510(k) Report

510(k) · Cleared

Recalled · 2004

Fast-Cath (TM) Intra-Cardiac Introducer

K973840510(k) TraditionalCleared May 21, 1998225-day reviewClass II21 CFR 870.1340Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Daig Corp. · DYB Introducer, Catheter

Cleared on May 21, 1998 after a 225-day review. The FDA summary names its predicate without a number. 2 of the 84 devices in its family, counting itself, have a Class I recall. 83 later devices descend from it. It has been recalled 2 times.

Decided
May 21, 1998
Review
225 days
Class
Class II
Product code
DYB
Recalls
2
Predicates
0

83 later devices descend from it, 3 citing it directly. 2 of the 84 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2004: 1 descendant cleared2006: 2 descendants cleared2005: 2 descendants cleared2019: 3 descendants cleared2012: 2 descendants cleared2010: 4 descendants cleared2007: 2 descendants cleared2017: 6 descendants cleared2009: 2 descendants cleared2016: 2 descendants cleared, 1 with a Class I recall12025: 4 descendants cleared2024: 8 descendants cleared2008: 3 descendants cleared2013: 4 descendants cleared2014: 3 descendants cleared2011: 4 descendants cleared2020: 3 descendants cleared2018: 4 descendants cleared2021: 5 descendants cleared2023: 5 descendants cleared2022: 6 descendants cleared, 1 with a Class I recall12015: 2 descendants cleared2026: 6 descendants cleared1995200020052010201520202025
  1. This deviceK973840 Fast-Cath (TM) Intra-Cardiac IntroducerDaig Corp. · 1998
  2. Descendant · cites it directly · 81 of its ownK042623 Agilis Steerable Catheter Introducer, Model 408304St. Jude Medical, Daig Division, Inc. · 2005 · Recalled
  3. Descendant · 2 generations on · 53 of its ownK061363 Agilis NXT Steerable IntroducerSt Jude Medical · 2006 · Recalled
  4. Descendant · 3 generations on · 36 of its ownK081645 Agilis NXT Steerable Introducer, Model G408324St Jude Medical · 2008
  5. Descendant · 2 generations on · 36 of its ownK052644 Swartz Braided Transseptal Guiding IntroducerSt Jude Medical · 2005
  6. Descendant · 3 generations on · 11 of its ownK070417 Across Transseptal Access System, 406977, 406978, 406979, 406980…St Jude Medical · 2007

Showing 5 of 83 descendants. See all descendants

Descendants cleared per year

2004: 12005: 22006: 22007: 22008: 32009: 22010: 42011: 42012: 22013: 42014: 32015: 22016: 22017: 62018: 42019: 32020: 32021: 52022: 62023: 52024: 82025: 42026: 6
2004high 8 (2024) · low 1 (2004)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 83 · See all
  • K042623Agilis Steerable Catheter Introducer, Model 408304St. Jude Medical, Daig Division, Inc. · 2005 · cites it directly81descendants
  • K061363Agilis NXT Steerable IntroducerSt Jude Medical · 2006 · 2 generations on53descendants
  • K081645Agilis NXT Steerable Introducer, Model G408324St Jude Medical · 2008 · 3 generations on36descendants
  • K052644Swartz Braided Transseptal Guiding IntroducerSt Jude Medical · 2005 · 2 generations on36descendants
  • K070417Across Transseptal Access System, 406977, 406978, 406979, 406980, 406981, 406982…St Jude Medical · 2007 · 3 generations on11descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K973840Fast-Cath (TM) Intra-Cardiac IntroducerDaig Corp.1998—
+1K042623Agilis Steerable Catheter Introducer, Model 408304St. Jude Medical, Daig Division, Inc.20051
+2K061363Agilis NXT Steerable IntroducerSt Jude Medical20061
+3K081645Agilis NXT Steerable Introducer, Model G408324St Jude Medical2008—
+2K052644Swartz Braided Transseptal Guiding IntroducerSt Jude Medical2005—
+3K070417Across Transseptal Access System, 406977, 406978, 406979, 406980, 406981…St Jude Medical2007—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200520
200620
200720
200830
200920
201040
201140
201220
201340
201430
201520
201621
201760
201840
201930
202030
202150
202261
202350
202480
202540
202660
FIG. 1 — Lineage of K973840, 1998–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DYB device

