Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series, and BF-160 Series

K023984510(k) TraditionalCleared Jan 31, 200360-day reviewClass II21 CFR 874.4680Recalls 45 (14 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

The Olympus Optical Co. · EOQ Bronchoscope (Flexible Or Rigid)

Cleared on Jan 31, 2003 as substantially equivalent to K910423 (Olympus Corp., 1991) and 1 other predicate after a 60-day review. Its family tree reaches back 2 generations to 1993. 13 of the 354 devices in its family, counting itself, have a Class I recall. 350 later devices descend from it. It has been recalled 45 times, including 14 Class I recalls.

Decided
Jan 31, 2003
Review
60 days
Class
Class II
Product code
EOQ
Recalls
45 (14 Class I)
Predicates
2

Its longest line of descent runs back 2 generations to K931154 (1993). 3 ancestors in all, the oldest K910423 (1991). 350 later devices descend from it, 9 citing it directly. 13 of the 354 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1991: 1 other ancestor2003: 2 descendants cleared2004: 3 descendants cleared2010: 6 descendants cleared2006: 3 descendants cleared, 1 with a Class I recall12005: 6 descendants cleared, 1 with a Class I recall12008: 7 descendants cleared2013: 7 descendants cleared, 1 with a Class I recall12009: 10 descendants cleared2012: 14 descendants cleared2014: 13 descendants cleared2023: 29 descendants cleared, 1 with a Class I recall12022: 20 descendants cleared2017: 9 descendants cleared2019: 25 descendants cleared, 2 with a Class I recall22021: 21 descendants cleared, 1 with a Class I recall12016: 10 descendants cleared2007: 2 descendants cleared2011: 4 descendants cleared2026: 27 descendants cleared2018: 16 descendants cleared2020: 28 descendants cleared, 1 with a Class I recall12024: 31 descendants cleared, 2 with a Class I recall22015: 13 descendants cleared2025: 44 descendants cleared, 2 with a Class I recall2K931154 · Evis 200 System · Olympus Corp. · 1993K963033 · BF 240/P240/IT240 Bronchovideoscope & Accessories · Olympus America, Inc. · 1996 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 9 descendants, 2 with a Class I recall22 generations forward: 22 descendants, 1 with a Class I recall13 generations forward: 55 descendants4 generations forward: 105 descendants, 2 with a Class I recall25 generations forward: 62 descendants, 1 with a Class I recall16 generations forward: 32 descendants, 1 with a Class I recall17 generations forward: 19 descendants8 generations forward: 12 descendants9 generations forward: 7 descendants10 generations forward: 7 descendants11 generations forward: 5 descendants, 1 with a Class I recall112 generations forward: 5 descendants, 4 with a Class I recall413 generations forward: 5 descendants14 generations forward: 3 descendants15 generations forward: 2 descendants K963033 · BF 240/P240/IT240 Bronchovideoscope & Accessories · Olympus America, Inc. · 1996 · recalledK931154 · Evis 200 System · Olympus Corp. · 19930+5+10+15
  1. Root · 2 generations backK931154 Evis 200 SystemOlympus Corp. · 1993
  2. 1 generation backK963033 BF 240/P240/IT240 Bronchovideoscope & AccessoriesOlympus America, Inc. · 1996 · Recalled
  3. This deviceK023984 Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series…The Olympus Optical Co. · 2003
  4. Descendant · cites it directly · 159 of its ownK031648 Visera Rhino-Laryngovideoscope Olympus Enf Type VOlympus Optical Co., Ltd. · 2003
  5. Descendant · cites it directly · 158 of its ownK061313 Evis Exera 180 SystemOlympus Medical Systems Corporation · 2006 · Recalled (Class I)
  6. Descendant · 2 generations on · 132 of its ownK100584 Evis Exeraii 180 SystemOlympus Medical Systems Corporation · 2010 · Recalled
  7. Descendant · cites it directly · 73 of its ownK042438 Superdimension/BronchusSuperdimension, Ltd. · 2004 · Recalled
  8. Descendant · cites it directly · 64 of its ownK130845 Ambu Ascope 3 5.0/2.2; Ambu Ascope 3 Slim 3.8/1.2; Ambu AviewAmbu A/S · 2013

