Sensis Vibe (VD15)

K223812510(k) TraditionalCleared Sep 15, 2023269-day reviewClass II21 CFR 870.1425Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · DQK Computer, Diagnostic, Programmable

Cleared on Sep 15, 2023 as substantially equivalent to K150493 (Siemens Medi Cal Solutions, Inc., 2015) and 1 other predicate after a 269-day review. Its family tree reaches back 8 generations to 2001. It has been recalled 3 times.

Decided
Sep 15, 2023
Review
269 days
Class
Class II
Product code
DQK
Recalls
3
Predicates
2

Its longest line of descent runs back 8 generations to K003122 (2001). 43 ancestors in all, the oldest K862740 (1986). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2015: 1 other ancestor2013: 3 other ancestors2000: 3 other ancestors2002: 2 other ancestors2001: 2 other ancestors1998: 5 other ancestors1994: 1 other ancestor2007: 1 other ancestor1997: 2 other ancestors1995: 4 other ancestors1993: 3 other ancestors1990: 2 other ancestors2005: 1 other ancestor1991: 2 other ancestors1986: 1 other ancestor1992: 2 other ancestorsReference device: K051217 · Datex-Ohmeda Prestin Module (Model Family E-Prestin, Including E-Prestin, E-Restin, E=Pretin, E-PP and E-PT/E-P) · GE Healthcare · 2006×3: 3 generations cleared close together, K003122 (2001), K012947 (2001), K020127 (2002) K022188 · Radianalyzer, Model 12710 · Radi Medical Systems AB · 2002K042628 · Modification to: Radianalyzer, Model 12711 · Radi Medical Systems AB · 2005K092105 · Radianalyzer Xpress · Radi Medical Systems AB · 2009K151613 · iLab Polaris Multi-Modality Guidance System · Boston Scientific Corporation · 2015K191008 · iLab Polaris Multi-Modality Guidance System · Boston Scientific Corporation · 2019 · recalled1990200020102020
  1. Root · 8 generations backK003122 Smartflow, Model Sfip 2000Florence Medical, Ltd. · 2001
  2. 7 generations backK012947 SmartflowFlorence Medical, Ltd. · 2001
  3. 6 generations backK020127 Smartflow Pulse Transmission CoefficientFlorence Medical, Ltd. · 2002
  4. 5 generations backK022188 Radianalyzer, Model 12710Radi Medical Systems AB · 2002
  5. 4 generations backK042628 Modification to: Radianalyzer, Model 12711Radi Medical Systems AB · 2005
  6. 3 generations backK092105 Radianalyzer XpressRadi Medical Systems AB · 2009
  7. 2 generations backK151613 iLab Polaris Multi-Modality Guidance SystemBoston Scientific Corporation · 2015
  8. 1 generation backK191008 iLab Polaris Multi-Modality Guidance SystemBoston Scientific Corporation · 2019 · Recalled
  9. This deviceK223812 Sensis Vibe (VD15)Siemens Medical Solutions USA, Inc. · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K003122Smartflow, Model Sfip 2000Florence Medical, Ltd.2001—
−7K012947SmartflowFlorence Medical, Ltd.2001—
−6K020127Smartflow Pulse Transmission CoefficientFlorence Medical, Ltd.2002—
−5K022188Radianalyzer, Model 12710Radi Medical Systems AB2002—
−4K042628Modification to: Radianalyzer, Model 12711Radi Medical Systems AB2005—
−3K092105Radianalyzer XpressRadi Medical Systems AB2009—
−2K151613iLab Polaris Multi-Modality Guidance SystemBoston Scientific Corporation2015—
−1K191008iLab Polaris Multi-Modality Guidance SystemBoston Scientific Corporation20191
0 (this device)K223812Sensis Vibe (VD15)Siemens Medical Solutions USA, Inc.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199020
199120
199220
199330
199410
199540
199720
199850
200030
200120
200220
200510
200710
201330
201510
FIG. 1 — Lineage of K223812, 2001–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQK device

