510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Tubing Connectors EC-2XXX Series

K833322510(k) TraditionalCleared Dec 16, 198381-day reviewClass II21 CFR 870.4290Recalls 22

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Gish Biomedical, Inc. · DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass

Cleared on Dec 16, 1983 after an 81-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 3 of the 14 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 22 times.

Decided
Dec 16, 1983
Review
81 days
Class
Class II
Product code
DTL
Recalls
22
Predicates
0

13 later devices descend from it, 2 citing it directly. 3 of the 14 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2005: 1 descendant cleared2009: 3 descendants cleared2008: 1 descendant cleared2006: 1 descendant cleared2014: 2 descendants cleared, 1 with a Class I recall12020: 1 descendant cleared, 1 with a Class I recall12011: 1 descendant cleared, 1 with a Class I recall12013: 1 descendant cleared2023: 1 descendant cleared2022: 1 descendant cleared19801990200020102020
  1. This deviceK833322 Tubing Connectors EC-2XXX SeriesGish Biomedical, Inc. · 1983
  2. Descendant · cites it directly · 12 of its ownK053025 Jostra HLM Tubing SetMaquet Cardiopulmonary, AG · 2005 · Recalled
  3. Descendant · 2 generations on · 8 of its ownK080592 HLM Tubing Set with Bioline CoatingMaquet Cardiopulmonary, AG · 2008 · Recalled
  4. Descendant · 3 generations on · 3 of its ownK090533 HLM Tubing Set with Softline CoatingMaquet Cardiopulmonary, AG · 2009
  5. Descendant · 3 generations on · 2 of its ownK200109 OXY-1 SystemAbiomed, Inc. · 2020 · Recalled (Class I)
  6. Descendant · 3 generations on · 2 of its ownK102726 Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · Recalled (Class I)

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

2005: 12006: 12008: 12009: 32011: 12013: 12014: 22020: 12022: 12023: 1
2005high 3 (2009) · low 0 (2007)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K053025Jostra HLM Tubing SetMaquet Cardiopulmonary, AG · 2005 · cites it directly12descendants
  • K080592HLM Tubing Set with Bioline CoatingMaquet Cardiopulmonary, AG · 2008 · 2 generations on8descendants
  • K090533HLM Tubing Set with Softline CoatingMaquet Cardiopulmonary, AG · 2009 · 3 generations on3descendants
  • K200109OXY-1 SystemAbiomed, Inc. · 2020 · 3 generations on2descendants
  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · 3 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K833322Tubing Connectors EC-2XXX SeriesGish Biomedical, Inc.1983—
+1K053025Jostra HLM Tubing SetMaquet Cardiopulmonary, AG20052
+2K080592HLM Tubing Set with Bioline CoatingMaquet Cardiopulmonary, AG2008221
+3K090533HLM Tubing Set with Softline CoatingMaquet Cardiopulmonary, AG2009—
+3K200109OXY-1 SystemAbiomed, Inc.2020Class I
+3K102726Cardiohelp SystemMaquet Cardiopulmonary, AG2011Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200610
200810
200930
201111
201310
201421
202011
202210
202310
FIG. 1 — Lineage of K833322, 1983–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (13)

See all 13

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K053025Jostra HLM Tubing SetMaquet Cardiopulmonary, AG200512
K092486Vortex Medical Angiovac Cardiopulmonary Bypass CircuitVortex Medical, Inc.20091

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 24, 2020Class IIZ-2448-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2443-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2449-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2452-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2450-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2454-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2462-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2451-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2458-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2447-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2455-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2456-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2457-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2459-2020Packaging process controlXXX
Apr 24, 2020Class IIZ-2444-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2446-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2461-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2445-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Apr 24, 2020Class IIZ-2460-2020Packaging process controlThe firm has identified through internal tests that the sterile barrier of certain Sterile Standalone Connectors may present a breach. The issue potentially affects component assembled with a specific packaging and is associated to a…
Nov 18, 2019Class IIIZ-2642-2020Packaging process controlSome packages of sterile disposable connectors contained a 3/8x1/4 reducer instead of the specified 1/2x1/2 connector. No significant risk to health is expected as the issue is easily detectable and the different shape would prevent use…
Sep 14, 2012Class IIZ-0157-2013Process controlDue to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.
Sep 14, 2012Class IIZ-0158-2013Process controlDue to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K200109OXY-1 SystemAbiomed, Inc. · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 4 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K053025 (Maquet Cardiopulmonary, AG, 2005) has gone furthest, with 12 descendants of its own.

How does its category compare?

All DTL devices

DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DTL

1in 2015

2007: 12008: 52009: 22010: 42012: 82013: 12014: 22015: 1
2007high 8 (2012) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTL

36in 2012

2008: 85 days median2010: 115 days median2012: 36 days median
2007high 115 (2010) · low 36 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTL

1in 2025

2008: 12010: 32011: 12012: 22013: 1, 1 Class I2014: 12017: 22022: 22025: 12026 (partial year): 3, 3 Class I
2007high 3 (2010) · low 0 (2007)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · DTL

110in 2025

2007: 372008: 162009: 622010: 1482011: 1602012: 532013: 672014: 802015: 892016: 672017: 1062018: 842019: 762020: 532021: 552022: 482023: 462024: 4522025: 110
2007high 452 (2024) · low 16 (2008)2025

MAUDE reports naming product code DTL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—037
2008585116
20092—062
201041153148
20110—1160
2012836253
20131—1 (1)67
20142—180
20151—089
20160—067
20170—2106
20180—084
20190—076
20200—053
20210—055
20220—248
20230—046
20240—0452
20250—1110
20260—3 (3)—
Cite this page

510(k) Report. “K833322: Tubing Connectors EC-2XXX Series.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K833322

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026