510(k) · Cleared
Recalled · Class I · 2025Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve Catheter (SCC-1) and IOPS Reverse Curve Catheter (RCC-1); IOPS Guidewire (ATW-2); IOPS Tracking Pad (TP-1); IOPS Guidewire Handle (SSH-1)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Centerline Biomedical, Inc. · DQK Computer, Diagnostic, Programmable
Cleared on Jun 21, 2023 as substantially equivalent to K190106 (Centerline Biomedical, Inc., 2019) after a 138-day review. Its family tree reaches back 14 generations to 1993. 6 of the 97 devices in its family, counting itself, have a Class I recall. 5 later devices descend from it. It has been recalled once, including a Class I recall.
- Decided
- Jun 21, 2023
- Review
- 138 days
- Class
- Class II
- Product code
- DQK
- Recalls
- 1 (1 Class I)
- Predicates
- 1
Its longest line of descent runs back 14 generations to K926514 (1993). 91 ancestors in all, the oldest K882061 (1988). 5 later devices descend from it, 3 citing it directly. 6 of the 97 devices in this family, counting this one, have a Class I recall.
- Root · 14 generations backK926514 Endoscopic Ultrasound System Type 20Olympus Corp. · 1993 · Recalled
- 10 generations between
- 3 generations backK091781 Guided Medical Positioning System (GMPS)Mediguide, Ltd. · 2009
- 2 generations backK120298 CPS Excel Mediguide Enabled GuidewireSt. Jude Medical, Cardiac Rhythm Management Divisi · 2012
- 1 generation backK190106 Intra-Operative Positioning System; Simple Curve Catheter, Reverse…Centerline Biomedical, Inc. · 2019 · Recalled (Class I)
- This deviceK230309 Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve…Centerline Biomedical, Inc. · 2023
- Descendant · cites it directly · 3 of its ownK242133 Intra-Operative Positioning System (IOPS®) (Fiducial Tracking Pad)…Centerline Biomedical, Inc. · 2024 · Recalled (Class I)
- Descendant · cites it directly · 3 of its ownK241243 Intra-Operative Positioning System (IOPS®) Viewpoint (Simple Curve…Centerline Biomedical, Inc. · 2024 · Recalled (Class I)
- Descendant · cites it directly · 2 of its ownK243842 Intra-Operative Positioning System (IOPS®)Centerline Biomedical, Inc. · 2025 · Recalled (Class I)
- Descendant · 2 generations on · 0 of its ownK261329 Intra-Operative Positioning System (IOPS) (MC-1);IOPS Viewpoint…Centerline Biomedical · 2026
- Descendant · 2 generations on · 0 of its ownK254089 IOPS Visionary System (MC-3); IOPS Viewpoint Simple Curve Catheter…Centerline Biomedical, Inc. · 2026
Show all 14 generations
- Root · 14 generations backK926514 Endoscopic Ultrasound System Type 20Olympus Corp. · 1993 · Recalled
- 13 generations backK951994 Olympus EU-M30 Endoscopic Ultrasound CenterOlympus America, Inc. · 1996
- 12 generations backK982323 Olympus UM 2R/3R Ultrasonic Probes and Associated Ancillary…Olympus America, Inc. · 1998
- 11 generations backK042438 Superdimension/BronchusSuperdimension, Ltd. · 2004 · Recalled
- 10 generations backK052260 Superdimension/Bronchus 4.1Superdimension, Ltd. · 2005 · Recalled
- 9 generations backK052852 Superdimension/Bronchus PremiumSuperdimension, Ltd. · 2005 · Recalled
- 8 generations backK062315 Superdimension/Bronchus Premium 2Superdimension, Ltd. · 2006 · Recalled
- 7 generations backK071473 Modification to Superdimension/Bronchus Premium 2Superdimension, Ltd. · 2007
- 6 generations backK080271 Inreach SystemSuperdimension, Ltd. · 2008
- 5 generations backK081379 Modification to Inreach SystemSuperdimension, Ltd. · 2008
- 4 generations backK083745 The IC-Pro System, Model Version 3.2Paieon, Inc. · 2009
- 3 generations backK091781 Guided Medical Positioning System (GMPS)Mediguide, Ltd. · 2009
- 2 generations backK120298 CPS Excel Mediguide Enabled GuidewireSt. Jude Medical, Cardiac Rhythm Management Divisi · 2012
- 1 generation backK190106 Intra-Operative Positioning System; Simple Curve Catheter, Reverse…Centerline Biomedical, Inc. · 2019 · Recalled (Class I)
