510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Genesis II Zirconium Femoral Component/Profix Zirconium Femoral Component

K962557510(k) TraditionalCleared Dec 5, 1996157-day reviewClass II21 CFR 888.3560Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smith & Nephew Richards, Inc. · JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Cleared on Dec 5, 1996 after a 157-day review. The FDA summary names its predicate without a number. 130 later devices descend from it. It has been recalled once.

Decided
Dec 5, 1996
Review
157 days
Class
Class II
Product code
JWH
Recalls
1
Predicates
0

130 later devices descend from it, 6 citing it directly.

Descendants, by year cleared 2005: 2 descendants cleared2011: 6 descendants cleared2010: 3 descendants cleared2012: 10 descendants cleared2007: 2 descendants cleared2021: 11 descendants cleared2015: 9 descendants cleared2022: 9 descendants cleared2009: 1 descendant cleared2023: 12 descendants cleared2006: 1 descendant cleared2008: 1 descendant cleared2013: 7 descendants cleared2014: 10 descendants cleared2017: 3 descendants cleared2020: 9 descendants cleared2019: 5 descendants cleared2018: 3 descendants cleared2026: 7 descendants cleared2024: 7 descendants cleared2016: 5 descendants cleared2025: 7 descendants cleared1995200020052010201520202025
  1. This deviceK962557 Genesis II Zirconium Femoral Component/Profix Zirconium Femoral…Smith & Nephew Richards, Inc. · 1996
  2. Descendant · cites it directly · 118 of its ownK042515 Smith & Nephew, Inc. High Performance KneeSmith & Nephew, Inc. · 2005 · Recalled
  3. Descendant · 2 generations on · 66 of its ownK071071 Crosslinked Polyethylene Articular InsertsSmith & Nephew, Inc. · 2007 · Recalled
  4. Descendant · 2 generations on · 45 of its ownK101499 Journey CR Knee SystemSmith & Nephew, Inc. · 2010
  5. Descendant · 2 generations on · 41 of its ownK094050 Itotal CR Knee Replacement SystemConformis, Inc. · 2010
  6. Descendant · 2 generations on · 41 of its ownK091014 Journey BCS Knee SystemSmith & Nephew, Inc. · 2009

Showing 5 of 130 descendants. See all descendants

Descendants cleared per year

2005: 22006: 12007: 22008: 12009: 12010: 32011: 62012: 102013: 72014: 102015: 92016: 52017: 32018: 32019: 52020: 92021: 112022: 92023: 122024: 72025: 72026: 7
2005high 12 (2023) · low 1 (2006)2026

Most influential descendants

Showing 5 of 130 · See all
  • K042515Smith & Nephew, Inc. High Performance KneeSmith & Nephew, Inc. · 2005 · cites it directly118descendants
  • K071071Crosslinked Polyethylene Articular InsertsSmith & Nephew, Inc. · 2007 · 2 generations on66descendants
  • K101499Journey CR Knee SystemSmith & Nephew, Inc. · 2010 · 2 generations on45descendants
  • K094050Itotal CR Knee Replacement SystemConformis, Inc. · 2010 · 2 generations on41descendants
  • K091014Journey BCS Knee SystemSmith & Nephew, Inc. · 2009 · 2 generations on41descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K962557Genesis II Zirconium Femoral Component/Profix Zirconium Femoral ComponentSmith & Nephew Richards, Inc.1996—
+1K042515Smith & Nephew, Inc. High Performance KneeSmith & Nephew, Inc.20053
+2K071071Crosslinked Polyethylene Articular InsertsSmith & Nephew, Inc.20075
+2K101499Journey CR Knee SystemSmith & Nephew, Inc.2010—
+2K094050Itotal CR Knee Replacement SystemConformis, Inc.2010—
+2K091014Journey BCS Knee SystemSmith & Nephew, Inc.2009—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200520
200610
200720
200810
200910
201030
201160
2012100
201370
2014100
201590
201650
201730
201830
201950
202090
2021110
202290
2023120
202470
202570
202670
FIG. 1 — Lineage of K962557, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JWH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JWH device

