgical Monopolar Curved Scissors, Model 400179; Tip Cover… | Intuitive Surgical, Inc. | 2005 | 2 |
| −2 | K050404 | Intuitive Surgical DA Vinci Surgical System and Endoscopic Instruments | Intuitive Surgical, Inc. | 2005 | — |
| −1 | K050802 | Modification to Intuitive Surgical DA Vinci Surgical System and Endoscopic… | Intuitive Surgical, Inc. | 2005 | — |
| 0 (this device) | K063220 | DA Vinci S Surgical System-V1.1, Model IS2000 | Intuitive Surgical, Inc. | 2006 | — |
| +1 | K081137 | Intuitive Surgical DA Vinci SI Surgical System: Model IS3000 | Intuitive Surgical, Inc. | 2009 | 25 |
| +1 | K090993 | Intuitive Surgical Endoscopic Instrument Control Systems, Models IS1200… | Intuitive Surgical, Inc. | 2009 | 3 |
| +2 | K123463 | Intuitive Surgical DA Vinci SI Surgical System Smartpedals | Intuitive Surgical, Inc. | 2012 | 2 |
| +2 | K131861 | DA Vinci Surgical System, Endowrist Instruments and Accessories | Intuitive Surgical, Inc. | 2014 | 80 |
| +2 | K101077 | DA Vinci Fluorescence Imaging Vision System, Model FF 100 | Intuitive Surgical, Inc. | 2011 | 2 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1981 | 1 | 0 |
| 1988 | 1 | 0 |
| 1991 | 2 | 0 |
| 1992 | 1 | 0 |
| 1993 | 2 | 0 |
| 1994 | 2 | 0 |
| 1995 | 2 | 0 |
| 1997 | 6 | 0 |
| 1998 | 2 | 0 |
| 1999 | 2 | 0 |
| 2001 | 5 | 0 |
| 2002 | 3 | 0 |
| 2004 | 3 | 0 |
| 2005 | 2 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2008 | 4 | 0 |
| 2009 | 2 | 0 |
| 2011 | 4 | 0 |
| 2012 | 4 | 0 |
| 2013 | 4 | 0 |
| 2014 | 9 | 0 |
| 2015 | 4 | 0 |
| 2016 | 9 | 0 |
| 2017 | 14 | 0 |
| 2018 | 16 | 0 |
| 2019 | 9 | 0 |
| 2020 | 17 | 0 |
| 2021 | 16 | 0 |
| 2022 | 18 | 0 |
| 2023 | 19 | 1 |
| 2024 | 22 | 0 |
| 2025 | 31 | 0 |
| 2026 | 22 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new NAY device
- #1 K232610 da Vinci Surgical System Model IS5000, da…score 32.9 · 24 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K050369 | Intuitive Surgical DA Vinci Surgical System, Model IS2000Summary page 1…Predicate Intuitive Surgical da Vinci S Surgical System (Model IS2000) and Devices EndoWrist Endoscopic Instruments (legally marketed under K050369, K02 1036, K050404 and K050802). Device The da Vinci S Surgical System, Model IS2000 VI.l… | 2005 |
| * | K021036 | Intuitive Surgical DA Vinci Surgical System, Model IS1000Summary page 1…ve Surgical da Vinci S Surgical System (Model IS2000) and Devices EndoWrist Endoscopic Instruments (legally marketed under K050369, K02 1036, K050404 and K050802). Device The da Vinci S Surgical System, Model IS2000 VI.l, consists of three Description… | 2002 |
| * | K050404 | Intuitive Surgical DA Vinci Surgical System and Endoscopic InstrumentsSummary page 1…l da Vinci S Surgical System (Model IS2000) and Devices EndoWrist Endoscopic Instruments (legally marketed under K050369, K02 1036, K050404 and K050802). Device The da Vinci S Surgical System, Model IS2000 VI.l, consists of three Description… | 2005 |
| * | K050802 | Modification to Intuitive Surgical DA Vinci Surgical System and Endoscopic InstrumentsSummary page 1…Surgical System (Model IS2000) and Devices EndoWrist Endoscopic Instruments (legally marketed under K050369, K02 1036, K050404 and K050802). Device The da Vinci S Surgical System, Model IS2000 VI.l, consists of three Description integrated… | 2005 |
What descends from it? (224)
See all 2247 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K081137 | Intuitive Surgical DA Vinci SI Surgical System: Model IS3000 | 2009 | 211 |
| K090993 | Intuitive Surgical Endoscopic Instrument Control Systems, Models IS1200, IS2000 and IS3000 | 2009 | 153 |
| K081207 | Intuitive Surgical DA Vinci S Surgical System, Model IS2000, with DA Vinci Connect and… | 2008 | 7 |
| K073112 | DA Vinci Sonicpro Cleaning System, Model 400253 | 2008 | 5 |
| K123329 | DA Vinvi Surgical System | 2014 | 3 |
