510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Exactech® Equinoxe® Reverse Humeral Liners and Humeral Adapter Trays

K223833510(k) TraditionalCleared Sep 14, 2023266-day reviewClass II21 CFR 888.3660Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Exactech, Inc. · PHX Shoulder Prosthesis, Reverse Configuration

Cleared on Sep 14, 2023 as substantially equivalent to K063569 (Exactech, Inc., 2007) and 5 other predicates after a 266-day review. Its family tree reaches back 17 generations to 1991. 1 of the 140 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled once.

Decided
Sep 14, 2023
Review
266 days
Class
Class II
Product code
PHX
Recalls
1
Predicates
6

Its longest line of descent runs back 17 generations to K911686 (1991). 137 ancestors in all, the oldest K790987 (1979). 2 later devices descend from it, 2 citing it directly. 1 of the 140 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2018: 2 other ancestors2010: 3 other ancestors2008: 5 other ancestors2017: 1 other ancestor2015: 1 other ancestor2013: 3 other ancestors2011: 4 other ancestors2006: 10 other ancestors2005: 7 other ancestors2014: 1 other ancestor2004: 7 other ancestors2003: 6 other ancestors2000: 9 other ancestors1999: 7 other ancestors1995: 4 other ancestors2012: 3 other ancestors2009: 1 other ancestor2007: 5 other ancestors2001: 9 other ancestors1998: 3 other ancestors1996: 8 other ancestors1980: 2 other ancestors1997: 2 other ancestors1994: 2 other ancestors1993: 1 other ancestor1992: 2 other ancestors1989: 1 other ancestor1987: 5 other ancestors1986: 1 other ancestor1979: 1 other ancestor1991: 2 other ancestors1988: 1 other ancestor2002: 1 other ancestor2026: 1 descendant cleared2025: 1 descendant clearedK911686 · Global Total Shoulder, Uncement Humeral Hemi-Shoul · Depuy, Inc. · 1991×2: 2 generations cleared close together, K984541 (1999), K992065 (1999) ×3: 3 generations cleared close together, K021478 (2003), K030941 (2004), K041873 (2004) ×2: 2 generations cleared close together, K052906 (2005), K060704 (2006) K063569 · Equinoxe Reverse Shoulder System · Exactech, Inc. · 2007 · recalled×2: 2 generations cleared close together, K073688 (2008), K080642 (2008) ×2: 2 generations cleared close together, K110599 (2011), K112615 (2011) K122698 · Aequalis Ascend Flex Shoulder System · Tornier S.A.S. · 2013 · recalledK163669 · Humeris Shoulder · Fx Solutions · 2017K192206 · 32mm Glenosphere and Humeral Cup · Fx Shoulder USA, Inc. · 2019 · recalledK193394 · Humeral Cup Stability & Humeral Cup 135/145o Stability · Fx Shoulder USA, Inc. · 202019801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 4 other ancestors2 generations back: 12 other ancestors3 generations back: 12 other ancestors4 generations back: 33 other ancestors5 generations back: 25 other ancestors6 generations back: 10 other ancestors7 generations back: 11 other ancestors8 generations back: 8 other ancestors9 generations back: 2 other ancestors10 generations back: 2 other ancestors11 generations back: 1 other ancestor 1 generation forward: 2 descendants K193394 · Humeral Cup Stability & Humeral Cup 135/145o Stability · Fx Shoulder USA, Inc. · 2020K192206 · 32mm Glenosphere and Humeral Cup · Fx Shoulder USA, Inc. · 2019 · recalledK163669 · Humeris Shoulder · Fx Solutions · 2017K122698 · Aequalis Ascend Flex Shoulder System · Tornier S.A.S. · 2013 · recalledK112615 · Aequalis Ascend Modular Reverse Shoulder System · Tornier, Inc. · 2011K110599 · Aequalis Ascend Modular Reverse Shoulder System · Tornier, Inc. · 2011K080642 · Comprehensive Reverse