- This deviceK230309 Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve…Centerline Biomedical, Inc. · 2023
Class I recalls in this family
- K190106 Intra-Operative Positioning System; Simple Curve…direct predicate · 2019
- K023984 Modification to Olympus Bronchoscopes BF-40…12 generations back · 2003
- K242133 Intra-Operative Positioning System (IOPS®)…cites it directly · 2024
- K241243 Intra-Operative Positioning System (IOPS®)…cites it directly · 2024
Most influential descendants
Showing 5 of 5 · See all- K242133Intra-Operative Positioning System (IOPS®) (Fiducial Tracking Pad); Intra-Operative…Centerline Biomedical, Inc. · 2024 · cites it directly3descendants
- K241243Intra-Operative Positioning System (IOPS®) Viewpoint (Simple Curve Catheters)…Centerline Biomedical, Inc. · 2024 · cites it directly3descendants
- K243842Intra-Operative Positioning System (IOPS®)Centerline Biomedical, Inc. · 2025 · cites it directly2descendants
- K261329Intra-Operative Positioning System (IOPS) (MC-1);IOPS Viewpoint Simple Curve Catheter…Centerline Biomedical · 2026 · 2 generations on0descendants
- K254089IOPS Visionary System (MC-3); IOPS Viewpoint Simple Curve Catheter, 75cm (C00751); IOPS…Centerline Biomedical, Inc. · 2026 · 2 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −14 | K926514 | Endoscopic Ultrasound System Type 20 | Olympus Corp. | 1993 | 1 |
| −13 | K951994 | Olympus EU-M30 Endoscopic Ultrasound Center | Olympus America, Inc. | 1996 | — |
| −12 | K982323 | Olympus UM 2R/3R Ultrasonic Probes and Associated Ancillary Equipment, for… | Olympus America, Inc. | 1998 | — |
| −11 | K042438 | Superdimension/Bronchus | Superdimension, Ltd. | 2004 | 1 |
| −10 | K052260 | Superdimension/Bronchus 4.1 | Superdimension, Ltd. | 2005 | 2 |
| −9 | K052852 | Superdimension/Bronchus Premium | Superdimension, Ltd. | 2005 | 1 |
| −8 | K062315 | Superdimension/Bronchus Premium 2 | Superdimension, Ltd. | 2006 | 1 |
| −7 | K071473 | Modification to Superdimension/Bronchus Premium 2 | Superdimension, Ltd. | 2007 | — |
| −6 | K080271 | Inreach System | Superdimension, Ltd. | 2008 | — |
| −5 | K081379 | Modification to Inreach System | Superdimension, Ltd. | 2008 | — |
| −4 | K083745 | The IC-Pro System, Model Version 3.2 | Paieon, Inc. | 2009 | — |
| −3 | K091781 | Guided Medical Positioning System (GMPS) | Mediguide, Ltd. | 2009 | — |
| −2 | K120298 | CPS Excel Mediguide Enabled Guidewire | St. Jude Medical, Cardiac Rhythm Management Divisi | 2012 | — |
| −1 | K190106 | Intra-Operative Positioning System; Simple Curve Catheter, Reverse Curve… | Centerline Biomedical, Inc. | 2019 | Class I |
| 0 (this device) | K230309 | Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve Catheter… | Centerline Biomedical, Inc. | 2023 | — |
| +1 | K242133 | Intra-Operative Positioning System (IOPS®) (Fiducial Tracking Pad)… | Centerline Biomedical, Inc. | 2024 | Class I |
| +1 | K241243 | Intra-Operative Positioning System (IOPS®) Viewpoint (Simple Curve Catheters)… | Centerline Biomedical, Inc. | 2024 | Class I |
| +1 | K243842 | Intra-Operative Positioning System (IOPS®) | Centerline Biomedical, Inc. | 2025 | Class I |
| +2 | K261329 | Intra-Operative Positioning System (IOPS) (MC-1);IOPS Viewpoint Simple Curve… | Centerline Biomedical | 2026 | — |
| +2 | K254089 | IOPS Visionary System (MC-3); IOPS Viewpoint Simple Curve Catheter, 75cm… | Centerline Biomedical, Inc. | 2026 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1988 | 1 | 0 |
| 1989 | 3 | 0 |
| 1991 | 2 | 0 |
| 1992 | 3 | 0 |
| 1993 | 3 | 0 |
| 1994 | 2 | 0 |
| 1995 | 8 | 0 |
| 1996 | 1 | 0 |
| 1997 | 2 | 0 |
| 1998 | 2 | 0 |
| 1999 | 3 | 0 |
| 2000 | 5 | 0 |
| 2001 | 4 | 0 |
| 2002 | 5 | 0 |
| 2003 | 8 | 1 |
| 2004 | 4 | 0 |
| 2005 | 2 | 0 |
| 2006 | 4 | 0 |
| 2007 | 3 | 0 |
| 2008 | 3 | 0 |
| 2009 | 1 | 0 |
| 2010 | 3 | 0 |
| 2012 | 2 | 0 |
| 2015 | 2 | 0 |
| 2016 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2024 | 2 | 2 |
| 2025 | 1 | 1 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
