Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE), Medsystem Iii Infusion (Msiii), Medsystem Patient C

K012383510(k) TraditionalCleared Oct 2, 200167-day reviewClass II21 CFR 880.5725Recalls 12 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Alaris Medical Systems, Inc. · FRN Pump, Infusion

Cleared on Oct 2, 2001 as substantially equivalent to K010924 (Abbott Laboratories, 2001) and 1 other predicate after a 67-day review. Its family tree reaches back 1 generation to 2000. 4 of the 16 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 12 times, including 3 Class I recalls.

Decided
Oct 2, 2001
Review
67 days
Class
Class II
Product code
FRN
Recalls
12 (3 Class I)
Predicates
2

Its longest line of descent runs back 1 generation to K002211 (2000). 2 ancestors in all. 13 later devices descend from it, 4 citing it directly. 4 of the 16 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor2009: 1 descendant cleared, 1 with a Class I recall12005: 1 descendant cleared2003: 1 descendant cleared2014: 2 descendants cleared, 1 with a Class I recall12015: 1 descendant cleared2008: 1 descendant cleared2016: 1 descendant cleared2022: 1 descendant cleared2023: 1 descendant cleared, 1 with a Class I recall12025: 2 descendants cleared2018: 1 descendant clearedK002211 · Colleague CX Volumetric Infusion Pump 2M8161 and Volumeteric Infusion Pump 2M8151 · Baxter Healthcare Corp · 2000 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 4 descendants, 2 with a Class I recall22 generations forward: 6 descendants, 1 with a Class I recall13 generations forward: 3 descendants K002211 · Colleague CX Volumetric Infusion Pump 2M8161 and Volumeteric Infusion Pump 2M8151 · Baxter Healthcare Corp · 2000 · recalled0
  1. Root · 1 generation backK002211 Colleague CX Volumetric Infusion Pump 2M8161 and Volumeteric…Baxter Healthcare Corp · 2000 · Recalled
  2. This deviceK012383 Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE)…Alaris Medical Systems, Inc. · 2001
  3. Descendant · cites it directly · 4 of its ownK050301 Mridium 3850 Mri Pump SystemIradimed Corporation · 2005 · Recalled
  4. Descendant · cites it directly · 3 of its ownK133532 Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Recalled (Class I)
  5. Descendant · cites it directly · 3 of its ownK030136 Drip AlertDrip Alert, Inc. · 2003
  6. Descendant · cites it directly · 2 of its ownK090087 Mridium 3860 Mri Infusion Pump/Monitoring SystemIradimed Corporation · 2009 · Recalled (Class I)
  7. Descendant · 2 generations on · 1 of its ownK211218 BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · Recalled (Class I)

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2003: 12005: 12008: 12009: 12014: 22015: 12016: 12018: 12022: 12023: 12025: 2
2003high 2 (2014) · low 0 (2004)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K050301Mridium 3850 Mri Pump SystemIradimed Corporation · 2005 · cites it directly4descendants
  • K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · cites it directly3descendants
  • K030136Drip AlertDrip Alert, Inc. · 2003 · cites it directly3descendants
  • K090087Mridium 3860 Mri Infusion Pump/Monitoring SystemIradimed Corporation · 2009 · cites it directly2descendants
  • K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K002211Colleague CX Volumetric Infusion Pump 2M8161 and Volumeteric Infusion Pump…Baxter Healthcare Corp20004
0 (this device)K012383Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE), Medsystem…Alaris Medical Systems, Inc.2001—
+1K050301Mridium 3850 Mri Pump SystemIradimed Corporation20052
+1K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc.2014Class I
+1K030136Drip AlertDrip Alert, Inc.2003—
+1K090087Mridium 3860 Mri Infusion Pump/Monitoring SystemIradimed Corporation2009Class I
+2K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc.2023Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200510
200810
200911
201421
201510
201610
201810
202210
202311
202520
FIG. 1 — Lineage of K012383, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K010924Modification to Plum XL Infusion Pump
Summary page 2
…AND EFFECTIVENESS, Continued ALARIS Medical Systems, Inc. Page 2 of 3 Predicate Device(s) ‘Manufacturer and Device S10(k)# | Date Cleared Abbott Plum XL Infusion Pump K010924 04/06/01 Colleague Volumetric and CX Volumetric Infusion K002211 08/28/00 Pumps ’ DEVICE DESCRIPTION ALARIS Medical Systems line…
Abbott Laboratories20016
*K002211Colleague CX Volumetric Infusion Pump 2M8161 and Volumeteric Infusion Pump 2M8151
Summary page 2
…of 3 Predicate Device(s) ‘Manufacturer and Device S10(k)# | Date Cleared Abbott Plum XL Infusion Pump K010924 04/06/01 Colleague Volumetric and CX Volumetric Infusion K002211 08/28/00 Pumps ’ DEVICE DESCRIPTION ALARIS Medical Systems line of infusion pumps are electrical volumetric infusion pumps that…
Baxter Healthcare Corp20004

