510(k) · Cleared
Recalled · 2009Daig Pacemaker Lead Adaptor
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Daig Corp. · DTD Pacemaker Lead Adaptor
Cleared on Mar 3, 1987 after a 21-day review. FDA has not posted a summary or statement for it, so its predicates are not known. It has been recalled 2 times.
- Decided
- Mar 3, 1987
- Review
- 21 days
- Class
- Class II
- Product code
- DTD
- Recalls
- 2
- Predicates
- 0
FDA has not posted a summary or statement for it, so its predicates are not known.
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DTD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 1, 2009 | Not classified | Z-0580-2010 | Component change control | The silicone tubing of affected leads was processed outside of validated parameters. Although the material remains biocompatible, test data regarding the performance characteristics are not available. |
| Jul 18, 2006 | Not classified | Z-0120-2007 | Other | Enpath Lead Adapters: The internal component (Connector Block) of the Lead Adapter used to secure a lead to the adapter has exhibited varying degress of radial cracks that could cause a temporary intermittent electrical signal disruption… |
No ancestor or descendant has a Class I recall.
How does its category compare?
All DTD devicesDTD Pacemaker Lead Adaptor: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DTD
1in 2015
2026 to Oct 25, outlined.
Median 510(k) review, days · DTD
Too few DTD 510(k)s in any year since 2007 to give a median.
Recalls per year · DTD
No recall names DTD since 2007.
Adverse event reports per year · DTD
32in 2025
MAUDE reports naming product code DTD, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 8 |
| 2008 | 0 | — | 0 | 8 |
| 2009 | 0 | — | 0 | 8 |
| 2010 | 0 | — | 0 | 36 |
| 2011 | 0 | — | 0 | 18 |
| 2012 | 0 | — | 0 | 19 |
| 2013 | 1 | — | 0 | 30 |
| 2014 | 0 | — | 0 | 21 |
| 2015 | 1 | — | 0 | 41 |
| 2016 | 0 | — | 0 | 20 |
| 2017 | 0 | — | 0 | 16 |
| 2018 | 0 | — | 0 | 29 |
| 2019 | 0 | — | 0 | 34 |
| 2020 | 0 | — | 0 | 19 |
| 2021 | 0 | — | 0 | 23 |
| 2022 | 0 | — | 0 | 25 |
| 2023 | 0 | — | 0 | 28 |
| 2024 | 0 | — | 0 | 35 |
| 2025 | 0 | — | 0 | 32 |
| 2026 | 0 | — | 0 | — |
https://510kreport.com/device/K870562 · printed Nov 6, 2026
Cite this page
510(k) Report. “K870562: Daig Pacemaker Lead Adaptor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K870562
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Page last updated Nov 3, 2026