510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Rosa Surgical Device

K101791510(k) TraditionalCleared Sep 23, 201087-day reviewClass II21 CFR 882.4560Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtech S.A.S · HAW Neurological Stereotaxic Instrument

Cleared on Sep 23, 2010 as substantially equivalent to K092239 (Medtech S.A, 2009) and 3 other predicates after an 87-day review. Its family tree reaches back 5 generations to 2000. 5 of the 146 devices in its family, counting itself, have a Class I recall. 126 later devices descend from it. It has been recalled 12 times.

Decided
Sep 23, 2010
Review
87 days
Class
Class II
Product code
HAW
Recalls
12
Predicates
4

Its longest line of descent runs back 5 generations to K002053 (2000). 19 ancestors in all, the oldest K811452 (1981). 126 later devices descend from it, 4 citing it directly. 5 of the 146 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 3 other ancestors2000: 1 other ancestor1999: 2 other ancestors2001: 2 other ancestors1997: 1 other ancestor2005: 2 other ancestors1987: 1 other ancestor1981: 1 other ancestor2003: 1 other ancestor2017: 2 descendants cleared2015: 2 descendants cleared, 1 with a Class I recall12016: 1 descendant cleared2019: 9 descendants cleared, 1 with a Class I recall12018: 1 descendant cleared, 1 with a Class I recall12021: 8 descendants cleared, 1 with a Class I recall12025: 27 descendants cleared2024: 20 descendants cleared2026: 24 descendants cleared2023: 19 descendants cleared2020: 2 descendants cleared, 1 with a Class I recall12022: 11 descendants clearedK002053 · Navitrak System-Optical Option, Model 900.004 · Orthosoft, Inc. · 2000K021760 · Navitrack System-Optical TKR CT-Less, Model 900.120 · Orthosoft, Inc. · 2002K043536 · Navitrack System - S&N Image Free Knee · Orthosoft, Inc. · 2005K060556 · Brigit Surgical Device · Zimmer, Inc. · 2006K092239 · Rosa Surgical Device, Model Rosa 1.1 · Medtech S.A · 2009 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors2 generations back: 3 other ancestors3 generations back: 6 other ancestors4 generations back: 1 other ancestor5 generations back: 1 other ancestor 1 generation forward: 4 descendants, 1 with a Class I recall12 generations forward: 5 descendants, 1 with a Class I recall13 generations forward: 44 descendants, 2 with a Class I recall24 generations forward: 23 descendants, 1 with a Class I recall15 generations forward: 20 descendants6 generations forward: 16 descendants7 generations forward: 6 descendants8 generations forward: 2 descendants9 generations forward: 2 descendants10 generations forward: 1 descendant11 generations forward: 1 descendant12 generations forward: 1 descendant13 generations forward: 1 descendant K092239 · Rosa Surgical Device, Model Rosa 1.1 · Medtech S.A · 2009 · recalledK060556 · Brigit Surgical Device · Zimmer, Inc. · 2006K043536 · Navitrack System - S&N Image Free Knee · Orthosoft, Inc. · 2005K021760 · Navitrack System-Optical TKR CT-Less, Model 900.120 · Orthosoft, Inc. · 2002K002053 · Navitrak System-Optical Option, Model 900.004 · Orthosoft, Inc. · 2000−50+5+10
  1. Root · 5 generations backK002053 Navitrak System-Optical Option, Model 900.004Orthosoft, Inc. · 2000
  2. 4 generations backK021760 Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc. · 2002
  3. 3 generations backK043536 Navitrack System - S&N Image Free KneeOrthosoft, Inc. · 2005
  4. 2 generations backK060556 Brigit Surgical DeviceZimmer, Inc. · 2006
  5. 1 generation backK092239 Rosa Surgical Device, Model Rosa 1.1Medtech S.A · 2009 · Recalled
  6. This deviceK101791 Rosa Surgical DeviceMedtech S.A.S · 2010
  7. Descendant · cites it directly · 109 of its ownK151511 ROSA SpineMedtech S.A · 2016 · Recalled
  8. Descendant · 2 generations on · 69 of its ownK171651 Excelsius GPSGlobus Medical, Inc. · 2017 · Recalled
  9. Descendant · 3 generations on · 42 of its ownK191100 ExcelsiusGPS Spine 1.1 Interbody ModuleGlobus Medical, Inc. · 2019
  10. Descendant · 2 generations on · 29 of its ownK182964 ROSA Knee SystemOrthosoft Inc (D/B/A Zimmer Cas) · 2019
  11. Descendant · 4 generations on · 23 of its ownK222270 HEDRON Cervical Spacers, HEDRON Lumbar Spacers, SABLE Expandable…Globus Medical, Inc. · 2022

