Vacutainer (R) Brand Plus PST (TM) Plasma Separation Tube

K945952510(k) TraditionalCleared Jan 18, 199543-day reviewClass II21 CFR 862.1675Recalls 10

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bd Becton Dickinson Vacutainer Systems Preanalytic · JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Cleared on Jan 18, 1995 after a 43-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 18 later devices descend from it. It has been recalled 10 times.

Decided
Jan 18, 1995
Review
43 days
Class
Class II
Product code
JKA
Recalls
10
Predicates
0

18 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 1998: 1 descendant cleared2026: 1 descendant cleared2002: 2 descendants cleared2017: 1 descendant cleared2015: 2 descendants cleared2011: 1 descendant cleared2005: 1 descendant cleared2001: 1 descendant cleared2007: 1 descendant cleared2020: 2 descendants cleared2016: 2 descendants cleared2023: 1 descendant cleared2019: 1 descendant cleared2022: 1 descendant cleared1995200020052010201520202025
  1. This deviceK945952 Vacutainer (R) Brand Plus PST (TM) Plasma Separation TubeBd Becton Dickinson Vacutainer Systems Preanalytic · 1995
  2. Descendant · cites it directly · 13 of its ownK972075 Vacutainer Brand PPT Plasma Preparation TubeBecton Dickinson Vacutainer Systems · 1998
  3. Descendant · 2 generations on · 6 of its ownK110701 Oragene-DX Ogd-500; Oragene-DX Ogd-575; Oragene-DX Oxd-525…DNA Genotek, Inc. · 2011
  4. Descendant · 3 generations on · 3 of its ownK141410 Oragene DX Ogd-500.001DNA Genotek, Inc. · 2015
  5. Descendant · 3 generations on · 2 of its ownK152556 Oragene DxDNA Genotek, Inc. · 2016
  6. Descendant · cites it directly · 2 of its ownK022130 BD Vacutainer Plus PST II TubeBecton, Dickinson & CO · 2002

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

1998: 12001: 12002: 22005: 12007: 12011: 12015: 22016: 22017: 12019: 12020: 22022: 12023: 12026: 1
1998high 2 (2002) · low 0 (1999)2026

Most influential descendants

Showing 5 of 18 · See all
  • K972075Vacutainer Brand PPT Plasma Preparation TubeBecton Dickinson Vacutainer Systems · 1998 · cites it directly13descendants
  • K110701Oragene-DX Ogd-500; Oragene-DX Ogd-575; Oragene-DX Oxd-525; Oragene-DX Oyd-500DNA Genotek, Inc. · 2011 · 2 generations on6descendants
  • K141410Oragene DX Ogd-500.001DNA Genotek, Inc. · 2015 · 3 generations on3descendants
  • K152556Oragene DxDNA Genotek, Inc. · 2016 · 3 generations on2descendants
  • K022130BD Vacutainer Plus PST II TubeBecton, Dickinson & CO · 2002 · cites it directly2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K945952Vacutainer (R) Brand Plus PST (TM) Plasma Separation TubeBd Becton Dickinson Vacutainer Systems Preanalytic1995—
+1K972075Vacutainer Brand PPT Plasma Preparation TubeBecton Dickinson Vacutainer Systems1998—
+2K110701Oragene-DX Ogd-500; Oragene-DX Ogd-575; Oragene-DX Oxd-525; Oragene-DX Oyd-500DNA Genotek, Inc.2011—
+3K141410Oragene DX Ogd-500.001DNA Genotek, Inc.2015—
+3K152556Oragene DxDNA Genotek, Inc.2016—
+1K022130BD Vacutainer Plus PST II TubeBecton, Dickinson & CO2002—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199810
200110
200220
200510
200710
201110
201520
201620
201710
201910
202020
202210
202310
202610
FIG. 1 — Lineage of K945952, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JKA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JKA device

  1. #1 K200027 Blood Collection Needle with/without Holder…score 15.6 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (18)

See all 18

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K972075Vacutainer Brand PPT Plasma Preparation TubeBecton Dickinson Vacutainer Systems199813
K022130BD Vacutainer Plus PST II TubeBecton, Dickinson & CO20022
K252040BD Vacutainer® Plasma Separator Tubes (PST™), BD Vacutainer® Sodium Heparin Blood…Becton, Dickinson and Company20260

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 12, 2019Class IIZ-2300-2019Device DesignFalse elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.
Jun 12, 2019Class IIZ-2304-2019Device DesignFalse elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.
Jun 12, 2019Class IIZ-2301-2019Device DesignFalse elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.
Jun 12, 2019Class IIZ-2305-2019Device DesignFalse elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.
Jun 12, 2019Class IIZ-2302-2019Device DesignFalse elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.
Jun 12, 2019Class IIZ-2303-2019Device DesignFalse elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.
Jun 12, 2019Class IIZ-2308-2019Device DesignFalse elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.
Mar 16, 2017Class IIZ-1746-2017Nonconforming Material/ComponentBD is initiating this product correction of multiple lots of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (13x75 mm x 4.5 mL). A limited number of tubes in these lots may exhibit stopper pullout from the tube as the…
Jan 16, 2017Class IIZ-1189-2017Process controlA portion of this lot of blood collection tubes was manufactured with approximately twice the expected amount of buffered sodium citrate additive, an anticoagulant.
Mar 5, 2012Class IIIZ-0911-2013Equipment maintenanceBD is conducting a recall of the BD Vacutainer Plus Plastic Plasma Separator Tube (PST) with Light Green/BD Homogard Closure due to a deformed top of the tube reservoir reported in two customer complaints.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K972075 (Becton Dickinson Vacutainer Systems, 1998) has gone furthest, with 13 descendants of its own.

How does its category compare?

All JKA devices

JKA Tubes, Vials, Systems, Serum Separators, Blood Collection: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JKA

7in 2025

2007: 42008: 22009: 22010: 42012: 12013: 32014: 12015: 12016: 42017: 22018: 12019: 62020: 62021: 42022: 92023: 172024: 62025: 72026 (partial year): 8
2007high 17 (2023) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JKA

115in 2025

2007: 281 days median2010: 28 days median2013: 244 days median2016: 116 days median2019: 181 days median2020: 181 days median2021: 217 days median2022: 164 days median2023: 189 days median2024: 240 days median2025: 115 days median2026 (partial year): 250 days median
2007high 281 (2007) · low 28 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JKA

13in 2025

2007: 32008: 32009: 32010: 42011: 142012: 22013: 12014: 12015: 62016: 12017: 22018: 22019: 92021: 32022: 22023: 22025: 132026 (partial year): 1
2007high 14 (2011) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JKA

6,089in 2025

2007: 732008: 782009: 422010: 372011: 672012: 542013: 332014: 682015: 412016: 652017: 2,0392018: 9,7312019: 6,4692020: 3,7742021: 3,6102022: 3,3752023: 4,4002024: 4,6092025: 6,089
2007high 9,731 (2018) · low 33 (2013)2025

MAUDE reports naming product code JKA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074281373
20082—378
20092—342
2010428437
20110—1467
20121—254
20133244133
20141—168
20151—641
20164116165
20172—22,039
20181—29,731
2019618196,469
2020618103,774
2021421733,610
2022916423,375
20231718924,400
2024624004,609
20257115136,089
202682501—
Cite this page

510(k) Report. “K945952: Vacutainer (R) Brand Plus PST (TM) Plasma Separation Tube.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K945952

Report an extraction error