510(k) Report

510(k) · Cleared

Cleared

VINScreen Urine Drug Test Cup; VINScreen Urine Drug Home Test Cup

K252867510(k) TraditionalCleared Oct 9, 202530-day reviewClass II21 CFR 862.3100

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Advin Biotech, Inc. · NFT Test, Amphetamine, Over The Counter

Cleared on Oct 9, 2025 as substantially equivalent to K244043 (Hangzhou AllTest Biotech Co., Ltd., 2025) after a 30-day review. Its family tree reaches back 10 generations to 2000.

Decided
Oct 9, 2025
Review
30 days
Class
Class II
Product code
NFT
Recalls
None
Predicates
1

Its longest line of descent runs back 10 generations to K001178 (2000). 20 ancestors in all, the oldest K971218 (1997). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2004: 1 other ancestor2002: 4 other ancestors2001: 1 other ancestor2000: 2 other ancestors1997: 2 other ancestorsK001178 · Cedia Dau 6-Acetylmorphine Assay · Microgenics Corp. · 2000K040316 · Cedia Buprenorphine Assay · Microgenics Corp. · 2004K042988 · Rapid One Step Buprenorphine Test Card · Tianjin New Bay Bioresearch Co., Ltd. · 2005K060527 · Instant-View Bup/Nbup Urine Test. Instant-Verdict Bup/Nbup Urine Test; Instant-Confirmatory Bup/Nbup Urine Test · Alfa Scientific Designs, Inc. · 2006K063545 · Multi-Drug of Abuse Urine Test,Urine Cup Test, Amphetamine Urine Test (Cassette, Dip-Strip), Cocaine, Methamphetamine · Alfa Scientific Designs, Inc. · 2007K152122 · Instant-View Multi-Drug Urine Test Cup (Home Use), Instant-View Multi-Drug Urine Test Panel (Home Use) · Alfa Scientific Designs, Inc. · 2016K173303 · INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use), INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use) · Alfa Scientific Designs, Inc. · 2018K182738 · Single and Multi-Drug Rapid Test Panel With Adulteration (Urine), Single and Multi-Drug Rapid Test Panel (Urine), Single and Multi-Drug Rapid Test Cup With Adulteration (Urine), Single and Multi-Drug Rapid Test Cup (Urine), Single Drug Rapid Test Dipstick (Urine), Single and Multi-Drug Home Rapid Test Panel (Urine), Single and Multi-Drug Home Rapid Test Cup (Urine), Single Drug Home Rapid Test Dipstick (Urine) · Hangzhou AllTest Biotech Co., Ltd. · 2019K233019 · AllTest Multi-Drug Rapid Test Cup; AllTest Multi-Drug Rapid Test Panel; AllTest Multi-Drug Rapid Test Cup Rx; AllTest Multi-Drug Rapid Test Panel Rx · Hangzhou AllTest Biotech Co., Ltd. · 2023K244043 · AllTest Multi-Drug Rapid Test Cup ; AllTest Multi-Drug Test Cup · Hangzhou AllTest Biotech Co., Ltd. · 20251995200020052010201520202025
Other ancestors, by generation 7 generations back: 8 other ancestors8 generations back: 2 other ancestors K244043 · AllTest Multi-Drug Rapid Test Cup ; AllTest Multi-Drug Test Cup · Hangzhou AllTest Biotech Co., Ltd. · 2025K233019 · AllTest Multi-Drug Rapid Test Cup; AllTest Multi-Drug Rapid Test Panel; AllTest Multi-Drug Rapid Test Cup Rx; AllTest Multi-Drug Rapid Test Panel Rx · Hangzhou AllTest Biotech Co., Ltd. · 2023K182738 · Single and Multi-Drug Rapid Test Panel With Adulteration (Urine), Single and Multi-Drug Rapid Test Panel (Urine), Single and Multi-Drug Rapid Test Cup With Adulteration (Urine), Single and Multi-Drug Rapid Test Cup (Urine), Single Drug Rapid Test Dipstick (Urine), Single and Multi-Drug Home Rapid Test Panel (Urine), Single and Multi-Drug Home Rapid Test Cup (Urine), Single Drug Home Rapid Test Dipstick (Urine) · Hangzhou AllTest Biotech Co., Ltd. · 2019K173303 · INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use), INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use) · Alfa Scientific Designs, Inc. · 2018K152122 · Instant-View Multi-Drug Urine Test Cup (Home Use), Instant-View Multi-Drug Urine Test Panel (Home Use) · Alfa Scientific Designs, Inc. · 2016K063545 · Multi-Drug of Abuse Urine Test,Urine Cup Test, Amphetamine Urine Test (Cassette, Dip-Strip), Cocaine, Methamphetamine · Alfa Scientific Designs, Inc. · 2007K060527 · Instant-View Bup/Nbup Urine Test. Instant-Verdict Bup/Nbup Urine Test; Instant-Confirmatory Bup/Nbup Urine Test · Alfa Scientific Designs, Inc. · 2006K042988 · Rapid One Step Buprenorphine Test Card · Tianjin New Bay Bioresearch Co., Ltd. · 2005K040316 · Cedia Buprenorphine Assay · Microgenics Corp. · 2004K001178 · Cedia Dau 6-Acetylmorphine Assay · Microgenics Corp. · 2000−10−50
