510(k) · Cleared
Recalled · 2008Steris System 1 Processor and 20 Sterilant
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Steris C/O Medical Device Consultants, Inc. · MED Sterilant, Medical Devices
Cleared on Sep 13, 1988 after a 264-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 11 later devices descend from it. It has been recalled 3 times.
- Decided
- Sep 13, 1988
- Review
- 264 days
- Class
- Class II
- Product code
- MED
- Recalls
- 3
- Predicates
- 0
11 later devices descend from it, 5 citing it directly.
- This deviceK875280 Steris System 1 Processor and 20 SterilantSteris C/O Medical Device Consultants, Inc. · 1988
- Descendant · cites it directly · 7 of its ownK972708 Endo-Spor Plus / HyprocideCottrell, Ltd. · 1999
- Descendant · cites it directly · 6 of its ownK013280 Sterilox Liquid Chemical Sterilant SystemSterilox Technologies, Inc. · 2002
- Descendant · cites it directly · 4 of its ownK090036 System 1.5 Sterile Processing SystemSTERIS Corporation · 2010 · Recalled
- Descendant · 2 generations on · 3 of its ownK101409 System IE Liquid Chemical Cterilant Processing SystemSTERIS Corporation · 2010 · Recalled
- Descendant · 2 generations on · 2 of its ownK102462 System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2010 · Recalled
Showing 5 of 11 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 11 · See all- K972708Endo-Spor Plus / HyprocideCottrell, Ltd. · 1999 · cites it directly7descendants
- K013280Sterilox Liquid Chemical Sterilant SystemSterilox Technologies, Inc. · 2002 · cites it directly6descendants
- K090036System 1.5 Sterile Processing SystemSTERIS Corporation · 2010 · cites it directly4descendants
- K101409System IE Liquid Chemical Cterilant Processing SystemSTERIS Corporation · 2010 · 2 generations on3descendants
- K102462System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2010 · 2 generations on2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K875280 | Steris System 1 Processor and 20 Sterilant | Steris C/O Medical Device Consultants, Inc. | 1988 | — |
| +1 | K972708 | Endo-Spor Plus / Hyprocide | Cottrell, Ltd. | 1999 | — |
| +1 | K013280 | Sterilox Liquid Chemical Sterilant System | Sterilox Technologies, Inc. | 2002 | — |
| +1 | K090036 | System 1.5 Sterile Processing System | STERIS Corporation | 2010 | 6 |
| +2 | K101409 | System IE Liquid Chemical Cterilant Processing System | STERIS Corporation | 2010 | 4 |
| +2 | K102462 | System 1E Liquid Chemical Sterilant Processing System | STERIS Corporation | 2010 | 5 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1999 | 1 | 0 |
| 2002 | 1 | 0 |
| 2005 | 1 | 0 |
| 2006 | 1 | 0 |
| 2007 | 1 | 0 |
| 2010 | 3 | 0 |
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2013 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MED? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MED device
- #1 K251666 SYSTEM 1E Liquid Chemical Sterilant Processing…score 13.8 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (11)
See all 115 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K972708 | Endo-Spor Plus / Hyprocide | 1999 | 7 |
| K013280 | Sterilox Liquid Chemical Sterilant System | 2002 | 6 |
| K090036 | System 1.5 Sterile Processing System | 2010 | 4 |
| K060458 | Manzi Mach 1 Instrument Cleaner-Processor System, Manzi Detergent MD10, Manzi Sterilant… | 2006 | 1 |
| K032959 | Cidex Opa Concentrate | 2005 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 4, 2008 | Not classified | Z-2462-2008 | Device Design | During ongoing product development testing, it was discovered that, when the two inflow barbs on adapter 201048 # 4 are in a position to face one direction, the top Barblock may interfere with the top surface of bottom barb of the… |
| Mar 15, 2004 | Not classified | Z-0765-04 | Process control | A defective port adapter may affect the sterilization process of the attached endoscope. |
| Mar 7, 2003 | Not classified | Z-0547-03 | Other | The Steris device may not attach properly to retrofitted Olympus bronchoscopes, and become loose during the sterile processing cycle. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K972708 (Cottrell, Ltd., 1999) has gone furthest, with 7 descendants of its own.
How does its category compare?
All MED devicesMED Sterilant, Medical Devices: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · MED
2in 2023
2026 to Oct 25, outlined.
Median 510(k) review, days · MED
84in 2019
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MED
2in 2011
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MED
9in 2025
MAUDE reports naming product code MED, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 3 | 34 |
| 2008 | 1 | — | 2 | 79 |
| 2009 | 0 | — | 0 | 158 |
| 2010 | 6 | 281 | 0 | 89 |
| 2011 | 0 | — | 2 | 106 |
| 2012 | 3 | 111 | 0 | 337 |
| 2013 | 1 | — | 0 | 140 |
| 2014 | 1 | — | 0 | 105 |
| 2015 | 0 | — | 0 | 110 |
| 2016 | 1 | — | 0 | 135 |
| 2017 | 1 | — | 0 | 74 |
| 2018 | 2 | — | 0 | 42 |
| 2019 | 4 | 84 | 0 | 35 |
| 2020 | 0 | — | 0 | 32 |
| 2021 | 2 | — | 0 | 31 |
| 2022 | 1 | — | 0 | 16 |
| 2023 | 2 | — | 0 | 9 |
| 2024 | 0 | — | 0 | 8 |
| 2025 | 0 | — | 0 | 9 |
| 2026 | 2 | — | 0 | — |
https://510kreport.com/device/K875280 · printed Nov 6, 2026
Cite this page
510(k) Report. “K875280: Steris System 1 Processor and 20 Sterilant.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K875280
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026