Showing 5 of 126 descendants. See all descendants

Descendants cleared per year

2015: 22016: 12017: 22018: 12019: 92020: 22021: 82022: 112023: 192024: 202025: 272026: 24
2015high 27 (2025) · low 1 (2016)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 126 · See all
  • K151511ROSA SpineMedtech S.A · 2016 · cites it directly109descendants
  • K171651Excelsius GPSGlobus Medical, Inc. · 2017 · 2 generations on69descendants
  • K191100ExcelsiusGPS Spine 1.1 Interbody ModuleGlobus Medical, Inc. · 2019 · 3 generations on42descendants
  • K182964ROSA Knee SystemOrthosoft Inc (D/B/A Zimmer Cas) · 2019 · 2 generations on29descendants
  • K222270HEDRON Cervical Spacers, HEDRON Lumbar Spacers, SABLE Expandable Spacer, ExcelsiusGPS…Globus Medical, Inc. · 2022 · 4 generations on23descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K002053Navitrak System-Optical Option, Model 900.004Orthosoft, Inc.2000—
−4K021760Navitrack System-Optical TKR CT-Less, Model 900.120Orthosoft, Inc.2002—
−3K043536Navitrack System - S&N Image Free KneeOrthosoft, Inc.2005—
−2K060556Brigit Surgical DeviceZimmer, Inc.2006—
−1K092239Rosa Surgical Device, Model Rosa 1.1Medtech S.A20099
0 (this device)K101791Rosa Surgical DeviceMedtech S.A.S2010—
+1K151511ROSA SpineMedtech S.A20163
+2K171651Excelsius GPSGlobus Medical, Inc.20171
+3K191100ExcelsiusGPS Spine 1.1 Interbody ModuleGlobus Medical, Inc.2019—
+2K182964ROSA Knee SystemOrthosoft Inc (D/B/A Zimmer Cas)2019—
+4K222270HEDRON Cervical Spacers, HEDRON Lumbar Spacers, SABLE Expandable Spacer…Globus Medical, Inc.2022—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198710
199710
199920
200010
200120
200310
200430
200520

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201521
201610
201720
201811
201991
202021
202181
2022110
2023190
2024200
2025270
2026240
FIG. 1 — Lineage of K101791, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HAW device

  1. #1 K253856 Geniant Cranial (Navigated Neurosurgical…score 15.0 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K092239Rosa Surgical Device, Model Rosa 1.1
Summary page 1
…HAW), 21 CFR Section 882.4560 classification name 3 Predicate devices ROSA Surgical Device, manufactured by MedTech SAS, K092239, cleared November 17, 2009 StealthStation Treatment Guidance Platform, manufactured by…
Medtech S.A20099
*K001801Steathstation Treatment Guidance Platform
Summary page 1
…09 StealthStation Treatment Guidance Platform, manufactured by Medtronic Surgical Navigation Technologies, K001801, cleared June 30, 2000 VectorVision Cranial / ENT, manufactured by…
Medtronic Surgical Navigation2000—
*K023651Vectorvision Cranial/Ent
Summary page 1
…ared June 30, 2000 VectorVision Cranial / ENT, manufactured by BrainLAB AG, K023651, cleared February 17, 2004 Frameless Neuromate Stereotactic System, manufactured by…
Brainlab AG20041
*K991081Frameless Neuromate
Summary page 1
…ruary 17, 2004 Frameless Neuromate Stereotactic System, manufactured by Integrated Surgical Systems, Inc., K991081, cleared June 25, 1999 4 Description ROSA Surgical Device is a computer con…
Integrated Surgical Systems, Inc.1999—

What descends from it? (126)

See all 126

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K151511ROSA SpineMedtech S.A2016109
K151359ROSA Brain Class IMedtech S.A201520
K172522Mazor XMazor Robotics, Ltd.20170
K143420SYMBIS Surgical SystemImris, Inc.20150

