510(k) Report

510(k) · Cleared

Recalled · 2010

Synchron CX4/CX5 Clinical Chemistry Analyzer Reage

K883181510(k) TraditionalCleared Sep 19, 198854-day reviewClass I21 CFR 862.1030Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Instruments, Inc. · CKA Nadh Oxidation/Nad Reduction, Alt/Sgpt

Cleared on Sep 19, 1988 after a 54-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 20 later devices descend from it. It has been recalled 2 times.

Decided
Sep 19, 1988
Review
54 days
Class
Class I
Product code
CKA
Recalls
2
Predicates
0

20 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2004: 3 descendants cleared1997: 3 descendants cleared2014: 1 descendant cleared2007: 2 descendants cleared2001: 1 descendant cleared2006: 1 descendant cleared2020: 3 descendants cleared2011: 1 descendant cleared2021: 1 descendant cleared2023: 3 descendants cleared2026: 1 descendant cleared1990200020102020
  1. This deviceK883181 Synchron CX4/CX5 Clinical Chemistry Analyzer ReageBeckman Instruments, Inc. · 1988
  2. Descendant · cites it directly · 17 of its ownK965240 Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin…Beckman Instruments, Inc. · 1997 · Recalled
  3. Descendant · 2 generations on · 11 of its ownK042291 Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc. · 2004 · Recalled
  4. Descendant · 3 generations on · 5 of its ownK183688 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
  5. Descendant · 3 generations on · 4 of its ownK183678 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
  6. Descendant · 3 generations on · 4 of its ownK191298 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020

Showing 5 of 20 descendants. See all descendants

Descendants cleared per year

1997: 32001: 12004: 32006: 12007: 22011: 12014: 12020: 32021: 12023: 32026: 1
1997high 3 (1997) · low 0 (1998)2026

Most influential descendants

Showing 5 of 20 · See all
  • K965240Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin Reagent, Synchron LX…Beckman Instruments, Inc. · 1997 · cites it directly17descendants
  • K042291Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc. · 2004 · 2 generations on11descendants
  • K183688i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 3 generations on5descendants
  • K183678i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 3 generations on4descendants
  • K191298i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 3 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K883181Synchron CX4/CX5 Clinical Chemistry Analyzer ReageBeckman Instruments, Inc.1988—
+1K965240Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin Reagent…Beckman Instruments, Inc.199712
+2K042291Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc.200434
+3K183688i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—
+3K183678i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—
+3K191298i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199730
200110
200430
200610
200720
201110
201410
202030
202110
202330
202610
FIG. 1 — Lineage of K883181, 1988–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CKA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (20)

See all 20

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K965240Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin Reagent, Synchron LX…Beckman Instruments, Inc.199717
K131189Unicel DXC Synchron Systems Glucose Reagent (Gluh)Beckman Coulter, Inc.20140
K041643Synchron Systems Phosphorus (PHS) ReagentBeckman Coulter, Inc.20040

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 10, 2010Not classifiedZ-1619-2011OtherThe recall was initiated because Cartridge Glucose (GLU) on UniCel DxC and Synchron LX instruments may give falsely high results for moderately hemolysed samples. The Impact: (1) Neonatal samples drawn by lancets are more likely to be…
Aug 26, 2009Not classifiedZ-1229-2012Nonconforming Material/ComponentA recall was initiated because Beckman has confirmed that the standardization of the Immuno Protein Calibrator Set has been re-established to IFCC Reference Preparation for Plasma Proteins, BCR-470 beginning with lot number M901330.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K965240 (Beckman Instruments, Inc., 1997) has gone furthest, with 17 descendants of its own.

How does its category compare?

All CKA devices

CKA Nadh Oxidation/Nad Reduction, Alt/Sgpt: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CKA

1in 2013

2012: 12013: 1
2007high 1 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CKA

Too few CKA 510(k)s in any year since 2007 to give a median.

Recalls per year · CKA

1in 2016

2007: 82010: 12016: 1
2007high 8 (2007) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CKA

8in 2025

2010: 82011: 392012: 82017: 22018: 12020: 32021: 22022: 22023: 42024: 102025: 8
2007high 39 (2011) · low 0 (2007)2025

MAUDE reports naming product code CKA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—80
20080—00
20090—00
20100—18
20110—039
20121—08
20131—00
20140—00
20150—00
20160—10
20170—02
20180—01
20190—00
20200—03
20210—02
20220—02
20230—04
20240—010
20250—08
20260—0—
Cite this page

510(k) Report. “K883181: Synchron CX4/CX5 Clinical Chemistry Analyzer Reage.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K883181

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026