510(k) · Cleared
Recalled · Class I · 2024Route 92 Medical Full Length 070 Reperfusion System and Aspiration Tubing Set
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Route 92 Medical, Inc. · NRY Catheter, Thrombus Retriever
Cleared on Apr 11, 2023 as substantially equivalent to K161064 (Penumbra, Inc., 2016) after a 139-day review. Its family tree reaches back 9 generations to 2001. 1 of the 13 devices in its family, counting itself, has a Class I recall. It has been recalled once, including a Class I recall.
- Decided
- Apr 11, 2023
- Review
- 139 days
- Class
- Class II
- Product code
- NRY
- Recalls
- 1 (1 Class I)
- Predicates
- 1
Its longest line of descent runs back 9 generations to K003410 (2001). 12 ancestors in all, the oldest K003086 (2001). No later device cites it as a predicate yet. 1 of the 13 devices in this family, counting this one, has a Class I recall.
- Root · 9 generations backK003410 Concentric Retriever, Model 90030Concentric Medical, Inc. · 2001 · Recalled
- 5 generations between
- 3 generations backK142458 Penumbra System ACEPenumbra, Inc. · 2015
- 2 generations backK152541 Penumbra System ACE 64 and ACE 68 Reperfusion CathetersPenumbra, Inc. · 2016
- 1 generation backK161064 Penumbra System ACE 68 Reperfusion CatheterPenumbra, Inc. · 2016
- This deviceK223530 Route 92 Medical Full Length 070 Reperfusion System and Aspiration…Route 92 Medical, Inc. · 2023
Show all 9 generations
- Root · 9 generations backK003410 Concentric Retriever, Model 90030Concentric Medical, Inc. · 2001 · Recalled
- 8 generations backK030476 Modified Concentric Retriever, Model 90037Concentric Medical, Inc. · 2003 · Recalled
- 7 generations backK033736 Merci Retriever, Models 90065, 90066Concentric Medical, Inc. · 2004
- 6 generations backK072718 Penumbra SystemPenumbra, Inc. · 2007 · Recalled
- 5 generations backK100769 Penumbra System Separator Flex (026,032,041,054), Models PSF026…Penumbra, Inc. · 2010
- 4 generations backK113163 Penumbra System MaxPenumbra, Inc. · 2011
- 3 generations backK142458 Penumbra System ACEPenumbra, Inc. · 2015
- 2 generations backK152541 Penumbra System ACE 64 and ACE 68 Reperfusion CathetersPenumbra, Inc. · 2016
- 1 generation backK161064 Penumbra System ACE 68 Reperfusion CatheterPenumbra, Inc. · 2016
- This deviceK223530 Route 92 Medical Full Length 070 Reperfusion System and Aspiration…Route 92 Medical, Inc. · 2023
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −9 | K003410 | Concentric Retriever, Model 90030 | Concentric Medical, Inc. | 2001 | 1 |
| −8 | K030476 | Modified Concentric Retriever, Model 90037 | Concentric Medical, Inc. | 2003 | 2 |
| −7 | K033736 | Merci Retriever, Models 90065, 90066 | Concentric Medical, Inc. | 2004 | — |
| −6 | K072718 | Penumbra System | Penumbra, Inc. | 2007 | 1 |
| −5 | K100769 | Penumbra System Separator Flex (026,032,041,054), Models PSF026, PSF032… | Penumbra, Inc. | 2010 | — |
| −4 | K113163 | Penumbra System Max | Penumbra, Inc. | 2011 | — |
| −3 | K142458 | Penumbra System ACE | Penumbra, Inc. | 2015 | — |
| −2 | K152541 | Penumbra System ACE 64 and ACE 68 Reperfusion Catheters | Penumbra, Inc. | 2016 | — |
| −1 | K161064 | Penumbra System ACE 68 Reperfusion Catheter | Penumbra, Inc. | 2016 | — |
| 0 (this device) | K223530 | Route 92 Medical Full Length 070 Reperfusion System and Aspiration Tubing Set | Route 92 Medical, Inc. | 2023 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2001 | 1 | 0 |
| 2009 | 1 | 0 |
| 2014 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new NRY device
- #1 K232971 APRO 55 Catheter and Alembic Aspiration Tubingscore 18.6 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Mar 8, 2024 | Class I | Z-1611-2024 | Process control | Their is the potential for distal tip separation at the proximal marker band of the delivery catheters |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 22 years and passes through 3 companies; a new generation was cleared every 1.8 years on median.
- 3 of the 12 ancestors have a recall of any class.
- 4 other devices also cite its predicate K161064.
How does its category compare?
All NRY devicesNRY Catheter, Thrombus Retriever: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.
FDA decisions per year · NRY
15in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · NRY
129in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · NRY
4in 2020
Class I (4 since 2007) Class II/III not classified
Adverse event reports per year · NRY
565in 2025
MAUDE reports naming product code NRY, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 5 | 94 | 0 | 13 |
| 2008 | 2 | — | 0 | 18 |
| 2009 | 2 | — | 0 | 57 |
| 2010 | 1 | — | 1 | 289 |
| 2011 | 2 | — | 0 | 263 |
| 2012 | 4 | 90 | 0 | 366 |
| 2013 | 0 | — | 1 | 405 |
| 2014 | 4 | 140 | 0 | 428 |
| 2015 | 6 | 128 | 0 | 598 |
| 2016 | 5 | 75 | 0 | 497 |
| 2017 | 3 | 185 | 4 | 432 |
| 2018 | 11 | 156 | 0 | 415 |
| 2019 | 7 | 89 | 0 | 650 |
| 2020 | 6 | 125 | 4 (4) | 703 |
| 2021 | 9 | 80 | 0 | 688 |
| 2022 | 3 | 129 | 0 | 549 |
| 2023 | 8 | 112 | 0 | 711 |
| 2024 | 8 | 165 | 0 | 641 |
| 2025 | 15 | 129 | 0 | 565 |
| 2026 | 14 | 147 | 2 | — |
What is it indicated for?
The Route 92 Medical Full Length 070 Reperfusion System and Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral – M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K223530 · printed Nov 6, 2026
Cite this page
510(k) Report. “K223530: Route 92 Medical Full Length 070 Reperfusion System and Aspiration Tubing Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K223530
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026