Unicel DXC 600 and 800 Synchron Systems

K042291510(k) TraditionalCleared Nov 12, 200480-day reviewClass II21 CFR 862.1600Recalls 34

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · CEM Electrode, Ion Specific, Potassium

Cleared on Nov 12, 2004 as substantially equivalent to K965240 (Beckman Instruments, Inc., 1997) and 1 other predicate after an 80-day review. Its family tree reaches back 2 generations to 1988. 11 later devices descend from it. It has been recalled 34 times.

Decided
Nov 12, 2004
Review
80 days
Class
Class II
Product code
CEM
Recalls
34
Predicates
2

Its longest line of descent runs back 2 generations to K883181 (1988). 5 ancestors in all. 11 later devices descend from it, 6 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2001: 1 other ancestor1995: 1 other ancestor1994: 1 other ancestor2006: 1 descendant cleared2020: 3 descendants cleared2011: 1 descendant cleared2007: 1 descendant cleared2021: 1 descendant cleared2023: 3 descendants cleared2026: 1 descendant clearedK883181 · Synchron CX4/CX5 Clinical Chemistry Analyzer Reage · Beckman Instruments, Inc. · 1988 · recalledK965240 · Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin Reagent, Synchron LX Creatinine Picric Reagent, Synchron L · Beckman Instruments, Inc. · 1997 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors 1 generation forward: 6 descendants2 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant K965240 · Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin Reagent, Synchron LX Creatinine Picric Reagent, Synchron L · Beckman Instruments, Inc. · 1997 · recalledK883181 · Synchron CX4/CX5 Clinical Chemistry Analyzer Reage · Beckman Instruments, Inc. · 1988 · recalled0
  1. Root · 2 generations backK883181 Synchron CX4/CX5 Clinical Chemistry Analyzer ReageBeckman Instruments, Inc. · 1988 · Recalled
  2. 1 generation backK965240 Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin…Beckman Instruments, Inc. · 1997 · Recalled
  3. This deviceK042291 Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc. · 2004
  4. Descendant · cites it directly · 5 of its ownK183688 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
  5. Descendant · cites it directly · 4 of its ownK183678 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
  6. Descendant · cites it directly · 4 of its ownK191298 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
  7. Descendant · 2 generations on · 3 of its ownK210958 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2021
  8. Descendant · 3 generations on · 1 of its ownK223755 i-STAT G cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2023

Showing 5 of 11 descendants. See all descendants

Descendants cleared per year

2006: 12007: 12011: 12020: 32021: 12023: 32026: 1
2006high 3 (2020) · low 0 (2008)2026

Most influential descendants

Showing 5 of 11 · See all
  • K183688i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · cites it directly5descendants
  • K183678i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · cites it directly4descendants
  • K191298i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · cites it directly4descendants
  • K210958i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2021 · 2 generations on3descendants
  • K223755i-STAT G cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2023 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K883181Synchron CX4/CX5 Clinical Chemistry Analyzer ReageBeckman Instruments, Inc.19882
−1K965240Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin Reagent…Beckman Instruments, Inc.199712
0 (this device)K042291Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc.2004—
+1K183688i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—
+1K183678i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—
+1K191298i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—
+2K210958i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2021—
+3K223755i-STAT G cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200710
201110
202030
202110
202330
202610
FIG. 1 — Lineage of K042291, 1988–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CEM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CEM device

  1. #1 K253491 ISE indirect K for Gen.2; ISE indirect Na for…score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K965240Synchron LX 20 Clinical Chemistry System, Synchron LX Albumin Reagent, Synchron LX…
Summary page 1
…ISYNCHRON PRO Systems LX®20 Beckman Coulter, Inc.* K(965240, K(011213 'Beckman Coulter, In…
Beckman Instruments, Inc.199712
*K011213Modification to Synchron LX 20 Clinical Chemistry System
Summary page 1
…LX®20 Beckman Coulter, Inc.* K(965240, K(011213 'Beckman Coulter, Inc., Brea, CA 5.0 Description: The UniCel [)xC 600 and 800 Systems are the next genera…
Beckman Coulter, Inc.200111

