510(k) Report

510(k) · Cleared

Cleared

ARIX Femur Nail System

K242247510(k) TraditionalCleared Apr 22, 2025265-day reviewClass II21 CFR 888.3020

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Jeil Medical Corporation · HSB Rod, Fixation, Intramedullary And Accessories

Cleared on Apr 22, 2025 as substantially equivalent to K200869 (Stryker Trauma GmbH, 2020) after a 265-day review. Its family tree reaches back 9 generations to 1997. 1 later device descends from it.

Decided
Apr 22, 2025
Review
265 days
Class
Class II
Product code
HSB
Recalls
None
Predicates
1

Its longest line of descent runs back 9 generations to K970097 (1997). 42 ancestors in all, the oldest K871539 (1987). 1 later device descends from it, 1 citing it directly.

  1. Root · 9 generations backK970097 Proximal Femoral Nail (PFN) SystemSynthes (Usa) · 1997
  2. 5 generations between
  3. 3 generations backK180436 T2 Alpha Tibia Nailing System, T2 Alpha Femur Antegrade GT/PF…Stryker Trauma GmbH · 2018
  4. 2 generations backK191271 T2 Alpha Femur Antegrade GT/PF Nailing System, T2 Alpha Tibia…Stryker Trauma GmbH · 2019
  5. 1 generation backK200869 Gamma3 SystemStryker Trauma GmbH · 2020
  6. This deviceK242247 ARIX Femur Nail SystemJeil Medical Corporation · 2025
  7. Descendant · cites it directly · 0 of its ownK252826 ARIX Femur Nail SystemJeil Medical Corporation · 2025
Show all 9 generations
  1. Root · 9 generations backK970097 Proximal Femoral Nail (PFN) SystemSynthes (Usa) · 1997
  2. 8 generations backK973240 Synthes Proximal Femoral Nail (PFN) System ModificationSynthes (Usa) · 1997
  3. 7 generations backK013563 Long Trochanteric Nail SystemDePuy Orthopaedics, Inc. · 2001 · Recalled
  4. 6 generations backK033329 Depuy Ace Universal and Troch Entry Femoral Nail SystemsDepuy, Inc. · 2003
  5. 5 generations backK083497 Zimmer Natural Nail System Piriformis Fossa and Greater Trochanter…Zimmer, Inc. · 2009 · Recalled
  6. 4 generations backK172774 T2 Alpha Femur Antegrade GT/PF Nailing System, IMN Screws System…Stryker Trauma GmbH · 2018
  7. 3 generations backK180436 T2 Alpha Tibia Nailing System, T2 Alpha Femur Antegrade GT/PF…Stryker Trauma GmbH · 2018
  8. 2 generations backK191271 T2 Alpha Femur Antegrade GT/PF Nailing System, T2 Alpha Tibia…Stryker Trauma GmbH · 2019
  9. 1 generation backK200869 Gamma3 SystemStryker Trauma GmbH · 2020
  10. This deviceK242247 ARIX Femur Nail SystemJeil Medical Corporation · 2025

Most influential descendants

Showing 1 of 1 · See all
  • K252826ARIX Femur Nail SystemJeil Medical Corporation · 2025 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K970097Proximal Femoral Nail (PFN) SystemSynthes (Usa)1997—
−8K973240Synthes Proximal Femoral Nail (PFN) System ModificationSynthes (Usa)1997—
−7K013563Long Trochanteric Nail SystemDePuy Orthopaedics, Inc.200154
−6K033329Depuy Ace Universal and Troch Entry Femoral Nail SystemsDepuy, Inc.2003—
−5K083497Zimmer Natural Nail System Piriformis Fossa and Greater Trochanter Antegrade…Zimmer, Inc.200916
−4K172774T2 Alpha Femur Antegrade GT/PF Nailing System, IMN Screws System, IMN…Stryker Trauma GmbH2018—
−3K180436T2 Alpha Tibia Nailing System, T2 Alpha Femur Antegrade GT/PF Nailing System…Stryker Trauma GmbH2018—
−2K191271T2 Alpha Femur Antegrade GT/PF Nailing System, T2 Alpha Tibia Nailing System…Stryker Trauma GmbH2019—
−1K200869Gamma3 SystemStryker Trauma GmbH2020—
0 (this device)K242247ARIX Femur Nail SystemJeil Medical Corporation2025—
+1K252826ARIX Femur Nail SystemJeil Medical Corporation2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199210
199620
199720
199830
199910
200010
200160
200220
200320
200440
200520
200810
201030
201310
201710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202510
FIG. 1 — Lineage of K242247, 1997–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HSB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HSB device

