510(k) Report

510(k) · Cleared

Cleared

Universal Spinal System

K252729510(k) TraditionalCleared Nov 18, 202582-day reviewClass II21 CFR 888.3070

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Double Medical Technology, Inc. · NKB Thoracolumbosacral Pedicle Screw System

Cleared on Nov 18, 2025 as substantially equivalent to K040072 (Medtronic Sofamor Danek, Inc., 2004) and 3 other predicates after an 82-day review. Its family tree reaches back 19 generations to 2004.

Decided
Nov 18, 2025
Review
82 days
Class
Class II
Product code
NKB
Recalls
None
Predicates
4

Its longest line of descent runs back 19 generations to K040583 (2004). 221 ancestors in all, the oldest K781443 (1978). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2004: 21 other ancestors1996: 5 other ancestors2011: 5 other ancestors2016: 2 other ancestors2014: 1 other ancestor2010: 8 other ancestors2009: 7 other ancestors2008: 8 other ancestors2006: 15 other ancestors2005: 22 other ancestors2003: 14 other ancestors2002: 10 other ancestors2000: 16 other ancestors1999: 10 other ancestors2013: 3 other ancestors2012: 3 other ancestors2007: 8 other ancestors2001: 20 other ancestors1998: 10 other ancestors1984: 1 other ancestor1997: 4 other ancestors1995: 5 other ancestors1978: 1 other ancestor1993: 2 other ancestors1985: 1 other ancestor×7: 7 generations cleared close together, K040583 (2004), K041118 (2004), K042025 (2004), K043151 (2005), K050809 (2005), K052187 (2005), K053101 (2006) ×2: 2 generations cleared close together, K091442 (2009), K091974 (2009) K102555 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2010×2: 2 generations cleared close together, K112473 (2011), K113174 (2011) K121680 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2012×3: 3 generations cleared close together, K131321 (2013), K132471 (2013), K140276 (2014) K142381 · Xia 3 Spinal System · Stryker Corp. · 2014K200245 · EXPEDIUM® Spine System; EXPEDIUM VERSE® Spine System · Medos International Sàrl · 2020K222473 · EXPEDIUM® Spine System, EXPEDIUM VERSE® Spine System: ALTALYNE™ Ultra Rods · Medos-International Sarl · 202219801990200020102020
Other ancestors, by generation 1 generation back: 2 other ancestors2 generations back: 2 other ancestors3 generations back: 22 other ancestors4 generations back: 22 other ancestors5 generations back: 53 other ancestors6 generations back: 57 other ancestors7 generations back: 30 other ancestors8 generations back: 11 other ancestors9 generations back: 3 other ancestors K222473 · EXPEDIUM® Spine System, EXPEDIUM VERSE® Spine System: ALTALYNE™ Ultra Rods · Medos-International Sarl · 2022K200245 · EXPEDIUM® Spine System; EXPEDIUM VERSE® Spine System · Medos International Sàrl · 2020K142381 · Xia 3 Spinal System · Stryker Corp. · 2014K140276 · CD Horizon Spinal System · Medtronic Sofamor Danek · 2014K132471 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2013K131321 · CD Horizon Spinal System · Medtronic Sofamor Danek USA · 2013K121680 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2012K113174 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2011 · recalledK112473 · CD Horizon Spinal System · Medtronic Sofamor Danek, Inc. · 2011K102555 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2010K091974 · Modification to CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2009 · recalledK091442 · CD Horizon Spinal System · Medtronic Sofamor Danek · 2009K053101 · CD Horizon Spinal System · Medtronic Sofamor Danek · 2006K052187 · Modification to: CD Horizon Spinal System · Medtronic Sofamor Danek · 2005K050809 · CD Horizon Spinal System · Medtronic Sofamor Danek USA, Inc. · 2005K043151 · Modification to: CD Horizon Spinal System · Medtronic Sofamor Danek · 2005K042025 · Modification to CD Horizon Spinal System · Medtronic Sofamor Danek, Inc. · 2004 · recalledK041118 · Modification to CD Horizon Spinal System · Medtronic Sofamor Danek · 2004K040583 · Modification to: CD Horizon Spinal System · Medtronic Sofamor Danek · 2004−15−10−50
  1. Root · 19 generations backK040583 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2004
  2. 15 generations between
  3. 3 generations backK142381 Xia 3 Spinal SystemStryker Corp. · 2014
