System IE Liquid Chemical Cterilant Processing System

K101409510(k) SpecialCleared Aug 2, 201075-day reviewClass II21 CFR 880.6885Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

STERIS Corporation · MED Sterilant, Medical Devices

Cleared on Aug 2, 2010 as substantially equivalent to K090036 (STERIS Corporation, 2010) after a 75-day review. Its family tree reaches back 4 generations to 1988. 3 later devices descend from it. It has been recalled 4 times.

Decided
Aug 2, 2010
Review
75 days
Class
Class II
Product code
MED
Recalls
4
Predicates
1

Its longest line of descent runs back 4 generations to K875280 (1988). 8 ancestors in all. 3 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1994: 3 other ancestors1995: 1 other ancestor2010: 1 descendant cleared2012: 1 descendant cleared2013: 1 descendant clearedK875280 · Steris System 1 Processor and 20 Sterilant · Steris C/O Medical Device Consultants, Inc. · 1988 · recalledK972708 · Endo-Spor Plus / Hyprocide · Cottrell, Ltd. · 1999K013280 · Sterilox Liquid Chemical Sterilant System · Sterilox Technologies, Inc. · 2002K090036 · System 1.5 Sterile Processing System · STERIS Corporation · 2010 · recalled1990200020102020
  1. Root · 4 generations backK875280 Steris System 1 Processor and 20 SterilantSteris C/O Medical Device Consultants, Inc. · 1988 · Recalled
  2. 3 generations backK972708 Endo-Spor Plus / HyprocideCottrell, Ltd. · 1999
  3. 2 generations backK013280 Sterilox Liquid Chemical Sterilant SystemSterilox Technologies, Inc. · 2002
  4. 1 generation backK090036 System 1.5 Sterile Processing SystemSTERIS Corporation · 2010 · Recalled
  5. This deviceK101409 System IE Liquid Chemical Cterilant Processing SystemSTERIS Corporation · 2010
  6. Descendant · cites it directly · 2 of its ownK102462 System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2010 · Recalled
  7. Descendant · cites it directly · 1 of its ownK113520 System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2012 · Recalled
  8. Descendant · cites it directly · 0 of its ownK131078 System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2013 · Recalled

Most influential descendants

Showing 3 of 3 · See all
  • K102462System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2010 · cites it directly2descendants
  • K113520System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2012 · cites it directly1descendant
  • K131078System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation · 2013 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K875280Steris System 1 Processor and 20 SterilantSteris C/O Medical Device Consultants, Inc.19883
−3K972708Endo-Spor Plus / HyprocideCottrell, Ltd.1999—
−2K013280Sterilox Liquid Chemical Sterilant SystemSterilox Technologies, Inc.2002—
−1K090036System 1.5 Sterile Processing SystemSTERIS Corporation20106
0 (this device)K101409System IE Liquid Chemical Cterilant Processing SystemSTERIS Corporation2010—
+1K102462System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation20105
+1K113520System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation20122
+1K131078System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation20132

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199430
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201210
201310
FIG. 1 — Lineage of K101409, 1988–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MED? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MED device

  1. #1 K251666 SYSTEM 1E Liquid Chemical Sterilant Processing…score 13.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090036System 1.5 Sterile Processing System
Summary page 2
…May 18, 2010 Page 8-1 STERIS SPECIAL 510(k) PREMARKET NOTIFICATION Modification of K090036 SYSTEM IIE Liquid Chemical Sterilant Processing System 1. Device Name Trade Name: SYSTEM lE Liq…
STERIS Corporation20106

Also cleared against K090036: 3 other devices →

What descends from it? (3)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K102462System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation20102
K113520System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation20121
K131078System 1E Liquid Chemical Sterilant Processing SystemSTERIS Corporation20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 21, 2013Class IIZ-0616-2014Device DesignCustomers indicated fluctuations in environmental conditions sites and gradual degradation/deformation of check valve o-rings following frequent contact with S40 sterilant use solution result in nuisance aborts.
Aug 1, 2012Class IIZ-2335-2012Nonconforming Material/ComponentThe firm initiated a recall after complaint investigations revealed that when the units are left on and not in use at night and/or on weekends the hose disconnects/separates resulting in water leakage that can cause damage to property.
Aug 11, 2011Not classifiedZ-3124-2011OtherSTERIS learned that SYSTEM 1E users may experience aborted cycles due to a failure within the UV light system.
Mar 14, 2011Not classifiedZ-1960-2011OtherThe printout, display and Operator Manual for the STERIS System 1E Liquid Chemical Sterilant Processing System are subject to recall/correction by the firm because in the event of an incomplete or aborted cycle, the language of the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 7.5 years on median.
  • 2 of the 8 ancestors have a recall of any class.
  • Of its descendants, K102462 (STERIS Corporation, 2010) has gone furthest, with 2 descendants of its own.
  • 3 other devices also cite its predicate K090036.

How does its category compare?

All MED devices

MED Sterilant, Medical Devices: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · MED

2in 2023

2007: 22008: 12010: 62012: 32013: 12014: 12016: 12017: 12018: 22019: 42021: 22022: 12023: 22026 (partial year): 2
2007high 6 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MED

84in 2019

2010: 281 days median2012: 111 days median2019: 84 days median
2007high 281 (2010) · low 84 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MED

2in 2011

2007: 32008: 22011: 2
2007high 3 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MED

9in 2025

2007: 342008: 792009: 1582010: 892011: 1062012: 3372013: 1402014: 1052015: 1102016: 1352017: 742018: 422019: 352020: 322021: 312022: 162023: 92024: 82025: 9
2007high 337 (2012) · low 8 (2024)2025

MAUDE reports naming product code MED, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—334
20081—279
20090—0158
20106281089
20110—2106
201231110337
20131—0140
20141—0105
20150—0110
20161—0135
20171—074
20182—042
2019484035
20200—032
20212—031
20221—016
20232—09
20240—08
20250—09
20262—0—

What is it indicated for?

The SYSTEM IE Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities. The SYSTEM IE Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 46 to 55°C, and rinses the load with…

Read the full statement (191 words)

The SYSTEM IE Liquid Chemical Sterilant Processing System is intended for liquid chemical sterilization of cleaned, immersible, and reusable critical and semi-critical heat-sensitive medical devices in healthcare facilities.

The SYSTEM IE Processor dilutes the S40 Sterilant Concentrate to its use dilution (>1820 mg/L peracetic acid), liquid chemically sterilizes the load during a controlled 6-minute exposure at 46 to 55°C, and rinses the load with extensively treated* potable water. The processed load should be used immediately.

The SYSTEM IE Processor uses only S40 Sterilant Concentrate to liquid chemically sterilize medical devices.

*The extensive treatment of EPA potable water consists of:

1. Pre-filtration through two pre-filters:

* Pre-filter 1 is a gross depth filter that removes approximately 5 micron or larger particles/contaminants.

* Pre-filter 2 is a surface filter that removes particles/contaminants > 0.1 micron.

2. UV Irradiation:

* During transit through the UV water treatment chamber, a UV dose sufficient to achieve a >6-log reduction of MS2 virus is delivered to the water.

3. 0.1 micron filtration:

* The water prepared by pre-filtration and UV irradiation is filtered through redundant, 0.1-micron (absolute rated) membranes to remove bacteria, fungi and protozoa >0.1 micron.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101409: System IE Liquid Chemical Cterilant Processing System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101409

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026