Showing 5 of 350 descendants. See all descendants

Descendants cleared per year

2003: 22004: 32005: 62006: 32007: 22008: 72009: 102010: 62011: 42012: 142013: 72014: 132015: 132016: 102017: 92018: 162019: 252020: 282021: 212022: 202023: 292024: 312025: 442026: 27
2003high 44 (2025) · low 2 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 350 · See all
  • K031648Visera Rhino-Laryngovideoscope Olympus Enf Type VOlympus Optical Co., Ltd. · 2003 · cites it directly159descendants
  • K061313Evis Exera 180 SystemOlympus Medical Systems Corporation · 2006 · cites it directly158descendants
  • K100584Evis Exeraii 180 SystemOlympus Medical Systems Corporation · 2010 · 2 generations on132descendants
  • K042438Superdimension/BronchusSuperdimension, Ltd. · 2004 · cites it directly73descendants
  • K130845Ambu Ascope 3 5.0/2.2; Ambu Ascope 3 Slim 3.8/1.2; Ambu AviewAmbu A/S · 2013 · cites it directly64descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K931154Evis 200 SystemOlympus Corp.1993—
−1K963033BF 240/P240/IT240 Bronchovideoscope & AccessoriesOlympus America, Inc.19966
0 (this device)K023984Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series, and BF-160…The Olympus Optical Co.2003—
+1K031648Visera Rhino-Laryngovideoscope Olympus Enf Type VOlympus Optical Co., Ltd.2003—
+1K061313Evis Exera 180 SystemOlympus Medical Systems Corporation2006Class I
+2K100584Evis Exeraii 180 SystemOlympus Medical Systems Corporation20109
+1K042438Superdimension/BronchusSuperdimension, Ltd.20041
+1K130845Ambu Ascope 3 5.0/2.2; Ambu Ascope 3 Slim 3.8/1.2; Ambu AviewAmbu A/S2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200320
200430
200561
200631
200720
200870
2009100
201060
201140
2012140
201371
2014130
2015130
2016100
201790
2018160
2019252
2020281
2021211
2022200
2023291
2024312
2025442
2026270
FIG. 1 — Lineage of K023984, 1993–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

13 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new EOQ device

  1. #1 K223299 Ambu® aScope™ 5 Broncho HD 5.0/2.2, Ambu®…score 18.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K910423Bronchoscope BF-N20
Summary page 1
…on: 21 CFR 874.4680 Bronchoscopes (flexible or rigid) and accessories, Class II *Predicate Device: BF-N20(#K910423) <Olynipus Bronchoscopes BF-240, 160 series> Trade Name: Olympus Bronchoscopes BF-240, 160 series Common…
Olympus Corp.19912
*K963033BF 240/P240/IT240 Bronchovideoscope & Accessories
Summary page 2
…and accessories, Class IT * Predicate Device: BF-240, BF-P24OYBF-1 T240 (#K963033) *The difference between the predicate devices and subjects devices is only a change in material of the Biopsy Port. The change…
Olympus America, Inc.19966

Also cleared against K910423: 1 other device →

What descends from it? (350)

See all 350

9 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K031648Visera Rhino-Laryngovideoscope Olympus Enf Type VOlympus Optical Co., Ltd.2003159
K061313Evis Exera 180 System Class IOlympus Medical Systems Corporation2006158
K042438Superdimension/BronchusSuperdimension, Ltd.200473
K130845Ambu Ascope 3 5.0/2.2; Ambu Ascope 3 Slim 3.8/1.2; Ambu AviewAmbu A/S201364
K033225Evis Exera Bronchofibervideoscope Olympus BF Type XP160F, Its Accessories and Ancillary…Olympus Optical Co., Ltd.200360
K082720Tracheal Intubation Fibervideoscope, Models LF-Y0004 and LF-Y0005Olympus Medical Systems Corporation20081
K093186Ambu Ascope and MonitorAmbu A/S20100
K050220Evis Exera Bronchovideoscope, Olympus XBF-160Y3AC, XBF-1T160Y3AC,XBF-Q160Y2AC Class IOlympus Medical Systems Corporation20050
K040940Bronchovideoscope Olympus XBF-D160HMOlympus Corporation20040

