510(k) Report

510(k) · Cleared

Cleared

PRIMA Humeral System; PRIMA TT Glenoid

K233712510(k) TraditionalCleared Jan 11, 202452-day reviewClass II21 CFR 888.3670

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Lima Corporate S.P.A. · MBF Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented

Cleared on Jan 11, 2024 as substantially equivalent to K222427 (Lima Corporate S.P.A., 2022) and 2 other predicates after a 52-day review. Its family tree reaches back 19 generations to 1991. 1 of the 260 devices in its family, counting itself, has a Class I recall. 1 later device descends from it.

Decided
Jan 11, 2024
Review
52 days
Class
Class II
Product code
MBF
Recalls
None
Predicates
3

Its longest line of descent runs back 19 generations to K911686 (1991). 258 ancestors in all, the oldest K790987 (1979). 1 later device descends from it, 1 citing it directly. 1 of the 260 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2022: 2 other ancestors2017: 3 other ancestors2019: 4 other ancestors2016: 3 other ancestors2014: 7 other ancestors2011: 12 other ancestors2010: 8 other ancestors2021: 1 other ancestor2013: 5 other ancestors2012: 8 other ancestors2009: 3 other ancestors2008: 11 other ancestors2006: 12 other ancestors2005: 12 other ancestors2004: 10 other ancestors1995: 11 other ancestors2020: 1 other ancestor2015: 1 other ancestor2007: 11 other ancestors2003: 12 other ancestors2001: 9 other ancestors2000: 14 other ancestors1999: 7 other ancestors1998: 10 other ancestors2018: 1 other ancestor2002: 9 other ancestors1997: 5 other ancestors1996: 14 other ancestors1993: 4 other ancestors1992: 7 other ancestors1989: 1 other ancestor1987: 5 other ancestors1986: 1 other ancestor1980: 2 other ancestors1979: 1 other ancestor1994: 7 other ancestors1991: 3 other ancestors1988: 1 other ancestor1981: 1 other ancestor2025: 1 descendant clearedK911686 · Global Total Shoulder, Uncement Humeral Hemi-Shoul · Depuy, Inc. · 1991×2: 2 generations cleared close together, K984541 (1999), K992065 (1999) ×3: 3 generations cleared close together, K021478 (2003), K030941 (2004), K041873 (2004) ×2: 2 generations cleared close together, K052906 (2005), K060704 (2006) K063569 · Equinoxe Reverse Shoulder System · Exactech, Inc. · 2007 · recalled×2: 2 generations cleared close together, K073688 (2008), K080642 (2008) ×2: 2 generations cleared close together, K110599 (2011), K112615 (2011) K122698 · Aequalis Ascend Flex Shoulder System · Tornier S.A.S. · 2013 · recalledK132285 · Aequalis Reversed Shoulder Prosthesis · Tornier S.A.S. · 2013K151293 · Aequalis Ascend Flex Shoulder System, Aequalis Reversed Shoulder Prosthesis · Tornier S.A.S. · 2015K190521 · Aequalis Ascend Flex Shoulder System · Tornier S.A.S. · 2019K201315 · PERFORM Humeral System – Stem · Tornier, Inc. · 2020 · recalledK212800 · PRIMA Humeral System and SMR Glenosphere Ø42 · Lima Corporate S.P.A. · 2022 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 11 other ancestors3 generations back: 23 other ancestors4 generations back: 30 other ancestors5 generations back: 51 other ancestors6 generations back: 42 other ancestors7 generations back: 32 other ancestors8 generations back: 22 other ancestors9 generations back: 17 other ancestors10 generations back: 9 other ancestors 1 generation forward: 1 descendant K212800 · PRIMA Humeral System and SMR Glenosphere Ø42 · Lima Corporate S.P.A. · 2022 · recalledK201315 · PERFORM Humeral System – Stem · Tornier, Inc. · 2020 · recalledK190521 · Aequalis Ascend Flex Shoulder System · Tornier S.A.S. · 2019K151293 · Aequalis Ascend Flex Shoulder System, Aequalis Reversed Shoulder Prosthesis · Tornier S.A.S. · 2015K132285 · Aequalis Reversed Shoulder Prosthesis · Tornier S.A.S. · 2013K122698 · Aequalis Ascend Flex Shoulder System · Tornier S.A.S. · 2013 · recalledK112615 · Aequalis Ascend Modular Reverse