510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Sensis

K150493510(k) TraditionalCleared Jun 30, 2015125-day reviewClass II21 CFR 870.1425Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medi Cal Solutions, Inc. · DQK Computer, Diagnostic, Programmable

Cleared on Jun 30, 2015 as substantially equivalent to K131812 (Siemens Medi Cal Solutions, Inc., 2013) and 1 other predicate after a 125-day review. Its family tree reaches back 6 generations to 1991. 1 later device descends from it. It has been recalled 11 times.

Decided
Jun 30, 2015
Review
125 days
Class
Class II
Product code
DQK
Recalls
11
Predicates
2

Its longest line of descent runs back 6 generations to K910772 (1991). 23 ancestors in all, the oldest K862740 (1986). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 3 other ancestors1998: 1 other ancestor1994: 1 other ancestor1997: 2 other ancestors1995: 4 other ancestors1993: 1 other ancestor1990: 2 other ancestors1991: 1 other ancestor1986: 1 other ancestor1992: 1 other ancestor2023: 1 descendant clearedK910772 · Propaq 102, 104 and 106 · Protocol Systems, Inc. · 1991×3: 3 generations cleared close together, K972020 (1998), K980879 (1998), K982776 (1998) K020440 · Newcor · Siemens Medical Systems, Inc. · 2002 · recalledK131812 · Sensis, Sensis Lite · Siemens Medi Cal Solutions, Inc. · 20131990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor3 generations back: 4 other ancestors4 generations back: 8 other ancestors5 generations back: 3 other ancestors6 generations back: 1 other ancestor 1 generation forward: 1 descendant K131812 · Sensis, Sensis Lite · Siemens Medi Cal Solutions, Inc. · 2013K020440 · Newcor · Siemens Medical Systems, Inc. · 2002 · recalledK982776 · 9303 Neonatal/Adult Vital Signs Monitor · Cas Medical Systems, Inc. · 1998K980879 · Oscillomate Nibp / SP02 Ems Monitor · Cas Medical Systems, Inc. · 1998K972020 · Oscillomate Non-Invasive Blood Pressure Monitor · Cas Medical Systems, Inc. · 1998K910772 · Propaq 102, 104 and 106 · Protocol Systems, Inc. · 1991−50
  1. Root · 6 generations backK910772 Propaq 102, 104 and 106Protocol Systems, Inc. · 1991
  2. 5 generations backK972020 Oscillomate Non-Invasive Blood Pressure MonitorCas Medical Systems, Inc. · 1998
  3. 4 generations backK980879 Oscillomate Nibp / SP02 Ems MonitorCas Medical Systems, Inc. · 1998
  4. 3 generations backK982776 9303 Neonatal/Adult Vital Signs MonitorCas Medical Systems, Inc. · 1998
  5. 2 generations backK020440 NewcorSiemens Medical Systems, Inc. · 2002 · Recalled
  6. 1 generation backK131812 Sensis, Sensis LiteSiemens Medi Cal Solutions, Inc. · 2013
  7. This deviceK150493 SensisSiemens Medi Cal Solutions, Inc. · 2015
  8. Descendant · cites it directly · 0 of its ownK223812 Sensis Vibe (VD15)Siemens Medical Solutions USA, Inc. · 2023 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K223812Sensis Vibe (VD15)Siemens Medical Solutions USA, Inc. · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K910772Propaq 102, 104 and 106Protocol Systems, Inc.1991—
−5K972020Oscillomate Non-Invasive Blood Pressure MonitorCas Medical Systems, Inc.1998—
−4K980879Oscillomate Nibp / SP02 Ems MonitorCas Medical Systems, Inc.1998—
−3K9827769303 Neonatal/Adult Vital Signs MonitorCas Medical Systems, Inc.1998—
−2K020440NewcorSiemens Medical Systems, Inc.200210
−1K131812Sensis, Sensis LiteSiemens Medi Cal Solutions, Inc.2013—
0 (this device)K150493SensisSiemens Medi Cal Solutions, Inc.2015—
+1K223812Sensis Vibe (VD15)Siemens Medical Solutions USA, Inc.20233

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199020
199110
199210
199310
199410
199540
199720
199810
200030

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202310
FIG. 1 — Lineage of K150493, 1991–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQK device

  1. #1 K213682 HemoSphere Advanced Monitor, HemoSphere…score 26.7 · 28 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K131812Sensis, Sensis Lite
Summary page 5
…Product Code: DQK 4. Legally Marketed Primary Predicate Device Trade Name: AXIOM Sensis 510(k) #: K131812 Clearance Date: July 25, 2013 Classification Name: Programmable Diagnostic Computer Classif…
Siemens Medi Cal Solutions, Inc.2013—
*K002137Cathcor Desktop
Summary page 5
…Recall Information: N.A. Legally Marketed Secondary Predicate Device Trade Name: Cathcor Desktop 510(k) #: K002137 Clearance Date: October 5, 2000 Classification Name: Programmable Diagnostic Computer Classif…
Siemens Elema AB20001

