510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 1000 ml/min); Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 750 ml/min)

K242735510(k) TraditionalCleared Dec 23, 2024103-day reviewClass II21 CFR 880.5725Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Belmont Medical Technologies · FRN Pump, Infusion

Cleared on Dec 23, 2024 as substantially equivalent to K141654 (Belmont Instrument Corporation, 2014) after a 103-day review. Its family tree reaches back 4 generations to 1985. It has been recalled 2 times.

Decided
Dec 23, 2024
Review
103 days
Class
Class II
Product code
FRN
Recalls
2
Predicates
1

Its longest line of descent runs back 4 generations to K852645 (1985). 5 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 4 generations backK852645 Fluid Management SystemHaemonetics Corp. · 1985
  2. 3 generations backK972284 Fluid Management SystemBelmont Instrument Corp. · 1997 · Recalled
  3. 2 generations backK032674 Belmont Fluid Management System, Model FMS2000Belmont Instrument Corp. · 2003 · Recalled
  4. 1 generation backK141654 The Belmont Rapid InfuserBelmont Instrument Corporation · 2014
  5. This deviceK242735 Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 1000 ml/min)…Belmont Medical Technologies · 2024
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K852645Fluid Management SystemHaemonetics Corp.1985—
−3K972284Fluid Management SystemBelmont Instrument Corp.19971
−2K032674Belmont Fluid Management System, Model FMS2000Belmont Instrument Corp.20031
−1K141654The Belmont Rapid InfuserBelmont Instrument Corporation2014—
0 (this device)K242735Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 1000 ml/min); Belmont Rapid…Belmont Medical Technologies2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200910
FIG. 1 — Lineage of K242735, 1985–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141654The Belmont Rapid Infuser
Summary page 5
…, Class II Classification Product Code: FRN Subsequent Product Code: LGZ Predicate Device: K141654: The Belmont Rapid Infuser This predicate has not been subject to a design-related recall. Devic…
Belmont Instrument Corporation2014—

Also cleared against K141654: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 31, 2025Class IIZ-1678-2025Process designPotential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.
Jan 2, 2025Class IIZ-1099-2025Process controlDisposable set may leak during priming from a female quick connector due to a crack in the female quick connector, may lead to a potential delay in treatment since the set will need to be replaced by another set

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 39 years and passes through 4 companies; a new generation was cleared every 10.8 years on median.
  • 3 of the 5 ancestors have a recall of any class.
  • 1 other device also cites its predicate K141654.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

The Belmont Rapid Infuser RI-2 is designed to be used in general operation in hospital or alternative care environments to

provide warmed blood and fluids to any patients ≥ 10 kg requiring warmed infusion from 2.5 mL/min to 1000 mL/min.

- Infusion of crystalloid, colloid, or blood product, including packed red blood cells, as volume replacement for patients

suffering from blood loss due to trauma or surgery.

- Infusion of warmed fluid to rewarm patients after surgery or for hypothermia.

- Infusion of warmed fluid for irrigation in urology procedures.

The 3.0L reservoir is an optional accessory for use in adults only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K242735: Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 1000 ml/min); Belmont Rapid Infuser, RI-2 (Rapid Infuser RI-2, 750 ml/min).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K242735

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026