510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Essenz HLM, Essenz ILBM

K232291510(k) SpecialCleared Aug 24, 202323-day reviewClass II21 CFR 870.4220Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Livanova Deutschland, GmbH · DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Cleared on Aug 24, 2023 as substantially equivalent to K221373 (Livanova Deutschland, GmbH, 2023) and 1 other predicate after a 23-day review. Its family tree reaches back 5 generations to 1992. It has been recalled once.

Decided
Aug 24, 2023
Review
23 days
Class
Class II
Product code
DTQ
Recalls
1
Predicates
2

Its longest line of descent runs back 5 generations to K915183 (1992). 21 ancestors in all, the oldest K871131 (1987). No later device cites it as a predicate yet.

  1. Root · 5 generations backK915183 Sarns 8000 Modular Perfusion System3M Health Care, Sarns · 1992 · Recalled
  2. 4 generations backK953901 Sarns 800 Roller Pump3M Health Care, Sarns · 1996
  3. 3 generations backK022947 Advanced Perfusion System 1 (100/120 Volt),Advanced Perfusion System…Terumo Cardiovascular Systems Corp. · 2003 · Recalled
  4. 2 generations backK082344 Stockert S5 System, Stockert Electrical Venous Occluder (Evo)Sorin Group Deutschland GmbH · 2008 · Recalled
  5. 1 generation backK221373 Essenz HLMLivanova Deutschland, GmbH · 2023
  6. This deviceK232291 Essenz HLM, Essenz ILBMLivanova Deutschland, GmbH · 2023
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K915183Sarns 8000 Modular Perfusion System3M Health Care, Sarns199245
−4K953901Sarns 800 Roller Pump3M Health Care, Sarns1996—
−3K022947Advanced Perfusion System 1 (100/120 Volt),Advanced Perfusion System 1…Terumo Cardiovascular Systems Corp.2003150
−2K082344Stockert S5 System, Stockert Electrical Venous Occluder (Evo)Sorin Group Deutschland GmbH20081
−1K221373Essenz HLMLivanova Deutschland, GmbH2023—
0 (this device)K232291Essenz HLM, Essenz ILBMLivanova Deutschland, GmbH2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199510
199620
199710
200010
200410
200610
200710
200920
201130
202120
FIG. 1 — Lineage of K232291, 1992–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DTQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTQ device

  1. #1 K221373 Essenz HLMscore 8.9 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K221373Essenz HLM
Summary page 3
…��� DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Submission Number (if known) I K221373 Indications for Use…
Livanova Deutschland, GmbH2023—
*K202154B-Capta
Summary page 6
…to its predicate Essenz HLM, K221373, LivaNova Deutschland GmbH. Essenz ILBM is substantially equivalent in function and intended use to its predicate B- Capta, K202154, LivaNova Deutschland GmbH. INDICATIONS FOR USE Essenz HLM is intended to be used during cardiopulmonary bypass for procedu…
Livanova Deutschland, GmbH2021—
†K210130Stockert S5 SystemLivanova Deutschland, GmbH2021—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 18, 2024Class IIZ-0571-2025Software change controlHeart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 4 companies; a new generation was cleared every 5.7 years on median.
  • 9 of the 21 ancestors have a recall of any class.

How does its category compare?

All DTQ devices

DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · DTQ

3in 2023

2007: 22008: 12009: 32010: 12011: 22014: 22015: 22017: 22018: 12019: 12021: 22023: 3
2007high 3 (2009) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTQ

292in 2023

2009: 86 days median2023: 292 days median
2007high 292 (2023) · low 86 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTQ

1in 2024

2007: 92008: 172009: 12010: 172011: 222012: 422013: 42014: 12018: 12019: 1, 1 Class I2022: 12023: 7, 1 Class I2024: 1
2007high 42 (2012) · low 0 (2015)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · DTQ

1,014in 2025

2007: 4202008: 5482009: 5892010: 8112011: 1,0812012: 1,6262013: 9062014: 1,2902015: 1,1842016: 7142017: 5272018: 6762019: 8572020: 6272021: 7292022: 7172023: 8182024: 8632025: 1,014
2007high 1,626 (2012) · low 420 (2007)2025

MAUDE reports naming product code DTQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—9420
20081—17548
20093861589
20101—17811
20112—221,081
20120—421,626
20130—4906
20142—11,290
20152—01,184
20160—0714
20172—0527
20181—1676
20191—1 (1)857
20200—0627
20212—0729
20220—1717
202332927 (1)818
20240—1863
20250—01,014
20260—0—

What is it indicated for?

Essenz HLM is intended to be used during cardiopulmonary bypass for procedures lasting six (6) hours or less.

Essenz ILBM is indicated for supplementary, in-line monitoring of the extracorporeal arterial oxygen partial pressure, venous oxygen saturation, venous hematocrit/hemoglobin, and arterial and venous temperature during cardiopulmonary bypass procedures up to six hours.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K232291: Essenz HLM, Essenz ILBM.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K232291

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026