| K070947 | Intuitive Surgical DA Vinci Surgical System and Endoscopic Instruments and Endowrist… | 2008 | 0 |
| K072627 | Intuitive Surgical DA Vinci and DA Vinci S Surgical System and Endoscopic Instruments… | 2008 | 0 |
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 6, 2012 | Not classified | Z-1202-2012 | Device Design | The holding brake, a component of the da Vinci S Surgical System, may allow passive uncontrolled motion due to gravity during specific power-off conditions. |
| Oct 4, 2007 | Not classified | Z-0152-2008 | Mixed-up of materials/components | Under-Rated Fuses -- The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this product with the… |
| Oct 4, 2007 | Not classified | Z-0151-2008 | Mixed-up of materials/components | Under-Rated Fuses--The Firm may have installed incorrect fuses. The daVinci Surgical System Vision Cart's isolation transformer requires a 10A fuse for 115V operation or a 5A fuse for 230V operation. Operation of this product with the use… |
| May 14, 2007 | Not classified | Z-1245-2007 | Process control | Product may malfunction and fail to start up on AC power. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K223095Senhance Surgical SystemAsensus Surgical, Inc. · 2023 · 3 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 12 years and passes through 2 companies; a new generation was cleared every 1.4 years on median.
- 6 of the 43 ancestors have a recall of any class.
- Of its descendants, K081137 (Intuitive Surgical, Inc., 2009) has gone furthest, with 211 descendants of its own.
- 19 other devices also cite its predicate K050369.
How does its category compare?
All NAY devicesNAY System, Surgical, Computer Controlled Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · NAY
16in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · NAY
117in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · NAY
1in 2025
Class I (2 since 2007) Class II/III not classified
Adverse event reports per year · NAY
48,456in 2025
MAUDE reports naming product code NAY, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 11 | 133 |
| 2008 | 4 | 189 | 3 | 224 |
| 2009 | 3 | 253 | 1 | 380 |
| 2010 | 1 | — | 0 | 623 |
| 2011 | 7 | 191 | 2 | 501 |
| 2012 | 4 | 106 | 1 | 1,552 |
| 2013 | 4 | 144 | 6 | 6,217 |
| 2014 | 9 | 143 | 25 | 6,135 |
| 2015 | 3 | 99 | 3 | 1,722 |
| 2016 | 8 | 197 | 1 | 1,077 |
| 2017 | 16 | 107 | 1 | 1,049 |
| 2018 | 15 | 152 | 2 | 1,074 |
| 2019 | 8 | 73 | 5 | 1,154 |
| 2020 | 9 | 212 | 0 | 1,558 |
| 2021 | 11 | 61 | 3 | 1,997 |
| 2022 | 6 | 57 | 3 | 5,727 |
| 2023 | 5 | 141 | 2 (1) | 10,898 |
| 2024 | 10 | 126 | 40 | 12,872 |
| 2025 | 16 | 117 | 1 | 48,456 |
| 2026 | 7 | 89 | 4 (1) | — |
What is it indicated for?
The Intuitive Surgical Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection…
Read the full statement (142 words)
The Intuitive Surgical Endoscopic Instrument Control System is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave ablation probes and accessories during urologic surgical procedures, general laparoscopic surgical procedures, gynecological laparoscopic surgical procedures, general non-cardiovascular thoracoscopic surgical procedures, and thoracoscopically assisted cardiotomy procedures. The system is indicated for adult and pediatric use. The system can also be employed, with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. It is intended for use by trained physicians in an operating room environment in accordance with the representative specific procedures set forth in the Professional Instructions for Use.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K063220 · printed Nov 6, 2026
Cite this page
510(k) Report. “K063220: DA Vinci S Surgical System-V1.1, Model IS2000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K063220
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026