Has anything in its family been recalled? (13 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 11, 2025Class IZ-0065-2026Device DesignAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Sep 11, 2025Class IZ-0045-2026Device DesignAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Sep 11, 2025Class IZ-0055-2026Device DesignAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Sep 11, 2025Class IZ-0058-2026Device DesignAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Sep 11, 2025Class IZ-0059-2026Device DesignAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Sep 11, 2025Class IZ-0054-2026Device DesignAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Sep 11, 2025Class IZ-0047-2026Device DesignAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Sep 11, 2025Class IZ-0064-2026Device DesignAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Sep 11, 2025Class IZ-0067-2026Device DesignAdditional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
Oct 12, 2023Class IZ-0189-2024Under Investigation by firmThere is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.
Sep 25, 2023Class IIZ-0327-2024Under Investigation by firmChannel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead…
Jul 27, 2023Class IIZ-2490-2023Labeling Change ControlComplaint of endoscope model becoming lodged in the endotracheal tube connector due to diameter being too large.
Jun 8, 2023Class IZ-2013-2023Under Investigation by firmThere have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
Jun 8, 2023Class IZ-2014-2023Under Investigation by firmThere have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
Jun 8, 2023Class IZ-2011-2023Under Investigation by firmThere have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
Jun 8, 2023Class IZ-2012-2023Under Investigation by firmThere have been complaints of endobronchial combustion during therapeutic laser procedures with the bronchoscope.
Mar 8, 2022Class IIZ-1047-2022OtherBronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using sterilization
Mar 8, 2022Class IIZ-1046-2022OtherBronchofiberscope Updated instructions on the use of certain chemicals to be used for reprocessing, conditions for ETO gas sterilization, rinsing steps after high-level disinfection, and a recommendation on using sterilization
Aug 17, 2021Class IIZ-2497-2021Under Investigation by firmA gluing step was not performed during the manufacturing of the affected device's insertion tube. There is a potential for the insertion tube to develop a leak, which poses an infection control risk.
Aug 16, 2021Class IIZ-2565-2021Reprocessing ControlsThe firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
Aug 16, 2021Class IIZ-2567-2021Reprocessing ControlsThe firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
Aug 16, 2021Class IIZ-2566-2021Reprocessing ControlsThe firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
Aug 16, 2021Class IIZ-2568-2021Reprocessing ControlsThe firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
Aug 16, 2021Class IIZ-2570-2021Reprocessing ControlsThe firm is issuing validated, new reprocessing instructions according to the FDA 2015 guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling."
Aug 8, 2018Class IIZ-3309-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3295-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3294-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3287-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3299-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3300-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3290-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3306-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3307-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3296-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3308-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3283-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3284-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3298-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3293-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3305-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3292-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3291-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3289-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3312-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.
Aug 8, 2018Class IIZ-3286-2018OtherThe attachment of non-Olympus accessories to the bronchoscope s instrument channel port may have resulted in more applied force than expected and lead to loosening of the instrument channel port.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K061313Evis Exera 180 SystemOlympus Medical Systems Corporation · 2006 · Cites it directly · ◆ in the 2006 descendants bar of FIG. 16 Class I recalls
  • K050220Evis Exera Bronchovideoscope, Olympus XBF-160Y3AC, XBF-1T160Y3AC,XBF-Q160Y2ACOlympus Medical Systems Corporation · 2005 · Cites it directly · ◆ in the 2005 descendants bar of FIG. 12 Class I recalls
  • K121959Evis Exera Iii Video SystemOlympus Medical Systems Corp. · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 18 Class I recalls
  • K183419EVIS EXERA III Bronchovideoscope Olympus BF-XT190Olympus Medical Systems Corp. · 2019 · 4 generations on · ◆ in the 2019 descendants bar of FIG. 12 Class I recalls
  • K250862Evis Exera Iii Bronchovideoscope (Olympus BF-XP190); Evis Exera Iii Bronchovideoscope (Olympus BF-P190)…Olympus Medical Systems Corp. · 2025 · 4 generations on · ◆ in the 2025 descendants bar of FIG. 14 Class I recalls
  • K201758EVIS EXERA III Bronchovideoscope Olympus BF-XP190, EVIS EXERA III Bronchovideoscope Olympus BF-P190Olympus Medical Systems Corp. · 2021 · 5 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
  • K193517ViziShot 2 FLEXOlympus Surgical Technologies America · 2020 · 6 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K190106Intra-Operative Positioning System; Simple Curve Catheter, Reverse Curve Catheter ; Angled Tip Guidewire…Centerline Biomedical, Inc. · 2019 · 11 generations on · ◆ in the 2019 descendants bar of FIG. 11 Class I recall
  • K242133Intra-Operative Positioning System (IOPS®) (Fiducial Tracking Pad); Intra-Operative Positioning System…Centerline Biomedical, Inc. · 2024 · 12 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K241243Intra-Operative Positioning System (IOPS®) Viewpoint (Simple Curve Catheters); Intra-Operative Positioning…Centerline Biomedical, Inc. · 2024 · 12 generations on · ◆ in the 2024 descendants bar of FIG. 11 Class I recall
  • K230309Intra-Operative Positioning System (IOPS®) (MC-1); IOPS Simple Curve Catheter (SCC-1) and IOPS Reverse Curve…Centerline Biomedical, Inc. · 2023 · 12 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall
  • K243842Intra-Operative Positioning System (IOPS®)Centerline Biomedical, Inc. · 2025 · 12 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 6.4 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K031648 (Olympus Optical Co., Ltd., 2003) has gone furthest, with 159 descendants of its own.
  • 1 other device also cites its predicate K910423.

