510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

MiniMed 780G insulin pump

K251032510(k) TraditionalCleared Jul 1, 202589-day reviewClass II21 CFR 880.5730Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Minimed, Inc. · QFG Alternate Controller Enabled Insulin Infusion Pump

Cleared on Jul 1, 2025 as substantially equivalent to K232380 (Tandem Diabetes Care, Inc., 2023) and 1 other predicate after an 89-day review. Its family tree reaches back 11 generations to 2001. 3 of the 28 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 6 times.

Decided
Jul 1, 2025
Review
89 days
Class
Class II
Product code
QFG
Recalls
6
Predicates
2

Its longest line of descent runs back 11 generations to K010362 (2001). 23 ancestors in all. 4 later devices descend from it, 3 citing it directly. 3 of the 28 devices in this family, counting this one, have a Class I recall.

Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor, 1 with a Class I recall12 generations back: 1 other ancestor, 1 with a Class I recall13 generations back: 1 other ancestor, 1 with a Class I recall14 generations back: 1 other ancestor5 generations back: 1 other ancestor6 generations back: 1 other ancestor8 generations back: 2 other ancestors9 generations back: 1 other ancestor10 generations back: 3 other ancestors 1 generation forward: 3 descendants2 generations forward: 1 descendant K242775 · InPen System App (MMT-8060 (iOS), MMT-8061 (Android)) · Medtronic Minimed · 2024 · recalledK201337 · InPen System · Companion Medical, Inc. · 2020 · recalledK190487 · InPen Dose Calculator · Companion Medical, Inc. · 2020 · recalledK160629 · InPen · Companion Medical, Inc. · 2016 · recalledK150910 · ACCU-CHEK Connect Diabetes Management App · Roche Diabetes Care, Inc. · 2015 · recalledK141929 · Accu-Chek Connect Diabetes Management App · Roche Diagnostics Corporation · 2015 · recalledK111353 · Accu-Check Combo System · Roche Diagnostics · 2012 · recalledK101299 · Accu-Chek Aviva Plus Blood Glucose Monitoring System · Roche Diagnostics · 2011K060620 · Accu-Chek Aviva Test Strips · Roche Diagnostics Corp. · 2006 · recalledK043474 · Accu-Chek Aviva System · Roche Diagnostics Corp. · 2005 · recalledK010362 · Accu-Chek Comfort Curve Test Strip · Roche Diagnostics Corp. · 2001 · recalled−10−50
  1. Root · 11 generations backK010362 Accu-Chek Comfort Curve Test StripRoche Diagnostics Corp. · 2001 · Recalled
  2. 7 generations between
  3. 3 generations backK190487 InPen Dose CalculatorCompanion Medical, Inc. · 2020 · Recalled
  4. 2 generations backK201337 InPen SystemCompanion Medical, Inc. · 2020 · Recalled
  5. 1 generation backK242775 InPen System App (MMT-8060 (iOS), MMT-8061 (Android))Medtronic Minimed · 2024 · Recalled
  6. This deviceK251032 MiniMed 780G insulin pumpMedtronic Minimed, Inc. · 2025
  7. Descendant · cites it directly · 1 of its ownK253701 SmartGuard Technology; Predictive Low Glucose TechnologyMedtronic Minimed · 2026
  8. Descendant · 2 generations on · 0 of its ownK261962 SmartGuard Technology; Predictive Low Glucose TechnologyMedtronic MiniMed · 2026
  9. Descendant · cites it directly · 0 of its ownK253743 MiniMed Flex pumpMedtronic Minimed, Inc. · 2026 · Recalled
  10. Descendant · cites it directly · 0 of its ownK253470 MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit…Medtronic Minimed, Inc. · 2026
Show all 11 generations
  1. Root · 11 generations backK010362 Accu-Chek Comfort Curve Test StripRoche Diagnostics Corp. · 2001 · Recalled
  2. 10 generations backK043474 Accu-Chek Aviva SystemRoche Diagnostics Corp. · 2005 · Recalled
  3. 9 generations backK060620 Accu-Chek Aviva Test StripsRoche Diagnostics Corp. · 2006 · Recalled
  4. 8 generations backK101299 Accu-Chek Aviva Plus Blood Glucose Monitoring SystemRoche Diagnostics · 2011
  5. 7 generations backK111353 Accu-Check Combo SystemRoche Diagnostics · 2012 · Recalled
  6. 6 generations backK141929 Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation · 2015 · Recalled
  7. 5 generations backK150910 ACCU-CHEK Connect Diabetes Management AppRoche Diabetes Care, Inc. · 2015 · Recalled
  8. 4 generations backK160629 InPenCompanion Medical, Inc. · 2016 · Recalled
  9. 3 generations backK190487 InPen Dose CalculatorCompanion Medical, Inc. · 2020 · Recalled
  10. 2 generations backK201337 InPen SystemCompanion Medical, Inc. · 2020 · Recalled
  11. 1 generation backK242775 InPen System App (MMT-8060 (iOS), MMT-8061 (Android))Medtronic Minimed · 2024 · Recalled
  12. This deviceK251032 MiniMed 780G insulin pumpMedtronic Minimed, Inc. · 2025