  1. #1 K200869 Gamma3 Systemscore 19.8 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K200869Gamma3 System
Summary page 6
…ation of Predicate Device(s) [21 CFR 807.92(a)(3)] The legally marketed device(s) to which substantial equivalence is claimed is/are: Primary Predicate Device K200869 - Gamma3 System, Stryker Trauma GmbH Reference Device K212266 - FIX-C PEEK Anterior Cervical Interbody System,…
Stryker Trauma GmbH2020—
†K212266FIX-C PEEK Anterior Cervical Interbody SystemJeil Medical Corporation2021—

Also cleared against K200869: 5 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K252826ARIX Femur Nail SystemJeil Medical Corporation20250

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 6 companies; a new generation was cleared every 1.9 years on median.
  • 21 of the 42 ancestors have a recall of any class.
  • Of its descendants, K252826 (Jeil Medical Corporation, 2025) has gone furthest, with 0 descendants of its own.
  • 5 other devices also cite its predicate K200869.

How does its category compare?

All HSB devices

HSB Rod, Fixation, Intramedullary And Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · HSB

15in 2025

2007: 82008: 112009: 182010: 142011: 162012: 172013: 182014: 192015: 222016: 82017: 232018: 172019: 132020: 172021: 102022: 122023: 112024: 132025: 152026 (partial year): 18
2007high 23 (2017) · low 8 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HSB

106in 2025

2007: 68 days median2008: 57 days median2009: 86 days median2010: 117 days median2011: 78 days median2012: 112 days median2013: 105 days median2014: 147 days median2015: 111 days median2016: 124 days median2017: 119 days median2018: 134 days median2019: 132 days median2020: 112 days median2021: 94 days median2022: 172 days median2023: 162 days median2024: 195 days median2025: 106 days median2026 (partial year): 151 days median
2007high 195 (2024) · low 57 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HSB

18in 2025

2008: 92009: 812010: 182011: 22012: 72013: 142014: 32015: 32016: 62017: 662018: 562019: 62020: 62021: 22022: 12023: 12024: 12025: 182026 (partial year): 1
2007high 81 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HSB

2,911in 2025

2007: 1622008: 1842009: 3342010: 4542011: 6622012: 1,0792013: 1,7732014: 1,4742015: 1,6712016: 1,9202017: 1,4472018: 3,5262019: 2,5172020: 3,2992021: 4,1112022: 2,7332023: 2,2192024: 2,8692025: 2,911
2007high 4,111 (2021) · low 162 (2007)2025

MAUDE reports naming product code HSB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20078680162
200811579184
2009188681334
20101411718454
201116782662
20121711271,079
201318105141,773
20141914731,474
20152211131,671
2016812461,920
201723119661,447
201817134563,526
20191313262,517
20201711263,299
2021109424,111
20221217212,733
20231116212,219
20241319512,869
202515106182,911
2026181511—

What is it indicated for?

ARIX Femur Nail System is indicated for the treatment of stable and unstable fractures as well as for stabilization of bones and correction of bone deformities in the intracapsular, trochanteric, subtrochanteric and shaft regions of the femur (including osteoporotic and osteopenic bone).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K242247: ARIX Femur Nail System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K242247

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026