510(k) · Cleared

Recalled · Class II · 2017

Beta Hip Prosthesis

K953337510(k) TraditionalCleared Jan 22, 1996189-day reviewClass II21 CFR 888.3358Recalls 15

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zimmer, Inc. · LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Cleared on Jan 22, 1996 after a 189-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 22 later devices descend from it. It has been recalled 15 times.

Decided
Jan 22, 1996
Review
189 days
Class
Class II
Product code
LPH
Recalls
15
Predicates
0

22 later devices descend from it, 4 citing it directly.

Descendants, by year cleared 1997: 3 descendants cleared2019: 1 descendant cleared2003: 2 descendants cleared2006: 2 descendants cleared2004: 1 descendant cleared2020: 2 descendants cleared2017: 1 descendant cleared2026: 3 descendants cleared2023: 2 descendants cleared2022: 1 descendant cleared2024: 1 descendant cleared2018: 1 descendant cleared2021: 1 descendant cleared2025: 1 descendant cleared1995200020052010201520202025
  1. This deviceK953337 Beta Hip ProsthesisZimmer, Inc. · 1996
  2. Descendant · cites it directly · 18 of its ownK973714 Versys Hip System Beaded Hip ProsthesisZimmer, Inc. · 1997 · Recalled
  3. Descendant · 2 generations on · 11 of its ownK042776 Versys Hip System Beaded Midcoat Low Head Center Hip Prosthesis…Zimmer, Inc. · 2004 · Recalled
  4. Descendant · 3 generations on · 10 of its ownK061786 Versys Fiber Metal Midcoat Low Head Center Hip ProsthesisZimmer, Inc. · 2006
  5. Descendant · 4 generations on · 9 of its ownK163577 Ecofit® Hip SystemImplantcast GmbH · 2017 · Recalled
  6. Descendant · 5 generations on · 5 of its ownK180263 EcoFit Vit E Acetabular SystemImplantcast GmbH · 2018

Showing 5 of 22 descendants. See all descendants

Descendants cleared per year

1997: 32003: 22004: 12006: 22017: 12018: 12019: 12020: 22021: 12022: 12023: 22024: 12025: 12026: 3
1997high 3 (1997) · low 0 (1998)2026

Most influential descendants

Showing 5 of 22 · See all
  • K973714Versys Hip System Beaded Hip ProsthesisZimmer, Inc. · 1997 · cites it directly18descendants
  • K042776Versys Hip System Beaded Midcoat Low Head Center Hip Prosthesis, Model 00-7840-XXX-XXZimmer, Inc. · 2004 · 2 generations on11descendants
  • K061786Versys Fiber Metal Midcoat Low Head Center Hip ProsthesisZimmer, Inc. · 2006 · 3 generations on10descendants
  • K163577Ecofit® Hip SystemImplantcast GmbH · 2017 · 4 generations on9descendants
  • K180263EcoFit Vit E Acetabular SystemImplantcast GmbH · 2018 · 5 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K953337Beta Hip ProsthesisZimmer, Inc.1996—
+1K973714Versys Hip System Beaded Hip ProsthesisZimmer, Inc.19978
+2K042776Versys Hip System Beaded Midcoat Low Head Center Hip Prosthesis, Model…Zimmer, Inc.20041
+3K061786Versys Fiber Metal Midcoat Low Head Center Hip ProsthesisZimmer, Inc.2006—
+4K163577Ecofit® Hip SystemImplantcast GmbH20171
+5K180263EcoFit Vit E Acetabular SystemImplantcast GmbH2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199730
200320
200410
200620
201710
201810
201910
202020
202110
202210
202320
202410
202510
202630
FIG. 1 — Lineage of K953337, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LPH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LPH device

  1. #1 K233498 Stryker Orthopaedics Hip Systems Labeling Updatescore 24.3 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (22)