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K253701SmartGuard Technology; Predictive Low Glucose TechnologyMedtronic Minimed · 2026 · cites it directly1descendant
  • K261962SmartGuard Technology; Predictive Low Glucose TechnologyMedtronic MiniMed · 2026 · 2 generations on0descendants
  • K253743MiniMed Flex pumpMedtronic Minimed, Inc. · 2026 · cites it directly0descendants
  • K253470MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit (MMT-1894)Medtronic Minimed, Inc. · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K010362Accu-Chek Comfort Curve Test StripRoche Diagnostics Corp.20016
−10K043474Accu-Chek Aviva SystemRoche Diagnostics Corp.200510
−9K060620Accu-Chek Aviva Test StripsRoche Diagnostics Corp.20061
−8K101299Accu-Chek Aviva Plus Blood Glucose Monitoring SystemRoche Diagnostics2011—
−7K111353Accu-Check Combo SystemRoche Diagnostics20123
−6K141929Accu-Chek Connect Diabetes Management AppRoche Diagnostics Corporation20151
−5K150910ACCU-CHEK Connect Diabetes Management AppRoche Diabetes Care, Inc.20156
−4K160629InPenCompanion Medical, Inc.20161
−3K190487InPen Dose CalculatorCompanion Medical, Inc.20201
−2K201337InPen SystemCompanion Medical, Inc.20202
−1K242775InPen System App (MMT-8060 (iOS), MMT-8061 (Android))Medtronic Minimed20243
0 (this device)K251032MiniMed 780G insulin pumpMedtronic Minimed, Inc.2025—
+1K253701SmartGuard Technology; Predictive Low Glucose TechnologyMedtronic Minimed2026—
+2K261962SmartGuard Technology; Predictive Low Glucose TechnologyMedtronic MiniMed2026—
+1K253743MiniMed Flex pumpMedtronic Minimed, Inc.20261
+1K253470MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit (MMT-1894)Medtronic Minimed, Inc.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200330
200510
200610
200810
201310
201910
202011
202211
202311

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202640
FIG. 1 — Lineage of K251032, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new QFG device

  1. #1 K253470 MiniMed 780G Insulin Pump (MMT-1884); MiniMed…score 14.0 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K232380t:slim X2 Insulin Pump with Interoperable Technology
Summary page 5
…Class II Predicate Device Information Product Code Predicate Device QFG Tandem t:slim X2 Insulin Pump with Interoperable Technology (K232380) NDC InPen Dose Calculator (K242775)…
Tandem Diabetes Care, Inc.2023Class I
*K242775InPen System App (MMT-8060 (iOS), MMT-8061 (Android))
Summary page 5
…oduct Code Predicate Device QFG Tandem t:slim X2 Insulin Pump with Interoperable Technology (K232380) NDC InPen Dose Calculator (K242775) Page 1 of 15…
Medtronic Minimed20243