  1. Root · 10 generations backK001178 Cedia Dau 6-Acetylmorphine AssayMicrogenics Corp. · 2000
  2. 6 generations between
  3. 3 generations backK182738 Single and Multi-Drug Rapid Test Panel With Adulteration (Urine)…Hangzhou AllTest Biotech Co., Ltd. · 2019
  4. 2 generations backK233019 AllTest Multi-Drug Rapid Test Cup; AllTest Multi-Drug Rapid Test…Hangzhou AllTest Biotech Co., Ltd. · 2023
  5. 1 generation backK244043 AllTest Multi-Drug Rapid Test Cup ; AllTest Multi-Drug Test CupHangzhou AllTest Biotech Co., Ltd. · 2025
  6. This deviceK252867 VINScreen Urine Drug Test Cup; VINScreen Urine Drug Home Test CupAdvin Biotech, Inc. · 2025
Show all 10 generations
  1. Root · 10 generations backK001178 Cedia Dau 6-Acetylmorphine AssayMicrogenics Corp. · 2000
  2. 9 generations backK040316 Cedia Buprenorphine AssayMicrogenics Corp. · 2004
  3. 8 generations backK042988 Rapid One Step Buprenorphine Test CardTianjin New Bay Bioresearch Co., Ltd. · 2005
  4. 7 generations backK060527 Instant-View Bup/Nbup Urine Test. Instant-Verdict Bup/Nbup Urine…Alfa Scientific Designs, Inc. · 2006
  5. 6 generations backK063545 Multi-Drug of Abuse Urine Test,Urine Cup Test, Amphetamine Urine…Alfa Scientific Designs, Inc. · 2007
  6. 5 generations backK152122 Instant-View Multi-Drug Urine Test Cup (Home Use), Instant-View…Alfa Scientific Designs, Inc. · 2016
  7. 4 generations backK173303 INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC…Alfa Scientific Designs, Inc. · 2018
  8. 3 generations backK182738 Single and Multi-Drug Rapid Test Panel With Adulteration (Urine)…Hangzhou AllTest Biotech Co., Ltd. · 2019
  9. 2 generations backK233019 AllTest Multi-Drug Rapid Test Cup; AllTest Multi-Drug Rapid Test…Hangzhou AllTest Biotech Co., Ltd. · 2023
  10. 1 generation backK244043 AllTest Multi-Drug Rapid Test Cup ; AllTest Multi-Drug Test CupHangzhou AllTest Biotech Co., Ltd. · 2025
  11. This deviceK252867 VINScreen Urine Drug Test Cup; VINScreen Urine Drug Home Test CupAdvin Biotech, Inc. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K001178Cedia Dau 6-Acetylmorphine AssayMicrogenics Corp.2000—
−9K040316Cedia Buprenorphine AssayMicrogenics Corp.2004—
−8K042988Rapid One Step Buprenorphine Test CardTianjin New Bay Bioresearch Co., Ltd.2005—
−7K060527Instant-View Bup/Nbup Urine Test. Instant-Verdict Bup/Nbup Urine Test…Alfa Scientific Designs, Inc.2006—
−6K063545Multi-Drug of Abuse Urine Test,Urine Cup Test, Amphetamine Urine Test…Alfa Scientific Designs, Inc.2007—
−5K152122Instant-View Multi-Drug Urine Test Cup (Home Use), Instant-View Multi-Drug…Alfa Scientific Designs, Inc.2016—
−4K173303INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use)…Alfa Scientific Designs, Inc.2018—
−3K182738Single and Multi-Drug Rapid Test Panel With Adulteration (Urine), Single and…Hangzhou AllTest Biotech Co., Ltd.2019—
−2K233019AllTest Multi-Drug Rapid Test Cup; AllTest Multi-Drug Rapid Test Panel…Hangzhou AllTest Biotech Co., Ltd.2023—
−1K244043AllTest Multi-Drug Rapid Test Cup ; AllTest Multi-Drug Test CupHangzhou AllTest Biotech Co., Ltd.2025—
0 (this device)K252867VINScreen Urine Drug Test Cup; VINScreen Urine Drug Home Test CupAdvin Biotech, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199720
200020
200110
200240
200410
FIG. 1 — Lineage of K252867, 2000–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NFT? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NFT device

