510(k) · Cleared
Recalled · Class II · 2017Versys Hip System Beaded Hip Prosthesis
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Zimmer, Inc. · LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Cleared on Dec 24, 1997 as substantially equivalent to K953337 (Zimmer, Inc., 1996) and 2 other predicates after an 86-day review. Its family tree reaches back 1 generation to 1994. 18 later devices descend from it. It has been recalled 8 times.
- Decided
- Dec 24, 1997
- Review
- 86 days
- Class
- Class II
- Product code
- LPH
- Recalls
- 8
- Predicates
- 3
Its longest line of descent runs back 1 generation to K931641 (1994). 3 ancestors in all. 18 later devices descend from it, 3 citing it directly.
- Root · 1 generation backK931641 Porocoat Porous CoatingDepuy, Inc. · 1994
- This deviceK973714 Versys Hip System Beaded Hip ProsthesisZimmer, Inc. · 1997
- Descendant · cites it directly · 11 of its ownK042776 Versys Hip System Beaded Midcoat Low Head Center Hip Prosthesis…Zimmer, Inc. · 2004 · Recalled
- Descendant · 2 generations on · 10 of its ownK061786 Versys Fiber Metal Midcoat Low Head Center Hip ProsthesisZimmer, Inc. · 2006
- Descendant · 3 generations on · 9 of its ownK163577 Ecofit® Hip SystemImplantcast GmbH · 2017 · Recalled
- Descendant · 4 generations on · 5 of its ownK180263 EcoFit Vit E Acetabular SystemImplantcast GmbH · 2018
- Descendant · 4 generations on · 4 of its ownK203420 EcoFit® Hip SystemImplantcast GmbH · 2021
Showing 5 of 18 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 18 · See all- K042776Versys Hip System Beaded Midcoat Low Head Center Hip Prosthesis, Model 00-7840-XXX-XXZimmer, Inc. · 2004 · cites it directly11descendants
- K061786Versys Fiber Metal Midcoat Low Head Center Hip ProsthesisZimmer, Inc. · 2006 · 2 generations on10descendants
- K163577Ecofit® Hip SystemImplantcast GmbH · 2017 · 3 generations on9descendants
- K180263EcoFit Vit E Acetabular SystemImplantcast GmbH · 2018 · 4 generations on5descendants
- K203420EcoFit® Hip SystemImplantcast GmbH · 2021 · 4 generations on4descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K931641 | Porocoat Porous Coating | Depuy, Inc. | 1994 | — |
| 0 (this device) | K973714 | Versys Hip System Beaded Hip Prosthesis | Zimmer, Inc. | 1997 | — |
| +1 | K042776 | Versys Hip System Beaded Midcoat Low Head Center Hip Prosthesis, Model… | Zimmer, Inc. | 2004 | 1 |
| +2 | K061786 | Versys Fiber Metal Midcoat Low Head Center Hip Prosthesis | Zimmer, Inc. | 2006 | — |
| +3 | K163577 | Ecofit® Hip System | Implantcast GmbH | 2017 | 1 |
| +4 | K180263 | EcoFit Vit E Acetabular System | Implantcast GmbH | 2018 | — |
| +4 | K203420 | EcoFit® Hip System | Implantcast GmbH | 2021 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1996 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2003 | 2 | 0 |
| 2004 | 1 | 0 |
| 2006 | 2 | 0 |
| 2017 | 1 | 0 |
| 2018 | 1 | 0 |
| 2020 | 2 | 0 |
| 2021 | 1 | 0 |
| 2022 | 1 | 0 |
| 2023 | 2 | 0 |
| 2024 | 1 | 0 |
| 2025 | 1 | 0 |
| 2026 | 3 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LPH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LPH device
- #1 K233498 Stryker Orthopaedics Hip Systems Labeling Updatescore 24.3 · 20 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K953337 | Beta Hip ProsthesisSummary page 2…emi-constrained porous-coated uncemented prosthesis éf\\é‘; A Bristol-Myers Squibb Company ey Predicate Devices: - Beta System, manufactured by Zimmer, K953337, cleared January 22, 1996 - AML Total Hip System with Porocoat, manufactured by DePuy, P820024, approved for use without bone ce… | 1996 |
| * | P820024 | Prosthesis, Hip, Hemi-Femoral, Metal FDA lists this PMA only through its supplements; the earliest we hold is P820024/S001 (Mar 16, 1984). Summary page 2…Predicate Devices: - Beta System, manufactured by Zimmer, K953337, cleared January 22, 1996 - AML Total Hip System with Porocoat, manufactured by DePuy, P820024, approved for use without bone cement August 19, 1993, (formerly called the Porocoat Modified Austin-Moore Total Hip Prosthesis);… | |