  4. 2 generations backK200245 EXPEDIUM® Spine System; EXPEDIUM VERSE® Spine SystemMedos International Sàrl · 2020
  5. 1 generation backK222473 EXPEDIUM® Spine System, EXPEDIUM VERSE® Spine System: ALTALYNE™…Medos-International Sarl · 2022
  6. This deviceK252729 Universal Spinal SystemDouble Medical Technology, Inc. · 2025
Show all 19 generations
  1. Root · 19 generations backK040583 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2004
  2. 18 generations backK041118 Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek · 2004
  3. 17 generations backK042025 Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc. · 2004 · Recalled
  4. 16 generations backK043151 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2005
  5. 15 generations backK050809 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2005
  6. 14 generations backK052187 Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek · 2005
  7. 13 generations backK053101 CD Horizon Spinal SystemMedtronic Sofamor Danek · 2006
  8. 12 generations backK091442 CD Horizon Spinal SystemMedtronic Sofamor Danek · 2009
  9. 11 generations backK091974 Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2009 · Recalled
  10. 10 generations backK102555 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2010
  11. 9 generations backK112473 CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc. · 2011
  12. 8 generations backK113174 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2011 · Recalled
  13. 7 generations backK121680 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2012
  14. 6 generations backK131321 CD Horizon Spinal SystemMedtronic Sofamor Danek USA · 2013
  15. 5 generations backK132471 CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc. · 2013
  16. 4 generations backK140276 CD Horizon Spinal SystemMedtronic Sofamor Danek · 2014
  17. 3 generations backK142381 Xia 3 Spinal SystemStryker Corp. · 2014
  18. 2 generations backK200245 EXPEDIUM® Spine System; EXPEDIUM VERSE® Spine SystemMedos International Sàrl · 2020
  19. 1 generation backK222473 EXPEDIUM® Spine System, EXPEDIUM VERSE® Spine System: ALTALYNE™…Medos-International Sarl · 2022
  20. This deviceK252729 Universal Spinal SystemDouble Medical Technology, Inc. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−19K040583Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek2004—
−18K041118Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek2004—
−17K042025Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc.20044
−16K043151Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek2005—
−15K050809CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.2005—
−14K052187Modification to: CD Horizon Spinal SystemMedtronic Sofamor Danek2005—
−13K053101CD Horizon Spinal SystemMedtronic Sofamor Danek2006—
−12K091442CD Horizon Spinal SystemMedtronic Sofamor Danek2009—
−11K091974Modification to CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.20091
−10K102555CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.2010—
−9K112473CD Horizon Spinal SystemMedtronic Sofamor Danek, Inc.2011—
−8K113174CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.20111
−7K121680CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.2012—
−6K131321CD Horizon Spinal SystemMedtronic Sofamor Danek USA2013—
−5K132471CD Horizon Spinal SystemMedtronic Sofamor Danek USA, Inc.2013—
−4K140276CD Horizon Spinal SystemMedtronic Sofamor Danek2014—
−3K142381Xia 3 Spinal SystemStryker Corp.2014—
−2K200245EXPEDIUM® Spine System; EXPEDIUM VERSE® Spine SystemMedos International Sàrl2020—
−1K222473EXPEDIUM® Spine System, EXPEDIUM VERSE® Spine System: ALTALYNE™ Ultra RodsMedos-International Sarl2022—
0 (this device)K252729Universal Spinal SystemDouble Medical Technology, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197810
198410
198510
199320
199550
199650
199740
1998100
1999100
2000160
2001200
2002100
2003140
2004210
2005220
2006150
200780
200880
200970
201080
201150
201230
201330
201410
201620
FIG. 1 — Lineage of K252729, 2004–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NKB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NKB device