  1. #1 K251051 AuST Steerable Sheathscore 18.0 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Fast-Cath™ Transseptal Catheter Introducer
Named without a submission number
—

What descends from it? (83)

See all 83

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042623Agilis Steerable Catheter Introducer, Model 408304St. Jude Medical, Daig Division, Inc.200581
K061015Fast-Cath Hemostasis Introducer and Transseptal IntroducerSt Jude Medical20069
K040552Model 3100 Series Sheath Introducer Kits, Models 3110, 3130Cryocor, Inc.20040

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 4, 2004Not classifiedZ-1426-04OtherSt. Jude Medical received two complaints from customers involving the Telesheath Left Atrial Introducer System side arm detaching from the hub of the inner sheath. Subsequent testing of Telesheath inventory at St. Jude Medical indicated…
Aug 4, 2004Not classifiedZ-1425-04OtherSt. Jude Medical received two complaints from customers involving the Telesheath Left Atrial Introducer System side arm detaching from the hub of the inner sheath. Subsequent testing of Telesheath inventory at St. Jude Medical indicated…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K152090Reprocessed Steerable IntroducerSterilmed, Inc. · 2016 · 3 generations on · ◆ in the 2016 descendants bar of FIG. 11 Class I recall
  • K220340Amplatzer Steerable Delivery SheathAbbott Medical · 2022 · 5 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K042623 (St. Jude Medical, Daig Division, Inc., 2005) has gone furthest, with 81 descendants of its own.

How does its category compare?

All DYB devices

DYB Introducer, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DYB

16in 2025

2007: 122008: 172009: 142010: 182011: 172012: 222013: 262014: 302015: 172016: 192017: 242018: 192019: 182020: 92021: 122022: 132023: 152024: 182025: 162026 (partial year): 17
2007high 30 (2014) · low 9 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DYB

63in 2025

2007: 95 days median2008: 37 days median2009: 52 days median2010: 123 days median2011: 100 days median2012: 128 days median2013: 118 days median2014: 122 days median2015: 129 days median2016: 95 days median2017: 70 days median2018: 156 days median2019: 47 days median2020: 49 days median2021: 78 days median2022: 143 days median2023: 55 days median2024: 56 days median2025: 63 days median2026 (partial year): 59 days median
2007high 156 (2018) · low 37 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DYB

3in 2025

2007: 152008: 92009: 252010: 182011: 42012: 32013: 5, 1 Class I2014: 22015: 2, 1 Class I2016: 45, 3 Class I2017: 22, 1 Class I2018: 12019: 22020: 21, 21 Class I2021: 62022: 22023: 102024: 22025: 32026 (partial year): 23, 21 Class I
2007high 45 (2016) · low 1 (2018)2026

Class I (48 since 2007) Class II/III not classified

Adverse event reports per year · DYB

4,539in 2025

2007: 3162008: 4232009: 3992010: 4452011: 6282012: 8982013: 7432014: 7802015: 9132016: 1,2342017: 1,6722018: 1,9032019: 2,4502020: 2,4652021: 2,8352022: 4,0842023: 3,7752024: 3,9312025: 4,539
2007high 4,539 (2025) · low 316 (2007)2025

MAUDE reports naming product code DYB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007129515316
200817379423
2009145225399
20101812318445
2011171004628
2012221283898
2013261185 (1)743
2014301222780
2015171292 (1)913
2016199545 (3)1,234
2017247022 (1)1,672
20181915611,903
2019184722,450
202094921 (21)2,465
2021127862,835
20221314324,084
20231555103,775
2024185623,931
2025166334,539
2026175923 (21)—

What is it indicated for?

The Fast-Cath™ Intra-Cardiac Introducer is indicated for use when introducing various cardiovascular catheters or biopsy devices into the heart.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K973840: Fast-Cath (TM) Intra-Cardiac Introducer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K973840

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026