510(k) · Cleared
Recalled · Class II · 2021EZ Administration Set
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Y · FPA Set, Administration, Intravascular
Cleared on Jan 10, 1994 after a 287-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 later devices descend from it. It has been recalled 2 times.
- Decided
- Jan 10, 1994
- Review
- 287 days
- Class
- Class II
- Product code
- FPA
- Recalls
- 2
- Predicates
- 0
2 later devices descend from it, 1 citing it directly.
- This deviceK931550 EZ Administration SetY · 1994
- Descendant · cites it directly · 1 of its ownK032147 Signature Edition Infusion Pump, Signature Edition Administration…Alaris Medical Systems, Inc. · 2003 · Recalled
- Descendant · 2 generations on · 0 of its ownK063288 Signature Edition Gold Infusion System, Model 7X3XCardinal Health 303,Inc. · 2006
Most influential descendants
Showing 2 of 2 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K931550 | EZ Administration Set | Y | 1994 | — |
| +1 | K032147 | Signature Edition Infusion Pump, Signature Edition Administration Sets, Models… | Alaris Medical Systems, Inc. | 2003 | 1 |
| +2 | K063288 | Signature Edition Gold Infusion System, Model 7X3X | Cardinal Health 303,Inc. | 2006 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2003 | 1 | 0 |
| 2006 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FPA device
- #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (2)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K032147 | Signature Edition Infusion Pump, Signature Edition Administration Sets, Models 70XX… | 2003 | 1 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 2, 2021 | Class II | Z-2465-2021 | Process control | Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not… |
| Aug 30, 2013 | Class II | Z-2279-2013 | Employee error | CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model 72313E, Lot Number 13016314, because of an incorrect filter was assembled on this lot. The correct filter is a 1.2 micron filter (blue color) but a 0.2 micron filter… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K032147 (Alaris Medical Systems, Inc., 2003) has gone furthest, with 1 descendant of its own.
How does its category compare?
All FPA devicesFPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FPA
11in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FPA
225in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FPA
98in 2025
Class I (17 since 2007) Class II/III not classified
Adverse event reports per year · FPA
34,506in 2025
MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 16 | 75 | 17 | 1,499 |
| 2008 | 18 | 89 | 15 | 1,941 |
| 2009 | 10 | 166 | 166 | 2,512 |
| 2010 | 23 | 120 | 58 | 2,537 |
| 2011 | 12 | 87 | 28 | 4,407 |
| 2012 | 21 | 96 | 16 (4) | 4,930 |
| 2013 | 13 | 90 | 64 | 4,651 |
| 2014 | 22 | 89 | 24 | 4,695 |
| 2015 | 17 | 123 | 6 | 6,625 |
| 2016 | 12 | 195 | 10 | 8,470 |
| 2017 | 17 | 240 | 13 | 6,462 |
| 2018 | 4 | 177 | 4 | 6,574 |
| 2019 | 7 | 161 | 13 | 11,227 |
| 2020 | 10 | 327 | 107 | 9,563 |
| 2021 | 13 | 186 | 1 | 10,414 |
| 2022 | 14 | 213 | 9 (5) | 13,393 |
| 2023 | 14 | 340 | 12 (4) | 10,916 |
| 2024 | 7 | 244 | 13 (1) | 47,210 |
| 2025 | 11 | 225 | 98 (1) | 34,506 |
| 2026 | 5 | 266 | 10 (2) | — |
https://510kreport.com/device/K931550 · printed Nov 6, 2026
Cite this page
510(k) Report. “K931550: EZ Administration Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K931550
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026