What descends from it? (13)

See all 13

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K050301Mridium 3850 Mri Pump SystemIradimed Corporation20054
K133532Alaris System with Guardrails Suite MX Class ICarefusion 303, Inc.20143
K030136Drip AlertDrip Alert, Inc.20033
K090087Mridium 3860 Mri Infusion Pump/Monitoring System Class IIradimed Corporation20092

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 4, 2020Class IZ-1360-2020Device Design1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion…
Feb 4, 2020Class IZ-1362-2020Device Design1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion…
Feb 4, 2020Class IZ-1361-2020Device Design1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion…
Jun 12, 2009Not classifiedZ-1790-2009Device DesignThere is a potential risk for medication not being delivered when IV tubing sets are occluded in the Alaris Pump module. This can occur if IV sets are set up in the infusion pumps but not used for an extended period of time. If users fail…
Jun 12, 2009Not classifiedZ-1791-2009Device DesignAffected Unit: Alaris PC Unit (Models 8000 and 8015) with software a versions 8 and above when used with the Alaris PCA module (Model 8120) There is a potential risk of over infusion related to a warning message on the user interface of…
Jun 12, 2009Not classifiedZ-1794-2009Device DesignThere is a potential risk of IUI connector failure in the Alaris System when these connectors become corroded or contaminated, which may result in an interruption of infusion with an audible and visual alarm. IUI connectors can become…
Mar 4, 2009Not classifiedZ-1237-2009Software designThe Alaris PCA module can potentially infuse above or below the intended infusion dose.
Mar 5, 2008Not classifiedZ-1712-2008Device DesignA channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due to failure of the U9 Socketed Integrated Circuits (U9 IC) on the display board of the…
Mar 5, 2008Not classifiedZ-1710-2008Device DesignA channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors due to failures of the U19 Socketed Integrated Circuits (U19 IC) on the logic board of the…
Mar 5, 2008Not classifiedZ-1711-2008Device DesignA channel error on an affected Alaris System module will stop an active infusion (or monitoring) with an audible and visual alarm. Channel errors with the Alaris System due to the Inter-Unit Interface (IUI) connectors on all Alaris System…
Aug 10, 2006Not classifiedZ-1484-06Software designThis recall was initiated because of a potential for over infusion with all models of the Alaris SE Pumps (formerly the Signature Edition Infusion Pumps) caused by key bounce.
Nov 10, 2004Not classifiedZ-0568-05OtherThe device modification is being implemented in response to information received from the field regarding a specific method of misloading the medication administration tubing set into the infusion system pump module. To date there have…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K090087Mridium 3860 Mri Infusion Pump/Monitoring SystemIradimed Corporation · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 11 Class I recall
  • K133532Alaris System with Guardrails Suite MXCarefusion 303, Inc. · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 128 Class I recalls
  • K211218BD Alaris System with Guardrails Suite MX v12.1.2Carefusion 303, Inc. · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 17 Class I recalls

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K050301 (Iradimed Corporation, 2005) has gone furthest, with 4 descendants of its own.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—

What is it indicated for?

ALARIS Medical Systems infusion pumps are intended for use in today’s growing professional healthcare environment including healthcare facilities, home care, and medical transport that utilize infusion pumps for the delivery of fluids, medications, blood and blood products. The following ALARIS Medical Systems infusion pumps are indicated for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous…

Read the full statement (136 words)

ALARIS Medical Systems infusion pumps are intended for use in today’s growing professional healthcare environment including healthcare facilities, home care, and medical transport that utilize infusion pumps for the delivery of fluids, medications, blood and blood products.

The following ALARIS Medical Systems infusion pumps are indicated for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural, enteral, or irrigation of fluid spaces:

IMED® Gemini Infusion Pump / Controllers model PC-1®

IMED® Gemini Infusion Pump / Controllers model PC-2®

IMED® Gemini Infusion Pump/Controller model PC-2TX®

IMED® Gemini Infusion Pump/Controller model PC-4*

IVAC® Signature Edition® Infusion Pumps

IVAC® MedSystem III®, Multi-channel Infusion Pumps

IVAC® MedSystem I1I® Multi-channel Infusion Pumps with DLE

ALARIS Medical Medley™ Patient Care System

Note: See the appropriate Directions for Use (DFU) for infusion system specifications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K012383: Gemini PC-1,PC-2, PC-2TX,PC-4, Signature Edition Infusion Pump (SE), Medsystem Iii Infusion (Msiii), Medsystem Patient C.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K012383

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026