| * | K931641 | Porocoat Porous CoatingSummary page 2…DePuy, P820024, approved for use without bone cement August 19, 1993, (formerly called the Porocoat Modified Austin-Moore Total Hip Prosthesis); Porocoat Porous Coating, K931641, cleared for use without bone cement March 21, 1994 - Prodigy Hip Prosthesis, manufactured by DePuy, K931641, cleared March 21,… | 1994 |
What descends from it? (18)
See all 183 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 29, 2017 | Class II | Z-1026-2018 | Packaging | The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag. |
| Nov 29, 2017 | Class II | Z-1023-2018 | Packaging | The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag. |
| Nov 29, 2017 | Class II | Z-1025-2018 | Packaging | The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag. |
| Nov 29, 2017 | Class II | Z-1022-2018 | Packaging | The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag. |
| Nov 29, 2017 | Class II | Z-1024-2018 | Packaging | The LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag. |
| Aug 1, 2017 | Class II | Z-0722-2018 | Package design/selection | The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier. |
| Jan 11, 2016 | Class II | Z-0848-2016 | Process control | LDPE bag containing the implant adheres to the highly polished implant surface. |
| May 19, 2014 | Class II | Z-1842-2014 | Nonconforming Material/Component | During routine inspection, beaded hip stem forgings received from supplier failed to meet ZES 2A-102. Processing of Beaded CoCrMo Alloy Hip Stem, material requirements per ATS #14-04818 and ATS #14-05549, respectively. |
The 5 latest of 8 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All LPH devicesLPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LPH
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LPH
153in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LPH
4in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LPH
6,163in 2025
MAUDE reports naming product code LPH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 10 | 116 | 16 | 509 |
| 2008 | 10 | 86 | 1 | 706 |
| 2009 | 3 | 112 | 2 | 1,156 |
| 2010 | 15 | 100 | 22 | 2,007 |
| 2011 | 16 | 110 | 26 | 3,320 |
| 2012 | 25 | 76 | 30 | 3,785 |
| 2013 | 11 | 135 | 9 | 5,002 |
| 2014 | 20 | 129 | 3 | 5,163 |
| 2015 | 18 | 111 | 1 | 5,830 |
| 2016 | 10 | 150 | 8 | 4,900 |
| 2017 | 12 | 121 | 14 | 5,312 |
| 2018 | 9 | 145 | 2 | 5,954 |
| 2019 | 22 | 164 | 8 | 7,130 |
| 2020 | 7 | 164 | 6 | 5,429 |
| 2021 | 8 | 410 | 4 | 5,978 |
| 2022 | 14 | 247 | 2 | 5,223 |
| 2023 | 10 | 111 | 2 | 5,345 |
| 2024 | 6 | 92 | 1 | 5,626 |
| 2025 | 12 | 153 | 4 | 6,163 |
| 2026 | 8 | 256 | 2 | — |
What is it indicated for?
The VerSys Beaded Hip Prosthesis is designed to achieve biologic fixation to bone and is indicated for: Total hip replacement for the following: severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, nonunion of previous fractures of the femur; congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, disability due…
Read the full statement (140 words)
The VerSys Beaded Hip Prosthesis is designed to achieve biologic fixation to bone and is indicated for:
Total hip replacement for the following: severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, nonunion of previous fractures of the femur; congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, disability due to previous fusion, previously failed endoprotheses and/or total hip components in the affected extremity, and acute femoral neck fractures.
Hemi-hip replacement for the following: fracture dislocation of the hip, elderly, debilitated patients when a total hip replacement is contraindicated, irreducible fractures in which adequate fixation cannot be obtained, certain high subcapital fractures and comminuted fractures, secondary avascular necrosis of the femoral head, pathological fractures of the femoral neck, and osteoarthritis in which the femoral head is primarily affected.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K973714 · printed Nov 4, 2026
Cite this page
510(k) Report. “K973714: Versys Hip System Beaded Hip Prosthesis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K973714
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026