Also cleared against K232380: 1 other device →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K253701SmartGuard Technology; Predictive Low Glucose TechnologyMedtronic Minimed20261
K253743MiniMed Flex pumpMedtronic Minimed, Inc.20260
K253470MiniMed 780G Insulin Pump (MMT-1884); MiniMed 780G Insulin Pump Kit (MMT-1894)Medtronic Minimed, Inc.20260

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 13, 2026Class IIZ-2980-2026Process designLoaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia…
Jul 13, 2026Class IIZ-2982-2026Process designLoaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia…
May 27, 2026Class IIZ-2698-2026Software designA software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by…
Feb 13, 2026Class IIZ-1741-2026Device DesignAll Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site…
Jan 30, 2026Class IIZ-1635-2026Software change controlThree software defects (Pump Error 53, BG check, and Critical Pump Error (Open Book Image)) that occurred as a result of software updates (versions 6.60, 6.61, and 6.62), which could result in a risk of hypoglycemia or hyperglycemia due…
Nov 2, 2025Class IIZ-1238-2026Process designA software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K232380t:slim X2 Insulin Pump with Interoperable TechnologyTandem Diabetes Care, Inc. · 2023 · Direct predicate · ◆ in the 2023 ancestors bar of FIG. 11 Class I recall
  • K203234t:slim X2 Insulin Pump with Interoperable Technology (with t:connect mobile app)Tandem Diabetes Care, Inc. · 2022 · 2 generations back · ◆ in the 2022 ancestors bar of FIG. 11 Class I recall
  • K201214t:slim X2 Insulin Pump with Interoperable TechnologyTandem Diabetes Care, Inc. · 2020 · 3 generations back · ◆ in the 2020 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 24 years and passes through 7 companies; a new generation was cleared every 1.3 years on median.
  • 18 of the 23 ancestors have a recall of any class; 3 have a Class I recall.
  • Of its descendants, K253701 (Medtronic Minimed, 2026) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K232380.

How does its category compare?

All QFG devices

QFG Alternate Controller Enabled Insulin Infusion Pump: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · QFG

5in 2025

2019: 22020: 12022: 22023: 92024: 52025: 52026 (partial year): 6
2007high 9 (2023) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QFG

44in 2025

2023: 150 days median2024: 89 days median2025: 44 days median2026 (partial year): 103 days median
2007high 150 (2023) · low 44 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · QFG

8so far in 2026, the most since 2007

2022: 12023: 1, 1 Class I2025: 3, 1 Class I2026 (partial year): 8, 1 Class I
2007high 8 (2026) · low 0 (2007)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · QFG

324,360in 2025

2020: 30,5222021: 74,7012022: 129,1192023: 188,0142024: 273,1802025: 324,360
2007high 324,360 (2025) · low 0 (2007)2025

MAUDE reports naming product code QFG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20170—00
20180—00
20192—00
20201—030,522
20210—074,701
20222—1129,119
202391501 (1)188,014
20245890273,180
20255443 (1)324,360
202661038 (1)—

What is it indicated for?

The MiniMed 780G insulin pump is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. The MiniMed 780G insulin pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The MiniMed 780G insulin pump contains a bolus…

Read the full statement (109 words)

The MiniMed 780G insulin pump is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.

The MiniMed 780G insulin pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices.

The MiniMed 780G insulin pump contains a bolus calculator that calculates an insulin dose based on user-entered data.

The MiniMed 780G insulin pump is indicated for use in individuals 7 years of age and older.

The MiniMed 780G insulin pump is intended for single patient use and requires a prescription.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K251032: MiniMed 780G insulin pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K251032

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026