510(k) · Cleared
Recalled · 2011Flo-Gard (R) 6301
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Baxter Healthcare Corp · FRN Pump, Infusion
Cleared on Feb 28, 1992 after an 80-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 later devices descend from it. It has been recalled 3 times.
- Decided
- Feb 28, 1992
- Review
- 80 days
- Class
- Class II
- Product code
- FRN
- Recalls
- 3
- Predicates
- 0
2 later devices descend from it, 2 citing it directly.
- This deviceK915523 Flo-Gard (R) 6301Baxter Healthcare Corp · 1992
- Descendant · cites it directly · 0 of its ownK032789 Replacement Pump Head Door Assembly, P1, Model BX10468, Replacement…American I.V. Products, Inc. · 2003
- Descendant · cites it directly · 0 of its ownK973640 TS2000 Medical Device MonitorTrincore Systems, Inc. · 1998
Most influential descendants
Showing 2 of 2 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K915523 | Flo-Gard (R) 6301 | Baxter Healthcare Corp | 1992 | — |
| +1 | K032789 | Replacement Pump Head Door Assembly, P1, Model BX10468, Replacement Pump Head… | American I.V. Products, Inc. | 2003 | — |
| +1 | K973640 | TS2000 Medical Device Monitor | Trincore Systems, Inc. | 1998 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1998 | 1 | 0 |
| 2003 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FRN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FRN device
- #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (2)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 26, 2011 | Not classified | Z-0134-2012 | Employee error | The Flo-Gard infusion Pump was released to the customer with OOS values for the Air Sensor Calibrations with the Air in Tubing test. |
| Jun 21, 2007 | Not classified | Z-1112-2007 | Other | Falsified Repair/Test and Inspection records: Pumps submitted for repair or service may not have been properly repaired or serviced before being returned to the customer. |
| Nov 27, 2006 | Not classified | Z-0633-2007 | Other | There is the potential for non-detection of, and no alarm for, upstream occlusion for Flo-Gard Volumetric Infusion Pumps, if an occlusion occurs above a flexible chamber such as a drip chamber, Buretrol or blood filter. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K032789 (American I.V. Products, Inc., 2003) has gone furthest, with 0 descendants of its own.
How does its category compare?
All FRN devicesFRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FRN
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FRN
172in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FRN
7in 2025
Class I (37 since 2007) Class II/III not classified
Adverse event reports per year · FRN
52,792in 2025
MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 102 | 15 | 1,086 |
| 2008 | 8 | 89 | 24 | 8,261 |
| 2009 | 9 | 66 | 102 | 1,952 |
| 2010 | 3 | 93 | 37 | 6,133 |
| 2011 | 3 | 103 | 69 | 41,367 |
| 2012 | 4 | 217 | 42 (2) | 15,482 |
| 2013 | 6 | 312 | 22 (3) | 40,082 |
| 2014 | 6 | 172 | 4 (3) | 94,180 |
| 2015 | 3 | 232 | 5 | 26,742 |
| 2016 | 3 | 301 | 7 (1) | 16,560 |
| 2017 | 3 | 305 | 13 | 12,060 |
| 2018 | 3 | 267 | 1 | 16,591 |
| 2019 | 2 | — | 18 (1) | 15,850 |
| 2020 | 1 | — | 16 (5) | 94,186 |
| 2021 | 2 | — | 3 (2) | 199,342 |
| 2022 | 6 | 268 | 2 | 888,020 |
| 2023 | 7 | 265 | 16 (10) | 165,838 |
| 2024 | 5 | 103 | 9 (4) | 57,730 |
| 2025 | 8 | 172 | 7 (4) | 52,792 |
| 2026 | 2 | — | 11 (2) | — |
https://510kreport.com/device/K915523 · printed Nov 3, 2026
Cite this page
510(k) Report. “K915523: Flo-Gard (R) 6301.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K915523
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026