Also cleared against K965240: 4 other devices →

What descends from it? (11)

See all 11

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K183688i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.20205
K183678i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.20204
K191298i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.20204
K103842Unicel DXC 600 Synchron System Software Version 5.0 and Unicel DXC 800 Synchron System…Beckman Coulter, Inc.20110
K071283Synchron Systems Creatinine ReagentBeckman Coulter, Inc.20070
K060935Synchron Systems Creatinine (Crea) ReagentBeckman Coulter, Inc.20060

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 29, 2020Class IIZ-2136-2020Process controlElectrolyte reference reagent, used for clinical laboratory testing to determine CO2 in human serum or plasma for diagnosis and treatment of potentially serious disorders associated with changes in the body acid-base balance, has a low pH…
Dec 5, 2018Class IIZ-1078-2019Under Investigation by firmWhen racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead…
Aug 21, 2014Class IIZ-2687-2014Under Investigation by firmBeckman Coulter is recalling the Sample Probe Cable because nonconforming sample probe cables may been installed on newly manufactured and/or recently serviced UniCel DxI 600 and 800 Access Immunoassay Systems, UniCel DxC 660i, 680i, and…
May 14, 2014Class IIZ-1737-2014Mixed-up of materials/componentsBeckman Coulter is initiating a field action because some Ultrasonic Transducer Kits were potentially assembled with incorrect mounting screws.
Apr 22, 2014Class IIZ-2272-2016Nonconforming Material/ComponentPossible leaking of No Foam solution from the No Foam Bottle Assembly on the UniCel DxC Synchron Systems.
Sep 3, 2012Class IIZ-0123-2013Software change controlBeckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be performed on certain DxC configurations with v5.0.11 software installed. If the Shuttle to Barcode Reader becomes misaligned…
Jul 6, 2012Class IIZ-0059-2013Nonconforming Material/ComponentThe recall was initiated because Beckman Coulter has confirmed that the current sample syringe pump drive assemblies installed on DxC Integrated ystem with UniCel Closted Tube Aliquotter (UCTA) serial numbers between 547 and 574 may cause…
Oct 13, 2011Class IIZ-2257-2012Software designThe recall was initiated because Beckman Coulter has received seven reports of a software anomaly for the UniCel Dxl Immunoassay Systems. If undetected, this anomaly may cause erroneous test results.
Oct 13, 2011Class IIZ-2259-2012Software designThe recall was initiated because Beckman Coulter has received seven reports of a software anomaly for the UniCel Dxl Immunoassay Systems. If undetected, this anomaly may cause erroneous test results.
Oct 13, 2011Class IIZ-2258-2012Software designThe recall was initiated because Beckman Coulter has received seven reports of a software anomaly for the UniCel Dxl Immunoassay Systems. If undetected, this anomaly may cause erroneous test results.
Aug 1, 2011Class IIZ-0347-2013Process designThe recall was initiated because Beckman Coulter has confirmed that the UniCel Dxl and DxC Access Immunoassay Systems had field modifications to the aspirate and dispense probe assemblies.
Jan 26, 2011Not classifiedZ-1938-2011PendingThe recall was initiated due to Beckman Coulter receiving a report of results that were associated with incorrect patient identifications when two sample racks advanced into the sample presentation area in one push. The first rack of the…
Jan 26, 2011Not classifiedZ-1940-2011PendingThe recall was initiated due to Beckman Coulter receiving a report of results that were associated with incorrect patient identifications when two sample racks advanced into the sample presentation area in one push. The first rack of the…
Jan 26, 2011Not classifiedZ-1939-2011PendingThe recall was initiated due to Beckman Coulter receiving a report of results that were associated with incorrect patient identifications when two sample racks advanced into the sample presentation area in one push. The first rack of the…
Nov 16, 2010Not classifiedZ-1972-2011PendingThe hose clamps in the DxC (with serial numbers less than 3412, labeled behind the lower right side instrument door) hydropneumatic system are sharp and pose a possible risk of injury. An operator could be cut by the hose clamps, which…
Oct 22, 2010Not classifiedZ-2075-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Oct 22, 2010Not classifiedZ-2077-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Oct 22, 2010Not classifiedZ-2076-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Oct 22, 2010Not classifiedZ-2078-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Oct 22, 2010Not classifiedZ-2080-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Oct 22, 2010Not classifiedZ-2081-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Oct 22, 2010Not classifiedZ-2074-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Oct 22, 2010Not classifiedZ-2079-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Oct 22, 2010Not classifiedZ-2073-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Aug 10, 2010Not classifiedZ-1619-2011OtherThe recall was initiated because Cartridge Glucose (GLU) on UniCel DxC and Synchron LX instruments may give falsely high results for moderately hemolysed samples. The Impact: (1) Neonatal samples drawn by lancets are more likely to be…
Jul 6, 2010Not classifiedZ-2686-2011Software designTest results may have incorrect ranges. If a default sample type setting on the tests screen is changed while running Unicel DXI system software version 4.3, the new setting is not saved to the system database. The changed sample type…
Apr 15, 2010Not classifiedZ-2388-2010OtherUniCel DxC and Synchron LX instruments produced inaccurate (low) GLUCm results. False low results could affect or delay diagnosis or treatment.
Jan 4, 2010Not classifiedZ-0863-2010Device DesignExcessive Buildup of protein, bacteria and sample tube additives in the ISE flow cell may cause erroneous NA (sodium) results.
Jul 23, 2009Not classifiedZ-2633-2011PendingThe recall was initiated because the Wash Concentrate Tube Assembly 142 may fail to deliver the appropriate amount of wash concentrate on some SYNCHRON DxC Systems due to an accumulation of crystals. The impact of this issue would…
Apr 5, 2007Not classifiedZ-1099-2008Software designIntermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags…
Apr 5, 2007Not classifiedZ-1101-2008Software designIntermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags…
Apr 5, 2007Not classifiedZ-1100-2008Software designIntermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags…
Apr 5, 2007Not classifiedZ-1097-2008Software designIntermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags…
Mar 15, 2006Not classifiedZ-0902-06OtherIt was confirmed that the ORDAC (Over Range Detection And Correction) feature (auto & manual) does not perform correctly for the glucose cup chemistry (GLUCm) when sample type of 'Other' is selected w/operating software versions 1.0 &…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 2 companies; a new generation was cleared every 8.5 years on median.
  • 4 of the 5 ancestors have a recall of any class.
  • Of its descendants, K183688 (Abbott Point of Care, Inc., 2020) has gone furthest, with 5 descendants of its own.
  • 4 other devices also cite its predicate K965240.

