510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Trilogy Acetabular System

K934765510(k) TraditionalCleared Apr 29, 1994206-day reviewClass II21 CFR 888.3358Recalls 16

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zimmer, Inc. · LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Cleared on Apr 29, 1994 after a 206-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 391 later devices descend from it. It has been recalled 16 times.

Decided
Apr 29, 1994
Review
206 days
Class
Class II
Product code
LPH
Recalls
16
Predicates
0

391 later devices descend from it, 12 citing it directly.

Descendants, by year cleared 2003: 1 descendant cleared2011: 9 descendants cleared2002: 1 descendant cleared2009: 3 descendants cleared1999: 3 descendants cleared2022: 25 descendants cleared2017: 16 descendants cleared2012: 25 descendants cleared2024: 26 descendants cleared2010: 12 descendants cleared2020: 26 descendants cleared2005: 3 descendants cleared2018: 23 descendants cleared2015: 15 descendants cleared2014: 16 descendants cleared2013: 21 descendants cleared2019: 33 descendants cleared2025: 38 descendants cleared2021: 21 descendants cleared2026: 27 descendants cleared2004: 4 descendants cleared2007: 2 descendants cleared2008: 1 descendant cleared2016: 15 descendants cleared2023: 24 descendants cleared2006: 1 descendant cleared1990200020102020
  1. This deviceK934765 Trilogy Acetabular SystemZimmer, Inc. · 1994
  2. Descendant · cites it directly · 248 of its ownK021891 Zimmer Trabecular Metal Modular Acetabular SystemZimmer, Inc. · 2002 · Recalled
  3. Descendant · cites it directly · 246 of its ownK091508 Zimmer Continuum Acetabular System, Zimmer Trilogy IT Acetabular…Zimmer, Inc. · 2009 · Recalled
  4. Descendant · cites it directly · 192 of its ownK990135 Trilogy Acetabular System Longevity Crosslinked PolyethyleneZimmer, Inc. · 1999
  5. Descendant · cites it directly · 161 of its ownK993082 Exactech Acumatch Integrated Hip System a-Series Porous Coated…Exactech, Inc. · 1999 · Recalled
  6. Descendant · 2 generations on · 145 of its ownK041906 Exactech 12/14 Total Hip SystemExactech, Inc. · 2004 · Recalled

Showing 5 of 391 descendants. See all descendants

Descendants cleared per year

1999: 32002: 12003: 12004: 42005: 32006: 12007: 22008: 12009: 32010: 122011: 92012: 252013: 212014: 162015: 152016: 152017: 162018: 232019: 332020: 262021: 212022: 252023: 242024: 262025: 382026: 27
1999high 38 (2025) · low 0 (2000)2026

Most influential descendants

Showing 5 of 391 · See all
  • K021891Zimmer Trabecular Metal Modular Acetabular SystemZimmer, Inc. · 2002 · cites it directly248descendants
  • K091508Zimmer Continuum Acetabular System, Zimmer Trilogy IT Acetabular SystemZimmer, Inc. · 2009 · cites it directly246descendants
  • K990135Trilogy Acetabular System Longevity Crosslinked PolyethyleneZimmer, Inc. · 1999 · cites it directly192descendants
  • K993082Exactech Acumatch Integrated Hip System a-Series Porous Coated Acetabular ComponentExactech, Inc. · 1999 · cites it directly161descendants
  • K041906Exactech 12/14 Total Hip SystemExactech, Inc. · 2004 · 2 generations on145descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K934765Trilogy Acetabular SystemZimmer, Inc.1994—
+1K021891Zimmer Trabecular Metal Modular Acetabular SystemZimmer, Inc.20023
+1K091508Zimmer Continuum Acetabular System, Zimmer Trilogy IT Acetabular SystemZimmer, Inc.20093
+1K990135Trilogy Acetabular System Longevity Crosslinked PolyethyleneZimmer, Inc.1999—
+1K993082Exactech Acumatch Integrated Hip System a-Series Porous Coated Acetabular…Exactech, Inc.19992
+2K041906Exactech 12/14 Total Hip SystemExactech, Inc.20041

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199930
200210
200310
200440
200530
200610
200720
200810
200930
2010120
201190
2012250
2013210
2014160
2015150
2016150
2017160
2018230
2019330
2020260
2021210
2022250
2023240
2024260
2025380
2026270
FIG. 1 — Lineage of K934765, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LPH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LPH device

  1. #1 K233498 Stryker Orthopaedics Hip Systems Labeling Updatescore 24.3 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (391)

