Polyfin Infusion Set MMT-106/MMT-107/MMT-133

K961474510(k) TraditionalCleared May 6, 199618-day reviewClass II21 CFR 880.5440

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Minimed · FPA Set, Administration, Intravascular

Cleared on May 6, 1996 as substantially equivalent to K831445 (Pacesetter Systems, 1983) and 1 other predicate after an 18-day review. Its family tree reaches back 1 generation to 1983. 20 later devices descend from it.

Decided
May 6, 1996
Review
18 days
Class
Class II
Product code
FPA
Recalls
None
Predicates
2

Its longest line of descent runs back 1 generation to K831445 (1983). 2 ancestors in all. 20 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1986: 1 other ancestor2002: 1 descendant cleared1997: 2 descendants cleared2011: 1 descendant cleared2013: 1 descendant cleared2008: 1 descendant cleared2012: 3 descendants cleared2004: 1 descendant cleared2014: 1 descendant cleared2021: 2 descendants cleared2015: 1 descendant cleared2016: 1 descendant cleared2022: 1 descendant cleared2018: 1 descendant cleared2024: 1 descendant cleared2023: 1 descendant cleared2025: 1 descendant clearedK831445 · Infusion Sets #'S 124 & 142 Personal · Pacesetter Systems · 198319801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 6 descendants3 generations forward: 5 descendants4 generations forward: 1 descendant5 generations forward: 2 descendants6 generations forward: 2 descendants7 generations forward: 1 descendant K831445 · Infusion Sets #'S 124 & 142 Personal · Pacesetter Systems · 19830+5
  1. Root · 1 generation backK831445 Infusion Sets #'S 124 & 142 PersonalPacesetter Systems · 1983
  2. This deviceK961474 Polyfin Infusion Set MMT-106/MMT-107/MMT-133Medtronic Minimed · 1996
  3. Descendant · cites it directly · 17 of its ownK020530 Evans Sub-Q, Model MC4206Evans Medical, Inc. · 2002
  4. Descendant · 2 generations on · 8 of its ownK140131 Sub-Q Subcutaneous Tissue Infusion SetEmed Technologies Corporation · 2014
  5. Descendant · 3 generations on · 6 of its ownK142319 SCIg60 Infuser, InfusetEmed Technologies Corporation · 2015
  6. Descendant · 2 generations on · 6 of its ownK082818 Subcutaneous Needle Infusion SetMarcal Medical, Inc. · 2008
  7. Descendant · 4 generations on · 5 of its ownK161906 SCIg60 InfuserEmed Technologies Corporation · 2016

Showing 5 of 20 descendants. See all descendants

Descendants cleared per year

1997: 22002: 12004: 12008: 12011: 12012: 32013: 12014: 12015: 12016: 12018: 12021: 22022: 12023: 12024: 12025: 1
1997high 3 (2012) · low 0 (1998)2026

Most influential descendants

Showing 5 of 20 · See all
  • K020530Evans Sub-Q, Model MC4206Evans Medical, Inc. · 2002 · cites it directly17descendants
  • K140131Sub-Q Subcutaneous Tissue Infusion SetEmed Technologies Corporation · 2014 · 2 generations on8descendants
  • K142319SCIg60 Infuser, InfusetEmed Technologies Corporation · 2015 · 3 generations on6descendants
  • K082818Subcutaneous Needle Infusion SetMarcal Medical, Inc. · 2008 · 2 generations on6descendants
  • K161906SCIg60 InfuserEmed Technologies Corporation · 2016 · 4 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K831445Infusion Sets #'S 124 & 142 PersonalPacesetter Systems1983—
0 (this device)K961474Polyfin Infusion Set MMT-106/MMT-107/MMT-133Medtronic Minimed1996—
+1K020530Evans Sub-Q, Model MC4206Evans Medical, Inc.2002—
+2K140131Sub-Q Subcutaneous Tissue Infusion SetEmed Technologies Corporation2014—
+3K142319SCIg60 Infuser, InfusetEmed Technologies Corporation2015—
+2K082818Subcutaneous Needle Infusion SetMarcal Medical, Inc.2008—
+4K161906SCIg60 InfuserEmed Technologies Corporation2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199720
200210
200410
200810
201110
201230
201310
201410
201510
201610
201810
202120
202210
202310
202410
202510
FIG. 1 — Lineage of K961474, 1983–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K831445Infusion Sets #'S 124 & 142 Personal
Our OCR of the scanned summary read K838445, which is not a number in FDA’s lists. Corrected to the number one digit away whose FDA name carries the summary’s words for the device, “Pacesetter infusion set, model 124 and 142”.
Pacesetter Systems1983—
*K861436Sof Set, Models 111 and 112
Summary page 1
…et, Models MMT-106, MMT-107, and MMT-133. C. Predicate device: Pacesetter infusion set, model 124 and 142 (K838445A); Sof-set® Infusion Set, Models MMT-111 and MMT-112 (K861436). D. Description of new device: This infusion set is intended for the administration of medicine, including insulin, from a port…
Pacesetter Systems1986—

What descends from it? (20)

See all 20

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K020530Evans Sub-Q, Model MC4206Evans Medical, Inc.200217
K964456Polyfin QR with Wings Infusion Sets, Models 365, 366, and 367Medtronic Minimed19970
K964455Polyfin with Wings Infusion Sets, Models 306, 307, 333Medtronic Minimed19970

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K020530 (Evans Medical, Inc., 2002) has gone furthest, with 17 descendants of its own.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—

What is it indicated for?

The intended use of the new device is subcutaneous infusion of medicine, including insulin, from an external pump, for up to 48 hours.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K961474: Polyfin Infusion Set MMT-106/MMT-107/MMT-133.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K961474

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026