Shoulder System · Tornier, Inc. · 2011K110599 · Aequalis Ascend Modular Reverse Shoulder System · Tornier, Inc. · 2011K080642 · Comprehensive Reverse Shoulder · Biomet, Inc. · 2008 · Class I recallK073688 · Equinoxe Reverse Shoulder System Fracture Humeral Adapter Tray and Locking Screw · Exactech, Inc. · 2008 · recalledK063569 · Equinoxe Reverse Shoulder System · Exactech, Inc. · 2007 · recalledK060704 · Zimmer Trabecular Metal Reverse Shoulder System, Sizes 8MM and 10MM · Zimmer, Inc. · 2006 · recalledK052906 · Zimmer Trabecular Metal Reverse Shoulder System · Zimmer, Inc. · 2005 · recalledK041873 · Aequalis Reversed Shoulder Prosthesis · Tornier · 2004K030941 · Aequalis Reversed Shoulder Prosthesis · Tornier · 2004K021478 · Delta Shoulder · Depuy, Inc. · 2003K992065 · Global Advantage Shoulder, Global Advantage Humeral Stem, Global Advantage Eccentric Head · DePuy Orthopaedics, Inc. · 1999K984541 · Global FX Humeral Stem, Global Advantage Humeral Head · DePuy Orthopaedics, Inc. · 1999K911686 · Global Total Shoulder, Uncement Humeral Hemi-Shoul · Depuy, Inc. · 1991−15−10−50
  1. Root · 19 generations backK911686 Global Total Shoulder, Uncement Humeral Hemi-ShoulDepuy, Inc. · 1991
  2. 15 generations between
  3. 3 generations backK190521 Aequalis Ascend Flex Shoulder SystemTornier S.A.S. · 2019
  4. 2 generations backK201315 PERFORM Humeral System – StemTornier, Inc. · 2020 · Recalled
  5. 1 generation backK212800 PRIMA Humeral System and SMR Glenosphere Ø42Lima Corporate S.P.A. · 2022 · Recalled
  6. This deviceK233712 PRIMA Humeral System; PRIMA TT GlenoidLima Corporate S.P.A. · 2024
  7. Descendant · cites it directly · 0 of its ownK251184 AltiVate Reverse Shoulder System; RSP Shoulder System; SMR Shoulder…Encore Medical L.P. · 2025
Show all 19 generations
  1. Root · 19 generations backK911686 Global Total Shoulder, Uncement Humeral Hemi-ShoulDepuy, Inc. · 1991
  2. 18 generations backK984541 Global FX Humeral Stem, Global Advantage Humeral HeadDePuy Orthopaedics, Inc. · 1999
  3. 17 generations backK992065 Global Advantage Shoulder, Global Advantage Humeral Stem, Global…DePuy Orthopaedics, Inc. · 1999
  4. 16 generations backK021478 Delta ShoulderDepuy, Inc. · 2003
  5. 15 generations backK030941 Aequalis Reversed Shoulder ProsthesisTornier · 2004
  6. 14 generations backK041873 Aequalis Reversed Shoulder ProsthesisTornier · 2004
  7. 13 generations backK052906 Zimmer Trabecular Metal Reverse Shoulder SystemZimmer, Inc. · 2005 · Recalled
  8. 12 generations backK060704 Zimmer Trabecular Metal Reverse Shoulder System, Sizes 8MM and 10MMZimmer, Inc. · 2006 · Recalled
  9. 11 generations backK063569 Equinoxe Reverse Shoulder SystemExactech, Inc. · 2007 · Recalled
  10. 10 generations backK073688 Equinoxe Reverse Shoulder System Fracture Humeral Adapter Tray and…Exactech, Inc. · 2008 · Recalled
  11. 9 generations backK080642 Comprehensive Reverse ShoulderBiomet, Inc. · 2008 · Recalled (Class I)
  12. 8 generations backK110599 Aequalis Ascend Modular Reverse Shoulder SystemTornier, Inc. · 2011
  13. 7 generations backK112615 Aequalis Ascend Modular Reverse Shoulder SystemTornier, Inc. · 2011
  14. 6 generations backK122698 Aequalis Ascend Flex Shoulder SystemTornier S.A.S. · 2013 · Recalled
  15. 5 generations backK132285 Aequalis Reversed Shoulder ProsthesisTornier S.A.S. · 2013
  16. 4 generations backK151293 Aequalis Ascend Flex Shoulder System, Aequalis Reversed Shoulder…Tornier S.A.S. · 2015
  17. 3 generations backK190521 Aequalis Ascend Flex Shoulder SystemTornier S.A.S. · 2019
  18. 2 generations backK201315 PERFORM Humeral System – StemTornier, Inc. · 2020 · Recalled
  19. 1 generation backK212800 PRIMA Humeral System and SMR Glenosphere Ø42Lima Corporate S.P.A. · 2022 · Recalled
  20. This deviceK233712 PRIMA Humeral System; PRIMA TT GlenoidLima Corporate S.P.A. · 2024