  1. #1 K213682 HemoSphere Advanced Monitor, HemoSphere…score 26.6 · 28 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K150493Sensis
Summary page 4
…CCK 4. Legally Marketed Primary Predicate Device Trade Name: Sensis Vibe (VC12) 510(k) Clearance K150493 Clearance Date June 30, 2015 Classification Name: Programmable diagnostic computer Classifi…
Siemens Medi Cal Solutions, Inc.201511
*K191008iLab Polaris Multi-Modality Guidance System
Summary page 5
…by any of the applicable issues. Predicate Device Trade Name: iLab Polaris-Modality Guidance System 510(k) Clearance K191008 Clearance Date July 02, 2019 Classification Name: Programmable diagnostic computer C…
Boston Scientific Corporation20191
†K051217Datex-Ohmeda Prestin Module (Model Family E-Prestin, Including E-Prestin, E-Restin…GE Healthcare2006—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 3, 2024Class IIZ-2438-2024Software designIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
Jul 3, 2024Class IIZ-2441-2024Software designIf the Sensis documentation functionality is used during adding of once-per-study reporting events (e.g., Type 1 events, as further defined in the administrator manual), the possibility exists that the application could crash.
Nov 29, 2023Class IIZ-0683-2024Software designThe possibility of the Sensis documentation functionality application to crash.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 3.7 years on median.
  • 7 of the 43 ancestors have a recall of any class.

How does its category compare?

All DQK devices

DQK Computer, Diagnostic, Programmable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · DQK

23so far in 2026, the most since 2007

2007: 132008: 42009: 122010: 92011: 122012: 72013: 132014: 92015: 72016: 152017: 72018: 142019: 152020: 162021: 162022: 192023: 212024: 132025: 222026 (partial year): 23
2007high 23 (2026) · low 4 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQK

95in 2025

2007: 83 days median2008: 70 days median2009: 93 days median2010: 64 days median2011: 69 days median2012: 74 days median2013: 68 days median2014: 90 days median2015: 125 days median2016: 190 days median2017: 146 days median2018: 150 days median2019: 113 days median2020: 123 days median2021: 132 days median2022: 136 days median2023: 133 days median2024: 100 days median2025: 95 days median2026 (partial year): 151 days median
2007high 190 (2016) · low 64 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQK

1in 2025

2007: 42008: 12009: 22010: 32011: 22012: 32013: 12014: 42015: 22017: 22018: 62019: 12020: 12021: 12022: 22023: 22024: 242025: 12026 (partial year): 2
2007high 24 (2024) · low 0 (2016)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQK

433in 2025

2007: 182008: 342009: 402010: 492011: 552012: 1292013: 1802014: 1152015: 1312016: 3402017: 2522018: 2612019: 2182020: 2152021: 3512022: 3192023: 3402024: 5572025: 433
2007high 557 (2024) · low 18 (2007)2025

MAUDE reports naming product code DQK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071383418
2008470134
20091293240
2010964349
20111269255
20127743129
201313681180
20149904115
201571252131
2016151900340
201771462252
2018141506261
2019151131218
2020161231215
2021161321351
2022191362319
2023211332340
20241310024557
202522951433
2026231512—

What is it indicated for?

The Sensis Vibe Hemo (VD15B) and Sensis Vibe Combo (VD15B) recording systems are intended to be used as a diagnostic and administrative tool supporting hemodynamic cardiac catheterizations and/or electrophysiology studies, for cardiac as well as interventional radiology and surgical studies. The system is equipped with modules, enabling various configurations ranging from a standalone acquisition unit with limited administrative functionality to…

Read the full statement (131 words)

The Sensis Vibe Hemo (VD15B) and Sensis Vibe Combo (VD15B) recording systems are intended to be used as a diagnostic and administrative tool supporting hemodynamic cardiac catheterizations and/or electrophysiology studies, for cardiac as well as interventional radiology and surgical studies. The system is equipped with modules, enabling various configurations ranging from a standalone acquisition unit with limited administrative functionality to multiunit installations with a common database and satellite workstations accessing the data using the administrative tools.

The device is intended to be used on either or both of the following populations:

1. Adult and pediatric populations requiring electrophysiology examinations, typically when the patient is suffering from cardiac arrhythmias.

2. Adult and pediatric populations requiring hemodynamic examinations, typically when the patient has a heart or vascular disease resulting in insufficient hemodynamic functionality.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K223812: Sensis Vibe (VD15).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K223812

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026