510(k) Report

510(k) · Cleared

Cleared

LINQ II™ Insertable Cardiac Monitor (ICM)

K240693510(k) SpecialCleared Mar 28, 202415-day reviewClass II21 CFR 870.1025

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic, Inc. · MXD Recorder, Event, Implantable Cardiac, (With Arrhythmia Detection)

Cleared on Mar 28, 2024 as substantially equivalent to K233320 (Medtronic, Inc., 2023) after a 15-day review. Its family tree reaches back 16 generations to 1987. 1 later device descends from it.

Decided
Mar 28, 2024
Review
15 days
Class
Class II
Product code
MXD
Recalls
None
Predicates
1

Its longest line of descent runs back 16 generations to K864889 (1987). 64 ancestors in all, the oldest K832113 (1983). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor2002: 2 other ancestors2001: 3 other ancestors2000: 3 other ancestors1999: 7 other ancestors2004: 2 other ancestors2003: 1 other ancestor1992: 2 other ancestors1998: 2 other ancestors1996: 3 other ancestors1995: 7 other ancestors1994: 3 other ancestors1986: 1 other ancestor2007: 1 other ancestor1997: 2 other ancestors1993: 2 other ancestors1991: 1 other ancestor1990: 1 other ancestor1989: 1 other ancestor1988: 1 other ancestor1983: 1 other ancestor1987: 1 other ancestor2026: 1 descendant clearedK864889 · Omni-Trak · Invivo Research Labs, Inc. · 1987K950264 · Maglife · Odam · 1996K023195 · Maglife C/Maglife C Plus · Schiller Medical S.A. · 2003K040915 · Magnitude 3150M Mri Patient Monitor, Model 3150M · Invivo Research, Inc. · 2004×2: 2 generations cleared close together, K050399 (2005), K053462 (2006) K071641 · Reveal XT; Insertable Cardiac Monitor (Model 9529), Patient Assistant (Model 9539) · Medtronic, Inc. · 2007K082475 · Reveal XT Insertable Cardiac Monitor, Model 9529, Revel DX Insertable Cardiac Monitor, Model 9528 · Medtronic, Inc. · 2008K103764 · Reveal XT Insertable Cardiac Monitor, Model 9529 and Reveal DX Insertable Cardiac Monitor, Model 9528 · Medtronic, Inc. · 2011K132649 · Reveal Linq Insertable Cardiac Monitor · Medtronic, Inc. · 2014 · recalledK150614 · Reveal LINQ · Medtronic, Inc. · 2015K160809 · Reveal LINQ · Medtronic, Inc. · 2016K162855 · Reveal LINQ · Medtronic, Inc. · 2017 · recalledK200795 · LINQ II Insertable Cardiac Monitor, LINQ Mobile Manager, Device Command Library, Instrument Command Library, LINQ Tool Kit · Medtronic, Inc. · 2020 · recalledK221962 · LINQ II Insertable Cardiac Monitor · Medtronic, Inc. · 2022 · recalledK233320 · LINQ II™ Insertable Cardiac Monitor (LNQ22) · Medtronic, Inc. · 2023 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 5 generations back: 5 other ancestors6 generations back: 11 other ancestors7 generations back: 12 other ancestors8 generations back: 16 other ancestors9 generations back: 4 other ancestors 1 generation forward: 1 descendant K233320 · LINQ II™ Insertable Cardiac Monitor (LNQ22) · Medtronic, Inc. · 2023 · recalledK221962 · LINQ II Insertable Cardiac Monitor · Medtronic, Inc. · 2022 · recalledK200795 · LINQ II Insertable Cardiac Monitor, LINQ Mobile Manager, Device Command Library, Instrument Command Library, LINQ Tool Kit · Medtronic, Inc. · 2020 · recalledK162855 · Reveal LINQ · Medtronic, Inc. · 2017 · recalledK160809 · Reveal LINQ · Medtronic, Inc. · 2016K150614 · Reveal LINQ · Medtronic, Inc. · 2015K132649 · Reveal Linq Insertable Cardiac Monitor · Medtronic, Inc. · 2014 · recalledK103764 · Reveal XT Insertable Cardiac Monitor, Model 9529 and Reveal DX Insertable Cardiac Monitor, Model 9528 · Medtronic, Inc. · 2011K082475 · Reveal XT Insertable Cardiac Monitor, Model 9529, Revel DX Insertable Cardiac Monitor, Model 9528 · Medtronic, Inc. · 2008K071641 · Reveal XT; Insertable Cardiac Monitor (Model 9529), Patient Assistant (Model 9539) · Medtronic, Inc. · 2007K053462 · 3160 Mri Patient Monitoring System · Invivo Corporation · 2006 · recalledK050399 · Mri Compatible Patient Monitoring System, Model 3160 · Invivo Research, Inc. · 2005K040915 · Magnitude 3150M Mri Patient Monitor, Model 3150M · Invivo Research, Inc. · 2004K023195 · Maglife C/Maglife C Plus · Schiller Medical S.A. · 2003K950264 · Maglife · Odam · 1996K864889 · Omni-Trak · Invivo Research Labs, Inc. · 1987−15−10−50
  1. Root · 16 generations backK864889 Omni-TrakInvivo Research Labs, Inc. · 1987
  2. 12 generations between
  3. 3 generations backK200795 LINQ II Insertable Cardiac Monitor, LINQ Mobile Manager, Device…Medtronic, Inc. · 2020 · Recalled
  4. 2 generations backK221962 LINQ II Insertable Cardiac MonitorMedtronic, Inc. · 2022 · Recalled
  5. 1 generation backK233320 LINQ II™ Insertable Cardiac Monitor (LNQ22)Medtronic, Inc. · 2023 · Recalled
  6. This deviceK240693 LINQ II™ Insertable Cardiac Monitor (ICM)Medtronic, Inc. · 2024
  7. Descendant · cites it directly · 0 of its ownK262735 LINQ II Insertable Cardiac Monitor (ICM) (LNQ22)Medtronic, Inc. · 2026
Show all 16 generations
  1. Root · 16 generations backK864889 Omni-TrakInvivo Research Labs, Inc. · 1987
  2. 15 generations backK950264 MaglifeOdam · 1996
  3. 14 generations backK023195 Maglife C/Maglife C PlusSchiller Medical S.A. · 2003
  4. 13 generations backK040915 Magnitude 3150M Mri Patient Monitor, Model 3150MInvivo Research, Inc. · 2004
  5. 12 generations backK050399 Mri Compatible Patient Monitoring System, Model 3160Invivo Research, Inc. · 2005
  6. 11 generations backK053462 3160 Mri Patient Monitoring SystemInvivo Corporation · 2006 · Recalled
  7. 10 generations backK071641 Reveal XT; Insertable Cardiac Monitor (Model 9529), Patient…Medtronic, Inc. · 2007
  8. 9 generations backK082475 Reveal XT Insertable Cardiac Monitor, Model 9529, Revel DX…Medtronic, Inc. · 2008
  9. 8 generations backK103764 Reveal XT Insertable Cardiac Monitor, Model 9529 and Reveal DX…Medtronic, Inc. · 2011
  10. 7 generations backK132649 Reveal Linq Insertable Cardiac MonitorMedtronic, Inc. · 2014 · Recalled
  11. 6 generations backK150614 Reveal LINQMedtronic, Inc. · 2015
  12. 5 generations backK160809 Reveal LINQMedtronic, Inc. · 2016
  13. 4 generations backK162855 Reveal LINQMedtronic, Inc. · 2017 · Recalled
  14. 3 generations backK200795 LINQ II Insertable Cardiac Monitor, LINQ Mobile Manager, Device…Medtronic, Inc. · 2020 · Recalled
  15. 2 generations backK221962 LINQ II Insertable Cardiac MonitorMedtronic, Inc. · 2022 · Recalled
  16. 1 generation backK233320 LINQ II™ Insertable Cardiac Monitor (LNQ22)Medtronic, Inc. · 2023 · Recalled
  17. This deviceK240693 LINQ II™ Insertable Cardiac Monitor (ICM)Medtronic, Inc. · 2024

