510(k) Report

510(k) · Cleared

Cleared

SURGIVISIO Device

K231886510(k) SpecialCleared Jul 27, 202330-day reviewClass II21 CFR 882.4560

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ecential Robotics · OLO Orthopedic Stereotaxic Instrument

Cleared on Jul 27, 2023 as substantially equivalent to K221028 (Ecential Robotics, 2022) and 3 other predicates after a 30-day review. Its family tree reaches back 9 generations to 2000. 6 later devices descend from it.

Decided
Jul 27, 2023
Review
30 days
Class
Class II
Product code
OLO
Recalls
None
Predicates
4

Its longest line of descent runs back 9 generations to K002053 (2000). 39 ancestors in all, the oldest K811452 (1981). 6 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2022: 2 other ancestors2021: 1 other ancestor2014: 2 other ancestors2004: 5 other ancestors2013: 1 other ancestor2003: 2 other ancestors2002: 1 other ancestor2001: 3 other ancestors1997: 2 other ancestors2006: 1 other ancestor2000: 4 other ancestors2005: 2 other ancestors1999: 2 other ancestors1987: 1 other ancestor1981: 1 other ancestor2025: 2 descendants cleared2024: 1 descendant cleared2026: 3 descendants clearedK002053 · Navitrak System-Optical Option, Model 900.004 · Orthosoft, Inc. · 2000K021760 · Navitrack System-Optical TKR CT-Less, Model 900.120 · Orthosoft, Inc. · 2002K043536 · Navitrack System - S&N Image Free Knee · Orthosoft, Inc. · 2005K060556 · Brigit Surgical Device · Zimmer, Inc. · 2006K092239 · Rosa Surgical Device, Model Rosa 1.1 · Medtech S.A · 2009 · recalledK101791 · Rosa Surgical Device · Medtech S.A.S · 2010 · recalledK151511 · ROSA Spine · Medtech S.A · 2016 · recalledK182848 · ROSA ONE Spine application · Medtech S.A · 2019K221028 · SURGIVISIO Device · Ecential Robotics · 202219801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 2 other ancestors3 generations back: 6 other ancestors4 generations back: 3 other ancestors5 generations back: 6 other ancestors6 generations back: 3 other ancestors7 generations back: 5 other ancestors8 generations back: 1 other ancestor9 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 1 descendant K221028 · SURGIVISIO Device · Ecential Robotics · 2022K182848 · ROSA ONE Spine application · Medtech S.A · 2019K151511 · ROSA Spine · Medtech S.A · 2016 · recalledK101791 · Rosa Surgical Device · Medtech S.A.S · 2010 · recalledK092239 · Rosa Surgical Device, Model Rosa 1.1 · Medtech S.A · 2009 · recalledK060556 · Brigit Surgical Device · Zimmer, Inc. · 2006K043536 · Navitrack System - S&N Image Free Knee · Orthosoft, Inc. · 2005K021760 · Navitrack System-Optical TKR CT-Less, Model 900.120 · Orthosoft, Inc. · 2002K002053 · Navitrak System-Optical Option, Model 900.004 · Orthosoft, Inc. · 2000−50
  1. Root · 9 generations backK002053 Navitrak System-Optical Option, Model 900.004Orthosoft, Inc. · 2000
  2. 5 generations between
  3. 3 generations backK151511 ROSA SpineMedtech S.A · 2016 · Recalled
  4. 2 generations backK182848 ROSA ONE Spine applicationMedtech S.A · 2019
  5. 1 generation backK221028 SURGIVISIO DeviceEcential Robotics · 2022
  6. This deviceK231886 SURGIVISIO DeviceEcential Robotics · 2023
  7. Descendant · cites it directly · 3 of its ownK242890 SPINEART Navigation Instrument SystemSpineart SA · 2025
  8. Descendant · 2 generations on · 2 of its ownK242933 SPINEART Navigation Instrument SystemSpineart SA · 2025
  9. Descendant · cites it directly · 2 of its ownK233228 Spine Navigation and Robotic-Assistance DeviceEcential Robotics · 2024 · Recalled
  10. Descendant · 2 generations on · 1 of its ownK261113 Op.n™ Spine Navigation and Robotic GuidanceEcential Robotics · 2026
  11. Descendant · 3 generations on · 1 of its ownK254158 SPINEART Navigation Instrument SystemSpineart SA · 2026
Show all 9 generations
  1. Root · 9 generations backK002053 Navitrak System-Optical Option, Model 900.004Orthosoft, Inc. · 2000
  2. 8 generations backK021760 Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc. · 2002
  3. 7 generations backK043536 Navitrack System - S&N Image Free KneeOrthosoft, Inc. · 2005
  4. 6 generations backK060556 Brigit Surgical DeviceZimmer, Inc. · 2006
  5. 5 generations backK092239 Rosa Surgical Device, Model Rosa 1.1Medtech S.A · 2009 · Recalled
  6. 4 generations backK101791 Rosa Surgical DeviceMedtech S.A.S · 2010 · Recalled
  7. 3 generations backK151511 ROSA SpineMedtech S.A · 2016 · Recalled
  8. 2 generations backK182848 ROSA ONE Spine applicationMedtech S.A · 2019
  9. 1 generation backK221028 SURGIVISIO DeviceEcential Robotics · 2022
  10. This deviceK231886 SURGIVISIO DeviceEcential Robotics · 2023

