510(k) Report

510(k) · Cleared

Cleared

Pentacam® Cornea OCT

K251848510(k) TraditionalCleared Mar 13, 2026270-day reviewClass II21 CFR 886.1850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Oculus Optikgeräte GmbH · MXK Device, Analysis, Anterior Segment

Cleared on Mar 13, 2026 as substantially equivalent to K222166 (Optovue, Inc., 2022) after a 270-day review. Its family tree reaches back 10 generations to 2002.

Decided
Mar 13, 2026
Review
270 days
Class
Class II
Product code
MXK
Recalls
None
Predicates
1

Its longest line of descent runs back 10 generations to K012727 (2002). 37 ancestors in all, the oldest K810880 (1981). No later device cites it as a predicate yet.

  1. Root · 10 generations backK012727 Humphrey OCT3Carl Zeiss Ophthalmic Systems, Inc. · 2002
  2. 6 generations between
  3. 3 generations backK153080 RTVue XR OCT Avanti with AngioVue SoftwareOptovue, Inc. · 2016
  4. 2 generations backK180660 AvantiOptovue, Inc. · 2018
  5. 1 generation backK222166 SOLIXOptovue, Inc. · 2022
  6. This deviceK251848 Pentacam® Cornea OCTOculus Optikgeräte GmbH · 2026
Show all 10 generations
  1. Root · 10 generations backK012727 Humphrey OCT3Carl Zeiss Ophthalmic Systems, Inc. · 2002
  2. 9 generations backK030433 Stratusoct with RNFL Normative DatabaseCarl Zeiss Meditec, Inc. · 2003 · Recalled
  3. 8 generations backK033123 Stratusoct with RNFL & Macula Normative DatabaseCarl Zeiss Meditec, Inc. · 2004 · Recalled
  4. 7 generations backK063343 Bioptigen Spectral Domain Ophthalmic Imaging SystemBioptigen, Inc. · 2006
  5. 6 generations backK071250 CA, Model: CamOptovue, Inc. · 2007
  6. 5 generations backK101505 Rtvue Model Rtvue 100Optovue, Inc. · 2010 · Recalled
  7. 4 generations backK120238 Rtvue XR OctOptovue, Inc. · 2012
  8. 3 generations backK153080 RTVue XR OCT Avanti with AngioVue SoftwareOptovue, Inc. · 2016
  9. 2 generations backK180660 AvantiOptovue, Inc. · 2018
  10. 1 generation backK222166 SOLIXOptovue, Inc. · 2022
  11. This deviceK251848 Pentacam® Cornea OCTOculus Optikgeräte GmbH · 2026
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K012727Humphrey OCT3Carl Zeiss Ophthalmic Systems, Inc.2002—
−9K030433Stratusoct with RNFL Normative DatabaseCarl Zeiss Meditec, Inc.20031
−8K033123Stratusoct with RNFL & Macula Normative DatabaseCarl Zeiss Meditec, Inc.20041
−7K063343Bioptigen Spectral Domain Ophthalmic Imaging SystemBioptigen, Inc.2006—
−6K071250CA, Model: CamOptovue, Inc.2007—
−5K101505Rtvue Model Rtvue 100Optovue, Inc.20101
−4K120238Rtvue XR OctOptovue, Inc.2012—
−3K153080RTVue XR OCT Avanti with AngioVue SoftwareOptovue, Inc.2016—
−2K180660AvantiOptovue, Inc.2018—
−1K222166SOLIXOptovue, Inc.2022—
0 (this device)K251848Pentacam® Cornea OCTOculus Optikgeräte GmbH2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
199110
199520
199610
199710
199830
199910
200110
200230
200320
200420
200510
200630
200710
200910
201010
201110
201310
FIG. 1 — Lineage of K251848, 2002–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MXK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MXK device

  1. #1 K252348 Tomey Optical Biometer OA-2000 (OA-2000)score 11.7 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K222166SOLIX
Summary page 5
…ber: 21 C.F.R. § 886.1570 Device class: II Product Code: OBO, MXK Predicate Devices Primary Predicate: Optovue, Inc., SOLIX (K222166) Reference Device: OCULUS Optikgeräte GmbH, Pentacam® AXL Wave (K201724) Pentacam® Cornea OCT…
Optovue, Inc.2022—
†K201724Pentacam AXL WaveOculus Optikgerate GmbH20202

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 5 companies; a new generation was cleared every 2.6 years on median.
  • 6 of the 37 ancestors have a recall of any class.

How does its category compare?

All MXK devices

MXK Device, Analysis, Anterior Segment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2026, this device’s year, is highlighted.

FDA decisions per year · MXK

1in 2025

2007: 12008: 12015: 12016: 12019: 22020: 12021: 12022: 12023: 12024: 12025: 12026 (partial year): 1
2007high 2 (2019) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MXK

Too few MXK 510(k)s in any year since 2007 to give a median.

Recalls per year · MXK

1in 2017

2017: 1
2007high 1 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MXK

1in 2025

2010: 12013: 12017: 22018: 22019: 22020: 22024: 12025: 1
2007high 2 (2017) · low 0 (2007)2025

MAUDE reports naming product code MXK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20081—00
20090—00
20100—01
20110—00
20120—00
20130—01
20140—00
20151—00
20161—00
20170—12
20180—02
20192—02
20201—02
20211—00
20221—00
20231—00
20241—01
20251—01
20261—0—

What is it indicated for?

The Pentacam® Cornea OCT is a multifunction ophthalmic imaging device designed to acquire images of the anterior segment using Scheimpflug imaging (digital CMOS camera with slit-lamp illumination) and OCT.

Scheimpflug imaging provides anterior segment visualization and quantitative anterior segment measurements.

The OCT component of the system uses broadband Multi-SLD and is intended for the in vivo imaging, cross-sectional, and the three-dimensional imaging and measurement of corneal epithelium.

The Penatcam® Cornea OCT is indicated for use as a diagnostic device to aid in the detection and management of ocular diseases affecting the anterior segment of the eye.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K251848: Pentacam® Cornea OCT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K251848

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026