Recalled · Class II · 2024CNS Interchangeable Bit Cranial Drill
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Clinical Neuro Systems, LLC · HBG Drills, Burrs, Trephines & Accessories (Manual)
Cleared on Jun 5, 1996 after a 77-day review. The FDA summary names its predicate without a number. 7 later devices descend from it. It has been recalled 39 times.
- Decided
- Jun 5, 1996
- Review
- 77 days
- Class
- Class II
- Product code
- HBG
- Recalls
- 39
- Predicates
- 0
7 later devices descend from it, 3 citing it directly.
- This deviceK961113 CNS Interchangeable Bit Cranial DrillClinical Neuro Systems, LLC · 1996
- Descendant · cites it directly · 4 of its ownK122456 Mrii Crainial DrillMri Interventions, Inc. · 2013
- Descendant · 2 generations on · 2 of its ownK151536 SmartTwist MRII Hand Drill, SmartTip MRII Drill bit Kit, 4.5mm, 6.0mmMri Interventions, Inc. · 2015
- Descendant · 2 generations on · 1 of its ownK160129 MRII Cranial Drill and AccessoriesMri Interventions, Inc. · 2016
- Descendant · cites it directly · 0 of its ownK252696 LegacyPhasor Health, LLC · 2025
- Descendant · cites it directly · 0 of its ownK252514 Evac-MriPhasor Health, LLC · 2025
Showing 5 of 7 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 7 · See all- K122456Mrii Crainial DrillMri Interventions, Inc. · 2013 · cites it directly4descendants
- K151536SmartTwist MRII Hand Drill, SmartTip MRII Drill bit Kit, 4.5mm, 6.0mmMri Interventions, Inc. · 2015 · 2 generations on2descendants
- K160129MRII Cranial Drill and AccessoriesMri Interventions, Inc. · 2016 · 2 generations on1descendant
- K252696LegacyPhasor Health, LLC · 2025 · cites it directly0descendants
- K252514Evac-MriPhasor Health, LLC · 2025 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K961113 | CNS Interchangeable Bit Cranial Drill | Clinical Neuro Systems, LLC | 1996 | — |
| +1 | K122456 | Mrii Crainial Drill | Mri Interventions, Inc. | 2013 | — |
| +2 | K151536 | SmartTwist MRII Hand Drill, SmartTip MRII Drill bit Kit, 4.5mm, 6.0mm | Mri Interventions, Inc. | 2015 | — |
| +2 | K160129 | MRII Cranial Drill and Accessories | Mri Interventions, Inc. | 2016 | — |
| +1 | K252696 | Legacy | Phasor Health, LLC | 2025 | — |
| +1 | K252514 | Evac-Mri | Phasor Health, LLC | 2025 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2013 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2016 | 1 | 0 |
| 2017 | 1 | 0 |
| 2025 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HBG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HBG device
- #1 K252696 Legacyscore 12.2 · 7 generations · no Class I recall
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | PS Medical Ventriculostomy Kit: Cranial Drill Named without a submission number | |
| * | — | Camino Laboratories Twist Drill Procedure Kit: Cranial Drill Named without a submission number | |
| * | — | Clinical Neuro Systems Cranial Drill Named without a submission number |
What descends from it? (7)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Apr 17, 2024 | Class II | Z-1873-2024 | Process control | Potential holes and tears in the sterile packaging (header bag) of the Cranial Access Kit may compromise the sterility of the device. |
| Jan 19, 2024 | Class II | Z-1197-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Jan 19, 2024 | Class II | Z-1199-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Jan 19, 2024 | Class II | Z-1201-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Jan 19, 2024 | Class II | Z-1200-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Jan 19, 2024 | Class II | Z-1196-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Jan 19, 2024 | Class II | Z-1194-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Jan 19, 2024 | Class II | Z-1203-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Jan 19, 2024 | Class II | Z-1195-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Jan 19, 2024 | Class II | Z-1193-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Jan 19, 2024 | Class II | Z-1198-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Jan 19, 2024 | Class II | Z-1202-2024 | Process control | A defect in the sterile packaging (header bag) of the Cranial Access Kit failing the required packaging integrity testing criteria. Therefore, product sterility maybe compromised. |
| Oct 19, 2023 | Class II | Z-0492-2024 | Process control | Defect in the outer packaging of the Cranial Access Kits can cause the packaging to split without any additional forces outside compromising the sterility of the device |
| Oct 19, 2023 | Class II | Z-0493-2024 | Process control | Defect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device |
| Oct 19, 2023 | Class II | Z-0494-2024 | Process control | Defect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device |