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2024: 12025: 22026: 3
2024high 3 (2026) · low 1 (2024)2026

Most influential descendants

Showing 5 of 6 · See all
  • K242890SPINEART Navigation Instrument SystemSpineart SA · 2025 · cites it directly3descendants
  • K242933SPINEART Navigation Instrument SystemSpineart SA · 2025 · 2 generations on2descendants
  • K233228Spine Navigation and Robotic-Assistance DeviceEcential Robotics · 2024 · cites it directly2descendants
  • K261113Op.n™ Spine Navigation and Robotic GuidanceEcential Robotics · 2026 · 2 generations on1descendant
  • K254158SPINEART Navigation Instrument SystemSpineart SA · 2026 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K002053Navitrak System-Optical Option, Model 900.004Orthosoft, Inc.2000—
−8K021760Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc.2002—
−7K043536Navitrack System - S&N Image Free KneeOrthosoft, Inc.2005—
−6K060556Brigit Surgical DeviceZimmer, Inc.2006—
−5K092239Rosa Surgical Device, Model Rosa 1.1Medtech S.A20099
−4K101791Rosa Surgical DeviceMedtech S.A.S201012
−3K151511ROSA SpineMedtech S.A20163
−2K182848ROSA ONE Spine applicationMedtech S.A2019—
−1K221028SURGIVISIO DeviceEcential Robotics2022—
0 (this device)K231886SURGIVISIO DeviceEcential Robotics2023—
+1K242890SPINEART Navigation Instrument SystemSpineart SA2025—
+2K242933SPINEART Navigation Instrument SystemSpineart SA2025—
+1K233228Spine Navigation and Robotic-Assistance DeviceEcential Robotics20241
+2K261113Op.n™ Spine Navigation and Robotic GuidanceEcential Robotics2026—
+3K254158SPINEART Navigation Instrument SystemSpineart SA2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198710
199720
199920
200040
200130
200210
200320
200450
200520
200610
201310
201420
202110
202220

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202410
202520
202630
FIG. 1 — Lineage of K231886, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OLO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OLO device

  1. #1 K240662 Spine Guidance Software (version 5.0); Xia 3/…score 23.7 · 15 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K221028SURGIVISIO Device
Summary page 4
…latory class II Primary product code OLO Legally marketed device SURGIVISIO Device by ECENTIAL ROBOTICS: to which equivalence is - Primary predicate K221028 claimed - Other predicates o K202547 o K2…
Ecential Robotics2022—
*K202547Surgivisio system
Summary page 4
…O Device by ECENTIAL ROBOTICS: to which equivalence is - Primary predicate K221028 claimed - Other predicates o K202547 o K220946 o K220627 Scope of the The subject de…
Surgivisio2021—
*K220946SURGIVISIO Device
Summary page 4
…valence is - Primary predicate K221028 claimed - Other predicates o K202547 o K220946 o K220627 Scope of the The subject device is the SURGIVISIO Device which submissi…
Ecential Robotics2022—
*K220627SURGIVISIO Device
Summary page 4
…laimed - Other predicates o K202547 o K220946 o K220627 Scope of the The subject device is the SURGIVISIO Device which submission enables 2D/3D medical imag…
Ecential Robotics2022—