How does its category compare?

All EOQ devices

EOQ Bronchoscope (Flexible Or Rigid): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · EOQ

20in 2025

2007: 22008: 22009: 32012: 22013: 42014: 32015: 62016: 32017: 42018: 72019: 112020: 102021: 152022: 112023: 122024: 132025: 202026 (partial year): 9
2007high 20 (2025) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · EOQ

147in 2025

2009: 24 days median2013: 178 days median2014: 302 days median2015: 206 days median2016: 262 days median2017: 136 days median2018: 212 days median2019: 147 days median2020: 103 days median2021: 169 days median2022: 134 days median2023: 160 days median2024: 130 days median2025: 147 days median2026 (partial year): 206 days median
2007high 302 (2014) · low 24 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · EOQ

25in 2025

2011: 12012: 42016: 12019: 12020: 12021: 92022: 52023: 18, 4 Class I2024: 42025: 25, 24 Class I
2007high 25 (2025) · low 0 (2007)2026

Class I (28 since 2007) Class II/III not classified

Adverse event reports per year · EOQ

6,546in 2025

2007: 392008: 872009: 1142010: 1002011: 962012: 932013: 932014: 982015: 1322016: 2972017: 3372018: 2792019: 6542020: 5152021: 1,3582022: 2,5412023: 5,0642024: 6,0512025: 6,546
2007high 6,546 (2025) · low 39 (2007)2025

MAUDE reports naming product code EOQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—039
20082—087
20093240114
20100—0100
20110—196
20122—493
20134178093
20143302098
201562060132
201632621297
201741360337
201872120279
2019111471654
2020101031515
20211516991,358
20221113452,541
20231216018 (4)5,064
20241313046,051
20252014725 (24)6,546
202692060—

What is it indicated for?

< BF-40 Series> These instruments have been designed to be used with an Olympus Light Source, documentation equipment, display monitor, Endo-Therapy Accessories and other ancillary equipment for endoscopic diagnostic and treatment within the airways and tracheobronchial tree. <BF-240 Series> This instrument has been designed to be used with an Olympus Video System Center, Light Source, Documentation Equipment, Display monitor, Endo-Therapy Accessories…

Read the full statement (122 words)

< BF-40 Series>

These instruments have been designed to be used with an Olympus Light Source, documentation equipment, display monitor, Endo-Therapy Accessories and other ancillary equipment for endoscopic diagnostic and treatment within the airways and tracheobronchial tree.

<BF-240 Series>

This instrument has been designed to be used with an Olympus Video System Center, Light Source, Documentation Equipment, Display monitor, Endo-Therapy Accessories, Electrosurgical Unit, and other ancillary equipment for endoscopic diagnosis and treatment within the airways and tracheobronchial tree.

<BF-160 Series>

These instrument have been designed to be used with an OLYMPUS video system center, light source, documentation equipment, video monitor, endo-therapy accessories (such as a biopsy forceps) and other ancillary equipment for endoscopy and endoscopic surgery within the airways and tracheobronchial tree.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K023984: Modification to Olympus Bronchoscopes BF-40 Series, BF-240 Series, and BF-160 Series.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K023984

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026