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K251184AltiVate Reverse Shoulder System; RSP Shoulder System; SMR Shoulder System; PRIMA…Encore Medical L.P. · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−19K911686Global Total Shoulder, Uncement Humeral Hemi-ShoulDepuy, Inc.1991—
−18K984541Global FX Humeral Stem, Global Advantage Humeral HeadDePuy Orthopaedics, Inc.1999—
−17K992065Global Advantage Shoulder, Global Advantage Humeral Stem, Global Advantage…DePuy Orthopaedics, Inc.1999—
−16K021478Delta ShoulderDepuy, Inc.2003—
−15K030941Aequalis Reversed Shoulder ProsthesisTornier2004—
−14K041873Aequalis Reversed Shoulder ProsthesisTornier2004—
−13K052906Zimmer Trabecular Metal Reverse Shoulder SystemZimmer, Inc.200512
−12K060704Zimmer Trabecular Metal Reverse Shoulder System, Sizes 8MM and 10MMZimmer, Inc.200610
−11K063569Equinoxe Reverse Shoulder SystemExactech, Inc.200713
−10K073688Equinoxe Reverse Shoulder System Fracture Humeral Adapter Tray and Locking ScrewExactech, Inc.20081
−9K080642Comprehensive Reverse ShoulderBiomet, Inc.2008Class I
−8K110599Aequalis Ascend Modular Reverse Shoulder SystemTornier, Inc.2011—
−7K112615Aequalis Ascend Modular Reverse Shoulder SystemTornier, Inc.2011—
−6K122698Aequalis Ascend Flex Shoulder SystemTornier S.A.S.20132
−5K132285Aequalis Reversed Shoulder ProsthesisTornier S.A.S.2013—
−4K151293Aequalis Ascend Flex Shoulder System, Aequalis Reversed Shoulder ProsthesisTornier S.A.S.2015—
−3K190521Aequalis Ascend Flex Shoulder SystemTornier S.A.S.2019—
−2K201315PERFORM Humeral System – StemTornier, Inc.20201
−1K212800PRIMA Humeral System and SMR Glenosphere Ø42Lima Corporate S.P.A.20222
0 (this device)K233712PRIMA Humeral System; PRIMA TT GlenoidLima Corporate S.P.A.2024—
+1K251184AltiVate Reverse Shoulder System; RSP Shoulder System; SMR Shoulder System…Encore Medical L.P.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198020
198110
198610
198750
198810
198910
199130
199270
199340
199470
1995110
1996140
199750
1998100
199970
2000140
200190
200290
2003120
2004100
2005120
2006120
2007110
2008110
200930
201080
2011120
201280
201350
201470
201510
201630
201730
201810
201940
202010
202110
202220

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K233712, 1991–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MBF device

  1. #1 K202716 Ignite Anatomic Shoulder Systemscore 20.3 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K222427PRIMA TT Glenoid
Summary page 6
…tical and Cancellous Screws L 20mm Description: This 510(k) submission aims at introducing new components to the previously cleared PRIMA Humeral (K212800) and Glenoid (K222427) systems. The PRIMA Short Stem Plus is introduced as part of the PRIMA Humeral System. As the already cleared PRIMA Short Stem, t…
Lima Corporate S.P.A.20223
*K212800PRIMA Humeral System and SMR Glenosphere Ø42
Summary page 6
…thesis Compressive Cortical and Cancellous Screws L 20mm Description: This 510(k) submission aims at introducing new components to the previously cleared PRIMA Humeral (K212800) and Glenoid (K222427) systems. The PRIMA Short Stem Plus is introduced as part of the PRIMA Humeral System. As the already clear…
Lima Corporate S.P.A.20222
*K172351AltiVate Reverse Humeral Stem, AltiVate Reverse Small Spacer, Altivate Reverse, Small…
Summary page 8
….p.A PRIMA Humeral System K212800 3 DJO Altivate Shoulder Prostheses Stem K172351 Summary of technology comparison: The intended use, design, and materials of the PRIMA Humeral System and PRIMA TT Glenoid are s…
Encore Medical L.P.20173

Also cleared against K222427: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K251184AltiVate Reverse Shoulder System; RSP Shoulder System; SMR Shoulder System; PRIMA…Encore Medical L.P.20250

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K080642Comprehensive Reverse ShoulderBiomet, Inc. · 2008 · 5 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 33 years and passes through 9 companies; a new generation was cleared every 1.2 years on median.
  • 87 of the 258 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K251184 (Encore Medical L.P., 2025) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K222427.