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K242890SPINEART Navigation Instrument SystemSpineart SA20253
K233228Spine Navigation and Robotic-Assistance DeviceEcential Robotics20242

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 5 companies; a new generation was cleared every 2.4 years on median.
  • 13 of the 39 ancestors have a recall of any class.
  • Of its descendants, K242890 (Spineart SA, 2025) has gone furthest, with 3 descendants of its own.

How does its category compare?

All OLO devices

OLO Orthopedic Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · OLO

40in 2025

2009: 112010: 92011: 112012: 72013: 72014: 102015: 102016: 142017: 332018: 242019: 352020: 342021: 282022: 402023: 452024: 442025: 402026 (partial year): 42
2007high 45 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OLO

119in 2025

2009: 126 days median2010: 113 days median2011: 138 days median2012: 129 days median2013: 137 days median2014: 119 days median2015: 176 days median2016: 120 days median2017: 91 days median2018: 97 days median2019: 95 days median2020: 89 days median2021: 91 days median2022: 89 days median2023: 90 days median2024: 90 days median2025: 119 days median2026 (partial year): 109 days median
2007high 176 (2015) · low 89 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OLO

1in 2025

2010: 12011: 12012: 12014: 32015: 22016: 32017: 22018: 12019: 12020: 62022: 52023: 7, 1 Class I2024: 42025: 12026 (partial year): 6
2007high 7 (2023) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · OLO

3,306in 2025

2008: 12009: 42010: 42011: 102012: 602013: 362014: 2452015: 4662016: 7502017: 1,0762018: 1,2302019: 1,5852020: 2,7462021: 1,6962022: 1,5682023: 1,4932024: 2,3532025: 3,306
2007high 3,306 (2025) · low 0 (2007)2025

MAUDE reports naming product code OLO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—01
20091112604
2010911314
201111138110
20127129160
20137137036
2014101193245
2015101762466
2016141203750
2017339121,076
2018249711,230
2019359511,585
2020348962,746
2021289101,696
2022408951,568
202345907 (1)1,493
2024449042,353
20254011913,306
2026421096—

What is it indicated for?

The SURGIVISIO Device is indicated to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical imaging of anatomical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures include procedures during which the spine, pelvis, or articulation structures are visualized The SURGIVISIO Device through its…

Read the full statement (236 words)

The SURGIVISIO Device is indicated to be used during surgical procedures in which the physician would benefit from

the visualization of 2D medical imaging and/or intraoperatively generated 3D medical imaging of anatomical structures or

objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures include procedures during

which the spine, pelvis, or articulation structures are visualized

The SURGIVISIO Device through its freehand navigation feature is indicated as an intraoperative guidance system to

enable open or percutaneous computer-assisted surgery.

It is indicated for conditions of the spine in which the use of stereotactic surgery may be appropriate, and where reference

to a rigid anatomical bony structure can be identified relative to the intraoperatively generated 3D image of the anatomy

provided by the SURGIVISIO Device.

It is indicated to precisely position the Navigated Targeting Needle SPX1 during general spinal procedures with a

posterior approach.

The SURGIVISIO Device through its robotic guidance feature is indicated for the positioning of instrument holders or tool guides to be used by surgeons to guide the Spine CoBot instruments during general spinal surgery.

Guidance is based on an intra-operative plan developed with three-dimensional imaging software based on intra-operative

3D images provided by the SURGIVISIO Device.

It is indicated for positioning of surgical instruments in vertebrae with a posterior approach in the thoracolumbar region.

The SPX1 Instrument and Spine CoBot instruments are designed to be used with the SURGIVISIO Device.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K231886: SURGIVISIO Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K231886

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026