510(k) Report

510(k) · Cleared

Cleared

CoverScan (CoverScan v1.1)

K230294510(k) SpecialCleared Mar 3, 202329-day reviewClass II21 CFR 892.2050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Perspectum · LLZ System, Image Processing, Radiological

Cleared on Mar 3, 2023 as substantially equivalent to K212565 (Perspectum, Ltd., 2022) after a 29-day review. Its family tree reaches back 15 generations to 1997. 3 of the 65 devices in its family, counting itself, have a Class I recall.

Decided
Mar 3, 2023
Review
29 days
Class
Class II
Product code
LLZ
Recalls
None
Predicates
1

Its longest line of descent runs back 15 generations to K971684 (1997). 64 ancestors in all, the oldest K896095 (1990). No later device cites it as a predicate yet. 3 of the 65 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2016: 1 other ancestor2014: 1 other ancestor2010: 2 other ancestors2013: 2 other ancestors2007: 3 other ancestors2012: 2 other ancestors2011: 3 other ancestors2009: 3 other ancestors2006: 2 other ancestors2005: 4 other ancestors2002: 3 other ancestors2008: 1 other ancestor2004: 3 other ancestors2000: 2 other ancestors1998: 2 other ancestors1994: 2 other ancestors2003: 2 other ancestors2001: 5 other ancestors1999: 2 other ancestors1996: 2 other ancestors1991: 1 other ancestor1990: 1 other ancestorK971684 · Magnetom Project 047 System (Numaris 3.5 Software) · Siemens Medical Solutions USA, Inc. · 1997 · recalledK032428 · Magnetom Avanto MR System · Siemens Medical Solutions USA, Inc. · 2003 · recalledK041112 · Magnetom Espree System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK052164 · Magnetom Systems with Software Syngo MR 2006A · Siemens Medical Solutions USA, Inc. · 2005K062454 · Magnetom Systems - I-Class and T-Class Releases · Siemens Medi Cal Solutions, Inc. · 2006K072237 · Magnetom Verio · Siemens Medical Solutions USA, Inc. · 2007 · Class I recallK082427 · Syngo MR B17 · Siemens Medical Solutions USA, Inc. · 2008K101347 · Magnetom Aera; Magnetom Skyra · Siemens Medical Solutions USA, Inc. · 2010 · Class I recallK111242 · Magnetom Aera and Magnetom Skyra · Siemens Medical Solutions USA, Inc. · 2011 · recalledK121434 · Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/Verio · Siemens Medi Cal Solutions, Inc. · 2012 · Class I recallK141977 · Magnetom Aera with Software Syngo MR E11A, Magnetom Skyra with Software Syngo MR E11A, Magnetom Skyra with 24 RF Channel · Siemens Medical Solutions USA, Inc. · 2014K143020 · LiverMultiScan · Mirada Medical, Ltd. · 2015K172685 · LiverMultiScan · Perspectum Diagnostics, Ltd. · 2017K190017 · LiverMultiScan (LMSv3) · Perspectum Diagnostics, Ltd. · 2019K212565 · CoverScan v1 · Perspectum, Ltd. · 202219901995200020052010201520202025
Other ancestors, by generation 2 generations back: 3 other ancestors3 generations back: 3 other ancestors4 generations back: 5 other ancestors5 generations back: 8 other ancestors6 generations back: 10 other ancestors7 generations back: 9 other ancestors8 generations back: 4 other ancestors9 generations back: 3 other ancestors10 generations back: 1 other ancestor11 generations back: 1 other ancestor12 generations back: 2 other ancestors K212565 · CoverScan v1 · Perspectum, Ltd. · 2022K190017 · LiverMultiScan (LMSv3) · Perspectum Diagnostics, Ltd. · 2019K172685 · LiverMultiScan · Perspectum Diagnostics, Ltd. · 2017K143020 · LiverMultiScan · Mirada Medical, Ltd. · 2015K141977 · Magnetom Aera with Software Syngo MR E11A, Magnetom Skyra with Software Syngo MR E11A, Magnetom Skyra with 24 RF Channel · Siemens Medical Solutions USA, Inc. · 2014K121434 · Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/Verio · Siemens Medi Cal Solutions, Inc. · 2012 · Class I recallK111242 · Magnetom Aera and Magnetom Skyra · Siemens Medical Solutions USA, Inc. · 2011 · recalledK101347 · Magnetom Aera; Magnetom Skyra · Siemens Medical Solutions USA, Inc. · 2010 · Class I recallK082427 · Syngo MR B17 · Siemens Medical Solutions USA, Inc. · 2008K072237 · Magnetom Verio · Siemens Medical Solutions USA, Inc. · 2007 · Class I recallK062454 · Magnetom Systems - I-Class and T-Class Releases · Siemens Medi Cal Solutions, Inc. · 2006K052164 · Magnetom Systems with Software Syngo MR 2006A · Siemens Medical Solutions USA, Inc. · 2005K041112 · Magnetom Espree System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK032428 · Magnetom Avanto MR System · Siemens Medical Solutions USA, Inc. · 2003 · recalledK971684 · Magnetom Project 047 System (Numaris 3.5 Software) · Siemens Medical Solutions USA, Inc. · 1997 · recalled−15−10−50
  1. Root · 15 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 11 generations between
  3. 3 generations backK172685 LiverMultiScanPerspectum Diagnostics, Ltd. · 2017
  4. 2 generations backK190017 LiverMultiScan (LMSv3)Perspectum Diagnostics, Ltd. · 2019
  5. 1 generation backK212565 CoverScan v1Perspectum, Ltd. · 2022
  6. This deviceK230294 CoverScan (CoverScan v1.1)Perspectum · 2023
Show all 15 generations
  1. Root · 15 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 14 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  3. 13 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  4. 12 generations backK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
  5. 11 generations backK062454 Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006
  6. 10 generations backK072237 Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · Recalled (Class I)
  7. 9 generations backK082427 Syngo MR B17Siemens Medical Solutions USA, Inc. · 2008
  8. 8 generations backK101347 Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · Recalled (Class I)
  9. 7 generations backK111242 Magnetom Aera and Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2011 · Recalled
  10. 6 generations backK121434 Software Syngo MR D13A for the Magnetom Systems…Siemens Medi Cal Solutions, Inc. · 2012 · Recalled (Class I)
  11. 5 generations backK141977 Magnetom Aera with Software Syngo MR E11A, Magnetom Skyra with…Siemens Medical Solutions USA, Inc. · 2014
  12. 4 generations backK143020 LiverMultiScanMirada Medical, Ltd. · 2015
  13. 3 generations backK172685 LiverMultiScanPerspectum Diagnostics, Ltd. · 2017
  14. 2 generations backK190017 LiverMultiScan (LMSv3)Perspectum Diagnostics, Ltd. · 2019
  15. 1 generation backK212565 CoverScan v1Perspectum, Ltd. · 2022
  16. This deviceK230294 CoverScan (CoverScan v1.1)Perspectum · 2023

