510(k) · Cleared
Recalled · Class II · 2017SFX-90 Table/Spotfilmer W/Fluoricon L200
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
General Electric Co. · JAA System, X-Ray, Fluoroscopic, Image-Intensified
Cleared on Jul 21, 1980 after a 48-day review. FDA has not posted a summary or statement for it, so its predicates are not known. It has been recalled 9 times.
- Decided
- Jul 21, 1980
- Review
- 48 days
- Class
- Class II
- Product code
- JAA
- Recalls
- 9
- Predicates
- 0
FDA has not posted a summary or statement for it, so its predicates are not known.
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JAA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JAA device
- #1 K253103 Nexus DRF Digital X-ray Imaging Systemscore 13.2 · 7 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 6, 2017 | Class II | Z-2179-2017 | Device Design | Reported incidents of a patient step detaching from the table. A fall from a patient step detaching while in use could result in an injury to a patient or operator. |
| Nov 21, 2016 | Class II | Z-0916-2017 | Nonconforming Material/Component | GE Healthcare is conducting a recall due to the potential fall of aged CRT monitors - Prestige 2, Prestilix, Prestige VH, Prestige SI, RFX/SFX, Legacy, Precision 500D, ADVANTX and Innova 2000 systems. |
| Nov 8, 2013 | Class II | Z-0804-2014 | Employee error | Missing screws in the overhead Video Monitor Suspension. GE discovered that the 4 outer screws that hold the overhead Video Monitor Suspension to the carriage were missing since installation of the system. GE is not aware of any fall of… |
| Aug 30, 2013 | Class II | Z-0012-2014 | Employee error | GE Healthcare has recently become aware of a potential safety issue involving the spotfilm device power assist handle. There is a potential for the internal bolts on the power assist handle to be loose or to loosen and/or fall out. This… |
| Mar 9, 2007 | Not classified | Z-0200-2009 | Component change control | GE Healthcare has recently become aware of a potential risk of monitor dropping associated with the monitor suspension of the Advantx system that may impact patient safety. It has been reported that suspensions manufactured after March 8… |
| Mar 30, 2004 | Not classified | Z-0852-04 | Other | The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall. |
| Mar 30, 2004 | Not classified | Z-0851-04 | Other | The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall. |
| Mar 30, 2004 | Not classified | Z-0850-04 | Other | The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall. |
| Mar 30, 2004 | Not classified | Z-0853-04 | Other | The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall. |
The 5 latest of 9 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
How does its category compare?
All JAA devicesJAA System, X-Ray, Fluoroscopic, Image-Intensified: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JAA
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JAA
89in 2024
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JAA
4in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JAA
33in 2025
MAUDE reports naming product code JAA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 35 | 9,517 |
| 2008 | 6 | 18 | 21 | 9,484 |
| 2009 | 0 | — | 18 | 12,040 |
| 2010 | 1 | — | 6 | 5,803 |
| 2011 | 4 | 48 | 7 | 14,000 |
| 2012 | 1 | — | 2 | 11,684 |
| 2013 | 6 | 129 | 3 | 13,706 |
| 2014 | 6 | 276 | 1 | 10,265 |
| 2015 | 3 | 197 | 1 | 2,922 |
| 2016 | 2 | — | 7 | 35 |
| 2017 | 1 | — | 3 | 56 |
| 2018 | 3 | 30 | 1 | 36 |
| 2019 | 6 | 49 | 2 | 49 |
| 2020 | 4 | 39 | 2 | 39 |
| 2021 | 6 | 104 | 2 | 35 |
| 2022 | 1 | — | 5 | 34 |
| 2023 | 2 | — | 9 | 41 |
| 2024 | 3 | 89 | 2 | 25 |
| 2025 | 2 | — | 4 | 33 |
| 2026 | 1 | — | 0 | — |
https://510kreport.com/device/K801320 · printed Nov 6, 2026