Siemens Medi Cal Solutions, Inc. · DQK Computer, Diagnostic, Programmable
Cleared on Jun 30, 2015 as substantially equivalent to K131812 (Siemens Medi Cal Solutions, Inc., 2013) and 1 other predicate after a 125-day review. Its family tree reaches back 6 generations to 1991. 1 later device descends from it. It has been recalled 11 times.
- Decided
- Jun 30, 2015
- Review
- 125 days
- Class
- Class II
- Product code
- DQK
- Recalls
- 11
- Predicates
- 2
Its longest line of descent runs back 6 generations to K910772 (1991). 23 ancestors in all, the oldest K862740 (1986). 1 later device descends from it, 1 citing it directly.
- Root · 6 generations backK910772 Propaq 102, 104 and 106Protocol Systems, Inc. · 1991
- 5 generations backK972020 Oscillomate Non-Invasive Blood Pressure MonitorCas Medical Systems, Inc. · 1998
- 4 generations backK980879 Oscillomate Nibp / SP02 Ems MonitorCas Medical Systems, Inc. · 1998
- 3 generations backK982776 9303 Neonatal/Adult Vital Signs MonitorCas Medical Systems, Inc. · 1998
- 2 generations backK020440 NewcorSiemens Medical Systems, Inc. · 2002 · Recalled
- 1 generation backK131812 Sensis, Sensis LiteSiemens Medi Cal Solutions, Inc. · 2013
- This deviceK150493 SensisSiemens Medi Cal Solutions, Inc. · 2015
- Descendant · cites it directly · 0 of its ownK223812 Sensis Vibe (VD15)Siemens Medical Solutions USA, Inc. · 2023 · Recalled
Most influential descendants
Showing 1 of 1 · See all- K223812Sensis Vibe (VD15)Siemens Medical Solutions USA, Inc. · 2023 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K910772 | Propaq 102, 104 and 106 | Protocol Systems, Inc. | 1991 | — |
| −5 | K972020 | Oscillomate Non-Invasive Blood Pressure Monitor | Cas Medical Systems, Inc. | 1998 | — |
| −4 | K980879 | Oscillomate Nibp / SP02 Ems Monitor | Cas Medical Systems, Inc. | 1998 | — |
| −3 | K982776 | 9303 Neonatal/Adult Vital Signs Monitor | Cas Medical Systems, Inc. | 1998 | — |
| −2 | K020440 | Newcor | Siemens Medical Systems, Inc. | 2002 | 10 |
| −1 | K131812 | Sensis, Sensis Lite | Siemens Medi Cal Solutions, Inc. | 2013 | — |
| 0 (this device) | K150493 | Sensis | Siemens Medi Cal Solutions, Inc. | 2015 | — |
| +1 | K223812 | Sensis Vibe (VD15) | Siemens Medical Solutions USA, Inc. | 2023 | 3 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1986 | 1 | 0 |
| 1990 | 2 | 0 |
| 1991 | 1 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1994 | 1 | 0 |
| 1995 | 4 | 0 |
| 1997 | 2 | 0 |
| 1998 | 1 | 0 |
| 2000 | 3 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2023 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DQK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DQK device
- #1 K213682 HemoSphere Advanced Monitor, HemoSphere…score 26.7 · 28 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K131812 | Sensis, Sensis LiteSummary page 5…Product Code: DQK 4. Legally Marketed Primary Predicate Device Trade Name: AXIOM Sensis 510(k) #: K131812 Clearance Date: July 25, 2013 Classification Name: Programmable Diagnostic Computer Classif… | 2013 |
| * | K002137 | Cathcor DesktopSummary page 5…Recall Information: N.A. Legally Marketed Secondary Predicate Device Trade Name: Cathcor Desktop 510(k) #: K002137 Clearance Date: October 5, 2000 Classification Name: Programmable Diagnostic Computer Classif… | 2000 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K223812 | Sensis Vibe (VD15) | 2023 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 16, 2022 | Class II | Z-0775-2023 | Software design | The firm will be performing a software update to address a software error which affects the listed products. This correction addresses four potential software issues: 1) "PASSWORD STORE CORRUPTED" error message during system boot; 2)… |
| Jul 12, 2022 | Class III | Z-1558-2022 | Device Design | The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting position deviates… |
| Jun 15, 2022 | Class II | Z-1426-2022 | Device Design | Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs… |
| Jun 15, 2022 | Class II | Z-1425-2022 | Device Design | Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs… |