How does its category compare?

All CEM devices

CEM Electrode, Ion Specific, Potassium: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CEM

1in 2025

2007: 12009: 12010: 32011: 32012: 42020: 12023: 22025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CEM

180in 2012

2010: 195 days median2011: 74 days median2012: 180 days median
2007high 195 (2010) · low 74 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CEM

4in 2025

2008: 12010: 122011: 62016: 12019: 12020: 32021: 12023: 22025: 4
2007high 12 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CEM

29in 2025

2007: 662008: 2452009: 792010: 1132011: 1262012: 452013: 372014: 42015: 72016: 62017: 322018: 132019: 252020: 272021: 182022: 162023: 302024: 422025: 29
2007high 245 (2008) · low 4 (2014)2025

MAUDE reports naming product code CEM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—066
20080—1245
20091—079
2010319512113
20113746126
20124180045
20130—037
20140—04
20150—07
20160—16
20170—032
20180—013
20190—125
20201—327
20210—118
20220—016
20232—230
20240—042
20251—429
20261—0—

What is it indicated for?

The UniCel DxC SYNCHRON Systems are fully automated, computer-controlled clinical chemistry analyzers intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, or cerebrospinal fluid, (sample type is chemistry dependent).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K042291: Unicel DXC 600 and 800 Synchron Systems.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K042291

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026