6 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DQK device
- #1 K213682 HemoSphere Advanced Monitor, HemoSphere…score 26.6 · 28 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K190106 | Intra-Operative Positioning System; Simple Curve Catheter, Reverse Curve Catheter…Summary page 4…) Centerline Biomedical Inc. is requesting marketing clearance for the labeling modification of the Intra-Operative Positioning System (IOPS®) previously cleared under K190106. A. Submitted By: Centerline Biomedical, Inc. 10000 Cedar Avenue… | 2019 |
What descends from it? (5)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (6 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 11, 2025 | Class I | Z-1760-2025 | Under Investigation by firm | Centerline Biomedical has determined that the coating on the guidewires for lot 2404-2005 may delaminate during use of the guidewire. |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K190106Intra-Operative Positioning System; Simple Curve Catheter, Reverse Curve Catheter ; Angled Tip Guidewire…Centerline Biomedical, Inc. · 2019 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
- K023984Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series, and BF-160 SeriesThe Olympus Optical Co. · 2003 · 12 generations back · ◆ in the 2003 ancestors bar of FIG. 114 Class I recalls
- K242133Intra-Operative Positioning System (IOPS®) (Fiducial Tracking Pad); Intra-Operative Positioning System…Centerline Biomedical, Inc. · 2024 · Cites it directly · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
- K241243Intra-Operative Positioning System (IOPS®) Viewpoint (Simple Curve Catheters); Intra-Operative Positioning…Centerline Biomedical, Inc. · 2024 · Cites it directly · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
- K243842Intra-Operative Positioning System (IOPS®)Centerline Biomedical, Inc. · 2025 · Cites it directly · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 30 years and passes through 7 companies; a new generation was cleared every 10 months on median.
- 19 of the 91 ancestors have a recall of any class; 2 have a Class I recall.
- Of its descendants, K242133 (Centerline Biomedical, Inc., 2024) has gone furthest, with 3 descendants of its own.
- 3 other devices also cite its predicate K190106.
How does its category compare?
All DQK devicesDQK Computer, Diagnostic, Programmable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.
FDA decisions per year · DQK
23so far in 2026, the most since 2007
2026 to Oct 25, outlined.
Median 510(k) review, days · DQK
95in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DQK
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DQK
433in 2025
MAUDE reports naming product code DQK, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 13 | 83 | 4 | 18 |
| 2008 | 4 | 70 | 1 | 34 |
| 2009 | 12 | 93 | 2 | 40 |
| 2010 | 9 | 64 | 3 | 49 |
| 2011 | 12 | 69 | 2 | 55 |
| 2012 | 7 | 74 | 3 | 129 |
| 2013 | 13 | 68 | 1 | 180 |
| 2014 | 9 | 90 | 4 | 115 |
| 2015 | 7 | 125 | 2 | 131 |
| 2016 | 15 | 190 | 0 | 340 |
| 2017 | 7 | 146 | 2 | 252 |
| 2018 | 14 | 150 | 6 | 261 |
| 2019 | 15 | 113 | 1 | 218 |
| 2020 | 16 | 123 | 1 | 215 |
| 2021 | 16 | 132 | 1 | 351 |
| 2022 | 19 | 136 | 2 | 319 |
| 2023 | 21 | 133 | 2 | 340 |
| 2024 | 13 | 100 | 24 | 557 |
| 2025 | 22 | 95 | 1 | 433 |
| 2026 | 23 | 151 | 2 | — |
What is it indicated for?
The IOPS® (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor equipped compatible catheters and guidewires used in endovascular interventions in the descending aorta. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K230309 · printed Nov 6, 2026
Cite this page
510(k) Report. “K230309: Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve Catheter (SCC-1) and IOPS Reverse Curve Catheter (RCC-1); IOPS Guidewire (ATW-2); IOPS Tracking Pad (TP-1); IOPS Guidewire Handle (SSH-1).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K230309
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026