| Oct 19, 2023 | Class II | Z-0495-2024 | Process control | Defect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device |
| Oct 19, 2023 | Class II | Z-0496-2024 | Process control | Defect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device |
| Oct 19, 2023 | Class II | Z-0491-2024 | Process control | Defect in the outer packaging of the Cranial Access Kits can cause the packaging to split without any additional forces outside compromising the sterility of the device |
| Oct 19, 2023 | Class II | Z-0499-2024 | Process control | Defect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device |
| Oct 19, 2023 | Class II | Z-0497-2024 | Process control | Defect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device |
| Oct 19, 2023 | Class II | Z-0498-2024 | Process control | Defect in the outer packaging of the Cranial Access Kits. can cause the packaging to split without any additional forces outside compromising the sterility of the device |
| Aug 24, 2016 | Class II | Z-2733-2016 | Packaging | Integra LifeSciences has been notified of a medical device recall by Adaptive Surgical, LLC involving their Riverlon" Nylon 3-0 Sutures that are packaged in Integras Cranial Access Kits. It was found that product seals may become open… |
| Sep 4, 2013 | Class II | Z-2260-2013 | Packaging | Integra is recalling certain lots of Cranial Access Kit due to a possibility that the seal integrity of the sterile package may become compromised during shipping. |
| Dec 21, 2011 | Not classified | Z-0749-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0750-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0752-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0756-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0757-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0748-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0751-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0755-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0758-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0759-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0753-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Dec 21, 2011 | Not classified | Z-0754-2012 | Labeling False and Misleading | Current package insert was revised to include additional warning information related to the potential hazards of an unapproved off-label use of Xylocaine. |
| Feb 22, 2011 | Not classified | Z-2194-2011 | Labeling mix-ups | Outer packaging of Combo Kit was mislabeled with an incorrect lot number. |
| Feb 22, 2011 | Not classified | Z-2171-2011 | Error in labeling | Outer box of Monitoring Kit was mislabeled with the incorrect part number. |
| Jul 13, 2009 | Not classified | Z-2084-2009 | Packaging process control | Lack of packaging seal integrity on various pain management kits may compromise sterility of contents. |
| Mar 19, 2003 | Not classified | Z-1097-04 | Other | The outer labeling of the convenience kit indicates that an additional bottle of Sodium Chloride (saline) solution is included in the Custom Kit 31176; however, an additional bottle of 1% Xylocaine was included in the Custom Kit. |
The 5 latest of 39 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K122456 (Mri Interventions, Inc., 2013) has gone furthest, with 4 descendants of its own.
How does its category compare?
All HBG devicesHBG Drills, Burrs, Trephines & Accessories (Manual): FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · HBG
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HBG
Too few HBG 510(k)s in any year since 2007 to give a median.
Recalls per year · HBG
9in 2023
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HBG
2in 2025
MAUDE reports naming product code HBG, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 1 | 0 |
| 2010 | 0 | — | 0 | 0 |
| 2011 | 0 | — | 14 | 0 |
| 2012 | 0 | — | 0 | 1 |
| 2013 | 1 | — | 0 | 1 |
| 2014 | 0 | — | 14 | 5 |
| 2015 | 1 | — | 0 | 3 |
| 2016 | 1 | — | 0 | 0 |
| 2017 | 2 | — | 0 | 6 |
| 2018 | 0 | — | 0 | 2 |
| 2019 | 0 | — | 0 | 3 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 1 |
| 2022 | 0 | — | 0 | 3 |
| 2023 | 0 | — | 9 | 3 |
| 2024 | 0 | — | 0 | 11 |
| 2025 | 1 | — | 0 | 2 |
| 2026 | 0 | — | 1 | — |
What is it indicated for?
The drill is intended to be used with an external drainage and monitoring system in selected patients to reduce intracranial pressure (ICP), to monitor CSF, to provide temporary drainage of CSF, and to monitor ICP. The drill is intended for single use only.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K961113 · printed Nov 5, 2026
Cite this page
510(k) Report. “K961113: CNS Interchangeable Bit Cranial Drill.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K961113