Most influential descendants

Showing 1 of 1 · See all
  • K262735LINQ II Insertable Cardiac Monitor (ICM) (LNQ22)Medtronic, Inc. · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−16K864889Omni-TrakInvivo Research Labs, Inc.1987—
−15K950264MaglifeOdam1996—
−14K023195Maglife C/Maglife C PlusSchiller Medical S.A.2003—
−13K040915Magnitude 3150M Mri Patient Monitor, Model 3150MInvivo Research, Inc.2004—
−12K050399Mri Compatible Patient Monitoring System, Model 3160Invivo Research, Inc.2005—
−11K0534623160 Mri Patient Monitoring SystemInvivo Corporation20062
−10K071641Reveal XT; Insertable Cardiac Monitor (Model 9529), Patient Assistant (Model…Medtronic, Inc.2007—
−9K082475Reveal XT Insertable Cardiac Monitor, Model 9529, Revel DX Insertable Cardiac…Medtronic, Inc.2008—
−8K103764Reveal XT Insertable Cardiac Monitor, Model 9529 and Reveal DX Insertable…Medtronic, Inc.2011—
−7K132649Reveal Linq Insertable Cardiac MonitorMedtronic, Inc.20141
−6K150614Reveal LINQMedtronic, Inc.2015—
−5K160809Reveal LINQMedtronic, Inc.2016—
−4K162855Reveal LINQMedtronic, Inc.20171
−3K200795LINQ II Insertable Cardiac Monitor, LINQ Mobile Manager, Device Command…Medtronic, Inc.20203
−2K221962LINQ II Insertable Cardiac MonitorMedtronic, Inc.20221
−1K233320LINQ II™ Insertable Cardiac Monitor (LNQ22)Medtronic, Inc.20231
0 (this device)K240693LINQ II™ Insertable Cardiac Monitor (ICM)Medtronic, Inc.2024—
+1K262735LINQ II Insertable Cardiac Monitor (ICM) (LNQ22)Medtronic, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
198610
198710
198810
198910
199010
199110
199220
199320
199430
199570
199630
199720
199820
199970
200030
200130
200220
200310
200420
200510
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K240693, 1987–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MXD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MXD device

  1. #1 K262735 LINQ II Insertable Cardiac Monitor (ICM) (LNQ22)score 16.7 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K233320LINQ II™ Insertable Cardiac Monitor (LNQ22)
Summary page 5
…Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm Predicate Device: LINQ™ II Insertable Cardiac Monitor (Model LNQ22) K233320 Device Description The LINQ II Insertable Cardiac Monitor (ICM) Model LNQ22 is a programmable device that continuously monitors…
Medtronic, Inc.20231

Also cleared against K233320: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K262735LINQ II Insertable Cardiac Monitor (ICM) (LNQ22)Medtronic, Inc.20260

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 37 years and passes through 6 companies; a new generation was cleared every 1.5 years on median.
  • 8 of the 64 ancestors have a recall of any class.
  • Of its descendants, K262735 (Medtronic, Inc., 2026) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K233320.

How does its category compare?

All MXD devices

MXD Recorder, Event, Implantable Cardiac, (With Arrhythmia Detection): FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · MXD

3in 2025

2014: 22015: 22016: 32017: 42019: 12020: 42021: 42022: 22023: 62024: 12025: 32026 (partial year): 2
2007high 6 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MXD

36in 2025

2016: 63 days median2017: 52 days median2020: 127 days median2021: 57 days median2023: 76 days median2025: 36 days median
2007high 127 (2020) · low 36 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MXD

5so far in 2026, the most since 2007

2022: 12023: 12024: 12026 (partial year): 5
2007high 5 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MXD

3,384in 2025

2013: 12015: 22016: 212017: 372018: 322019: 702020: 2282021: 6942022: 1,3622023: 1,7662024: 3,0192025: 3,384
2007high 3,384 (2025) · low 0 (2007)2025

MAUDE reports naming product code MXD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—01
20142—00
20152—02
2016363021
2017452037
20180—032
20191—070
202041270228
20214570694
20222—11,362
202367611,766
20241—13,019
202533603,384
20262—5—

What is it indicated for?

The LINQ II ICM is an insertable automatically-activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in adult patients, and in pediatric patients who are at least 2 years old, in the following cases:

• patients with clinical syndromes or situations at increased risk of cardiac arrhythmias

• patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K240693: LINQ II™ Insertable Cardiac Monitor (ICM).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K240693

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026