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 8, 2018Class IIZ-2108-2018No Marketing ApplicationReplacement of units lacking an updated device approval.
May 27, 2016Class IIZ-0105-2018Process controlInsufficient amount of washers in the Telescopic Arm, allowing some movement of the device.
Apr 8, 2016Class IIZ-0115-2018Software designPotential for software to change the final tool orientation for the command position without command.
Jan 20, 2016Class IIZ-3117-2017Software designComplaints were received reporting the system would freeze/shut down while in Fulgurate mode.
Nov 6, 2014Class IIZ-0160-2018Device DesignPotential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).
Sep 5, 2014Class IIZ-0184-2018Software designThe software issue described was corrected in the modification to the MXTTOUT controller parameter settings.
Jun 4, 2014Class IIZ-0161-2018Software Design ChangeSoftware corrections reactivating the cooperative endoscopy mode.
Feb 14, 2014Class IIZ-0102-2018Error in labelingAn undetected shift between the information displayed in the navigation software and the actual patient anatomy
Dec 18, 2013Class IIZ-0096-2018Device DesignPossible break in the connector of the Force Sensor.
Feb 10, 2013Class IIZ-0178-2018Software designIssue with optional neuro-endoscopy module detected under specific conditions during internal testing.
Feb 5, 2013Class IIZ-0099-2018Device DesignObserved instability of the device.
Apr 19, 2012Class IIZ-0100-2018Device DesignPotential failure of the optical distance sensor due to cable disconnection.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K151359ROSA BrainMedtech S.A · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 11 Class I recall
  • K172444ROSA BRAIN (v3.0.0.5)Medtech S.A · 2018 · 2 generations on · ◆ in the 2018 descendants bar of FIG. 11 Class I recall
  • K211616ExcelsiusHubGlobus Medical, Inc. · 2021 · 3 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
  • K182417ROSA ONE Brain ApplicationMedtech S.A · 2019 · 3 generations on · ◆ in the 2019 descendants bar of FIG. 11 Class I recall
  • K200511ROSA ONE Brain applicationMedtech S.A · 2020 · 4 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 10 years and passes through 4 companies; a new generation was cleared every 2.1 years on median.
  • 3 of the 19 ancestors have a recall of any class.
  • Of its descendants, K151511 (Medtech S.A, 2016) has gone furthest, with 109 descendants of its own.

How does its category compare?

All HAW devices

HAW Neurological Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · HAW

9in 2025

2007: 152008: 122009: 82010: 62011: 42012: 82013: 32014: 52015: 102016: 92017: 72018: 62019: 102020: 52021: 92022: 72023: 142024: 72025: 92026 (partial year): 5
2007high 15 (2007) · low 3 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HAW

163in 2025

2007: 107 days median2008: 112 days median2009: 221 days median2010: 138 days median2011: 66 days median2012: 178 days median2013: 172 days median2014: 238 days median2015: 232 days median2016: 175 days median2017: 226 days median2018: 112 days median2019: 147 days median2020: 88 days median2021: 168 days median2022: 128 days median2023: 142 days median2024: 107 days median2025: 163 days median2026 (partial year): 162 days median
2007high 238 (2014) · low 66 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HAW

2in 2025

2007: 22008: 102009: 62010: 282011: 52012: 42013: 6, 1 Class I2014: 52015: 32016: 52017: 22018: 2, 2 Class I2019: 4, 1 Class I2020: 12021: 4, 2 Class I2022: 12023: 4, 1 Class I2024: 2, 2 Class I2025: 22026 (partial year): 6
2007high 28 (2010) · low 1 (2020)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · HAW

3,425in 2025

2007: 492008: 532009: 962010: 2992011: 1,7252012: 9862013: 1,2902014: 1,6062015: 1,7672016: 1,8282017: 3,5052018: 4,4392019: 6,0622020: 3,4802021: 2,8272022: 1,9292023: 2,5102024: 3,7552025: 3,425
2007high 6,062 (2019) · low 49 (2007)2025

MAUDE reports naming product code HAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715107249
2008121121053
20098221696
2010613828299
201146651,725
201281784986
201331726 (1)1,290
2014523851,606
20151023231,767
2016917551,828
2017722623,505
201861122 (2)4,439
2019101474 (1)6,062
202058813,480
202191684 (2)2,827
2022712811,929
2023141424 (1)2,510
202471072 (2)3,755
2025916323,425
202651626—

What is it indicated for?

ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide.

Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fidlucial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments.

It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101791: Rosa Surgical Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101791

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026