510(k) · Cleared
Recalled · Class II · 2025Endoscope and Accessories
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Olympus Corp. · GCP Endoscope, Ac-Powered And Accessories
Cleared on Apr 2, 1979 after an 81-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 26 later devices descend from it. It has been recalled 6 times.
- Decided
- Apr 2, 1979
- Review
- 81 days
- Class
- Class II
- Product code
- GCP
- Recalls
- 6
- Predicates
- 0
26 later devices descend from it, 5 citing it directly.
- This deviceK790071 Endoscope and AccessoriesOlympus Corp. · 1979
- Descendant · cites it directly · 18 of its ownK093665 Endo SmartcapByrne Medical, Inc. · 2009 · Recalled
- Descendant · 2 generations on · 7 of its ownK102855 Universal Irrigation Solution HybridByrne Medical, Inc. · 2010
- Descendant · 2 generations on · 3 of its ownK103696 Erbeflo Clevercap Hybrid Tubing/Cap Sets for Olympus Models 160 and…Erbe USA, Inc. · 2011
- Descendant · 3 generations on · 2 of its ownK132340 Erbeflo Clevercap Hybrid Tubing/Cap Set for Pentax Scopes, Erbeflo…Erbe USA, Inc. · 2013 · Recalled
- Descendant · cites it directly · 2 of its ownK992141 Olympus Working Element for ProbesOlympus Winter & Ibe GmbH · 1999
Showing 5 of 26 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 26 · See all- K093665Endo SmartcapByrne Medical, Inc. · 2009 · cites it directly18descendants
- K102855Universal Irrigation Solution HybridByrne Medical, Inc. · 2010 · 2 generations on7descendants
- K103696Erbeflo Clevercap Hybrid Tubing/Cap Sets for Olympus Models 160 and 180 Series ScopesErbe USA, Inc. · 2011 · 2 generations on3descendants
- K132340Erbeflo Clevercap Hybrid Tubing/Cap Set for Pentax Scopes, Erbeflo Clevercap Hybrid CO2…Erbe USA, Inc. · 2013 · 3 generations on2descendants
- K992141Olympus Working Element for ProbesOlympus Winter & Ibe GmbH · 1999 · cites it directly2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K790071 | Endoscope and Accessories | Olympus Corp. | 1979 | — |
| +1 | K093665 | Endo Smartcap | Byrne Medical, Inc. | 2009 | 7 |
| +2 | K102855 | Universal Irrigation Solution Hybrid | Byrne Medical, Inc. | 2010 | — |
| +2 | K103696 | Erbeflo Clevercap Hybrid Tubing/Cap Sets for Olympus Models 160 and 180 Series… | Erbe USA, Inc. | 2011 | — |
| +3 | K132340 | Erbeflo Clevercap Hybrid Tubing/Cap Set for Pentax Scopes, Erbeflo Clevercap… | Erbe USA, Inc. | 2013 | 1 |
| +1 | K992141 | Olympus Working Element for Probes | Olympus Winter & Ibe GmbH | 1999 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1996 | 1 | 0 |
| 1998 | 1 | 0 |
| 1999 | 2 | 0 |
| 2005 | 1 | 0 |
| 2006 | 1 | 0 |
| 2009 | 1 | 0 |
| 2010 | 1 | 0 |
| 2011 | 1 | 0 |
| 2013 | 1 | 0 |
| 2014 | 1 | 0 |
| 2016 | 1 | 0 |
| 2017 | 1 | 0 |
| 2019 | 2 | 0 |
| 2020 | 2 | 0 |
| 2021 | 1 | 0 |
| 2022 | 3 | 0 |
| 2023 | 4 | 0 |
| 2024 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in GCP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (26)
See all 265 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 11, 2025 | Class II | Z-1454-2026 | Under Investigation by firm | Complaints of the ceramic tip of the resection sheath breaking have been received. |
| Sep 13, 2024 | Class II | Z-0351-2025 | Under Investigation by firm | Olympus is removing the statement of compatibility with a GreenLight Laser for BPH therapy from the Instruction for Use (IFU) due to the potential for a damaged tip during use of a laser probe. |
| Sep 13, 2024 | Class II | Z-1437-2026 | Under Investigation by firm | Product being removed due to incompatibility when used in conjunction with a GreenLight Laser for BPH therapy which could damage the tip of the device. |
| Aug 18, 2017 | Class II | Z-0696-2018 | Component change control | Olympus has received complaints about fragments of adhesive which detached from inside the working channel of the referenced products. Cracking, chipping, missing pieces, and delamination of the adhesive have also been reported. In one… |
| Apr 19, 2017 | Class II | Z-2629-2017 | Under Investigation by firm | Loop wires breaking at the distal end of the Monopolar HF Resection Electrodes. |
| Feb 8, 2010 | Not classified | Z-1163-2010 | Nonconforming Material/Component | weak weld seams |
The 5 latest of 6 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K093665 (Byrne Medical, Inc., 2009) has gone furthest, with 18 descendants of its own.
How does its category compare?
All GCP devicesGCP Endoscope, Ac-Powered And Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · GCP
FDA has decided no GCP device since 2007.
Median 510(k) review, days · GCP
Too few GCP 510(k)s in any year since 2007 to give a median.
Recalls per year · GCP
1in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · GCP
180in 2025
MAUDE reports naming product code GCP, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 0 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 4 |
| 2016 | 0 | — | 0 | 13 |
| 2017 | 0 | — | 1 | 117 |
| 2018 | 0 | — | 0 | 26 |
| 2019 | 0 | — | 0 | 19 |
| 2020 | 0 | — | 0 | 18 |
| 2021 | 0 | — | 0 | 12 |
| 2022 | 0 | — | 0 | 26 |
| 2023 | 0 | — | 0 | 100 |
| 2024 | 0 | — | 1 | 108 |
| 2025 | 0 | — | 0 | 180 |
| 2026 | 0 | — | 0 | — |
https://510kreport.com/device/K790071 · printed Nov 6, 2026
Cite this page
510(k) Report. “K790071: Endoscope and Accessories.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K790071
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026