Clinical Neuro Syst Monitorr Ext CSF Drain/Moni SY

K920156510(k) TraditionalCleared Mar 16, 199262-day reviewClass II21 CFR 882.5550Recalls 1 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Clinical Neuro Systems, LLC · JXG Shunt, Central Nervous System And Components

Cleared on Mar 16, 1992 after a 62-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 4 of the 8 devices in its family, counting itself, have a Class I recall. 7 later devices descend from it. It has been recalled once, including a Class I recall.

Decided
Mar 16, 1992
Review
62 days
Class
Class II
Product code
JXG
Recalls
1 (1 Class I)
Predicates
0

7 later devices descend from it, 4 citing it directly. 4 of the 8 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2006: 1 descendant cleared, 1 with a Class I recall12019: 1 descendant cleared2002: 1 descendant cleared, 1 with a Class I recall11999: 1 descendant cleared2007: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant cleared2020: 1 descendant cleared1990200020102020
  1. This deviceK920156 Clinical Neuro Syst Monitorr Ext CSF Drain/Moni SYClinical Neuro Systems, LLC · 1992
  2. Descendant · cites it directly · 4 of its ownK022554 Monitorr Icp External CSF Drainage and Monitoring SystemIntegra Neurosciences · 2002 · Recalled (Class I)
  3. Descendant · 2 generations on · 2 of its ownK072929 Limitorr Volume Limiting External CSF Drainage and Monitoring…Integra LifeSciences Corporation · 2007 · Recalled (Class I)
  4. Descendant · cites it directly · 1 of its ownK191684 MoniTorr ICP External Drainage and Monitoring System, LimiTorr…Integra LifeSciences Corporation · 2019
  5. Descendant · cites it directly · 1 of its ownK983799 Medtronic PS Medical Niveau Drainage System, Medtronic PS Medical…Medtronic PS Medical · 1999
  6. Descendant · 2 generations on · 0 of its ownK221694 External Drainage SystemJmed (Shenzhen) Technology Limited · 2023

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

1999: 12002: 12006: 12007: 12019: 12020: 12023: 1
1999high 1 (1999) · low 0 (2000)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 7 · See all
  • K022554Monitorr Icp External CSF Drainage and Monitoring SystemIntegra Neurosciences · 2002 · cites it directly4descendants
  • K072929Limitorr Volume Limiting External CSF Drainage and Monitoring System, Models INS9020…Integra LifeSciences Corporation · 2007 · 2 generations on2descendants
  • K191684MoniTorr ICP External Drainage and Monitoring System, LimiTorr Volume Limiting External…Integra LifeSciences Corporation · 2019 · cites it directly1descendant
  • K983799Medtronic PS Medical Niveau Drainage System, Medtronic PS Medical Niveau Drainage System…Medtronic PS Medical · 1999 · cites it directly1descendant
  • K221694External Drainage SystemJmed (Shenzhen) Technology Limited · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K920156Clinical Neuro Syst Monitorr Ext CSF Drain/Moni SYClinical Neuro Systems, LLC1992—
+1K022554Monitorr Icp External CSF Drainage and Monitoring SystemIntegra Neurosciences2002Class I
+2K072929Limitorr Volume Limiting External CSF Drainage and Monitoring System, Models…Integra LifeSciences Corporation2007Class I
+1K191684MoniTorr ICP External Drainage and Monitoring System, LimiTorr Volume Limiting…Integra LifeSciences Corporation2019—
+1K983799Medtronic PS Medical Niveau Drainage System, Medtronic PS Medical Niveau…Medtronic PS Medical1999—
+2K221694External Drainage SystemJmed (Shenzhen) Technology Limited2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
200211
200611
200711
201910
202010
202310
FIG. 1 — Lineage of K920156, 1992–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXG device

  1. #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (7)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K022554Monitorr Icp External CSF Drainage and Monitoring System Class IIntegra Neurosciences20024
K191684MoniTorr ICP External Drainage and Monitoring System, LimiTorr Volume Limiting External…Integra LifeSciences Corporation20191
K983799Medtronic PS Medical Niveau Drainage System, Medtronic PS Medical Niveau Drainage System…Medtronic PS Medical19991
K062599Monitorr Icp External Drainage and Monitoring Systems Class IIntegra LifeSciences Corporation20060

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 8, 2019Class IZ-1310-2019Device DesignThe firm has identified a complaint trend regarding breakage of the LimiTorr Transducer Mount stopcock and MoniTorr Panel Mount stopcock.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K062599Monitorr Icp External Drainage and Monitoring SystemsIntegra LifeSciences Corporation · 2006 · Cites it directly · ◆ in the 2006 descendants bar of FIG. 12 Class I recalls
  • K022554Monitorr Icp External CSF Drainage and Monitoring SystemIntegra Neurosciences · 2002 · Cites it directly · ◆ in the 2002 descendants bar of FIG. 12 Class I recalls
  • K072929Limitorr Volume Limiting External CSF Drainage and Monitoring System, Models INS9020, INS9030Integra LifeSciences Corporation · 2007 · 2 generations on · ◆ in the 2007 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K022554 (Integra Neurosciences, 2002) has gone furthest, with 4 descendants of its own.

How does its category compare?

All JXG devices

JXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JXG

8in 2025

2007: 32008: 62009: 52010: 32011: 72012: 32013: 32015: 62016: 62017: 72018: 62019: 42020: 22021: 12022: 42023: 52024: 32025: 82026 (partial year): 2
2007high 8 (2025) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXG

255in 2025

2007: 184 days median2008: 60 days median2009: 106 days median2010: 30 days median2011: 136 days median2012: 78 days median2013: 108 days median2015: 276 days median2016: 143 days median2017: 90 days median2018: 102 days median2019: 101 days median2022: 101 days median2023: 134 days median2024: 90 days median2025: 255 days median
2007high 276 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JXG

3in 2024

2008: 12009: 302011: 52012: 12013: 142014: 62017: 12020: 82023: 22024: 32026 (partial year): 1
2007high 30 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXG

1,198in 2025

2007: 6492008: 6302009: 6292010: 8122011: 7872012: 9772013: 1,0042014: 1,2512015: 9842016: 1,0912017: 1,3352018: 1,4842019: 1,2952020: 1,2062021: 1,4672022: 1,3292023: 1,2212024: 1,2142025: 1,198
2007high 1,484 (2018) · low 629 (2009)2025

MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731840649
20086601630
2009510630629
20103300812
201171365787
20123781977
20133108141,004
20140—61,251
201562760984
2016614301,091
201779011,335
2018610201,484
2019410101,295
20202—81,206
20211—01,467
2022410101,329
2023513421,221
202439031,214
2025825501,198
20262—1—
Cite this page

510(k) Report. “K920156: Clinical Neuro Syst Monitorr Ext CSF Drain/Moni SY.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K920156

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026