Shoulder · Biomet, Inc. · 2008 · Class I recallK073688 · Equinoxe Reverse Shoulder System Fracture Humeral Adapter Tray and Locking Screw · Exactech, Inc. · 2008 · recalledK063569 · Equinoxe Reverse Shoulder System · Exactech, Inc. · 2007 · recalledK060704 · Zimmer Trabecular Metal Reverse Shoulder System, Sizes 8MM and 10MM · Zimmer, Inc. · 2006 · recalledK052906 · Zimmer Trabecular Metal Reverse Shoulder System · Zimmer, Inc. · 2005 · recalledK041873 · Aequalis Reversed Shoulder Prosthesis · Tornier · 2004K030941 · Aequalis Reversed Shoulder Prosthesis · Tornier · 2004K021478 · Delta Shoulder · Depuy, Inc. · 2003K992065 · Global Advantage Shoulder, Global Advantage Humeral Stem, Global Advantage Eccentric Head · DePuy Orthopaedics, Inc. · 1999K984541 · Global FX Humeral Stem, Global Advantage Humeral Head · DePuy Orthopaedics, Inc. · 1999K911686 · Global Total Shoulder, Uncement Humeral Hemi-Shoul · Depuy, Inc. · 1991−15−10−50
  1. Root · 17 generations backK911686 Global Total Shoulder, Uncement Humeral Hemi-ShoulDepuy, Inc. · 1991
  2. 13 generations between
  3. 3 generations backK163669 Humeris ShoulderFx Solutions · 2017
  4. 2 generations backK192206 32mm Glenosphere and Humeral CupFx Shoulder USA, Inc. · 2019 · Recalled
  5. 1 generation backK193394 Humeral Cup Stability & Humeral Cup 135/145o StabilityFx Shoulder USA, Inc. · 2020
  6. This deviceK223833 Exactech® Equinoxe® Reverse Humeral Liners and Humeral Adapter TraysExactech, Inc. · 2023
  7. Descendant · cites it directly · 0 of its ownK260583 Equinoxe® Shoulder SystemExactech, Inc. · 2026
  8. Descendant · cites it directly · 0 of its ownK243448 Equinoxe® Shoulder SystemExactech, Inc. · 2025
Show all 17 generations
  1. Root · 17 generations backK911686 Global Total Shoulder, Uncement Humeral Hemi-ShoulDepuy, Inc. · 1991
  2. 16 generations backK984541 Global FX Humeral Stem, Global Advantage Humeral HeadDePuy Orthopaedics, Inc. · 1999
  3. 15 generations backK992065 Global Advantage Shoulder, Global Advantage Humeral Stem, Global…DePuy Orthopaedics, Inc. · 1999
  4. 14 generations backK021478 Delta ShoulderDepuy, Inc. · 2003
  5. 13 generations backK030941 Aequalis Reversed Shoulder ProsthesisTornier · 2004
  6. 12 generations backK041873 Aequalis Reversed Shoulder ProsthesisTornier · 2004
  7. 11 generations backK052906 Zimmer Trabecular Metal Reverse Shoulder SystemZimmer, Inc. · 2005 · Recalled
  8. 10 generations backK060704 Zimmer Trabecular Metal Reverse Shoulder System, Sizes 8MM and 10MMZimmer, Inc. · 2006 · Recalled
  9. 9 generations backK063569 Equinoxe Reverse Shoulder SystemExactech, Inc. · 2007 · Recalled
  10. 8 generations backK073688 Equinoxe Reverse Shoulder System Fracture Humeral Adapter Tray and…Exactech, Inc. · 2008 · Recalled
  11. 7 generations backK080642 Comprehensive Reverse ShoulderBiomet, Inc. · 2008 · Recalled (Class I)
  12. 6 generations backK110599 Aequalis Ascend Modular Reverse Shoulder SystemTornier, Inc. · 2011
  13. 5 generations backK112615 Aequalis Ascend Modular Reverse Shoulder SystemTornier, Inc. · 2011
  14. 4 generations backK122698 Aequalis Ascend Flex Shoulder SystemTornier S.A.S. · 2013 · Recalled
  15. 3 generations backK163669 Humeris ShoulderFx Solutions · 2017
  16. 2 generations backK192206 32mm Glenosphere and Humeral CupFx Shoulder USA, Inc. · 2019 · Recalled
  17. 1 generation backK193394 Humeral Cup Stability & Humeral Cup 135/145o StabilityFx Shoulder USA, Inc. · 2020
  18. This deviceK223833 Exactech® Equinoxe® Reverse Humeral Liners and Humeral Adapter TraysExactech, Inc. · 2023

