510(k) · Cleared
Recalled · 2006Signature Edition Infusion Pump, Signature Edition Administration Sets, Models 70XX, 71XX, 72XX, 7XXXX Series
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Alaris Medical Systems, Inc. · FRN Pump, Infusion
Cleared on Jul 25, 2003 as substantially equivalent to K931549 (Y, 1993) and 1 other predicate after an 11-day review. Its family tree reaches back 1 generation to 1993. 1 later device descends from it. It has been recalled once.
- Decided
- Jul 25, 2003
- Review
- 11 days
- Class
- Class II
- Product code
- FRN
- Recalls
- 1
- Predicates
- 2
Its longest line of descent runs back 1 generation to K931549 (1993). 2 ancestors in all. 1 later device descends from it, 1 citing it directly.
- Root · 1 generation backK931549 Model EZ-1 Sing Chan/EZ-2 Dual Channel Infus PumpY · 1993 · Recalled
- This deviceK032147 Signature Edition Infusion Pump, Signature Edition Administration…Alaris Medical Systems, Inc. · 2003
- Descendant · cites it directly · 0 of its ownK063288 Signature Edition Gold Infusion System, Model 7X3XCardinal Health 303,Inc. · 2006
Most influential descendants
Showing 1 of 1 · See all- K063288Signature Edition Gold Infusion System, Model 7X3XCardinal Health 303,Inc. · 2006 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K931549 | Model EZ-1 Sing Chan/EZ-2 Dual Channel Infus Pump | Y | 1993 | 1 |
| 0 (this device) | K032147 | Signature Edition Infusion Pump, Signature Edition Administration Sets, Models… | Alaris Medical Systems, Inc. | 2003 | — |
| +1 | K063288 | Signature Edition Gold Infusion System, Model 7X3X | Cardinal Health 303,Inc. | 2006 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1994 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2006 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FRN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FRN device
- #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K931549 | Model EZ-1 Sing Chan/EZ-2 Dual Channel Infus PumpSummary page 1…cessories, FRN (880.5R5) Administration Sets, Intravascular, FPA (880-5440:) PREDICATE DEVICES: EZ 1 and EZ 2 Infusion Pumps, K93 1549 EZ Administration Sets, K93 1550 DEVICE DESCRIPTION This device is essentially the same as the or… | 1993 |
| * | K931550 | EZ Administration SetSummary page 1…ration Sets, Intravascular, FPA (880-5440:) PREDICATE DEVICES: EZ 1 and EZ 2 Infusion Pumps, K93 1549 EZ Administration Sets, K93 1550 DEVICE DESCRIPTION This device is essentially the same as the originally submitted devices (the predicate devices), filed as Mod… | 1994 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K063288 | Signature Edition Gold Infusion System, Model 7X3X | 2006 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 10, 2006 | Not classified | Z-1484-06 | Software design | This recall was initiated because of a potential for over infusion with all models of the Alaris SE Pumps (formerly the Signature Edition Infusion Pumps) caused by key bounce. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All FRN devicesFRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FRN
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FRN
172in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FRN
7in 2025
Class I (37 since 2007) Class II/III not classified
Adverse event reports per year · FRN
52,792in 2025
MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 12 | 102 | 15 | 1,086 |
| 2008 | 8 | 89 | 24 | 8,261 |
| 2009 | 9 | 66 | 102 | 1,952 |
| 2010 | 3 | 93 | 37 | 6,133 |
| 2011 | 3 | 103 | 69 | 41,367 |
| 2012 | 4 | 217 | 42 (2) | 15,482 |
| 2013 | 6 | 312 | 22 (3) | 40,082 |
| 2014 | 6 | 172 | 4 (3) | 94,180 |
| 2015 | 3 | 232 | 5 | 26,742 |
| 2016 | 3 | 301 | 7 (1) | 16,560 |
| 2017 | 3 | 305 | 13 | 12,060 |
| 2018 | 3 | 267 | 1 | 16,591 |
| 2019 | 2 | — | 18 (1) | 15,850 |
| 2020 | 1 | — | 16 (5) | 94,186 |
| 2021 | 2 | — | 3 (2) | 199,342 |
| 2022 | 6 | 268 | 2 | 888,020 |
| 2023 | 7 | 265 | 16 (10) | 165,838 |
| 2024 | 5 | 103 | 9 (4) | 57,730 |
| 2025 | 8 | 172 | 7 (4) | 52,792 |
| 2026 | 2 | — | 11 (2) | — |
What is it indicated for?
The Signature Edition Infusion System (infusion pumps and administration sets) is intended for use in today’s growing professional healthcare environment including healthcare facilities, home care, and medical transport that utilize infusion systems for the delivery of fluids, medications, blood and blood products.
The Signature Edition Infusion System is indicated for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural, centeral, or irrigation of fluid spaces.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K032147 · printed Nov 6, 2026
Cite this page
510(k) Report. “K032147: Signature Edition Infusion Pump, Signature Edition Administration Sets, Models 70XX, 71XX, 72XX, 7XXXX Series.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K032147
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026