510(k) · Cleared
ClearedMR DWI/FLAIR Measurement V1.0
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Olea Medical · LLZ System, Image Processing, Radiological
Cleared on Apr 26, 2023 as substantially equivalent to K152602 (Olea Medical, 2016) after a 57-day review. Its family tree reaches back 12 generations to 1997. 3 of the 49 devices in its family, counting itself, have a Class I recall.
- Decided
- Apr 26, 2023
- Review
- 57 days
- Class
- Class II
- Product code
- LLZ
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 12 generations to K971684 (1997). 48 ancestors in all, the oldest K896095 (1990). No later device cites it as a predicate yet. 3 of the 49 devices in this family, counting this one, have a Class I recall.
- Root · 12 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
- 8 generations between
- 3 generations backK121434 Software Syngo MR D13A for the Magnetom Systems…Siemens Medi Cal Solutions, Inc. · 2012 · Recalled (Class I)
- 2 generations backK130749 Syngo.MR General; Syngo.MR Cardiology; Syngo.MR VascularSiemens Medical Solutions USA, Inc. · 2013
- 1 generation backK152602 Olea Sphere V3.0Olea Medical · 2016
- This deviceK230552 MR DWI/FLAIR Measurement V1.0Olea Medical · 2023
Show all 12 generations
- Root · 12 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
- 11 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
- 10 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
- 9 generations backK052164 Magnetom Systems with Software Syngo MR 2006ASiemens Medical Solutions USA, Inc. · 2005
- 8 generations backK062454 Magnetom Systems - I-Class and T-Class ReleasesSiemens Medi Cal Solutions, Inc. · 2006
- 7 generations backK072237 Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · Recalled (Class I)
- 6 generations backK082427 Syngo MR B17Siemens Medical Solutions USA, Inc. · 2008
- 5 generations backK101347 Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · Recalled (Class I)
- 4 generations backK111242 Magnetom Aera and Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2011 · Recalled
- 3 generations backK121434 Software Syngo MR D13A for the Magnetom Systems…Siemens Medi Cal Solutions, Inc. · 2012 · Recalled (Class I)
- 2 generations backK130749 Syngo.MR General; Syngo.MR Cardiology; Syngo.MR VascularSiemens Medical Solutions USA, Inc. · 2013
- 1 generation backK152602 Olea Sphere V3.0Olea Medical · 2016
- This deviceK230552 MR DWI/FLAIR Measurement V1.0Olea Medical · 2023
Class I recalls in this family
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −12 | K971684 | Magnetom Project 047 System (Numaris 3.5 Software) | Siemens Medical Solutions USA, Inc. | 1997 | 4 |
| −11 | K032428 | Magnetom Avanto MR System | Siemens Medical Solutions USA, Inc. | 2003 | 4 |
| −10 | K041112 | Magnetom Espree System | Siemens Medical Solutions USA, Inc. | 2004 | 4 |
| −9 | K052164 | Magnetom Systems with Software Syngo MR 2006A | Siemens Medical Solutions USA, Inc. | 2005 | — |
| −8 | K062454 | Magnetom Systems - I-Class and T-Class Releases | Siemens Medi Cal Solutions, Inc. | 2006 | — |
| −7 | K072237 | Magnetom Verio | Siemens Medical Solutions USA, Inc. | 2007 | Class I |
| −6 | K082427 | Syngo MR B17 | Siemens Medical Solutions USA, Inc. | 2008 | — |
| −5 | K101347 | Magnetom Aera; Magnetom Skyra | Siemens Medical Solutions USA, Inc. | 2010 | Class I |
| −4 | K111242 | Magnetom Aera and Magnetom Skyra | Siemens Medical Solutions USA, Inc. | 2011 | 3 |
| −3 | K121434 | Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/Verio | Siemens Medi Cal Solutions, Inc. | 2012 | Class I |
| −2 | K130749 | Syngo.MR General; Syngo.MR Cardiology; Syngo.MR Vascular | Siemens Medical Solutions USA, Inc. | 2013 | — |
| −1 | K152602 | Olea Sphere V3.0 | Olea Medical | 2016 | — |
| 0 (this device) | K230552 | MR DWI/FLAIR Measurement V1.0 | Olea Medical | 2023 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1990 | 1 | 0 |
| 1991 | 1 | 0 |
| 1994 | 1 | 0 |
| 1996 | 1 | 0 |
| 1998 | 2 | 0 |
| 1999 | 2 | 0 |
| 2000 | 2 | 0 |
| 2001 | 4 | 0 |
| 2002 | 3 | 0 |
| 2004 | 1 | 0 |
| 2005 | 3 | 0 |
| 2006 | 1 | 0 |
| 2007 | 2 | 0 |
| 2008 | 1 | 0 |
| 2009 | 3 | 0 |
| 2010 | 1 | 0 |
| 2011 | 3 | 0 |
| 2012 | 2 | 0 |
| 2013 | 1 | 0 |
| 2014 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LLZ device
- #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K152602 | Olea Sphere V3.0Summary page 4…system (PACS) Regulation Name: Medical Image Management and Processing System Regulatory Class: 21 CFR 892.2050 Product Code: LLZ Predicate Device: Olea Sphere V3.0 (K152602) Product name Olea Sphere V3.0 510(k) number K152602 Trend name Olea Sphere V3.0 Classificati… | 2016 |
| † | K223502 | MR Diffusion Perfusion Mismatch V1.0 | 2023 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (3 Class I)
No FDA recall names this device.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K121434Software Syngo MR D13A for the Magnetom Systems Aera/Skyra/Avanto/VerioSiemens Medi Cal Solutions, Inc. · 2012 · 2 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
- K101347Magnetom Aera; Magnetom SkyraSiemens Medical Solutions USA, Inc. · 2010 · 4 generations back · ◆ on the line of descent in FIG. 12 Class I recalls
- K072237Magnetom VerioSiemens Medical Solutions USA, Inc. · 2007 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 26 years and passes through 3 companies; a new generation was cleared every 1.1 years on median.