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K260583Equinoxe® Shoulder SystemExactech, Inc. · 2026 · cites it directly0descendants
  • K243448Equinoxe® Shoulder SystemExactech, Inc. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−17K911686Global Total Shoulder, Uncement Humeral Hemi-ShoulDepuy, Inc.1991—
−16K984541Global FX Humeral Stem, Global Advantage Humeral HeadDePuy Orthopaedics, Inc.1999—
−15K992065Global Advantage Shoulder, Global Advantage Humeral Stem, Global Advantage…DePuy Orthopaedics, Inc.1999—
−14K021478Delta ShoulderDepuy, Inc.2003—
−13K030941Aequalis Reversed Shoulder ProsthesisTornier2004—
−12K041873Aequalis Reversed Shoulder ProsthesisTornier2004—
−11K052906Zimmer Trabecular Metal Reverse Shoulder SystemZimmer, Inc.200512
−10K060704Zimmer Trabecular Metal Reverse Shoulder System, Sizes 8MM and 10MMZimmer, Inc.200610
−9K063569Equinoxe Reverse Shoulder SystemExactech, Inc.200713
−8K073688Equinoxe Reverse Shoulder System Fracture Humeral Adapter Tray and Locking ScrewExactech, Inc.20081
−7K080642Comprehensive Reverse ShoulderBiomet, Inc.2008Class I
−6K110599Aequalis Ascend Modular Reverse Shoulder SystemTornier, Inc.2011—
−5K112615Aequalis Ascend Modular Reverse Shoulder SystemTornier, Inc.2011—
−4K122698Aequalis Ascend Flex Shoulder SystemTornier S.A.S.20132
−3K163669Humeris ShoulderFx Solutions2017—
−2K19220632mm Glenosphere and Humeral CupFx Shoulder USA, Inc.20191
−1K193394Humeral Cup Stability & Humeral Cup 135/145o StabilityFx Shoulder USA, Inc.2020—
0 (this device)K223833Exactech® Equinoxe® Reverse Humeral Liners and Humeral Adapter TraysExactech, Inc.2023—
+1K260583Equinoxe® Shoulder SystemExactech, Inc.2026—
+1K243448Equinoxe® Shoulder SystemExactech, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198020
198610
198750
198810
198910
199120
199220
199310
199420
199540
199680
199720
199830
199970
200090
200190
200210
200360
200470
200570
2006100
200750
200850
200910
201030
201140
201230
201330
201410
201510
201710
201820

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
202610
FIG. 1 — Lineage of K223833, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PHX device

  1. #1 K212737 INHANCE™ Reverse Shoulder Systemscore 28.4 · 15 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K063569Equinoxe Reverse Shoulder System
Summary page 7
…IVALENCE IS CLAIMED 510k Number(s) Trade or Proprietary or Model Name Manufacturer Primary Predicate K063569, K082702, K093275, K180632, Equinoxe Reverse Shoulder System Exactech, Inc. K182536…
Exactech, Inc.200713
*K082702Exactech Equinoxe Reverse Shoulder System+15MM Humeral Adapter Tray
Summary page 7
…IS CLAIMED 510k Number(s) Trade or Proprietary or Model Name Manufacturer Primary Predicate K063569, K082702, K093275, K180632, Equinoxe Reverse Shoulder System Exactech, Inc. K182536…
Exactech, Inc.2008—
*K093275Exactech Equinoxe Reverse Shoulder System 36MM Glenosphere and Humeral Liner
Summary page 7
…D 510k Number(s) Trade or Proprietary or Model Name Manufacturer Primary Predicate K063569, K082702, K093275, K180632, Equinoxe Reverse Shoulder System Exactech, Inc. K182536…
Exactech, Inc.20103
*K180632Exactech Equinoxe Small Reverse Shoulder System
Summary page 7
…mber(s) Trade or Proprietary or Model Name Manufacturer Primary Predicate K063569, K082702, K093275, K180632, Equinoxe Reverse Shoulder System Exactech, Inc. K182536 Addit…
Exactech, Inc.20183
*K182536Exactech Equinoxe Small Reverse Shoulder Glenospheres , Exactech Equinoxe Reverse…
Summary page 7
…Primary Predicate K063569, K082702, K093275, K180632, Equinoxe Reverse Shoulder System Exactech, Inc. K182536 Additional Predicate…
Exactech, Inc.20183
*K193394Humeral Cup Stability & Humeral Cup 135/145o Stability
Summary page 7
…tech, Inc. K182536 Additional Predicate FX Shoulder K193394 Humeral Cup Stability & Humeral Cup 135/145° Stability USA NON-CLINICAL TESTING The proposed Equinoxe Reverse Hum…
Fx Shoulder USA, Inc.2020—

Also cleared against K063569: 15 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K260583Equinoxe® Shoulder SystemExactech, Inc.20260
K243448Equinoxe® Shoulder SystemExactech, Inc.20250

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 15, 2025Class IIZ-2562-2025Process designReverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K080642Comprehensive Reverse ShoulderBiomet, Inc. · 2008 · 6 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 32 years and passes through 10 companies; a new generation was cleared every 1.0 years on median.
  • 40 of the 137 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K260583 (Exactech, Inc., 2026) has gone furthest, with 0 descendants of its own.
  • 15 other devices also cite its predicate K063569.