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−15K971684Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc.19974
−14K032428Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc.20034
−13K041112Magnetom Espree SystemSiemens Medical Solutions USA, Inc.20044
−12K052164Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc.2005—
−11K062454Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc.2006—
−10K072237Magnetom VerioSiemens Medical Solutions USA, Inc.2007Class I
−9K082427Syngo MR B17Siemens Medical Solutions USA, Inc.2008—
−8K101347Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc.2010Class I
−7K111242Magnetom Aera and Magnetom SkyraSiemens Medical Solutions USA, Inc.20113
−6K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc.2012Class I
−5K141977Magnetom Aera with Software Syngo MR E11A, Magnetom Skyra with Software Syngo…Siemens Medical Solutions USA, Inc.2014—
−4K143020LiverMultiScanMirada Medical, Ltd.2015—
−3K172685LiverMultiScanPerspectum Diagnostics, Ltd.2017—
−2K190017LiverMultiScan (LMSv3)Perspectum Diagnostics, Ltd.2019—
−1K212565CoverScan v1Perspectum, Ltd.2022—
0 (this device)K230294CoverScan (CoverScan v1.1)Perspectum2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199110
199420
199620
199820
199920
200020
200150
200230
200320
200430
200540
200620
200730
200810
200930
201020
201130
201220
201320
201410
201610
FIG. 1 — Lineage of K230294, 1997–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K212565CoverScan v1
Summary page 4
…Ioan Wigley Subject and Predicate Device Predicate Device Subject Device 510(k) number K212565 Unknown Legal Manufacturer Perspectum Ltd. Perspectum Ltd. Owner/Ope…
Perspectum, Ltd.2022—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc. · 2012 · 3 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K101347Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · 5 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
  • K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 26 years and passes through 6 companies; a new generation was cleared every 1.1 years on median.
  • 18 of the 64 ancestors have a recall of any class; 3 have a Class I recall.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

CoverScan is a medical image management and processing software package that allows the display, analysis and post-processing of DICOM compliant medical images and MR data. CoverScan provides both viewing and analysis capabilities to ascertain quantified metrics of multiple organs such as the heart, lungs, liver, spleen, pancreas and kidney. CoverScan provides measurements in different organs to be used for the assessment of longitudinal and transversal…

Read the full statement (143 words)

CoverScan is a medical image management and processing software package that allows the

display, analysis and post-processing of DICOM compliant medical images and MR data.

CoverScan provides both viewing and analysis capabilities to ascertain quantified metrics of multiple

organs such as the heart, lungs, liver, spleen, pancreas and kidney.

CoverScan provides measurements in different organs to be used for the assessment of longitudinal

and transversal relaxation time (T1, srT1, cT1, T2), fat content (proton density fat fraction or PDFF)

and metrics of organ function (e.g., left ventricular ejection fraction and lung fractional area change

on deep inspiration).

These metrics derived from the images, when interpreted by a licensed physician, yield information

that may assist in diagnosis, clinical management and monitoring of patients.

CoverScan is not intended for asymptomatic screening. This device is intended for use with

Siemens 1.5T and Siemens 3T MRI scanners.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K230294: CoverScan (CoverScan v1.1).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K230294

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026