510(k) Report

510(k) · Cleared

Cleared

D·Kutting™ LL Peripheral Scoring Balloon Dilatation Catheter

K242254510(k) TraditionalCleared Apr 21, 2025264-day reviewClass II21 CFR 870.1250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dk Medical Technology Co., Ltd. · PNO Catheter, Percutaneous, Cutting/Scoring

Cleared on Apr 21, 2025 as substantially equivalent to K103459 (C.R. Bard, Inc., 2010) after a 264-day review. Its family tree reaches back 7 generations to 2000.

Decided
Apr 21, 2025
Review
264 days
Class
Class II
Product code
PNO
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K001134 (2000). 26 ancestors in all, the oldest K971010 (1997). No later device cites it as a predicate yet.

  1. Root · 7 generations backK001134 Gazelle Balloon Dilatation CatheterBoston Scientific Corp · 2000
  2. 6 generations backK021735 Boston Scientific Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Scimed, Inc. · 2002 · Recalled
  3. 5 generations backK050389 Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Corp · 2005
  4. 4 generations backK053116 Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-XxxxxxBoston Scientific Corp · 2005
  5. 3 generations backK072283 Dorado Pta Balloon Dilatation CatheterBard Peripheral Vascular, Inc. · 2007 · Recalled
  6. 2 generations backK083657 Conquest Pta Balloon Dilatation CatheterC.R. Bard, Inc. · 2008 · Recalled
  7. 1 generation backK103459 Vascutrak Pta Dilatation CatheterC.R. Bard, Inc. · 2010
  8. This deviceK242254 D·Kutting™ LL Peripheral Scoring Balloon Dilatation CatheterDk Medical Technology Co., Ltd. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K001134Gazelle Balloon Dilatation CatheterBoston Scientific Corp2000—
−6K021735Boston Scientific Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Scimed, Inc.200223
−5K050389Ultra-Soft SV Balloon Dilatation CatheterBoston Scientific Corp2005—
−4K053116Sterling Otw Pta Balloon Dilatation Catheter, Model H74939032-XxxxxxBoston Scientific Corp2005—
−3K072283Dorado Pta Balloon Dilatation CatheterBard Peripheral Vascular, Inc.200715
−2K083657Conquest Pta Balloon Dilatation CatheterC.R. Bard, Inc.20082
−1K103459Vascutrak Pta Dilatation CatheterC.R. Bard, Inc.2010—
0 (this device)K242254D·Kutting™ LL Peripheral Scoring Balloon Dilatation CatheterDk Medical Technology Co., Ltd.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
200020
200110
200230
200330
200410
200540
200620
200710
200810
FIG. 1 — Lineage of K242254, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PNO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PNO device

  1. #1 K261778 JADE SCORE PTA Scoring Balloon Dilatation…score 14.1 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K103459Vascutrak Pta Dilatation Catheter
Summary page 5
…VASCUTRAK PTA DILATATION CATHETER (VASCUTRAK) Device classification name: catheter, percutaneous, cutting/scoring 510(k) number: K103459 Regulation number: 21 CFR Part 870.1250 Product code: PNO Classification: Class II…
C.R. Bard, Inc.2010—
†K1512532cm Peripheral Cutting Balloon Microsurgical Dilatation CatheterBoston Scientific Limited2015—

Also cleared against K103459: 5 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 5 companies; a new generation was cleared every 1.9 years on median.
  • 5 of the 26 ancestors have a recall of any class.
  • 5 other devices also cite its predicate K103459.

How does its category compare?

All PNO devices

PNO Catheter, Percutaneous, Cutting/Scoring: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · PNO

2in 2025

2007: 32008: 42009: 12010: 32011: 22012: 12013: 12014: 12015: 42016: 12017: 32019: 12020: 42021: 12022: 22024: 42025: 22026 (partial year): 3
2007high 4 (2008) · low 0 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PNO

246in 2024

2007: 61 days median2008: 28 days median2010: 23 days median2015: 84 days median2017: 74 days median2020: 94 days median2024: 246 days median2026 (partial year): 110 days median
2007high 246 (2024) · low 23 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PNO

No recall names PNO since 2007.

Adverse event reports per year · PNO

511in 2025

2016: 92017: 262018: 552019: 812020: 1212021: 902022: 752023: 3042024: 5372025: 511
2007high 537 (2024) · low 0 (2007)2025

MAUDE reports naming product code PNO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200736100
200842800
20091—00
201032300
20112—00
20121—00
20131—00
20141—00
201548400
20161—09
2017374026
20180—055
20191—081
20204940121
20211—090
20222—075
20230—0304
202442460537
20252—0511
202631100—

What is it indicated for?

The D-KuttingTm LL Peripheral Scoring Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty (PTA) in the peripheral vasculature, including the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in the coronary vasculature.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K242254: D·Kutting™ LL Peripheral Scoring Balloon Dilatation Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K242254

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026