  1. #1 K223494 CD Horizon Spinal Systemscore 40.4 · 30 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040072Modification to CD Horizon Spinal System
Summary page 5
…II Classification panel Orthopedic 3. Predicate Device Primary predicate device: K040072 Medtronic CD HORIZON Spinal System Legally marketed Additional predicate devices: device(s) to which…
Medtronic Sofamor Danek, Inc.2004—
*K113174CD Horizon Spinal System
Summary page 5
…K040072 Medtronic CD HORIZON Spinal System Legally marketed Additional predicate devices: device(s) to which K113174 Medtronic CD HORIZON® Spinal System equivalence is claimed K222473 Expedium Spine System…
Medtronic Sofamor Danek USA, Inc.20111
*K222473EXPEDIUM® Spine System, EXPEDIUM VERSE® Spine System: ALTALYNE™ Ultra Rods
Summary page 5
…Additional predicate devices: device(s) to which K113174 Medtronic CD HORIZON® Spinal System equivalence is claimed K222473 Expedium Spine System K963655 CD HORIZON™ Anterior Spinal System Reason for 510(k)…
Medos-International Sarl2022—
*K963655CD Horizon Anterior Spinal System
Summary page 5
…K113174 Medtronic CD HORIZON® Spinal System equivalence is claimed K222473 Expedium Spine System K963655 CD HORIZON™ Anterior Spinal System Reason for 510(k) New device(Implant) submission 4. Device…
Sofamor Danek USA,Inc.1996—

Also cleared against K040072: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 8 companies; a new generation was cleared every 5 months on median.
  • 24 of the 221 ancestors have a recall of any class.
  • 1 other device also cites its predicate K040072.

How does its category compare?

All NKB devices

NKB Thoracolumbosacral Pedicle Screw System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · NKB

24in 2025

2007: 382008: 362009: 442010: 372011: 482012: 602013: 352014: 462015: 552016: 502017: 632018: 562019: 392020: 482021: 322022: 342023: 442024: 392025: 242026 (partial year): 30
2007high 63 (2017) · low 24 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NKB

61in 2025

2007: 87 days median2008: 59 days median2009: 71 days median2010: 111 days median2011: 84 days median2012: 79 days median2013: 73 days median2014: 90 days median2015: 90 days median2016: 84 days median2017: 69 days median2018: 61 days median2019: 84 days median2020: 68 days median2021: 61 days median2022: 88 days median2023: 72 days median2024: 57 days median2025: 61 days median2026 (partial year): 91 days median
2007high 111 (2010) · low 57 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NKB

1in 2025

2007: 42008: 52009: 332010: 42011: 32012: 132014: 3, 1 Class I2015: 32016: 12017: 12018: 32019: 12020: 12022: 52023: 22024: 32025: 12026 (partial year): 3
2007high 33 (2009) · low 0 (2013)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · NKB

1,171in 2025

2007: 342008: 822009: 1712010: 4622011: 5772012: 1,7472013: 1,9622014: 1,6712015: 1,4462016: 1,5852017: 1,6182018: 1,6842019: 1,9152020: 2,0762021: 1,7452022: 1,3392023: 9192024: 1,6142025: 1,171
2007high 2,076 (2020) · low 34 (2007)2025

MAUDE reports naming product code NKB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073887434
20083659582
2009447133171
2010371114462
201148843577
20126079131,747
2013357301,962
201446903 (1)1,671
2015559031,446
2016508411,585
2017636911,618
2018566131,684
2019398411,915
2020486812,076
2021326101,745
2022348851,339
202344722919
2024395731,614
2025246111,171
202630913—

What is it indicated for?

The Universal Spinal System is intended for non-cervical posterior and anterolateral fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e.…

Read the full statement (216 words)

The Universal Spinal System is intended for non-cervical posterior and anterolateral fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

When used for posterior, non-cervical, pedicle screw fixation in pediatric patients, the Universal Spinal System is indicated as an adjunct to fusion to treat progressive spinal deformities (i.e. scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the Universal Spinal System is intended to treat pediatric patients diagnosed with spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. This system is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The HOBBIT Growth Connectors are indicated in patients under 10 years of age with potential for additional spinal growth who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis, as part of a growing rod construct.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K252729: Universal Spinal System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K252729

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026