510(k) Report

510(k) · Cleared

Cleared

Codman Libertís™ 1.5 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821761); Codman Libertís™ 1.9 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821762)

K243552510(k) TraditionalCleared Aug 6, 2025264-day reviewClass II21 CFR 882.5550

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Integra Lifesciences Production Corporation · JXG Shunt, Central Nervous System And Components

Cleared on Aug 6, 2025 as substantially equivalent to K233448 (Integra Lifesciences Production Corporation, 2023) after a 264-day review. Its family tree reaches back 3 generations to 2002.

Decided
Aug 6, 2025
Review
264 days
Class
Class II
Product code
JXG
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K021653 (2002). 4 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 3 generations backK021653 Codman Bactiseal Evd Catheter SetCodman & Shurtleff, Inc. · 2002
  2. 2 generations backK090348 Codman Bactiseal Evd Catheter Set, Codman Bactiseal Clear Evd…Codman & Shurtleff, Inc. · 2009
  3. 1 generation backK233448 Bactiseal EVD Catheter Sets; Bactiseal Clear EVD Catheter SetIntegra Lifesciences Production Corporation · 2023 · Recalled
  4. This deviceK243552 Codman Libertís™ 1.5 mm EVD Catheter with Bactiseal® and Endexo®…Integra Lifesciences Production Corporation · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K021653Codman Bactiseal Evd Catheter SetCodman & Shurtleff, Inc.2002—
−2K090348Codman Bactiseal Evd Catheter Set, Codman Bactiseal Clear Evd Catheter SetCodman & Shurtleff, Inc.2009—
−1K233448Bactiseal EVD Catheter Sets; Bactiseal Clear EVD Catheter SetIntegra Lifesciences Production Corporation20233
0 (this device)K243552Codman Libertís™ 1.5 mm EVD Catheter with Bactiseal® and Endexo® Technology…Integra Lifesciences Production Corporation2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200310
FIG. 1 — Lineage of K243552, 2002–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JXG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXG device

  1. #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K233448Bactiseal EVD Catheter Sets; Bactiseal Clear EVD Catheter Set
Summary page 5
…cate Device Bactiseal EVD Catheter Sets and Bactiseal Clear EVD Catheter Set (K233448) 807.92(a) (4) Device Description The Codman LibertísTM EVD Catheter with Bactiseal® and Endexo® Technology (LibertísTM EVD C…
Integra Lifesciences Production Corporation20233

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 2 companies; a new generation was cleared every 6.6 years on median.
  • 2 of the 4 ancestors have a recall of any class.

How does its category compare?

All JXG devices

JXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · JXG

8in 2025

2007: 32008: 62009: 52010: 32011: 72012: 32013: 32015: 62016: 62017: 72018: 62019: 42020: 22021: 12022: 42023: 52024: 32025: 82026 (partial year): 2
2007high 8 (2025) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXG

255in 2025

2007: 184 days median2008: 60 days median2009: 106 days median2010: 30 days median2011: 136 days median2012: 78 days median2013: 108 days median2015: 276 days median2016: 143 days median2017: 90 days median2018: 102 days median2019: 101 days median2022: 101 days median2023: 134 days median2024: 90 days median2025: 255 days median
2007high 276 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JXG

3in 2024

2008: 12009: 302011: 52012: 12013: 142014: 62017: 12020: 82023: 22024: 32026 (partial year): 1
2007high 30 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXG

1,198in 2025

2007: 6492008: 6302009: 6292010: 8122011: 7872012: 9772013: 1,0042014: 1,2512015: 9842016: 1,0912017: 1,3352018: 1,4842019: 1,2952020: 1,2062021: 1,4672022: 1,3292023: 1,2212024: 1,2142025: 1,198
2007high 1,484 (2018) · low 629 (2009)2025

MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731840649
20086601630
2009510630629
20103300812
201171365787
20123781977
20133108141,004
20140—61,251
201562760984
2016614301,091
201779011,335
2018610201,484
2019410101,295
20202—81,206
20211—01,467
2022410101,329
2023513421,221
202439031,214
2025825501,198
20262—1—

What is it indicated for?

The Codman Libertis ™ EVD Catheter with Bactiseal® and Endexo® Technology is indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K243552: Codman Libertís™ 1.5 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821761); Codman Libertís™ 1.9 mm EVD Catheter with Bactiseal® and Endexo® Technology with Luer Connection (821762).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K243552

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026