510(k) Report

510(k) · Cleared

Recalled · 2011

Jostra HL-20 Integrated Perfusion System

K943803510(k) TraditionalCleared Jan 30, 1996544-day reviewClass II21 CFR 870.4220Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Janus Biomedical, Inc. · DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Cleared on Jan 30, 1996 after a 544-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 of the 5 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 5 times.

Decided
Jan 30, 1996
Review
544 days
Class
Class II
Product code
DTQ
Recalls
5
Predicates
0

4 later devices descend from it, 2 citing it directly. 2 of the 5 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1999: 1 descendant cleared2011: 1 descendant cleared, 1 with a Class I recall12013: 1 descendant cleared2014: 1 descendant cleared, 1 with a Class I recall11995200020052010201520202025
  1. This deviceK943803 Jostra HL-20 Integrated Perfusion SystemJanus Biomedical, Inc. · 1996
  2. Descendant · cites it directly · 2 of its ownK102726 Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · Recalled (Class I)
  3. Descendant · 2 generations on · 0 of its ownK133598 Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · Recalled (Class I)
  4. Descendant · 2 generations on · 0 of its ownK133265 Quadrox-IR Adult and Small AdultMaquet Cardiopulmonary, AG · 2013
  5. Descendant · cites it directly · 0 of its ownK984338 Jostra HL-20 Twin PumpJostra Bentley · 1999

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · cites it directly2descendants
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 2 generations on0descendants
  • K133265Quadrox-IR Adult and Small AdultMaquet Cardiopulmonary, AG · 2013 · 2 generations on0descendants
  • K984338Jostra HL-20 Twin PumpJostra Bentley · 1999 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K943803Jostra HL-20 Integrated Perfusion SystemJanus Biomedical, Inc.1996—
+1K102726Cardiohelp SystemMaquet Cardiopulmonary, AG2011Class I
+2K133598Cardiohelp SystemMaquet Cardiopulmonary, AG2014Class I
+2K133265Quadrox-IR Adult and Small AdultMaquet Cardiopulmonary, AG2013—
+1K984338Jostra HL-20 Twin PumpJostra Bentley1999—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
201111
201310
201411
FIG. 1 — Lineage of K943803, 1996–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTQ device

  1. #1 K221373 Essenz HLMscore 8.9 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K102726Cardiohelp System Class IMaquet Cardiopulmonary, AG20112
K984338Jostra HL-20 Twin PumpJostra Bentley19990

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 20, 2011Not classifiedZ-3057-2011Device DesignHL 20 Consoles for 4-pump and 5-pump configuration can malfunction if an excessive amount of liquid is used causing circuit board malfunction.
Jun 20, 2011Not classifiedZ-3084-2011Device DesignMaquet HL 20 Twin Pump Modules may unintentionally increase speeds for a 1.5-2 second period of time before returning to its original speed.
Sep 28, 2010Not classifiedZ-1394-2011Component change controlA wrong ball bearing was installed in the pump heads of product. This may make the pump head wear out and cease function.
Oct 3, 2006Not classifiedZ-0159-2007OtherThe HL-20's pump head on the roller pump has four pairs of tube guide rolls that keep the tube in place in the raceway and roll along the inserted tube as the pump head turns. If dirt and other particles accumulate in the tube guide…
Sep 1, 2006Not classifiedZ-0158-2007OtherThe hand crank necessary for the emergency manual operation of the pump is too wide diametrically to fit into the holes which they were designed to fill.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K102726 (Maquet Cardiopulmonary, AG, 2011) has gone furthest, with 2 descendants of its own.

How does its category compare?

All DTQ devices

DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DTQ

3in 2023

2007: 22008: 12009: 32010: 12011: 22014: 22015: 22017: 22018: 12019: 12021: 22023: 3
2007high 3 (2009) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTQ

292in 2023

2009: 86 days median2023: 292 days median
2007high 292 (2023) · low 86 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTQ

1in 2024

2007: 92008: 172009: 12010: 172011: 222012: 422013: 42014: 12018: 12019: 1, 1 Class I2022: 12023: 7, 1 Class I2024: 1
2007high 42 (2012) · low 0 (2015)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · DTQ

1,014in 2025

2007: 4202008: 5482009: 5892010: 8112011: 1,0812012: 1,6262013: 9062014: 1,2902015: 1,1842016: 7142017: 5272018: 6762019: 8572020: 6272021: 7292022: 7172023: 8182024: 8632025: 1,014
2007high 1,626 (2012) · low 420 (2007)2025

MAUDE reports naming product code DTQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—9420
20081—17548
20093861589
20101—17811
20112—221,081
20120—421,626
20130—4906
20142—11,290
20152—01,184
20160—0714
20172—0527
20181—1676
20191—1 (1)857
20200—0627
20212—0729
20220—1717
202332927 (1)818
20240—1863
20250—01,014
20260—0—
Cite this page

510(k) Report. “K943803: Jostra HL-20 Integrated Perfusion System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K943803

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026