| Apr 13, 2022 | Class II | Z-1243-2022 | Software Design Change | Siemens has become aware of three potential software issues with AXIOM Sensis or Sensis/ Sensis Lite systems. This may lead to a hazardous situation for patients if treatment cannot be continued on the system and treatment needs to be… |
| Nov 23, 2021 | Class II | Z-0499-2022 | Software design | Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic… |
| Oct 5, 2021 | Class II | Z-0226-2022 | Software design | System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances, this may result in a… |
| Mar 4, 2021 | Class II | Z-1434-2021 | Software design | Due to the configuration of certain Windows Service Permissions within the operating systems of the Sensis/ Sensis Vibe computer, there is a risk for exposure of sensitive information, manipulation of data, or Denial of Service attacks… |
| Jan 27, 2021 | Class II | Z-1173-2021 | Software design | System may sporadically freeze (lock-up) during operation or while being in an idle state, no longer possible to interact with the system |
| Jul 24, 2020 | Class II | Z-2921-2020 | Process design | The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning… |
| Nov 29, 2018 | Class II | Z-0936-2019 | Software design | A software error may result in a system crash. The system must be restarted before the clinical procedure can be continued. The ablation treatment must be performed with a different system or without the use of the interface. |
The 5 latest of 11 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 24 years and passes through 4 companies; a new generation was cleared every 4.1 years on median.
- 3 of the 23 ancestors have a recall of any class.
- Of its descendants, K223812 (Siemens Medical Solutions USA, Inc., 2023) has gone furthest, with 0 descendants of its own.
How does its category compare?
All DQK devicesDQK Computer, Diagnostic, Programmable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.
FDA decisions per year · DQK
23so far in 2026, the most since 2007
2026 to Oct 25, outlined.
Median 510(k) review, days · DQK
95in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DQK
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DQK
433in 2025
MAUDE reports naming product code DQK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 13 | 83 | 4 | 18 |
| 2008 | 4 | 70 | 1 | 34 |
| 2009 | 12 | 93 | 2 | 40 |
| 2010 | 9 | 64 | 3 | 49 |
| 2011 | 12 | 69 | 2 | 55 |
| 2012 | 7 | 74 | 3 | 129 |
| 2013 | 13 | 68 | 1 | 180 |
| 2014 | 9 | 90 | 4 | 115 |
| 2015 | 7 | 125 | 2 | 131 |
| 2016 | 15 | 190 | 0 | 340 |
| 2017 | 7 | 146 | 2 | 252 |
| 2018 | 14 | 150 | 6 | 261 |
| 2019 | 15 | 113 | 1 | 218 |
| 2020 | 16 | 123 | 1 | 215 |
| 2021 | 16 | 132 | 1 | 351 |
| 2022 | 19 | 136 | 2 | 319 |
| 2023 | 21 | 133 | 2 | 340 |
| 2024 | 13 | 100 | 24 | 557 |
| 2025 | 22 | 95 | 1 | 433 |
| 2026 | 23 | 151 | 2 | — |
What is it indicated for?
The Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology as well as surgical studies. The system is equipped by modules, enabling various configurations ranging from a stand-alone acquisition unit with limited administrative functionality to multiunit installations with a common database and satellite…
Read the full statement (124 words)
The Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology as well as surgical studies. The system is equipped by modules, enabling various configurations ranging from a stand-alone acquisition unit with limited administrative functionality to multiunit installations with a common database and satellite workstations accessing the data using the administrative tools.
The device is intended to be used on either or both of the following populations:
1. Adult and pediatric populations requiring electrophysiology examinations, typically when the patient is suffering from cardiac arrhythmias.
2. Adult and pediatric populations requiring hemodynamic examinations, typically when the patient has a heart or vascular disease resulting in insufficient hemodynamic functionality.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K150493 · printed Nov 4, 2026
Cite this page
510(k) Report. “K150493: Sensis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K150493