How does its category compare?

All MBF devices

MBF Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · MBF

2in 2025

2007: 22008: 12009: 12012: 22013: 12014: 12017: 22018: 12019: 22020: 22021: 12022: 42023: 42024: 22025: 22026 (partial year): 1
2007high 4 (2022) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MBF

144in 2023

2022: 164 days median2023: 144 days median
2007high 164 (2022) · low 144 (2023)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MBF

1in 2023

2009: 12017: 12018: 12019: 22022: 82023: 1
2007high 8 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MBF

255in 2025

2007: 22008: 72009: 162010: 242011: 322012: 302013: 402014: 282015: 492016: 852017: 1932018: 992019: 472020: 242021: 362022: 1142023: 1432024: 1642025: 255
2007high 255 (2025) · low 2 (2007)2025

MAUDE reports naming product code MBF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—02
20081—07
20091—116
20100—024
20110—032
20122—030
20131—040
20141—028
20150—049
20160—085
20172—1193
20181—199
20192—247
20202—024
20211—036
202241648114
202341441143
20242—0164
20252—0255
20261—0—

What is it indicated for?

The PRIMA Humeral System is intended for partial or total, primary or revision, shoulder joint replacement in skeletally mature patients. The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The PRIMA Short Stem and PRIMA Short Stem Plus are intended for use in cementless and cemented applications, at the discretion of the surgeon. The PRIMA…

Read the full statement (401 words)

The PRIMA Humeral System is intended for partial or total, primary or revision, shoulder joint replacement in skeletally mature patients. The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device. The PRIMA Short Stem and PRIMA Short Stem Plus are intended for use in cementless and cemented applications, at the discretion of the surgeon.

The PRIMA Anatomic implant is indicated for partial or total, primary or revision shoulder joint replacement, in patients suffering from pain and disability due to:

• Non-inflammatory degenerative joint disease (i.e. osteoarthritis),

• Inflammatory arthritis of the glenohumeral join including rheumatoid arthritis,

• Avascular necrosis of the humeral head,

• Traumatic/post-traumatic arthritis,

• Fractures of the humeral head where adequate fixation can be achieved and adequate bone stock remains,

• Post-fracture deformity with intact rotator cuff, where adequate fixation can be achieved and adequate bone stock remains,

• Cuff tear arthropathy (CTA Heads only).

The PRIMA Reverse implant is indicated for primary reverse total shoulder replacement or for revision when converting an anatomic PRIMA arthroplasty to a reverse total shoulder arthroplasty (i.e. in case of cuff tear arthropathy or in a grossly rotator cuff deficiency joint with severe arthropathy).

Revision surgery with retention of the PRIMA Short Stem and PRIMA Short Stem Plus are intended as conversion surgery from anatomic to reverse, where the stem is stable, well positioned and tissue integrated. Other revisions of the humeral prosthesis part should be treated with traditional shoulder prostheses.

The PRIMA reverse implant is indicated for patients suffering from pain and disability due to:

• Rotator cuff tear arthropathy,

• Osteoarthritis with rotator cuff tear,

• Rheumatoid arthritis with rotator cuff tear,

• Massive irreparable rotator cuff tear,

• Avascular necrosis of the humeral head,

• Correction of functional deformity, where adequate fixation can be achieved and adequate bone stock remains,

• Fractures of the humeral head where adequate fixation can be achieved and adequate bone stock remains.

The PRIMA Glenoid System is indicated for primary, fracture or revision total shoulder replacement in a grossly rotator cuff deficient joint with severe arthropathy (disabled shoulder). The patient's joint must be anatomically and structurally suited to receive the selected implants and a functional deltoid muscle is necessary to use the device.

The PRIMA Glenoid System components are intended for uncemented use with the addition of screw fixation.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K233712: PRIMA Humeral System; PRIMA TT Glenoid.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K233712

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026