  1. #1 K244043 AllTest Multi-Drug Rapid Test Cup ; AllTest…score 20.3 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K244043AllTest Multi-Drug Rapid Test Cup ; AllTest Multi-Drug Test Cup
Summary page 7
…862.3870, Cannabinoids Test System Toxicology Cannabinoids (THC) 6. Predicate Devices: AllTest Multi-Drug Rapid Test Cup (K244043) 7. Intended Use The VINScreen Urine Drug Home Test Cup is lateral flow immunoassays for rapid detection of multiple…
Hangzhou AllTest Biotech Co., Ltd.2025—

Also cleared against K244043: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 5 companies; a new generation was cleared every 1.2 years on median.
  • None of the 20 ancestors has a recall.
  • 2 other devices also cite its predicate K244043.

How does its category compare?

All NFT devices

NFT Test, Amphetamine, Over The Counter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · NFT

8in 2025

2017: 12018: 62019: 12020: 22021: 12022: 32023: 12024: 42025: 82026 (partial year): 5
2007high 8 (2025) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NFT

37in 2025

2018: 56 days median2022: 63 days median2024: 31 days median2025: 37 days median2026 (partial year): 35 days median
2007high 63 (2022) · low 31 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NFT

No recall names NFT since 2007.

Adverse event reports per year · NFT

No MAUDE report names NFT since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20171—00
201865600
20191—00
20202—00
20211—00
202236300
20231—00
202443100
202583700
20265350—

What is it indicated for?

The VINScreen Urine Drug Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below: Abbreviation Analyte Calibrator Cut-off (ng/mL) 6AM 6-Acetylmorphine 6-Monoacetylmorphine 10 AMP Amphetamine d-Amphetamine 500 or 1000 BAR Secobarbital Secobarbital 300 BUP Buprenorphine Buprenorphine 10 BZO Oxazepam Oxazepam 300 COC Cocaine…

Read the full statement (356 words)

The VINScreen Urine Drug Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

Abbreviation Analyte Calibrator Cut-off (ng/mL)

6AM 6-Acetylmorphine 6-Monoacetylmorphine 10

AMP Amphetamine d-Amphetamine 500 or 1000

BAR Secobarbital Secobarbital 300

BUP Buprenorphine Buprenorphine 10

BZO Oxazepam Oxazepam 300

COC Cocaine Benzoylecgonine 150 or 300

EDDP EDDP 2-ethylidene-1,5-dimethyl- 300

3,3-diphenylpyrrolidine

FEN Fentanyl Fentanyl 1

FEN Norfentanyl Norfentanyl 5

MDMA Ecstasy d,l-Methylenedioxymethamphetamine 500

MET Methamphetamine d-Methamphetamine 500 or 1000

MOR Morphine Morphine 300

MTD Methadone d/l-Methadone 300

OPI Opiates Morphine 2000

OXY Oxycodone Oxycodone 100

PCP Phencyclidine Phencyclidine 25

PPX Propoxyphene Propoxyphene 300

TCA Nortriptyline Nortriptyline 1000

THC Marijuana 11-nor-Δ9-THC-9 COOH 20 or 50

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This in vitro diagnostic device provides only a preliminary test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

The VINScreen Urine Drug Home Test Cup is lateral flow immunoassays for rapid detection of multiple commonly abused drugs in human urine. The detectable drugs and their cutoff concentrations are listed below:

Drug Cut-off (ng/mL) Abbreviation

6-monoacetylmorphine 10 6AM

Amphetamine 500 / 1000 AMP

Secobarbital 300 BAR

Buprenorphine 10 BUP

Oxazepam 300 BZO

Cocaine 150 / 300 COC

EDDP 300 EDDP

Fentanyl 1 FEN

Norfentanyl 5 FEN

Ecstasy 500 MDMA

Methamphetamine 500 / 1000 MET

Methadone 300 MTD

Opiates 300 / 2000 MOR/OPI

Oxycodone 100 OXY

Phencyclidine 25 PCP

Propoxyphene 300 PPX

Nortriptyline 1000 TCA

Marijuana 20 / 50 THC

The single or multi-test cup can include any combination of the analytes listed above, with and without on-board adulteration tests. However, only one cut-off concentration can be included per analyte per device.

This device provides only a preliminary test result. A more specific alternate method must be used in order to obtain a confirmed analytical test result. GC/MS or LC/MS is the preferred confirmatory method.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K252867: VINScreen Urine Drug Test Cup; VINScreen Urine Drug Home Test Cup.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K252867

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026