- 18 of the 48 ancestors have a recall of any class; 3 have a Class I recall.
- 10 other devices also cite its predicate K152602.
How does its category compare?
All LLZ devicesLLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.
FDA decisions per year · LLZ
58in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LLZ
173in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LLZ
11in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LLZ
116in 2025
MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 76 | 36 | 20 | 152 |
| 2008 | 94 | 52 | 32 | 154 |
| 2009 | 70 | 62 | 21 | 145 |
| 2010 | 51 | 98 | 17 | 145 |
| 2011 | 106 | 79 | 30 | 454 |
| 2012 | 82 | 87 | 12 | 266 |
| 2013 | 95 | 84 | 64 | 1,130 |
| 2014 | 78 | 89 | 6 | 15,631 |
| 2015 | 64 | 90 | 16 | 211 |
| 2016 | 82 | 90 | 12 | 290 |
| 2017 | 78 | 107 | 5 | 104 |
| 2018 | 104 | 105 | 7 | 72 |
| 2019 | 74 | 107 | 2 | 88 |
| 2020 | 81 | 115 | 6 | 82 |
| 2021 | 83 | 116 | 3 | 70 |
| 2022 | 65 | 137 | 4 | 28 |
| 2023 | 87 | 157 | 8 | 26 |
| 2024 | 62 | 131 | 6 | 42 |
| 2025 | 58 | 173 | 11 | 116 |
| 2026 | 45 | 110 | 14 | — |
What is it indicated for?
MR DWI/FLAIR Measurement V1.0 is an image processing application indicated for use in the analysis of: (1) MR Diffusion-weighted imaging (DWI) (2) MR FLAIR images. The device is intended to be used by trained professionals with medical imaging education including, but not limited to, physicians and medical technicians in the imaging assessment workflow: • computation of the map relative to the water diffusion, i.e., ADC map; • extraction and…
Read the full statement (224 words)
MR DWI/FLAIR Measurement V1.0 is an image processing application indicated for use in the analysis of:
(1) MR Diffusion-weighted imaging (DWI)
(2) MR FLAIR images.
The device is intended to be used by trained professionals with medical imaging education including, but not limited to, physicians and medical technicians in the imaging assessment workflow:
• computation of the map relative to the water diffusion, i.e., ADC map;
• extraction and communication of metrics derived from the above map, i.e., hypointense area on ADC, and FLAIR series as well as ratios with contralateral information on FLAIR images.
The results of MR DWI/FLAIR Measurement V1.0 are intended to be used in conjunction with other patient information and, based on professional judgment, to assist the clinician in the medical imaging assessment. Trained professionals are responsible for viewing the full set of native images per the standard of care.
The device does not alter the original medical image. MR DWI/FLAIR Measurement V1.0 is not intended to be used as a standalone diagnostic device and shall not be used to take decisions with diagnosis or therapeutic purposes. Patient management decisions should not solely be based on MR DWI/FLAIR Measurement V1.0 results.
MR DWI/FLAIR Measurement V1.0 can be integrated and deployed through technical platforms, responsible for transferring, storing, converting formats, notifying of detected image variations and display of DICOM imaging data.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K230552 · printed Nov 3, 2026
Cite this page
510(k) Report. “K230552: MR DWI/FLAIR Measurement V1.0.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K230552
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026