510(k) · Cleared
Recalled · Class II · 2025Stryker Strykeflow Suction Irrigator (Modification)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Stryker Endoscopy · GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered
Cleared on Nov 29, 1995 after a 118-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 11 later devices descend from it. It has been recalled 22 times.
- Decided
- Nov 29, 1995
- Review
- 118 days
- Class
- Class II
- Product code
- GCX
- Recalls
- 22
- Predicates
- 0
11 later devices descend from it, 2 citing it directly.
- This deviceK954726 Stryker Strykeflow Suction Irrigator (Modification)Stryker Endoscopy · 1995
- Descendant · cites it directly · 5 of its ownK963765 Stryker Strykeflow Electrocautery ProbesStryker Endoscopy · 1997 · Recalled
- Descendant · cites it directly · 4 of its ownK042454 Stryker Cassette PumpStryker Endoscopy · 2004 · Recalled
- Descendant · 2 generations on · 3 of its ownK122619 Epix Irrigation Pump, Epix Suction Irrigation Tubing SetApplied Medical · 2012
- Descendant · 3 generations on · 2 of its ownK180735 Endomat SelectKarl Storz SE & CO. KG · 2018
- Descendant · 4 generations on · 1 of its ownK201355 Endomat Select, IBS Shaver Software, LicenseKarl Storz SE & CO. KG · 2020
Showing 5 of 11 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 11 · See all- K963765Stryker Strykeflow Electrocautery ProbesStryker Endoscopy · 1997 · cites it directly5descendants
- K042454Stryker Cassette PumpStryker Endoscopy · 2004 · cites it directly4descendants
- K122619Epix Irrigation Pump, Epix Suction Irrigation Tubing SetApplied Medical · 2012 · 2 generations on3descendants
- K180735Endomat SelectKarl Storz SE & CO. KG · 2018 · 3 generations on2descendants
- K201355Endomat Select, IBS Shaver Software, LicenseKarl Storz SE & CO. KG · 2020 · 4 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K954726 | Stryker Strykeflow Suction Irrigator (Modification) | Stryker Endoscopy | 1995 | — |
| +1 | K963765 | Stryker Strykeflow Electrocautery Probes | Stryker Endoscopy | 1997 | 1 |
| +1 | K042454 | Stryker Cassette Pump | Stryker Endoscopy | 2004 | 5 |
| +2 | K122619 | Epix Irrigation Pump, Epix Suction Irrigation Tubing Set | Applied Medical | 2012 | — |
| +3 | K180735 | Endomat Select | Karl Storz SE & CO. KG | 2018 | — |
| +4 | K201355 | Endomat Select, IBS Shaver Software, License | Karl Storz SE & CO. KG | 2020 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1997 | 1 | 0 |
| 2004 | 1 | 0 |
| 2005 | 1 | 0 |
| 2010 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2013 | 1 | 0 |
| 2018 | 1 | 0 |
| 2019 | 1 | 0 |
| 2020 | 1 | 0 |
| 2026 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in GCX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (11)
See all 112 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 18, 2025 | Class II | Z-2171-2025 | Under Investigation by firm | Due to complaints their is the potential that irrigation solution may leaked into the handpiece and battery pack causing it to malfunction. |
| Nov 17, 2017 | Class II | Z-0835-2018 | Under Investigation by firm | Routine bioburden testing of certain lots were found to have levels higher than internal acceptable rates. |
| Mar 27, 2013 | Class II | Z-1225-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1220-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1235-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1224-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1226-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1223-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1233-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1229-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1237-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1219-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1218-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1232-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1234-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1231-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1228-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1230-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1222-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1227-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1236-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
| Mar 27, 2013 | Class II | Z-1221-2013 | Error in labeling | The pre-vacuum steam sterilization (wrapped method) temperature in the Instructions for Use incorrectly listed 121-123¿C , instead of the correct 132-133¿C. |
The 5 latest of 22 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K963765 (Stryker Endoscopy, 1997) has gone furthest, with 5 descendants of its own.
How does its category compare?
All GCX devicesGCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · GCX
FDA has decided no GCX device since 2007.
Median 510(k) review, days · GCX
Too few GCX 510(k)s in any year since 2007 to give a median.
Recalls per year · GCX
1in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · GCX
274in 2025
MAUDE reports naming product code GCX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 24 |
| 2008 | 0 | — | 0 | 44 |
| 2009 | 0 | — | 0 | 40 |
| 2010 | 0 | — | 0 | 51 |
| 2011 | 0 | — | 0 | 67 |
| 2012 | 0 | — | 0 | 242 |
| 2013 | 0 | — | 1 | 47 |
| 2014 | 0 | — | 0 | 32 |
| 2015 | 0 | — | 0 | 52 |
| 2016 | 0 | — | 0 | 99 |
| 2017 | 0 | — | 1 | 45 |
| 2018 | 0 | — | 0 | 50 |
| 2019 | 0 | — | 0 | 65 |
| 2020 | 0 | — | 0 | 65 |
| 2021 | 0 | — | 0 | 65 |
| 2022 | 0 | — | 0 | 72 |
| 2023 | 0 | — | 3 | 59 |
| 2024 | 0 | — | 1 | 65 |
| 2025 | 0 | — | 0 | 274 |
| 2026 | 0 | — | 3 | — |
https://510kreport.com/device/K954726 · printed Nov 6, 2026
Cite this page
510(k) Report. “K954726: Stryker Strykeflow Suction Irrigator (Modification).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K954726
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026