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K223812Sensis Vibe (VD15)Siemens Medical Solutions USA, Inc.20230

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 16, 2022Class IIZ-0775-2023Software designThe firm will be performing a software update to address a software error which affects the listed products. This correction addresses four potential software issues: 1) "PASSWORD STORE CORRUPTED" error message during system boot; 2)…
Jul 12, 2022Class IIIZ-1558-2022Device DesignThe Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting position deviates…
Jun 15, 2022Class IIZ-1426-2022Device DesignSensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs…
Jun 15, 2022Class IIZ-1425-2022Device DesignSensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs…
Apr 13, 2022Class IIZ-1243-2022Software Design ChangeSiemens has become aware of three potential software issues with AXIOM Sensis or Sensis/ Sensis Lite systems. This may lead to a hazardous situation for patients if treatment cannot be continued on the system and treatment needs to be…
Nov 23, 2021Class IIZ-0499-2022Software designSoftware error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic…
Oct 5, 2021Class IIZ-0226-2022Software designSystem operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances, this may result in a…
Mar 4, 2021Class IIZ-1434-2021Software designDue to the configuration of certain Windows Service Permissions within the operating systems of the Sensis/ Sensis Vibe computer, there is a risk for exposure of sensitive information, manipulation of data, or Denial of Service attacks…
Jan 27, 2021Class IIZ-1173-2021Software designSystem may sporadically freeze (lock-up) during operation or while being in an idle state, no longer possible to interact with the system
Jul 24, 2020Class IIZ-2921-2020Process designThe connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning…
Nov 29, 2018Class IIZ-0936-2019Software designA software error may result in a system crash. The system must be restarted before the clinical procedure can be continued. The ablation treatment must be performed with a different system or without the use of the interface.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 4 companies; a new generation was cleared every 4.1 years on median.
  • 3 of the 23 ancestors have a recall of any class.
  • Of its descendants, K223812 (Siemens Medical Solutions USA, Inc., 2023) has gone furthest, with 0 descendants of its own.

How does its category compare?

All DQK devices

DQK Computer, Diagnostic, Programmable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · DQK

23so far in 2026, the most since 2007

2007: 132008: 42009: 122010: 92011: 122012: 72013: 132014: 92015: 72016: 152017: 72018: 142019: 152020: 162021: 162022: 192023: 212024: 132025: 222026 (partial year): 23
2007high 23 (2026) · low 4 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQK

95in 2025

2007: 83 days median2008: 70 days median2009: 93 days median2010: 64 days median2011: 69 days median2012: 74 days median2013: 68 days median2014: 90 days median2015: 125 days median2016: 190 days median2017: 146 days median2018: 150 days median2019: 113 days median2020: 123 days median2021: 132 days median2022: 136 days median2023: 133 days median2024: 100 days median2025: 95 days median2026 (partial year): 151 days median
2007high 190 (2016) · low 64 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQK

1in 2025

2007: 42008: 12009: 22010: 32011: 22012: 32013: 12014: 42015: 22017: 22018: 62019: 12020: 12021: 12022: 22023: 22024: 242025: 12026 (partial year): 2
2007high 24 (2024) · low 0 (2016)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQK

433in 2025

2007: 182008: 342009: 402010: 492011: 552012: 1292013: 1802014: 1152015: 1312016: 3402017: 2522018: 2612019: 2182020: 2152021: 3512022: 3192023: 3402024: 5572025: 433
2007high 557 (2024) · low 18 (2007)2025

MAUDE reports naming product code DQK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071383418
2008470134
20091293240
2010964349
20111269255
20127743129
201313681180
20149904115
201571252131
2016151900340
201771462252
2018141506261
2019151131218
2020161231215
2021161321351
2022191362319
2023211332340
20241310024557
202522951433
2026231512—

What is it indicated for?

The Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology as well as surgical studies. The system is equipped by modules, enabling various configurations ranging from a stand-alone acquisition unit with limited administrative functionality to multiunit installations with a common database and satellite…

Read the full statement (124 words)

The Sensis recording system is intended to be used as a diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology as well as surgical studies. The system is equipped by modules, enabling various configurations ranging from a stand-alone acquisition unit with limited administrative functionality to multiunit installations with a common database and satellite workstations accessing the data using the administrative tools.

The device is intended to be used on either or both of the following populations:

1. Adult and pediatric populations requiring electrophysiology examinations, typically when the patient is suffering from cardiac arrhythmias.

2. Adult and pediatric populations requiring hemodynamic examinations, typically when the patient has a heart or vascular disease resulting in insufficient hemodynamic functionality.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K150493: Sensis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K150493

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026