  1. #1 K221479 Persona Personalized Knee Systemscore 22.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Genesis I C/R and LDK Femoral Component
Named without a submission number
—
*—Profix C/R Femoral Component
Named without a submission number
—
*—Genesis I C/R Femoral Component
Named without a submission number
—

What descends from it? (130)

See all 130

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042515Smith & Nephew, Inc. High Performance KneeSmith & Nephew, Inc.2005118
K093746Verilast Wear Claims for the Legion Primary Knee SystemSmith & Nephew, Inc.201029
K112941Legion Narrow Oxinium Cruciate Retaining and Posterior Stabilized Fenoral ComponentsSmith & Nephew, Inc.20118
K071220Columbus Total Knee System asAesculap Implant Systems, Inc.20077
K121393Smith & Nephew, Inc. Total Knee System InstrumentationSmith & Nephew, Inc.20125
K052265Smith & Nephew Competitor Duo Knee FemoralSmith & Nephew, Inc.20052

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 2, 2024Class IIZ-0281-2025Packaging change controlComplaint received indicating that a package contained a LEGION NARROW PS OXIN SZ 5N RIGHT implant instead of a GENESIS II POSTERIOR STABILIZED FEMORAL SIZE 5 LEFT as described on the product label.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K042515 (Smith & Nephew, Inc., 2005) has gone furthest, with 118 descendants of its own.

How does its category compare?

All JWH devices

JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JWH

20in 2025

2007: 182008: 172009: 112010: 192011: 252012: 332013: 312014: 332015: 272016: 182017: 252018: 252019: 282020: 222021: 242022: 172023: 202024: 192025: 202026 (partial year): 12
2007high 33 (2012) · low 11 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JWH

55in 2025

2007: 74 days median2008: 81 days median2009: 168 days median2010: 85 days median2011: 87 days median2012: 85 days median2013: 128 days median2014: 98 days median2015: 87 days median2016: 88 days median2017: 118 days median2018: 90 days median2019: 133 days median2020: 59 days median2021: 61 days median2022: 56 days median2023: 57 days median2024: 89 days median2025: 55 days median2026 (partial year): 63 days median
2007high 168 (2009) · low 55 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JWH

2in 2025

2007: 142008: 102009: 1092010: 322011: 272012: 852013: 312014: 132015: 112016: 172017: 612018: 142019: 52020: 22021: 82022: 122023: 672024: 72025: 22026 (partial year): 4
2007high 109 (2009) · low 2 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JWH

9,306in 2025

2007: 1,3052008: 1,4832009: 1,8702010: 3,2862011: 5,1332012: 5,2032013: 6,1352014: 6,7222015: 6,8812016: 7,5862017: 12,2632018: 14,0712019: 14,9872020: 12,5922021: 9,4082022: 8,3642023: 8,9472024: 10,8282025: 9,306
2007high 14,987 (2019) · low 1,305 (2007)2025

MAUDE reports naming product code JWH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071874141,305
20081781101,483
2009111681091,870
20101985323,286
20112587275,133
20123385855,203
201331128316,135
20143398136,722
20152787116,881
20161888177,586
2017251186112,263
201825901414,071
201928133514,987
20202259212,592
2021246189,408
20221756128,364
20232057678,947
20241989710,828
2025205529,306
202612634—

What is it indicated for?

Genesis II and Profix Knee Systems are indicated for:

1. Rheumatoid arthritis.

2. Post-traumatic arthritis, osteoarthritis, or degenerative arthritis in older patients whose age, weight, and activity level are compatible with an adequate long-term resulit.

3. Failed osteotomies, unicompartmental replacement, or total knee replacement.

Genesis II and Profix Knee Systems are indicated for use only with cement and are single use devices.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K962557: Genesis II Zirconium Femoral Component/Profix Zirconium Femoral Component.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K962557

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026