How does its category compare?

All PHX devices

PHX Shoulder Prosthesis, Reverse Configuration: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · PHX

9in 2025

2007: 52008: 62009: 12010: 82011: 122012: 122013: 82014: 122015: 42016: 42017: 112018: 142019: 92020: 132021: 212022: 92023: 92024: 92025: 92026 (partial year): 18
2007high 21 (2021) · low 1 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PHX

72in 2025

2007: 97 days median2008: 79 days median2010: 124 days median2011: 94 days median2012: 114 days median2013: 143 days median2014: 110 days median2015: 206 days median2016: 132 days median2017: 141 days median2018: 127 days median2019: 144 days median2020: 143 days median2021: 143 days median2022: 198 days median2023: 154 days median2024: 139 days median2025: 72 days median2026 (partial year): 130 days median
2007high 206 (2015) · low 72 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PHX

3in 2025

2015: 22018: 12019: 32020: 32021: 82023: 72024: 12025: 32026 (partial year): 7
2007high 8 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PHX

2,848in 2025

2015: 142016: 52017: 252018: 3232019: 5042020: 9862021: 1,4042022: 1,3422023: 1,4012024: 2,1422025: 2,848
2007high 2,848 (2025) · low 0 (2007)2025

MAUDE reports naming product code PHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200759700
200867900
20091—00
2010812400
2011129400
20121211400
2013814300
20141211000
20154206214
2016413205
201711141025
2018141271323
201991443504
2020131433986
20212114381,404
2022919801,342
2023915471,401
2024913912,142
202597232,848
2026181307—

What is it indicated for?

The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi-arthroplasty is determined by the surgeon to be the preferred method of treatment. • The cemented primary humeral stem, long/revision stem, fracture stems, and all Equinoxe glenoids are intended for cemented fixation. • The press-fit humeral stems are intended for press-fit applications…

Read the full statement (414 words)

The Equinoxe Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases or fractures of the glenohumeral joint where total or hemi-arthroplasty is determined by the surgeon to be the preferred method of treatment.

• The cemented primary humeral stem, long/revision stem, fracture stems, and all Equinoxe glenoids are intended for cemented fixation.

• The press-fit humeral stems are intended for press-fit applications but may be used with bone cement at the discretion the surgeon.

• The reverse humeral components are intended to be used in cemented applications or in revision cases when the humeral component is well-fixed/stable, as deemed by the orthopaedic surgeon.

• Humeral Heads are intended for use in cemented and press-fit applications.

Clinical indications for the PRIMARY (P), LONG/REVISION (L), and FRACTURE (F) humeral components are as follows:

P L F Indications

√ √ Rheumatoid arthritis, osteoarthritis, osteonecrosis or post-traumatic degenerative problems

√ √ Congenital abnormalities in the skeletally mature

√ Primary and secondary necrosis of the humeral head.

√ √ Humeral head fracture with displacement of the tuberosities

√ √ Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable

√ √ Revisions of humeral prostheses when other treatments or devices have failed (where adequate fixation can be achieved)

√ Displaced three-part and four-part upper humeral fractures

√ Spiral and other fractures of the mid-humerus (in combination with glenohumeral degenerative diseases)

√ Revision of failed previous reconstructions when distal anchorage is required

√ √ To restore mobility from previous procedures (e.g. previous fusion)

The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.

The Equinoxe Platform Fracture Stem is indicated for use in skeletally mature individuals with acute fracture of the proximal humerus and displacement of the tuberosities, displaced 3- and 4-part fractures of the proximal humerus (hemi-arthroplasty), or acute fracture of the proximal humerus with failure of the glenohumeral joint (primary total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is also indicated for acute fracture of the proximal humerus in combination with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff resulting in superior migration of the humeral head (reverse total shoulder arthroplasty). The Equinoxe Platform Fracture Stem is indicated for cemented use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K223833: Exactech® Equinoxe® Reverse Humeral Liners and Humeral Adapter Trays.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K223833

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026