See all 391

12 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K021891Zimmer Trabecular Metal Modular Acetabular SystemZimmer, Inc.2002248
K091508Zimmer Continuum Acetabular System, Zimmer Trilogy IT Acetabular SystemZimmer, Inc.2009246
K990135Trilogy Acetabular System Longevity Crosslinked PolyethyleneZimmer, Inc.1999192
K993082Exactech Acumatch Integrated Hip System a-Series Porous Coated Acetabular ComponentExactech, Inc.1999161
K030153Zca All-Poly Acetabular Cup, Snap-in, Model 8005-946/958-32Zimmer, Inc.200317
K170444PROCOTYL® PRIME Acetabular Cup SystemMicroport Orthopedics, Inc.201716
K112898Delta TT Cups, Delta TT Liners, Bone ScrewsLima Corporate S.P.A.201212
K241636MobileLink Acetabular Cup System - Line Extension (Multiple)Waldemar Link GmbH & Co. KG20242
K213842ONVOY Acetabular SystemGlobus Medical, Inc.20221
K092443Interface Acetabular Cup LinersOmni Life Science, Inc.20091

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 4, 2016Class IIZ-1684-2016Packaging change controlA small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
Jan 11, 2016Class IIZ-1145-2016PackagingAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Jan 11, 2016Class IIZ-1123-2016PackagingAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Jan 11, 2016Class IIZ-1161-2016PackagingAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Jan 11, 2016Class IIZ-1115-2016PackagingAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Dec 4, 2014Class IIZ-0875-2015Under Investigation by firmZimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but…
Nov 21, 2014Class IIZ-0824-2015Equipment maintenanceAnalysis of returned complaint devices, product in inventory, and device manufacturing records has identified tool marks isolated to one specific machining mill to be the cause of the fractures. .
Jun 12, 2014Class IIZ-2032-2014Process designTwo lots of screws were commingled. Screws with etching and machining for lot number 62629432 were packaged in bags labeled for lot number 62628759; and screws with etching and machining for lot number 62628759 were packaged in bags…
Oct 11, 2013Class IIZ-1024-2014PackagingZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the…
Sep 14, 2012Class IIZ-0771-2013Process controlSome units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of…
Sep 14, 2012Class IIZ-0773-2013Process controlSome units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of…
Sep 14, 2012Class IIZ-0769-2013Process controlSome units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of…
Sep 14, 2012Class IIZ-0775-2013Process controlSome units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of…
Sep 14, 2012Class IIZ-0772-2013Process controlSome units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of…
Oct 19, 2007Not classifiedZ-0620-2008Process controlParts missing and incomplete etch: The units may not contain the locking ring or the etched alignment marks on the rim.
Jun 26, 2006Not classifiedZ-1344-06OtherLack of assurance of sterility, as the sterile barrier packaging may have been compromised.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K021891 (Zimmer, Inc., 2002) has gone furthest, with 248 descendants of its own.

How does its category compare?

All LPH devices

LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LPH

12in 2025

2007: 102008: 102009: 32010: 152011: 162012: 252013: 112014: 202015: 182016: 102017: 122018: 92019: 222020: 72021: 82022: 142023: 102024: 62025: 122026 (partial year): 8
2007high 25 (2012) · low 3 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LPH

153in 2025

2007: 116 days median2008: 86 days median2009: 112 days median2010: 100 days median2011: 110 days median2012: 76 days median2013: 135 days median2014: 129 days median2015: 111 days median2016: 150 days median2017: 121 days median2018: 145 days median2019: 164 days median2020: 164 days median2021: 410 days median2022: 247 days median2023: 111 days median2024: 92 days median2025: 153 days median2026 (partial year): 256 days median
2007high 410 (2021) · low 76 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LPH

4in 2025

2007: 162008: 12009: 22010: 222011: 262012: 302013: 92014: 32015: 12016: 82017: 142018: 22019: 82020: 62021: 42022: 22023: 22024: 12025: 42026 (partial year): 2
2007high 30 (2012) · low 1 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LPH

6,163in 2025

2007: 5092008: 7062009: 1,1562010: 2,0072011: 3,3202012: 3,7852013: 5,0022014: 5,1632015: 5,8302016: 4,9002017: 5,3122018: 5,9542019: 7,1302020: 5,4292021: 5,9782022: 5,2232023: 5,3452024: 5,6262025: 6,163
2007high 7,130 (2019) · low 509 (2007)2025

MAUDE reports naming product code LPH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071011616509
200810861706
2009311221,156
201015100222,007
201116110263,320
20122576303,785
20131113595,002
20142012935,163
20151811115,830
20161015084,900
201712121145,312
2018914525,954
20192216487,130
2020716465,429
2021841045,978
20221424725,223
20231011125,345
202469215,626
20251215346,163
202682562—
Cite this page

510(k) Report. “K934765: Trilogy Acetabular System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K934765

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026