510(k) · Cleared
ClearedRmi Internal Mammary Artery Cannula
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Research Medical, Inc. · DQR Cannula, Catheter
Cleared on Mar 26, 1996 after a 75-day review. The FDA summary names its predicate without a number.
- Decided
- Mar 26, 1996
- Review
- 75 days
- Class
- Class II
- Product code
- DQR
- Recalls
- None
- Predicates
- 0
The FDA summary names its predicate without a number.
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DQR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | DLP I.M.A. Cannula, Code #31001 Named without a submission number |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
How does its category compare?
All DQR devicesDQR Cannula, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DQR
1in 2013
2026 to Oct 25, outlined.
Median 510(k) review, days · DQR
Too few DQR 510(k)s in any year since 2007 to give a median.
Recalls per year · DQR
1in 2020
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DQR
6in 2025
MAUDE reports naming product code DQR, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 17 |
| 2008 | 1 | — | 0 | 8 |
| 2009 | 0 | — | 0 | 27 |
| 2010 | 0 | — | 0 | 24 |
| 2011 | 0 | — | 0 | 7 |
| 2012 | 0 | — | 0 | 7 |
| 2013 | 1 | — | 0 | 6 |
| 2014 | 0 | — | 0 | 8 |
| 2015 | 0 | — | 0 | 40 |
| 2016 | 0 | — | 0 | 75 |
| 2017 | 0 | — | 0 | 29 |
| 2018 | 0 | — | 0 | 17 |
| 2019 | 0 | — | 0 | 16 |
| 2020 | 0 | — | 1 | 4 |
| 2021 | 0 | — | 0 | 11 |
| 2022 | 0 | — | 0 | 27 |
| 2023 | 0 | — | 0 | 36 |
| 2024 | 0 | — | 0 | 7 |
| 2025 | 0 | — | 0 | 6 |
| 2026 | 0 | — | 0 | — |
https://510kreport.com/device/K960098 · printed Nov 4, 2026
Cite this page
510(k) Report. “K960098: Rmi Internal Mammary Artery Cannula.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K960098
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026