510(k) Report

510(k) · Cleared

Cleared

ICS Dizcovery (1091)

K242198510(k) TraditionalCleared Feb 14, 2025203-day reviewClass II21 CFR 882.1460

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Natus Medical Denmark Aps · GWN Nystagmograph

Cleared on Feb 14, 2025 as substantially equivalent to K151504 (Gn Otometrics, 2015) after a 203-day review. Its family tree reaches back 3 generations to 1999.

Decided
Feb 14, 2025
Review
203 days
Class
Class II
Product code
GWN
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K991497 (1999). 6 ancestors in all, the oldest K891008 (1989). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2013: 1 other ancestor1997: 1 other ancestor1989: 1 other ancestorK991497 · Ics Medical Chartr Eng/VNG Diagnostic System · Ics Medical Corp. · 1999K061791 · Focus VNG/Eng Type 1068 · Gn Otometrics A/S · 2006K151504 · ICS Impulse · Gn Otometrics · 20151990200020102020
Other ancestors, by generation 2 generations back: 1 other ancestor3 generations back: 2 other ancestors K151504 · ICS Impulse · Gn Otometrics · 2015K061791 · Focus VNG/Eng Type 1068 · Gn Otometrics A/S · 2006K991497 · Ics Medical Chartr Eng/VNG Diagnostic System · Ics Medical Corp. · 19990
  1. Root · 3 generations backK991497 Ics Medical Chartr Eng/VNG Diagnostic SystemIcs Medical Corp. · 1999
  2. 2 generations backK061791 Focus VNG/Eng Type 1068Gn Otometrics A/S · 2006
  3. 1 generation backK151504 ICS ImpulseGn Otometrics · 2015
  4. This deviceK242198 ICS Dizcovery (1091)Natus Medical Denmark Aps · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K991497Ics Medical Chartr Eng/VNG Diagnostic SystemIcs Medical Corp.1999—
−2K061791Focus VNG/Eng Type 1068Gn Otometrics A/S2006—
−1K151504ICS ImpulseGn Otometrics2015—
0 (this device)K242198ICS Dizcovery (1091)Natus Medical Denmark Aps2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199710
201310
FIG. 1 — Lineage of K242198, 1999–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GWN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GWN device

  1. #1 K260566 SMARTVERTIGO™ Vestibular Assessment Systemscore 12.1 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K151504ICS Impulse
Summary page 6
…tagmograph Regulation Number: 21CFR882.1460 Classification Name: Nystagmograph Product code: GWN Device Class: II Predicate Device: ICS Impulse 510(k) number: K151504 Natus Medical Incorporated | natus.com…
Gn Otometrics2015—

Also cleared against K151504: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 4 companies; a new generation was cleared every 9.2 years on median.
  • None of the 6 ancestors has a recall.
  • 2 other devices also cite its predicate K151504.

How does its category compare?

All GWN devices

GWN Nystagmograph: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · GWN

3in 2025

2007: 22008: 12009: 12013: 32015: 32016: 22017: 42018: 32019: 12020: 22023: 12025: 32026 (partial year): 1
2007high 4 (2017) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GWN

203in 2025

2013: 88 days median2015: 165 days median2017: 160 days median2018: 87 days median2025: 203 days median
2007high 203 (2025) · low 87 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GWN

1in 2013

2013: 1
2007high 1 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GWN

1in 2025

2012: 22018: 12019: 52021: 12022: 12023: 32024: 42025: 1
2007high 5 (2019) · low 0 (2007)2025

MAUDE reports naming product code GWN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20081—00
20091—00
20100—00
20110—00
20120—02
201338810
20140—00
2015316500
20162—00
2017416000
201838701
20191—05
20202—00
20210—01
20220—01
20231—03
20240—04
2025320301
20261—0—

What is it indicated for?

ICS Dizcovery is used in the assessment of the vestibular-ocular reflex (VOR) and nystagmus by measuring, recording, displaying, and analyzing eye and head movements. The 1091 ICS Dizcovery is used in the assessment of patients with complaints of a vestibular nature such as dizziness, disequilibrium, and vertigo. The ICS Dizcovery type 1091 does not treat or diagnose the patient; the diagnosis is determined by the credentialed physician. ICS Dizcovery…

Read the full statement (166 words)

ICS Dizcovery is used in the assessment of the vestibular-ocular reflex (VOR) and nystagmus by measuring, recording, displaying, and analyzing eye and head movements.

The 1091 ICS Dizcovery is used in the assessment of patients with complaints of a vestibular nature such as dizziness, disequilibrium, and vertigo. The ICS Dizcovery type 1091 does not treat or diagnose the patient; the diagnosis is determined by the credentialed physician.

ICS Dizcovery system is intended to be used by qualified medical personnel. Typical device users are Neurologists, Ears, Nose, and Throat specialists (ENTs), Audiologists, Physical Therapists, and Technicians supervised by one of the four mentioned typical users. Professionals with knowledge of diagnosing balance disorders. Users are assumed to have prior knowledge of the medical and scientific facts underlying the procedures offered by the ICS Dizcovery system.

The intended patient population are children in the age of 10 to 18 and adults in age range from 18 to 99 years with a complaint of dizziness, balance disorder, or vestibular disease.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K242198: ICS Dizcovery (1091).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K242198

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026