510(k) · Cleared
Recalled · 2003Gish Reliance Oxygenator (Model Hfo)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Gish Biomedical, Inc. · DTZ Oxygenator, Cardiopulmonary Bypass
Cleared on Dec 5, 1996 as substantially equivalent to K923272 (Cobe Laboratories, Inc., 1993) and 1 other predicate after a 227-day review. Its family tree reaches back 1 generation to 1992. 10 later devices descend from it. It has been recalled once.
- Decided
- Dec 5, 1996
- Review
- 227 days
- Class
- Class II
- Product code
- DTZ
- Recalls
- 1
- Predicates
- 2
Its longest line of descent runs back 1 generation to K922933 (1992). 2 ancestors in all. 10 later devices descend from it, 3 citing it directly.
- Root · 1 generation backK922933 Sorin Monolyrh Integrated Membrane LungSorin Biomedica, Fiat, USA, Inc. · 1992
- This deviceK961530 Gish Reliance Oxygenator (Model Hfo)Gish Biomedical, Inc. · 1996
- Descendant · cites it directly · 6 of its ownK082403 Medos Hilite Oxygenator, Model 7000 & 7000 LTGish Biomedical, Inc. · 2009
- Descendant · 2 generations on · 5 of its ownK133261 Medos Hilite 7000 & 7000 LT OxygenatorMedos Medizintechnik AG · 2014
- Descendant · 3 generations on · 3 of its ownK191407 Novalung SystemFresenius Medical Care Renal Therapies Group, LLC · 2020 · Recalled
- Descendant · cites it directly · 2 of its ownK023381 Vision Hollow Fiber Oxygenator with Guardian CoatingGish Biomedical, Inc. · 2003 · Recalled
- Descendant · 4 generations on · 1 of its ownK240880 VitalFlow Set with Balance BiosurfaceMedtronic, Inc. · 2024
Showing 5 of 10 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 10 · See all- K082403Medos Hilite Oxygenator, Model 7000 & 7000 LTGish Biomedical, Inc. · 2009 · cites it directly6descendants
- K133261Medos Hilite 7000 & 7000 LT OxygenatorMedos Medizintechnik AG · 2014 · 2 generations on5descendants
- K191407Novalung SystemFresenius Medical Care Renal Therapies Group, LLC · 2020 · 3 generations on3descendants
- K023381Vision Hollow Fiber Oxygenator with Guardian CoatingGish Biomedical, Inc. · 2003 · cites it directly2descendants
- K240880VitalFlow Set with Balance BiosurfaceMedtronic, Inc. · 2024 · 4 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K922933 | Sorin Monolyrh Integrated Membrane Lung | Sorin Biomedica, Fiat, USA, Inc. | 1992 | — |
| 0 (this device) | K961530 | Gish Reliance Oxygenator (Model Hfo) | Gish Biomedical, Inc. | 1996 | — |
| +1 | K082403 | Medos Hilite Oxygenator, Model 7000 & 7000 LT | Gish Biomedical, Inc. | 2009 | — |
| +2 | K133261 | Medos Hilite 7000 & 7000 LT Oxygenator | Medos Medizintechnik AG | 2014 | — |
| +3 | K191407 | Novalung System | Fresenius Medical Care Renal Therapies Group, LLC | 2020 | 2 |
| +1 | K023381 | Vision Hollow Fiber Oxygenator with Guardian Coating | Gish Biomedical, Inc. | 2003 | 2 |
| +4 | K240880 | VitalFlow Set with Balance Biosurface | Medtronic, Inc. | 2024 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1993 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2003 | 2 | 0 |
| 2009 | 2 | 0 |
| 2014 | 1 | 0 |
| 2020 | 1 | 0 |
| 2024 | 2 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DTZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DTZ device
- #1 K240934 Quantum Perfusion Blood Oxygenator ECC…score 14.3 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K923272 | Cobe CMS-HF Hollow Fiber Membrane OxygenatorSummary page 1…4.553.7390 (FAX) Jack W. Brown President DEVICE NAME MODEL NUMBER Gish Reliance™ Oxygenator HFO NAME OF PREDICATE OR LEGALLY MARKETED DEVICE COBE Optima Oxygenator (K923272) and Sorin MONOLYTH Oxygenator (K922933). DESCRIPTION OF DEVICE The Gish Reliance™ Oxygenator is designed to control blood gase… | 1993 |
| * | K922933 | Sorin Monolyrh Integrated Membrane LungSummary page 1…t DEVICE NAME MODEL NUMBER Gish Reliance™ Oxygenator HFO NAME OF PREDICATE OR LEGALLY MARKETED DEVICE COBE Optima Oxygenator (K923272) and Sorin MONOLYTH Oxygenator (K922933). DESCRIPTION OF DEVICE The Gish Reliance™ Oxygenator is designed to control blood gases in the extracorporeal circuit during c… | 1992 |
What descends from it? (10)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Mar 31, 2003 | Not classified | Z-0780-03 | Other | Two reports of leakage. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- None of the 2 ancestors has a recall.
- Of its descendants, K082403 (Gish Biomedical, Inc., 2009) has gone furthest, with 6 descendants of its own.
How does its category compare?
All DTZ devicesDTZ Oxygenator, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DTZ
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DTZ
124in 2024
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DTZ
1in 2025
Class I (3 since 2007) Class II/III not classified
Adverse event reports per year · DTZ
976in 2025
MAUDE reports naming product code DTZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 4 | 55 | 7 | 134 |
| 2008 | 3 | 88 | 5 | 105 |
| 2009 | 9 | 210 | 0 | 52 |
| 2010 | 2 | — | 1 | 119 |
| 2011 | 4 | 160 | 13 | 190 |
| 2012 | 9 | 90 | 14 | 215 |
| 2013 | 9 | 28 | 0 | 346 |
| 2014 | 5 | 28 | 2 | 764 |
| 2015 | 9 | 134 | 0 | 658 |
| 2016 | 3 | 62 | 1 | 443 |
| 2017 | 2 | — | 0 | 388 |
| 2018 | 3 | 72 | 2 (2) | 320 |
| 2019 | 5 | 90 | 1 | 489 |
| 2020 | 5 | 93 | 3 | 442 |
| 2021 | 5 | 336 | 1 (1) | 595 |
| 2022 | 2 | — | 6 | 559 |
| 2023 | 5 | 145 | 0 | 738 |
| 2024 | 3 | 124 | 0 | 535 |
| 2025 | 2 | — | 1 | 976 |
| 2026 | 3 | 137 | 0 | — |
What is it indicated for?
The Gish Reliance™ Oxygenator is indicated for use in procedures requiring the extracorporeal oxygenation of, and carbon dioxide removal from, human blood. It is designed to operate at blood flow rates of one half (0.5) to eight (8.0) liters per minute for periods of up to six (6.0) hours.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K961530 · printed Nov 4, 2026
Cite this page
510(k) Report. “K961530: Gish Reliance Oxygenator (Model Hfo).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K961530
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026