See all 22

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K973714Versys Hip System Beaded Hip ProsthesisZimmer, Inc.199718
K183457Zimmer Biomet 12/14 CoCr Femoral Head and Freedom HeadZimmer, Inc.20190
K972690Zirconia Ceramic Femoral HeadZimmer, Inc.19970
K971752Zirconia Ceramic Femoral Head (12/14 Taper)Zimmer, Inc.19970

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 1, 2017Class IIZ-0722-2018Package design/selectionThe products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
Feb 28, 2017Class IIZ-2577-2017PackagingZimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
Feb 28, 2017Class IIZ-2579-2017PackagingZimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
Feb 28, 2017Class IIZ-2578-2017PackagingZimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
Feb 28, 2017Class IIZ-2580-2017PackagingZimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
Feb 28, 2017Class IIZ-2581-2017PackagingZimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
Jan 11, 2016Class IIZ-0849-2016Process controlLDPE bag containing the implant adheres to the highly polished implant surface.
Jan 11, 2016Class IIZ-1127-2016PackagingAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
Oct 11, 2013Class IIZ-1031-2014PackagingZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the…
Oct 11, 2013Class IIZ-1034-2014PackagingZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the…
Aug 30, 2013Class IIZ-0070-2014Package design/selectionThe low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
Jul 5, 2010Not classifiedZ-2291-2010OtherZimmer, Inc. has determined that there is a potential for a package to be labeled as a 28 mm femoral head but contain a 32 mm femoral head. In addition, there is a potential for the patient record label to be incorrect.
Jul 5, 2010Not classifiedZ-2290-2010OtherZimmer, Inc. has determined that there is a potential for a package to be labeled as a 28 mm femoral head but contain a 32 mm femoral head. In addition, there is a potential for the patient record label to be incorrect.
Aug 3, 2009Not classifiedZ-0198-2010Manufacturing material removalManufacturing material may be present on the surface of the implant.
Apr 15, 2005Not classifiedZ-1066-05OtherMay be mislabeled as to dimensions; may be 12/14 taper, 28mm dia., neck length +3.5mm.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K973714 (Zimmer, Inc., 1997) has gone furthest, with 18 descendants of its own.

How does its category compare?

All LPH devices

LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LPH

12in 2025

2007: 102008: 102009: 32010: 152011: 162012: 252013: 112014: 202015: 182016: 102017: 122018: 92019: 222020: 72021: 82022: 142023: 102024: 62025: 122026 (partial year): 8
2007high 25 (2012) · low 3 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LPH

153in 2025

2007: 116 days median2008: 86 days median2009: 112 days median2010: 100 days median2011: 110 days median2012: 76 days median2013: 135 days median2014: 129 days median2015: 111 days median2016: 150 days median2017: 121 days median2018: 145 days median2019: 164 days median2020: 164 days median2021: 410 days median2022: 247 days median2023: 111 days median2024: 92 days median2025: 153 days median2026 (partial year): 256 days median
2007high 410 (2021) · low 76 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LPH

4in 2025

2007: 162008: 12009: 22010: 222011: 262012: 302013: 92014: 32015: 12016: 82017: 142018: 22019: 82020: 62021: 42022: 22023: 22024: 12025: 42026 (partial year): 2
2007high 30 (2012) · low 1 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LPH

6,163in 2025

2007: 5092008: 7062009: 1,1562010: 2,0072011: 3,3202012: 3,7852013: 5,0022014: 5,1632015: 5,8302016: 4,9002017: 5,3122018: 5,9542019: 7,1302020: 5,4292021: 5,9782022: 5,2232023: 5,3452024: 5,6262025: 6,163
2007high 7,130 (2019) · low 509 (2007)2025

MAUDE reports naming product code LPH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071011616509
200810861706
2009311221,156
201015100222,007
201116110263,320
20122576303,785
20131113595,002
20142012935,163
20151811115,830
20161015084,900
201712121145,312
2018914525,954
20192216487,130
2020716465,429
2021841045,978
20221424725,223
20231011125,345
202469215,626
20251215346,163
202682562—
Cite this page

510(k) Report. “K953337: Beta Hip Prosthesis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K953337

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026