510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Pentacam AXL Wave

K201724510(k) TraditionalCleared Oct 21, 2020120-day reviewClass II21 CFR 886.1850Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Oculus Optikgerate GmbH · MXK Device, Analysis, Anterior Segment

Cleared on Oct 21, 2020 as substantially equivalent to K152311 (Oculus Optikgerate GmbH, 2016) and 1 other predicate after a 120-day review. Its family tree reaches back 8 generations to 1996. It has been recalled 2 times.

Decided
Oct 21, 2020
Review
120 days
Class
Class II
Product code
MXK
Recalls
2
Predicates
2

Its longest line of descent runs back 8 generations to K955442 (1996). 26 ancestors in all, the oldest K781994 (1979). No later device cites it as a predicate yet.

  1. Root · 8 generations backK955442 Kowa SC-1200 Photo Slit Lamp SystemKowa Optimed, Inc. · 1996
  2. 7 generations backK991284 Anterior Eye-Segment Analysis SystemNidek, Inc. · 1999
  3. 6 generations backK012873 Documenting Laser Slit Lamp, DP 2010Bausch & Lomb, Inc. · 2001
  4. 5 generations backK030393 Optical Low Coherence Reflectometry (Olcr) PachymeterHaag-Streit Holding AG · 2003
  5. 4 generations backK082891 Lenstar, Model LS900Haag-Streit AG · 2009 · Recalled
  6. 3 generations backK101182 Iolmaster 500Carl Zeiss Meditec, AG · 2010 · Recalled
  7. 2 generations backK122418 Iolmaster 500Carl Zeiss Meditec, AG · 2013 · Recalled
  8. 1 generation backK152311 Pentacam AXLOculus Optikgerate GmbH · 2016 · Recalled
  9. This deviceK201724 Pentacam AXL WaveOculus Optikgerate GmbH · 2020
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K955442Kowa SC-1200 Photo Slit Lamp SystemKowa Optimed, Inc.1996—
−7K991284Anterior Eye-Segment Analysis SystemNidek, Inc.1999—
−6K012873Documenting Laser Slit Lamp, DP 2010Bausch & Lomb, Inc.2001—
−5K030393Optical Low Coherence Reflectometry (Olcr) PachymeterHaag-Streit Holding AG2003—
−4K082891Lenstar, Model LS900Haag-Streit AG20093
−3K101182Iolmaster 500Carl Zeiss Meditec, AG20104
−2K122418Iolmaster 500Carl Zeiss Meditec, AG20134
−1K152311Pentacam AXLOculus Optikgerate GmbH20163
0 (this device)K201724Pentacam AXL WaveOculus Optikgerate GmbH2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198610
198720
199010
199320
199420
199620
199710
199910
200010
200320
201010
201510
FIG. 1 — Lineage of K201724, 1996–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MXK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MXK device

  1. #1 K251848 Pentacam® Cornea OCTscore 13.7 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K152311Pentacam AXL
Summary page 4
….F.R. § 886.1850 Device class: II Product Code: MXK, HJO Predicate Devices Primary Predicate: OCULUS Optikgeräte GmbH, Pentacam AXL (K152311) Secondary Predicate: LUNEAU SAS, VX120 (K143086) Page 1 of 10…
Oculus Optikgerate GmbH20163
*K143086VX120 Ophthalmic Diagnostic Device
Summary page 4
…Product Code: MXK, HJO Predicate Devices Primary Predicate: OCULUS Optikgeräte GmbH, Pentacam AXL (K152311) Secondary Predicate: LUNEAU SAS, VX120 (K143086) Page 1 of 10…
Luneau Sas2015—

Also cleared against K152311: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 8, 2023Class IIZ-2482-2023Process controlTheir is a potential that optical devices with insufficient anti-reflective coating may lead to incorrect axial length measurements.
Jul 8, 2022Class IIZ-1672-2022Software designDue to a software issue, IOL calculator printout often does not accurately reflect the alignment axis and incision position when planning toric IOLs

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 7 companies; a new generation was cleared every 2.8 years on median.
  • 6 of the 26 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K152311.

How does its category compare?

All MXK devices

MXK Device, Analysis, Anterior Segment: FDA decisions, review times, recalls and adverse event reports over 20 years. 2020, this device’s year, is highlighted.

FDA decisions per year · MXK

1in 2025

2007: 12008: 12015: 12016: 12019: 22020: 12021: 12022: 12023: 12024: 12025: 12026 (partial year): 1
2007high 2 (2019) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MXK

Too few MXK 510(k)s in any year since 2007 to give a median.

Recalls per year · MXK

1in 2017

2017: 1
2007high 1 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MXK

1in 2025

2010: 12013: 12017: 22018: 22019: 22020: 22024: 12025: 1
2007high 2 (2017) · low 0 (2007)2025

MAUDE reports naming product code MXK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20081—00
20090—00
20100—01
20110—00
20120—00
20130—01
20140—00
20151—00
20161—00
20170—12
20180—02
20192—02
20201—02
20211—00
20221—00
20231—00
20241—01
20251—01
20261—0—

What is it indicated for?

The Pentacam AXL Wave is intended to image the anterior segment of the eye which includes the cornea, pupil, anterior chamber and lens. It is indicated for the evaluation of - corneal shape, - condition of the lens (opaque crystalline lens), - the anterior chamber angle, - anterior chamber depth, - the volume of the anterior chamber, - anterior or posterior cortical opacity, - the location of cataracts using cross slit imaging with densitometry, -…

Read the full statement (113 words)

The Pentacam AXL Wave is intended to image the anterior segment of the eye which includes the cornea, pupil, anterior chamber and lens. It is indicated for the evaluation of

- corneal shape,

- condition of the lens (opaque crystalline lens),

- the anterior chamber angle,

- anterior chamber depth,

- the volume of the anterior chamber,

- anterior or posterior cortical opacity,

- the location of cataracts using cross slit imaging with densitometry,

- corneal thickness,

- axial length,

- white-to-white distance,

- optical aberrations of the eye,

- and retroillumination imaging.

The Pentacam AXL Wave also performs calculations to assist physicians in determining the power of the intraocular lens for implantation.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K201724: Pentacam AXL Wave.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K201724

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026